The EndoFLIP® system is an endoscopically placed device indicated for use in patients fitted with a gastric band. The device is intended to estimate the size of the stoma produced by the gastric band in a clinical setting. EndoFLIP® EF-620 catheter is indicated for use in measuring the size of a gastric sleeve created during bariatric surgery. It is suitable for diameter measurements for 22 to 60 French sleeves. The EndoFLIP® system is indicated for use in a clinical setting as a pressure and dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity. Note: EndoFLIP is a measurement system. It is not intended to perform a diagnostic test. The EF-325L has a locating LED used in conjunction with the shaft markings to locate the balloon midpoint.
Device Story
EndoFLIP® system uses balloon catheter inserted endoscopically to measure lumen diameter and pressure; catheter attached to syringe pre-filled with diluted saline; balloon inflated to user-programmed volume; system measures diameter at 8 or 16 points along balloon length; EF-325L model includes internal red LED for transillumination to assist surgeon in locating balloon midpoint; used in hospitals/surgery centers by physicians; output provides estimated diameter and pressure readings; assists in bariatric stoma/sleeve sizing and esophageal evaluation; device is measurement tool, not diagnostic test.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including accelerated aging/shelf-life testing and comparison of design/materials to predicate devices.
Technological Characteristics
System consists of measuring console and single-use/reusable balloon catheters. Materials in contact with patient are identical to predicates. Sensing via electrodes/pressure sensors; EF-325L includes low-voltage LED for transillumination. Diameter range 5-25mm or 7-20mm; resolution 0.1mm; accuracy ±1mm. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for patients undergoing bariatric surgery (gastric band or sleeve) or patients with esophageal disorders/symptoms of esophageal sensory hypersensitivity. Contraindicated where endoscopy is contraindicated.
Regulatory Classification
Identification
A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.
{1}------------------------------------------------
| 510(k) Summary<br>Page 1 of 7 | |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| | JUL 1 7 2013 |
| Date Prepared | 28-Jun-13 |
| Company | Crospon Ltd.<br>Galway Business Park<br>Dangan<br>Galway, Ireland<br>Tel - 011 [353] (91) 519882<br>Fax - 011 [353] (91) 519873 |
| Official Contact: | John O'Dea PhD |
| Proprietary or Trade Name: | EndoFLIP® Catheters<br>EF-325, EF-325L, EF-325N,<br>BF-325, BF-325N<br>EF-620 |
| Common/Usual Name: | Gastrointestinal motility monitoring system |
| Classification / CFR: | FFX / CFR 876.1725 / Class 2 |
| Device: | EndoFLIP® System and Catheter |
| Predicate Devices: | K092850 - Crospon - EndoFLIP®<br>K102214 - Crospon - EndoFLIP®<br>K120997 – Crospon – EndoFLIP® |
#### Device Description:
The EndoFLIP® comprises a measuring system and a single use or reusable catheter to assist in measuring the lumen diameter at 16 points (EF family) or 8 points (BF family). The EndoFLIP® comprises a measuring system and catheter to assist in measuring lumens. In practice, the EndoFLIP® balloon catheter is attached to a syringe, pre-filled with a diluted saline solution, which is inserted into the syringe pump on the front of the EndoFLIP® system. The deflated balloon is inserted under direct laparoscopic visualization. Once the balloon has been correctly located, it is then inflated with the diluted saline solution to a user programmed volume.
#### Modifications of these devices vs. Predicates:
The following is a list of the changes to the EndoFLIP® catheters.
- (1) Addition of a model designated as EF-325L that has a locating LED (colored red) positioned inside the wire lumen at the balloon midpoint. It is used in conjunction with the shaft markings to locate the balloon midpoint.
{2}------------------------------------------------
### 510(k) Summary
Page 2 of 7 28-Jun-13
- (2) Extension of the shelf life from 1 year to 2 years for the following products:
- EF-325 and BF-325 family of catheters a.
- EF-620 family of catheters b.
#### Indications for Use:
- K 092850 The EndoFLIP® system is an endoscopically placed device indicated for use . in patients fitted with a gastric band. The device is intended to estimate the size of the stoma produced by the gastric band in a clinical setting.
- . K 102214 - EndoFLIP® EF-620 catheter is indicated for use in measuring the size of a gastric sleeve created during bariatric surgery. It is suitable for diameter measurements for 22 to 60 French sleeves.
- . K 120997 - The EndoFLIP® system is indicated for use in a clinical setting as a pressure and dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity.
- . Note: EndoFLIP is a measurement system. It is not intended to perform a diagnostic test. The EF-325L has a locating LED used in conjunction with the shaft markings to locate the balloon midpoint.
| Patient Population: | Patients undergoing bariatric surgery |
|---------------------|--------------------------------------------------------------------------------|
| Environment of Use: | Hospitals and Surgery Centers |
| Contraindications: | The EndoFLIP® System is contraindicated where endoscopy is<br>contraindicated. |
#### Predicate Device Comparison:
Table 1 compares the Catheters, while Table 2 compares the EF-325L to the Predicate. One will note that the difference between EF-325 and BF-325 is that they are identical except in the number of electrodes and sensors.
