K113402 · Solohealth, Inc. · DXN · May 30, 2012 · Cardiovascular
Device Facts
Record ID
K113402
Device Name
SOLOHEALTH STATION
Applicant
Solohealth, Inc.
Product Code
DXN · Cardiovascular
Decision Date
May 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The SoloHealth Station is intended to be used by the general public so that a user can measure his/her own blood pressure and pulse rate and his/her own weight. Additionally, the SoloHealth Station is intended to screen adults for clarity of central vision. SoloHealth Station does not provide a general screening of visual function and does not provide a diagnosis of eye health or other disease. The SoloHealth Station only screens clarity of central vision. Users should consult their personal physicians if they have concerns regarding their eyesight.
Device Story
SoloHealth Station is an automated, kiosk-based system for use in high-traffic indoor commercial areas. It measures blood pressure and pulse rate via an oscillometric cuff, weight via a scale, and screens central vision clarity. The user interacts with the device via an LCD touchscreen and voice instructions. Data can be stored, accessed via a website, or emailed to the user. The device provides health data to the user, who is advised to consult a physician for interpretation or concerns. It serves as a point-of-care screening tool for the general public to monitor health parameters.
Clinical Evidence
Bench testing only. Compliance with AAMI/ANSI SP10:2002/(R)2008 and A1:2003 for blood pressure measurement was verified via the module supplier and confirmed on the final kiosk design. Limited confirmatory testing on human subjects ensured the alternate cuff did not affect results. Vision screening validation demonstrated similarity to licensed optometrist screening methods.
Technological Characteristics
Kiosk-based system with LCD touchscreen, metal housing, and latex-free polyester cuff. Measures blood pressure (oscillometric), weight, and central vision. Connectivity includes web-based data storage and email. Electrical safety per IEC 60601-1:1988/A1:1991/A2:1995 and IEC 60601-2-30:1999. EMC per IEC 60601-1-2:2001. Software is automated, rule-based system for data acquisition and user guidance.
Indications for Use
Indicated for general public use to measure blood pressure, pulse rate, and weight, and to screen adults for clarity of central vision. Not for general visual function screening or diagnosis of eye health/disease. Users advised to consult physicians for eyesight concerns.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size.
January 19, 2022
Solohealth, Inc. % Jeff Rongero Third Party Official Underwriters Laboratories, Inc. 12 Laboratory Dr. Research Triangle, North Carolina 27709
Re: K113402
Trade/Device Name: Solohealth Station Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, HOX
Dear Jeff Rongero:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 30, 2012. Specifically, FDA is updating this letter to reflect the correct product code.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, Jennifer.Shih(@fda.hhs.gov.
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 3 0 2012
SoloHealth, Inc. c/o Mr. Jeff D. Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709
Re: K113402
Trade Name: SoloHealth Station Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Codes: DRG, HOX Dated: May 24, 2012 Received: May 25, 2012
# Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Mr. Jeff D. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
# 510(k) Number (if known): K113402
Device Name: SoloHealth Station
## Indications for Use:
The SoloHealth Station is intended to be used by the general public so that a user can measure his/her own blood pressure and pulse rate and his/her own weight. Additionally, the SoloHealth Station is intended to screen adults for clarity of central vision. SoloHealth Station does not provide a general screening of visual function and does not provide a diagnosis of eye health or other disease. The SoloHealth Station only screens clarity of central vision. Users should consult their personal physicians if they have concerns regarding their eyesight.
| Prescription Use | No |
|-----------------------------|----|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | Yes |
|------------------------|-----|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Cardiovascular Devices
| 510(k) Number | K113402 |
|---------------|---------|
|---------------|---------|
{5}------------------------------------------------
SoloHealth Station
# K113402
MAY 3 0 2012
# 510(k) Summary SoloHealth Station K113402 May 18, 2012
| Company: | SoloHealth, Inc.<br>11555 Medlock Bridge, Suite 190<br>Duluth, GA 30097<br>Phone: (770) 622-4158<br>Fax: (770) 622-4122 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration: | 3007156596 |
| Primary Contact: | Kimberly Strohkirch<br>Memphis Regulatory Consulting, LLC<br>Phone: (901) 361-2037 |
| Company Contact: | Stephen Kendig<br>Chief Operating Officer |
| Trade Name: | Solo Health Station |
| Common Name: | Automated Blood Pressure Monitor |
| Classification: | Class II |
| Regulation Number: | 870.1130 - Non-invasive blood pressure measurement system |
| Panel: | Cardiovascular |
| Product Code: | DXN |
#### Device Description:
The SoloHealth Station is an automated system for measuring blood pressure and pulse rate designed to be used by the general public in indoor high-traffic commercial areas. It is completely automatic, and measures blood pressure by the oscillometric method. The user is guided by a series for interactive screen and voice instructions. Additionally, the SoloHealth Station measures weight, screens clarity of central vision, and does a risk factor analysis. Users are advised to consult a physician. Upon completion, data may be stored by the user and accessed via a website or sent via electronic mail to the user.
