K111906 · Navilyst Medical, Inc. · LJS · Jul 29, 2011 · General Hospital
Device Facts
Record ID
K111906
Device Name
XCELA HYBRID PICC WITH PASV VALVE TECHNOLOGY
Applicant
Navilyst Medical, Inc.
Product Code
LJS · General Hospital
Decision Date
Jul 29, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Xcela™ Hybrid PICC with PASV™ Valve Technology is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the Xcela™ Hybrid PICC with PASV™ Valve Technology is 6 mL/sec.
Device Story
Xcela™ Hybrid PICC is a flexible, radiopaque, triple-lumen catheter with PASV™ (Pressure Activated Safety Valve) technology; features reverse tapered shaft for insertion site hemostasis; includes suture wings and extension tubes with luer lock adapters. Device provides peripheral access to central venous system for fluid/medication administration, blood sampling, and power injection of contrast media (up to 6 mL/sec). Non-valved lumens support central venous pressure monitoring. Used in clinical settings by healthcare professionals. Output is direct venous access; healthcare providers use device to deliver therapies or monitor pressure. Benefits include reduced bleeding at insertion site via tapered shaft and valve-controlled access.
Clinical Evidence
Bench testing only. Performance evaluation conducted per EN ISO 10555-1:2009 and EN ISO 10555-3:1997. Testing included central venous pressure monitoring, luer connection, assembly leak/aspiration strength, priming volume, power injection flow rates (up to 6 mL/sec), static burst, tensile strength, gravity flow rate, and long-term extension tube/clamp life. All tests passed.
Technological Characteristics
Flexible radiopaque catheter; triple-lumen configuration; PASV™ pressure-activated valves; reverse tapered shaft. Materials and design consistent with ISO 10555-1 and ISO 10555-3 standards. Connectivity: None (mechanical device). Sterilization: Sterile.
Indications for Use
Indicated for patients requiring short- or long-term peripheral access to the central venous system for IV therapy (fluids, medications, nutrients), blood sampling, power injection of contrast media, and central venous pressure monitoring (non-valved lumens only).
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
# K 111906
# 510(k) Summary for the XCELA™ HYBRID PICC WITH PASV™ VALVE TECHNOLOGY
Date prepared: 11-May-2011
#### A. Sponsor
Navilyst Medical, Inc 26 Forest Street Marlborough, MA 01752
#### B. Contact
Lorraine M. Hanley VP Global Regulatory Affairs 508-658-7945
and
Wanda Carpinella Sr. Mgr. Global Regulatory Affairs 508-658-7929
#### C. Device Name
Trade Name: Common/Usual name: Classification Name:
Classification Panel:
#### D. Predicate Device(s)
Common/Usual name: Classification Name:
Premarket Notification(s):
#### XCELA™ HYBRID PICC WITH PASVTM VALVE TECHNOLOGY
Peripherally Inserted Central Catheter (PICC) Short and Long-Term Intravascular Catheter 21CFR8880.5970. Class II General Hospital
Peripherally Inserted Central Catheter (PICC) Short and Long-Term Intravascular Catheter 21CFR§880.5970, Class II K101326, K091261, K093366
#### E. Device Description
The Xcela™ Hybrid PICC with PASVTM Valve Technology are flexible radiopaque catheters with suture wings for catheter securement, extension tubes(s) which connect to proximally located luer lock adapter(s) with two pressure activated safety valves (PASV™), available in a triple lumen configuration; a reverse tapered shaft to aid in staunching bleeding at the insertion site and a non-valved lumen for central venous pressure monitoring.
{1}------------------------------------------------
The lumens are differentiated by proximally located colored extension tube clamps and/or colored luer adaptors, with gauge descriptor, if the lumen is rated for power injection the maximum power injection flow rates, and "NO CT" for non-power injectable lumens.
The Xcela™ Hybrid PICC with PASV™ Valve Technology is designed with the option of being used with power injectors for the administration of contrast media for imaging studies such as Computerized Tomography (CT) scans and Magnetic Resonance Imaging (MRIs).
The catheters are available as single, sterile packages with a variety of procedural accessories in standard kit configurations and as a convenience to suite specific clinical needs.
#### F. Intended Use
The Xcela™ Hybrid PICC with PASV™ Valve Technology is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the Xcela™ Hybrid PICC with PASV™ Valve Technology is 6 mL/sec.
#### G. Summary of Similarities and Differences in Technological Characteristics and Performance
The proposed device has similar materials, design, components and technological characteristics as the predicate intravascular catheters.
