ORTHOCORD

K123668 · Depuy Mitek, Inc., A Johnson and Johnson Company · NEW · Feb 26, 2013 · General, Plastic Surgery

Device Facts

Record IDK123668
Device NameORTHOCORD
ApplicantDepuy Mitek, Inc., A Johnson and Johnson Company
Product CodeNEW · General, Plastic Surgery
Decision DateFeb 26, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4840
Device ClassClass 2
AttributesTherapeutic

Indications for Use

ORTHOCORD Suture is indicated for use in general soft tissue approximation, and/or ligation, including orthopaedic procedures.

Device Story

Small Size Orthocord Suture is a synthetic, sterile, braided, composite suture. It consists of dyed absorbable polydioxanone (PDS) and undyed non-absorbable polyethylene, coated with a copolymer of 90% caprolactone and 10% glycolide. The device is used by surgeons in clinical or operating room settings for soft tissue approximation and ligation. It is provided in various sizes (0, 2-0, 3-0, 4-0) and configurations, including single-armed and loop formats. The suture provides mechanical tissue support during healing. It is also utilized as a component in specific anchor configurations (Minilok and Microfix).

Clinical Evidence

Bench testing only. Non-clinical laboratory testing demonstrated conformance to USP monograph requirements for absorbable sutures, confirming mechanical performance and material safety consistent with predicate devices.

Technological Characteristics

Synthetic, braided, composite suture. Materials: absorbable polydioxanone (PDS) and non-absorbable polyethylene. Coating: 90% caprolactone and 10% glycolide copolymer. Sizes: 0, 2-0, 3-0, 4-0. Configurations: single-armed and loop. Sterilized.

Indications for Use

Indicated for general soft tissue approximation and/or ligation, including orthopaedic procedures, in patients requiring surgical suture fixation.

Regulatory Classification

Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123668 page 1/2 ### SECTION 2 - 510(k) SUMMARY ### Small Size Orthocord® Suture FEB 2 6 2013 | Submitter's<br>Name and<br>Address | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 | |----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Tatyana Korsunsky<br>Regulatory Affairs Specialist<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA | | | Telephone: 508-828-3122<br>Facsimile: 508-977-6911<br>e-mail: tkorsuns@its.jnj.com | | Name of<br>Medical Device | Proprietary Name: ORTHOCORD®<br>Classification Name: Suture, surgical, absorbable, polydioxanone;<br>Suture, nonabsorbable, synthetic, polyethylene<br>Common Name:<br>Suture | | Substantial<br>Equivalence<br>Facility | The Small Size Orthocord® Suture is substantially equivalent to:<br>• K040004: Orthocord® Suture (USP Size 2, Violet) (April 13, 2004)<br>• K043298: Orthocord® Suture (USP Size 2, Blue) (Dec 10, 2004)<br>• K071257: Minilok Anchor with #2-0 Orthocord® Suture (Jun 29, 2007)<br>• K080352: Microfix Anchor with #3-0 and 4-0 Orthocord® Suture (Mar 12, 2008)<br>• K946173: Ethibond Suture (January 09, 1995)<br>• K022715: Vicryl Suture (December 19, 2002)<br>• K122374, K071622, K041553: Fiberwire FiberLoop Suture (September 25, 2012;<br>July 03, 2007; December 10, 2004) | | Device<br>Classification | Class II:<br>• Sec. 878.4840 Absorbable polydioxanone surgical suture, product code NEW;<br>• Sec. 878.5000 Nonabsorbable poly(ethylene terephthalate) surgical suture,<br>product code GAT. | | Device<br>Description | Orthocord® Suture is a synthetic, sterile, braided, composite suture composed of dyed<br>(D&C Violet #2 or D&C Blue #6) absorbable polydioxanone (PDS) and un-dyed non-<br>absorbable polyethylene. The partially absorbable suture is coated with a copolymer of<br>90% caprolactone and 10% glycolide. | | Indications for<br>Use | ORTHOCORD Suture is indicated for use in general soft tissue approximation, and/or<br>ligation, including orthopaedic procedures. | | Safety and<br>Performance | The determination of substantial equivalence for this device was based on a detailed<br>device description, and conformance to consensus and voluntary standards. Non-<br>clinical laboratory testing was performed demonstrating that the device conformed to<br>the USP monograph for absorbable sutures, except for oversized diameter.<br>The proposed Small Size Orthocord sutures are comprised of the same materials and<br>are similar design to the predicate Size 2 Orthocord suture (K040004/K043298) with<br>differences in size and braid construction. The proposed device, Small Size Orthocord,<br>is currently offered in an anchor configuration of the DePuy Mitek's Minilok<br>(K071257) and Microfix (K080352) anchors.<br>Based on the indications for use, technological characteristics, and comparison to<br>predicate devices, the ORTHOCORD suture has been shown to be substantially<br>equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act. | . {1}------------------------------------------------ K123668 page 2/2 Small Size Orthocord Suture Sizes and Configurations: Size 0 Orthocord Suture; (single armed) Size 2-0 Orthocord Suture; (single armed) Size 3-0 Orthocord Suture; (single armed) Size 4-0 Orthocord Suture; (single armed) Size 4-0 Orthocord Suture in a Loop (both ends of suture swaged to one needle) . {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the eagle. The eagle is depicted in black, and the text is also in black. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 DePuy Mitek Incorporated., A Johnson and Johnson Company % Ms. Tatyana Korsunsky Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts, 02767 Letter dated: February 26, 2013 Re: K123668 Trade/Device Name: Small Size ORTHOCORD® Suture Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable polydioxanone surgical suture Regulatory Class: Class II Product Code: NEW, GAT Dated: January 24, 2013 Received: January 28, 2013 Dear Ms. Korsunsky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Tatyana Korsunsky forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Mark N.Melkerson -S Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K123668 Device Name: Small Size ORTHOCORD® Suture Indications for Use: The Small Size ORTHOCORD® Suture is indicated for: ORTHOCORD Suture is indicated for use in general soft tissue approximation, and/or ligation, including orthopaedic procedures. Prescription Use _____________________________________________________________________________________________________________________________________________________________ . AND/OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # David Kraúse (Division Sign-Off) Division of Surgical Devices 510(k) Number: K123668 : . Page 1 of_1 22 | P a g e
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...