The EF-325L is identical to the EF-325 other than the addition of a locating LED mounted on the shaft inside the balloon.
Note that the reference to "Nasal" is the tip design (deflective tip) and not to how the catheter is placed. All of these catheters are placed orally.
The catheters for use with the EndoFLIP® system are viewed as substantially equivalent to the predicate catheters because:
{3}------------------------------------------------
#### 510(k) Summary Page 3 of 7 28-Jun-13
#### Indications -
The indications for use are identical to the predicates. Discussion - The indications for use are unchanged from the predicates, EndoFLIP® - K092850, K102214, and K120997.
#### Technology -
The technology, construction and design are unchanged. Discussion - The technology is unchanged from the predicates, EndoFLJP® - K092850, K102214, and K120997.
#### Environment of Use -
The environments of use - Hospital and surgery centers - are identical. Discussion - The environments of use are unchanged and identical to the predicates, EndoFLIP® - K092850, K102214, and K120997.
## Patient Population -
The patient populations are identical and unchanged.
K092850 - Patients undergoing gastric band surgery and post-operative band adjustment
K 102214 - Patients undergoing bariatric procedures
K 120997 - Patients with esophageal disorders
Discussion - The patient population is unchanged and identical to the predicates, EndoFLIP® -K092850, K102214, and K120997.
#### Non-clinical Testing Summary -
We performed:
- Shelf-life Test .
Packaging is identical to the predicate devices. The additional accelerated aging testing demonstrated that they performed to their specifications and thus are substantially equivalent to the predicate devices,
#### Materials -
The materials in contact with the patient are identical to predicate. Discussion - The materials are unchanged and identical to the predicates, EndoFLIP® -K092850, K102214, and K120997.
{4}------------------------------------------------
## 510(k) Summary Page 4 of 7 28-Jun-13
# Table 1 – Comparison of Proposed Device vs. Predicate
| | EndoFLIP® system with catheter<br>K092850, K102214, K120997 | Proposed<br>EndoFLIP® Catheters |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Indications for Use | K092850<br>The EndoFLIP® system is an<br>endoscopically placed device<br>indicated for use in patients fitted<br>with a gastric band. The device is<br>intended to estimate the size of the<br>stoma produced by the gastric band in<br>a clinical setting. | Identical, unchanged |
| | K102214<br>EndoFLIP® EF-620 catheter is<br>indicated for use in measuring the<br>size of a gastric sleeve created during<br>bariatric surgery. It is suitable for<br>diameter measurements for 22 to 60<br>French sleeves. | |
| | K120997<br>The EndoFLIP® system is indicated<br>for use in a clinical setting as a<br>pressure and dimension measurement<br>device and as an adjunct to other<br>methods in the comprehensive<br>evaluation of patients with symptoms<br>consistent with esophageal sensory<br>hypersensitivity. | |
| | Note: EndoFLIP® is a measurement<br>system. It is not intended to perform a<br>diagnostic test. | |
| Environments of use | Hospital and surgery centers | Identical, unchanged |
{5}------------------------------------------------
# 510(k) Summary
28-Jun-13
| EndoFLIP® system with catheter | Proposed |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K092850, K102214, K120997 | EndoFLIP® Catheters |
| K092850 - Patients undergoing | Identical, unchanged |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | Identical, unchanged |
| | |
| | |
| Prescription use | Identical, unchanged |
| Provides an Estimated Diameter | Identical, unchanged |
| (Dest) of the balloon at 16 or 8 points | |
| along its length when inflated with | |
| saline solution. | |
| | |
| Records balloon pressure. | |
| All materials have passed | Identical to K092850, unchanged |
| | |
| | |
| | Identical, unchanged |
| | |
| | |
| | Accessories are supplied non-sterile, |
| | and are single patient use, disposable |
| | Identical, unchanged |
| | |
| | |
| | |
| | |
| integer | |
| | gastric band surgery and post-<br>operative band adjustment<br>K 102214 - Patients undergoing<br>bariatric procedures<br>K 120997 - Patients with esophageal<br>disorders<br>The EndoFLIP® System is<br>contraindicated where endoscopy is<br>contraindicated.<br>biocompatibility tests in accordance<br>with ISO 10993-1<br>Only operates with EndoFLIP®<br>system<br>Accessories are supplied non-sterile.<br>and are single patient use, disposable<br>Range: 5 to 25 mm<br>and 7 to 20 mm<br>Resolution: 0.1 mm<br>Accuracy: ± 1mm (at 95%<br>confidence) rounded to nearest |
{6}------------------------------------------------
#### 510(k) Summary Page 6 of 7 28-Jun-13
### Table 2 - Locating LED Comparison to Predicate
| | Proposed EndoFLIP®<br>EF-325L catheter | Predicate Stryker InfraVision™<br>K983220 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | All of the EndoFLIP indications for<br>use under K092850, K102214, and<br>K120997 plus the new feature<br><br>Transillumination is intended to help<br>the surgeon with placement of EF-<br>325L catheter | Specific indication for helping to<br>locate or position devices<br><br>Esophageal transilluminating dilation<br>device with a reusable illumination<br>source, and a combined single use<br>disposable fiberoptic light guide,<br>nasogastric tube and balloon dilation<br>system. Intended to transilluminate<br>and dilate the esophagus during<br>fundoplication procedure.<br><br>Transillumination is intended to help<br>the surgeon placement of medical<br>devices within body tissue. |
| Environments of use | Hospital and surgery centers | Identical |
| Patient Population | Identical to K092850, K102214,<br>K120997 | Similar |
| Contraindications | The EndoFLIP® System is<br>contraindicated where endoscopy is<br>contraindicated. | Not stated |
| Prescription/OTC | Prescription use | Prescription use |
| Principle of Operation | LED to assist with transillumination<br>LED located inside balloon | Light source for transilluminating<br>Separate light source |
| Materials used | The LED is located inside the<br>catheter balloon and is isolated from<br>the patient | N/A |
The Locating LED feature does not alter the primary indications for use for the EF-325 catheter. It is to assist the user in locating the balloon mid-point. As the primary indications for use has not changed this feature can be as being substantially equivalent to the predicate because:
#### Indications -
The indications for use are identical to the predicate catheter and the predicate Stryker InfraVision™ system (K983220).