The predicate devices are offered in sizes 3 F to 6 F outside diameter, single and multilumen (dual and triple) models, whereas the proposed device is offered in 6 F outside diameter. multi-lumen (triple); both predicate and proposed devices offer PASV valved and non-valved models, and are for optional use to power inject contrast media.
#### H. Substantial Equivalence
Based on responses to questions posed in FDA's 510(k) Decision Making Tree, the proposed device is determined to be substantially equivalent to the predicate devices.
{2}------------------------------------------------
#### I. Performance Data
The performance evaluation of the Xcela™ Hybrid PICC with PASV™ included testing conducting in accordance with the following FDA guidance document and international standards with successful results:
- EN ISO 10555-1:2009, Sterile, Single use intravascular catheters Part 1: General . Requirements
- EN ISO 10555-3:1997 COR 2002, Sterile, Single-Use Intravascular Catheters Part . 3: Central Venous Catheters
- . FDA's "Guidance on Premarket Notification [510(K)] Submissions for Short-Term and Long-Term Intravascular Catheters dated March 16, 1995"
The testing performed is summarized below:
## Performance Testing – Proposed 6F Triple Lumen Hybrid PICC with PASV™
| Test | Test Objective | Test Result |
|------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| 1. | Product Labeling: Verify catheter is labeled correctly. | PASS |
| 2. | Central Venous Pressure Monitoring: Verify that catheter can monitor central venous pressure (Non-<br>Valved Lumen Only) | PASS |
| 3. | Luer Connection:<br>Verify luer to EN 1707 | PASS |
| 4. | Assembly Leak Strength: Verify the catheter assembly does not leak simulating the method in ISO<br>10555-1 clause 4.6.1. | PASS |
| 5. | Assembly Aspiration Strength (Closed Ended): Verify catheter assembly does not leak simulating the<br>method in ISO 10555-1 clause 4.6.2. | PASS |
| 6. | Assembly Aspiration Strength (Open Ended):<br>Verify the catheter assembly does not leak simulating the method in ISO 10555-1 clause 4.6.2. | PASS |
| 7. | Priming Volume:<br>Verify the priming volume of the catheter | PASS |
| 8. | Power Injection Flow Rate: Verify the catheter can meet its labeled flow rate. | PASS |
| 9. | Power Injection Multiple Injections: Verify catheter can withstand multiple simulated use injections at<br>the maximum target flow rate. | PASS |
| 10. | Static Burst: Verify catheter does not leak/burst when taken to failure below the minimum requirement. | PASS |
| 11. | Oversleeve to PASVTM Valve Housing and Hybrid Luer Tensile Strength:<br>Verify force at break of the valve housing to oversleeve and hybrid luer to oversleeve simulating the<br>method in ISO 10555-1 clause 4.5 and ISO 10555-3 clause 4.7. | PASS |
| 12. | | Gravity Flow Rate:<br>Verify gravity flow rate of the catheter simulating the method in ISO 10555-3 clause 4.6. |
| 13. | Long term Extension Tube Life (Non-valved lumen):<br>Verify effectiveness of extension tube to undergo repeat clamping without cracking or leaking | PASS |
| 14. | | Long Term Clamp Life (Non-Valved lumen):<br>Verify effectiveness of clamp to undergo repeat clamping without cracking. |
| 15. | Long Term Clamp Life Compatibility with Extension Tube (non-Valved lumen):<br>Verify effectiveness of clamp and extension tube to not leak after repeat clamping. | PASS |
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings. To the left of the eagle is a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Lorraine M. Hanley Vice President, Global Regulatory Affairs Navilyst Medical, Incorporated 26 Forest Street Marlborough, Massachusetts 01752
JUL 29 2011
Re: K111906
Trade/Device Name: Xcela™ Hybrid PICC with PASV™ Valve Technology Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous Implanted Long-term Intravascular Catherter Regulatory Class: II Product Code: LJS Dated: June 30, 2011 Received: July 5, 2011
Dear Ms. Hanley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Hanley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Wang
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
### Indications for Use
510(k) Number (if Known):
Device Name:
Xcela™ Hybrid PICC with PASV™ Valve Technology
Indications for Use:
The Xcela™ Hybrid PICC with PASV™ Valve Technology is indicated for short- or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients, the sampling of blood, and for power injection of contrast media. Non-valved lumens are indicated for central venous pressure monitoring. The maximum power injection flow rate for the Xcela™ Hybrid PICC with PASV™ Valve Technology is 6 mL/sec.
Prescription Use (21 CFR 801 Subpart D) 2 And/Or
AND/OR Over-The-Counter Use: (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
A. St
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.