{6}------------------------------------------------
#### Indications for use:
The SoloHealth Station is intended to be used by the general public so that a user can measure his/her own blood pressure and pulse rate and his/her own weight. Additionally, the SoloHealth Station is intended to screen adults for clarity of central vision. SoloHealth Station does not provide a general screening of visual function and does not provide a diagnosis of eye health or other disease. The SoloHealth Station only screens clarity of central vision. Users should consult their personal physicians if they have concerns regarding their eyesight.
#### Substantial Equivalence:
The SoloHealth Station is substantially similar to predicate devices currently on the market. These devices are the LifeClinic® 2400 (K040562), Computerized Screening, Inc CSI Model 5k Managed Health System Kiosk (K093389) and Xperex Health Check Kiosk (K063008) as shown in Table 1. The central vision screening of the SoloHealth Station is similar to Accutome Eye Charts, Vimetrics Central Vision Analyzer 1000, and Optec 5000 Series Vision Tester shown in Table 2. The SoloHealth Station has the following main components: Blood Pressure Application, Weight Measuring Application, Vision Test, and Test Results. Biocompatibility for the blood pressure cuff is established utilizing Xperex's Health Check Kiosk (K063008).
{7}------------------------------------------------
:
:
| | SoloHealth<br>Station | LifeClinic® 2400<br>(K040562) | Computerized<br>Screening, Inc<br>CSI Model 5k<br>Managed<br>Health System<br>Kiosk<br>(K093389) | Xperex Inc.<br>Health Check<br>Kiosk<br>(K063008) | Comparison |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | General public to<br>measure blood<br>pressure and<br>pulse rate;<br>Weight<br>measurement<br>and screen<br>clarity of central<br>vision | General public to<br>measure blood<br>pressure and<br>pulse rate; Not a<br>diagnostic device | General public<br>to measure<br>blood pressure<br>and pulse rate;<br>Not a<br>diagnostic<br>device | General public to<br>measure blood<br>pressure and<br>pulse rate; Not a<br>diagnostic device | Similar to LifeClinic<br>2400, CSI Model 5k<br>and Xperex Health<br>Check Kiosk |
| Indications<br>for Use | The SoloHealth<br>Station is<br>intended to be<br>used by the<br>general public so<br>that a user can<br>measure his/her<br>own blood<br>pressure and<br>pulse rate and<br>his/her own<br>weight.<br>Additionally, the<br>SoloHealth<br>Station is<br>intended to<br>screen adults for<br>clarity of central<br>vision.<br>SoloHealth<br>Station does not<br>provide a<br>general<br>screening of<br>visual function<br>and does not<br>provide a<br>diagnosis of eye<br>health or other<br>disease. The<br>SoloHealth<br>Station only | The Lifeclinic<br>2400 is intended<br>to be used by<br>the general<br>public so that a<br>user can<br>measure his/her<br>own blood<br>pressure and<br>pulse rate. It is<br>not a diagnostic<br>device, and only<br>furnished data<br>so that the users<br>can consult their<br>personal<br>physician. | The CSI Model<br>5K Managed<br>Health System<br>Kiosk is<br>intended for<br>use by the<br>general public<br>to measure<br>blood pressure,<br>pulse and<br>weight. It is not<br>intended to be<br>a diagnostic<br>device; it only<br>provides data<br>on blood<br>pressure, heart<br>rate and weight<br>and users are<br>advised to<br>consult a<br>physician. | The Health<br>Check Kiosk is<br>intended to be<br>used by the<br>general public so<br>that a user can<br>measure health<br>parameters such<br>as weight, body<br>fat, blood<br>pressure and<br>pulse rate in<br>public places<br>and/or<br>corporate<br>environments. It<br>is not for<br>diagnostic use. | Similar to LifeClinic<br>2400, CSI Model 5k<br>and Xperex Health<br>Check Kiosk |
| | SoloHealth<br>Station | LifeClinic® 2400<br>(K040562) | Computerized<br>Screening, Inc<br>CSI Model 5k<br>Managed<br>Health System<br>Kiosk<br>(K093389) | Xperex Inc.<br>Health Check<br>Kiosk<br>(K063008) | Comparison |