For the Locating LED feature Transillumination is intended to help the surgeon with placement of EF-325L catheter K983220 is indicated for helping to locate or position devices
Note: EndoFLIP® is a measurement system. It is not intended to perform a diagnostic test.
{7}------------------------------------------------
# 510(k) Summary
Page 7 of 7 28-Jun-13
Discussion - The indications for use are unchanged for the catheter from the predicates, EndoFLJP® - K092850. K102214, and K120997 and the Locating LED feature is similar to the predicate Stryker InfraVision (K983220).
# Technology -
The technology, construction and design are unchanged. The addition of a low voltage LED with the pressure sensor inside the balloon does not alter the construction.
Discussion - The technology is unchanged from the predicates, EndoFLIP® - K092850, K102214, and K120997 and the addition of a LED for a light source is similar to the predicate Stryker InfraVision (K983220).
## Materials -
The materials in contact with the patient are identical to predicate. Discussion - The materials are unchanged and identical to the predicates, EndoFLIP® - K092850, K102214, and K120997.
## Environment of Use -
The environments of use - Hospital and surgery centers - are identical. Discussion - The environments of use are unchanged and identical to the predicates, EndoFLIP® -K092850, K102214, and K120997.
#### Patient Population -
The patient populations are identical and unchanged.
K092850 Patients undergoing gastric band surgery and post-operative band adjustment K102214 Patients undergoing bariatric procedures K120997 Patients with esophageal disorders
Discussion - The patient population is unchanged and identical to the predicates. EndoFLIP® -K092850, K102214, and K120997.
# Non-clinical Testing Summary -
There was no additional non-clinical testing for the locating LED feature.
# Substantial Equivalence Conclusion -
The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to substantially equivalent.
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### July 17, 2013
Crospon Ltd. % Paul Dryden Regulatory Consultant 24301 Woodsage Dr. Bonita Springs, FL 34134-2958
Re: K130906
Trade/Device Name: EndoFLIP® System with catheters (EF-325, EF325N, EF-325L, BF-325 BF-325N, EF-620) Regulation Number: 21 CFR §876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: II Product Code: FFX Dated: June 28, 2013 Received: July 1, 2013
Dear Paul Dryden,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, derestere, manns of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and advises, 200 Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{9}------------------------------------------------
Page 2 - Paul Dryden
You must comply with all the Act's requirements, including, but not limited to: registration and Iisting (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
#### Indications for Use Statement
Page 1 of 1
510(k) Number:
K130906
Device Name:
EndoFLIP® System with catheters o EF-325 o EF-325N o EF-325L o BF-325 o BF-325N o EF-620
#### Indications for Use:
- The EndoFLIP® system is an endoscopically placed device indicated for use in . patients fitted with a gastric band. The device is intended to estimate the size of the stoma produced by the gastric band in a clinical setting.
- EndoFLIP® EF-620 catheter is indicated for use in measuring the size of a gastric . sleeve created during bariatric surgery. It is suitable for diameter measurements for 22 to 60 French sleeves.
- The EndoFLIP® system is indicated for use in a clinical setting as a pressure and . dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity.
Note: EndoFLIP is a measurement system. It is not intended to perform a diagnostic test. The EF-325L has a locating LED used in conjunction with the shaft markings to locate the balloon midpoint.
| Prescription Use XX | or | Over-the-counter use |
|-----------------------------|----|------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|-------------------------------------|--------------------------------------------------------|
| Benjamin R. Fisher -S | |
| | 2013.07.07.116 447-23 -04'00' |
| (Division Sign-Off) | |
| | Division of Reproductive, Gastro-Renal, and |
| Urological Devices<br>510(k) Number | K130906 |
Page 4.3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.