STAYHEALTHY HEALTH KIOSK, STAYHEALTHY HEALTHCENTER KIOSK

K123539 · Stayhealthy, Inc. · DXN · Apr 22, 2013 · Cardiovascular

Device Facts

Record IDK123539
Device NameSTAYHEALTHY HEALTH KIOSK, STAYHEALTHY HEALTHCENTER KIOSK
ApplicantStayhealthy, Inc.
Product CodeDXN · Cardiovascular
Decision DateApr 22, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The Stayhealthy SH650 is intended to be used by the general public so that the user can measure his/her own blood pressure, heart rate, percent lean body mass, percent body fat, percent total body water, body weight, and if the user may have color blindness. It is not a diagnostic device and only furnishes data so that users can consult their personal physicians or other healthcare professional. The user can also choose to print the data or send the data to a personal physician or a healthcare professional.

Device Story

Stayhealthy SH650/SH650-C is a public-access health kiosk for self-measurement of physiological parameters. Inputs include blood pressure (via cuff), heart rate, body composition (via bioelectrical impedance analysis), weight (via scale), and color blindness screening (via visual test). Device processes inputs to provide quantitative health data to the user. Used in public settings; operated by the user. Output is provided via printout or electronic transmission to a healthcare professional. Device serves as a screening tool to facilitate consultation with physicians; it does not provide medical diagnoses. Benefits include increased access to health monitoring for the general public.

Clinical Evidence

No clinical data provided in the document; substantial equivalence determination based on regulatory review of device specifications and intended use.

Technological Characteristics

Kiosk-based system for physiological monitoring. Includes noninvasive blood pressure cuff, bioelectrical impedance analysis (BIA) for body composition, digital scale for weight, and visual testing module for color blindness. Connectivity includes printing and electronic data transmission capabilities. Class II device.

Indications for Use

Indicated for general public use for self-measurement of blood pressure, heart rate, body weight, body composition (percent lean body mass, body fat, total body water), and color blindness screening. Not for diagnostic use.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 22, 2013 Stayhealthy, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313 Re: K123539 > Trade/Device Name: Stayhealthy Kiosk SH650 and SH650-C Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN, MNW and HIT Dated: March 22, 2013 Received: March 25, 2013 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Mr. Mark Job Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Owen-PFaris -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use Page _________________________________________________________________________________________________________________________________________________________________________ 510(k) Number (if known): Device Name: Stayhealthy SH650 Health Kiosk Indications for Use: The Stayhealthy SH650 is intended to be used by the general public so that the user can measure his/her own blood pressure, heart rate, percent lean body mass, percent body fat, percent total body water, body weight, and if the user may have color blindness. It is not a diagnostic device and only furnishes data so that users can consult their personal physicians or other healthcare professional. The user can also choose to print the data or send the data to a personal physician or a healthcare professional. | Prescription Use | OR | Over-the-Counter Use_X_ | |----------------------------|----|-------------------------| | (Per 21 CFR 810.subpart D) | | (21 CFR 801 Subpart C) | ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Owen P. Faris -S 2013.04.22 15:53:30 -04'00' Page 1 of 1
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%