ORTHOCORD SUTURE

K040004 · Depuy Mitek · NEW · Apr 13, 2004 · General, Plastic Surgery

Device Facts

Record IDK040004
Device NameORTHOCORD SUTURE
ApplicantDepuy Mitek
Product CodeNEW · General, Plastic Surgery
Decision DateApr 13, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4840
Device ClassClass 2
AttributesTherapeutic

Intended Use

ORTHOCORD Suture is indicated for use in general soft tissue approximation and/or ligation, including use in orthopedic surgeries.

Device Story

ORTHOCORD Suture is a synthetic, sterile, braided composite suture used for soft tissue approximation and ligation in orthopedic surgeries. It consists of dyed (D&C Violet #2) absorbable polydioxanone (PDS) and un-dyed non-absorbable polyethylene, coated with a copolymer of 90% caprolactone and 10% glycolide. The device is intended for use by surgeons in clinical settings.

Clinical Evidence

Bench testing only. The device demonstrated conformance to the USP monograph for absorbable sutures.

Technological Characteristics

Synthetic, sterile, braided composite suture. Materials: polydioxanone (PDS) and polyethylene. Coating: 90% caprolactone and 10% glycolide copolymer. Dyed with D&C Violet #2. Conforms to USP monograph for absorbable sutures.

Indications for Use

Indicated for general soft tissue approximation and/or ligation, including orthopedic surgical procedures.

Regulatory Classification

Identification

An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KO4OOOY P/2 ## APR 1 3 2004 ## SECTION 2 – 510(k) SUMMARY ## ORTHOCORD Suture | Submitter's Name and<br>Address: | DePuy Mitek<br>a Johnson & Johnson company<br>249 Vanderbilt Avenue<br>Norwood, MA 02062 | |----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Ruth C. Forstadt<br>Senior Regulatory Affairs Associate<br>DePuy Mitek<br>a Johnson & Johnson company<br>249 Vanderbilt Avenue<br>Norwood, MA 02062<br>Telephone: 781-251-3188<br>Facsimile: 781-278-9578<br>e-mail: rcforstad@ethus.jnj.com | | Name of Medical Device | Classification Name:<br>Non-absorbable poly(ethylene terephthalate)<br>surgical suture; and<br>Common/Usual Name:<br>Suture, Surgical, Absorbable, Polydioxanone<br>Suture<br>Proprietary Name:<br>ORTHOCORD Suture | | Substantial Equivalence | ORTHOCORD Suture is substantially equivalent to:<br>PDS II suture (N18331) manufactured by Ethicon, Inc. and FiberWire<br>suture (K010673) manufactured by Arthrex, Inc. | | Device Classification | Sutures are classified by the FDA as Class II Medical Devices.<br>PDS Suture carries an FDA product code NEW, and is classified as<br>absorbable surgical suture, polydiaxanone under 21 CFR 878.4840.<br>Polyethylene suture carries an FDA product code GAT, and is classified<br>under 21 CFR 878.5000. | | Device Description | ORTHOCORD suture is a synthetic, sterile, braided composite suture<br>composed of dyed (D&C Violet #2) absorbable polydioxnone (PDS)<br>and un-dyed non-absorbable polyethylene. The partially absorbable<br>suture is coated with a copolymer of 90% caprolactone and 10%<br>glycolide. | Confidential {1}------------------------------------------------ KOYOOOY P2/2 ORTHOCORD Suture is indicated for use in general soft tissue Indications for Use approximation and/or ligation, including use in orthopedic surgeries. The determination of substantial equivalence for this device was based Safety and Performance on a detailed device description, and conformance to consensus and voluntary standards. Non-clinical laboratory testing was performed demonstrating that the device conformed to the USP monograph for absorbable sutures. Based on the indications for use, technological characteristics, and comparison to predicate devices, the Mitek ORTHOCORD Suture has been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. Inside the circle is an abstract image of a bird-like figure. Public Health Service APR 1 3 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Ruth C. Forstadt Senior Regulatory Affairs Associate DePuy Mitek 249 Vanderbilt Avenue Norwood, Massachusetts 02062 Re: K040004 Trade/Device Name: Orthocord Suture Regulation Number: 21 CFR 878.4840, 878.5000 Regulation Name: Absorbable PDS suture, Nonabsorbable polyethylene suture Regulatory Class: II Product Code: NEW, GAT Dated: March 31, 2004 Received: April 1, 2004 Dear Mrs. Forstadt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Ruth C. Forstadt This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entifled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Koyoooy ## Indications for Use 510(k) Number (if known): K040000 Device Name: ORTHOCORD Suture Indications for Use: ORTHOCORD Suture is indicated for use in general soft tissue approximation and/or ligation, including use in orthopedic surgeries. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) · (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) uriam C. Provost (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page 1 of ____________________________________________________________________________________________________________________________________________________________________ **510(k) Number** K040004
Innolitics
510(k) Summary
Decision Summary
Classification Order