The PROFEMUR® RENAISSANCE® Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and, 4. revision procedures where other treatments or devices have failed. The PROFEMUR® RENAISSANCE® Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.
Device Story
Monolithic femoral hip stem for uncemented total hip arthroplasty; manufactured from forged titanium alloy (ASTM F620) with titanium plasma spray coating (ASTM F1580). Device provides reduced and standard proximal flare options and neutral/varus neck offsets. Features modified impaction slot oriented 90 degrees from predicate. Implanted by orthopedic surgeons in clinical settings. Functions as structural component in hip replacement; provides mechanical stability for femoral head attachment. Benefits patients via pain relief and improved hip function.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by nonclinical bench testing, including proximal and distal fatigue testing (ISO 7206-4, 6, 8) and range of motion evaluation (ISO 21535).
Technological Characteristics
Monolithic femoral stem; forged titanium alloy (ASTM F620); titanium plasma spray coating (ASTM F1580). Uncemented prosthesis. Dimensions: 7 reduced flare sizes (10-16), 9 standard flare sizes (10-18). Mechanical fixation via press-fit. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory/inflammatory degenerative joint disease (e.g., osteoarthritis, RA, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), functional deformity correction, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{1}------------------------------------------------
K130984 (pg 1/3)
PROFEMUR® RENAISSANCE® Classic Stems Special 510(k) 510(k) Summary
Image /page/0/Picture/2 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a graphic of two angled lines that appear to be overlapping. The text and graphic are black against a white background.
#### 510(k) Summary
# MAY 2 4 2013
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the PROFEMUR® RENAISSANCE® Classic Stems.
(a)(1) Submitted By:
Wright Medical Technology, Inc. 5677 Airline Rd Arlington, TN 38002 (901) 867-4146
Date:
Contact Person:
(a)(2) Proprietary Name of Modified Device:
Common Name:
Classification Name and Reference:
April 03, 2013
Yuan Li. PhD Regulatory Affairs Specialist II ·
PROFEMUR® RENAISSANCE® Classic Stems
Femoral Hip Stem
888.3353 LZO Hip joint metal/ceramic/polymer, cemented or non-porous, uncemented prosthesis Class II
888.3350 JDI Hip joint metal/polymer semiconstrained cemented prosthesis Class II
Subject Product Code and Panel Code:
(a)(3) Predicate Devices:
Orthopedics/87/JDI/LZO
PROFEMUR® RENAISSANCE® Hip Stem, K051995
{2}------------------------------------------------
PROFEMUR® RENAISSANCE® Classic Stems Special 510(k) 510(k) Summary
#### (a)(4) Device Description
The purpose of this submission is to add a monolithic femoral neck option to the PROFEMUR® RENAISSANCE® Hip System. The PROFEMUR® RENAISSANCE® Classic Stems are monolithic stems manufactured from a forged titanium alloy (ASTM F620) and designed for use in uncemented total hip arthroplasty. The PROFEMUR® RENAISSANCE® Classic Stems offer both reduced flare stems (7 sizes, 10-16), and standard flare stems (9 sizes, 10 - 18) and are coated with titanium plasma spray conforming to ASTM F1580, provided by plasma vendors cleared in K123688. The subject device offers two neck offset options, Standard (neutral) and Extended (varus), which are identical to the predicate stems used with modular necks cleared in K100866 and K091423; as well as two proximal flare options, Standard and Reduced. The stem geometry, neck length, and neck offset are designed to create options identical to those available with the predicate modular device.
The subject devices feature one minor additional change in that they have an impaction slot that is oriented 90 degrees from the impaction feature of the predicate device. This is viewed a minor design change, as the rounded slot is not a worse case compared to the predicate design. This modification to the impaction slot was cleared in K123434 and K123688 with similar devices. In addition, these modifications were evaluated by proximal and distal fatigue testing. These modifications do not affect the intended use or alter the fundamental technology of the device.
Instruments used with the system are identical to ones for the predicate system, except for the stem impactor which was reviewed in K123434 and K123688. The subject devices may also be implanted with instruments for alternative surgical approaches, as cleared in K121221 and K122382. Wright has previously provided validation for the cleaning/sterilization of instruments in previous reviews (K122218).
#### (a)(5) Indications for Use
The PROFEMUR® RENAISSANCE® Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed.
The PROFEMUR® RENAISSANCE® Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.
The indications for use can be also found in the package insert previously cleared in
Headquarters Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 901.867.9971 phone www.wmt.com
Tab 007: Page 2 of 3
{3}------------------------------------------------
PROFEMUR® RENAISSANCE® Classic Stems Special 510(k) 510(k) Summary
Image /page/2/Picture/1 description: The image shows the word "WRIGHT." in large, bold, sans-serif font. Above the word "WRIGHT." is the text "K130984 (pg 3/3)". Below the word "WRIGHT." is a graphic of two overlapping shapes.
K123434 and K123688.
# (a)(6) Technological Characteristics of the Device
The indications for use of the PROFEMUR® RENAISSANCE® Classic Stems are identical to those for the predicate device (K051995). The subject devices are made from an identical titanium alloy (ASTM F620) and possess an identical titanium plasma spray coating (ASTM 1580) as the predicate device. The PROFEMUR® RENAISSANCE® Classic Stems are designed to provide identical range of sizes, neck lengths, and neck offsets as those provided by the modular predicate with the exception of featuring a monolithic neck design and a slightly modified impaction slot.
#### (b)(1) Nonclinical Testing
The PROFEMUR® RENAISSANCE® Classic Stems were evaluated by proximal and distal fatigue tests in accordance with ISO 7206-4, 6, and 8 and satisfied the acceptance criteria of each. Range of motion was evaluated in accordance with ISO 21535 and the subject device satisfied the acceptance criteria. The titanium plasma spray coating is applied according to WMT specification and is identical to that used in the predicate device (K051995).
# (b)(2) Clinical Testing
Clinical data was not provided for the subject devices.
#### (b)(3) Conclusions
The indications for use and fundamental scientific technology of the PROFEMUR® RENAISSANCE® Classic Stems are identical to those of the predicate device. The design features and materials of the subject devices are substantially equivalent to those of the predicate devices. The safety and effectiveness of the PROFEMUR® RENAISSANCE® Classic Stems is adequately supported by the substantial equivalence information, materials information, and nonclinical testing data provided within this premarket notification.
Headquarters Arlington, TN 38002 901.867.9971 phone Wright Medical Technology, Inc. 5677 Airline Road www.wmt.com
Tab 007: Page 3 of 3
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2013
Wright Medical Technology, Incorporated % Dr. Yuan Li Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
Re: K130984
Trade/Device Name: PROFEMUR® RENAISSANCE® Classic Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, JDI Dated: April 25, 2013 Received: April 26, 2013
Dear Dr. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
# Page 2 - Dr. Yuan Li
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin I. Keith
For
Mark N. Melkerson . Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K130984 (pg 1/1)
Device Name: PROFEMUR® RENAISSANCE® Classic Stems
Indications for Use:
The PROFEMUR® RENAISSANCE® Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:
- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed.
The PROFEMUR® RENAISSANCE® Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth - Brank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.