| | screens clarity of<br>central vision.<br>Users should<br>consult their<br>personal<br>physicians if they<br>have concerns<br>regarding their<br>eyesight. | | | | |
| Intended<br>Population | General Public | General Public | General Public | General Public | Identical |
| Human<br>Factors | Shock Hazard<br>tested per IEC<br>60601-1:1988<br>and IEC 60601-2-<br>30:1999; Blood<br>Pressure Risks<br>tested per AAMI<br>SP10 | Blood Pressure<br>Risks tested per<br>AAMI SP10 | Blood Pressure<br>Risks tested per<br>AAMI SP10 | Not available | Risks addressed<br>per current<br>industry standards |
| Hardware<br>Design | LCD Interface to<br>control blood<br>pressure and<br>pulse rate<br>measuring<br>device. A stop<br>button stops the<br>blood pressure<br>rate and<br>measuring<br>device. | Start/stop<br>button to control<br>blood pressure<br>and pulse rate<br>measuring<br>device | LCD Interface to<br>control blood<br>pressure and<br>pulse rate<br>measuring<br>device | Start/stop<br>button to control<br>blood pressure<br>and pulse rate<br>measuring<br>device | SoloHealth Kiosk<br>uses an LCD screen<br>as opposed to<br>directions on the<br>device for the user<br>to control the<br>device. A stop<br>button stops the<br>blood pressure<br>rate and measuring<br>device. No<br>additional risks for<br>control method. |
| Software<br>Design | The SoloHealth<br>Station uses<br>software as an<br>automated<br>system for<br>measuring<br>blood pressure<br>and pulse rate.<br>It is completely<br>automatic, and<br>measures blood<br>pressure by the | The Lifeclinic<br>2400 uses<br>software as an<br>automated<br>system for<br>measuring<br>blood pressure<br>and pulse rate.<br>It is completely<br>automatic, and<br>measures blood<br>pressure by the | The CSI 5k<br>Kiosk uses<br>software as an<br>automated<br>system for<br>measuring<br>blood pressure<br>and pulse rate.<br>It is completely<br>automatic, and<br>measures<br>blood pressure | Xperex Health<br>Check Kiosk uses<br>software as an<br>automated<br>system for<br>measuring<br>blood pressure<br>and pulse rate.<br>It is completely<br>automatic, and<br>measures blood<br>pressure by the | The software<br>program used by<br>the predicate is not<br>available for direct<br>comparison.<br>However, both<br>systems provide<br>blood pressure<br>measurements in<br>an automatic way<br>using software. |
| | SoloHealth<br>Station | LifeClinic® 2400<br>(K040562) | Computerized<br>Screening, Inc<br>CSI Model 5k<br>Managed<br>Health System<br>Kiosk<br>(K093389) | Xperex Inc.<br>Health Check<br>Kiosk<br>(K063008) | Comparison |
| | oscillometric<br>method. The<br>user is guided<br>by a series of<br>interactive<br>screens and<br>voice<br>instructions. | oscillometric<br>method. The<br>user is guided by<br>a series of<br>interactive<br>screens and<br>voice<br>instructions. | by the<br>oscillometric<br>method. The<br>user completes<br>functions by<br>pushing labeled<br>buttons. | oscillometric<br>method. The<br>user is guided by<br>a series of<br>interactive<br>screens and<br>voice<br>instructions. | |
| Dimensions<br>and Weight | 2 ft wide x 3 ft<br>deep x 5 feet tall | Not available | Not available | Not available | Dimensions of<br>predicate devices<br>are not available<br>for comparison. |
| Performance<br>Standards<br>Met | AAMI Standards<br>IEC Standards | AAMI Standards | AAMI Standard<br>IEC Standards | AAMI Standards | Identical AAMI<br>Standards.<br>SoloHealth Kiosk<br>also meets<br>Electrical Safety<br>Standards. |
| Materials | Commercially<br>available<br>materials<br>including soda<br>lime glass for<br>touchscreen,<br>metal housing,<br>and latex-free<br>polyester thread<br>for cuff | Not available | Not available | Latex-free<br>polyester thread<br>for cuff. The rest<br>of the material is<br>not available. | Materials of<br>predicate not<br>available for<br>comparison.<br>Materials of<br>SoloHealth Station<br>currently<br>commercially<br>available. |
| Cleaning/Disi<br>nfection/Ster<br>ilization | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Identical |
| Biocompatibi<br>lity | Commercially<br>available<br>materials<br>including soda<br>lime glass for<br>touchscreen,<br>metal housing,<br>and latex-free<br>polyester thread<br>for cuff. Blood<br>pressure cuff is<br>identical to cuff<br>used in Xperex | None listed in<br>510(k) Summary | None listed in<br>510(k)<br>Summary | None listed in<br>510(k) Summary | Materials of<br>SoloHealth are<br>widely available<br>commercially and<br>do not necessitate<br>biocompatibility<br>testing. |
| | SoloHealth<br>Station | LifeClinic® 2400<br>(K040562) | Computerized<br>Screening, Inc<br>CSI Model 5k<br>Managed<br>Health System<br>Kiosk<br>(K093389) | Xperex Inc.<br>Health Check<br>Kiosk<br>(K063008) | Comparison |
| | Corp's Health<br>Check Kiosk. | | | | |
| Electromagne<br>tic<br>Compatibility | Meets IEC<br>60601-1-2:2001 | None listed in<br>510(k) Summary | Meets IEC<br>60601-1-2 | None listed in<br>510(k) Summary | The SoloHealth<br>Station is tested to<br>meet<br>electromagnetic<br>compatibility<br>requirements. |
| Components | Blood Pressure<br>and Pulse Rate;<br>Weight; Vision | Blood Pressure<br>and Pulse Rate | Blood Pressure<br>and Pulse Rate;<br>Weight | Blood Pressure<br>and Pulse Rate | The SoloHealth<br>Station has<br>additional<br>functionality of<br>vision screening via<br>standard eye chart,<br>which is not a<br>medical device. |
| User<br>Interaction | Interactive<br>screens and<br>voice<br>instructions | Interactive<br>screens and<br>voice<br>instructions | Push buttons | Interactive<br>screens and<br>voice<br>instructions | Identical to<br>LifeClinic 2400 |
# Table 1. Detailed comparison of the subject and predicate devices.
{8}------------------------------------------------
.
.
.
ﺘ
{9}------------------------------------------------
:
{10}------------------------------------------------
:
.
.
,
{11}------------------------------------------------
SoloHealth Station
Table 2. Similar Vision Screening Devices
| | Subject Device | | Similar Vision Screening Devices | | | |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SoloHealth Station | Eye Chart | Vimetrics Central Vision<br>Analyzer 1000 (CVA-1000) | Optec 5000 Series<br>Vision Tester | Detailed Comparison | |
| 510(k) | Subject Device K113402 | None | K100095 | None located | Comparison not applicable. | |
| Manufacturer<br>Establishment<br>Registration | SoloHealth | Accutome | Vimetrics, LLC | Stereo Optical | | |
| | 3007156596 | 2521877 | None found in FDA database | 1419226 | | |
| Primary Device<br>Listing | DXN, Automated Blood<br>Pressure Monitor<br>HOX, Chart, Visual<br>Acuity | HOX, Chart,<br>Visual Acuity | HOX, Chart, Visual Acuity | HOX, Chart, Visual<br>Acuity; HIT, Tester,<br>Vision Color | The SoloHealth Station<br>central vision screening is<br>similar to the visual acuity<br>chart of the Eye chart,<br>Central Vision Analyzer and<br>Optec 5000. | |
| Classification | Class II | Class I Exempt | Class I, requiring premarket<br>notification | Class I Exempt | The SoloHealth Station is<br>Class II and is subject to<br>510(k) clearance. The vision<br>screening component of the<br>SoloHealth Station is similar<br>to the Eye Chart, Central<br>Vision Analyzer, and Optec<br>5000 that are Class I. | |
| Intended Use | General public to<br>measure blood pressure<br>and pulse rate; Weight<br>measurement and<br>screen clarity of central<br>vision | To test visual<br>acuity | The CVA-1000 is intended for<br>use under the direct<br>supervision<br>of an ophthalmologist or<br>optometrist in the<br>measurement of vision at<br>fixation in one<br>or both eyes, with or without<br>optical correction. | To test visual<br>acuity | The SoloHealth Station is<br>similar to all similar devices<br>and is intended to screen for<br>clarity of central vision. | |
| Indications | The SoloHealth Station | No labeling | The OVA-1000 is intended | No labeling | The indication for screening | |
| | Intended Population | Subject Device | Similar Vision Screening Devices | | | |
| | | SoloHealth Station | Eye Chart | Vimetrics Central Vision Analyzer 1000 (CVA-1000) | Optec 5000 Series Vision Tester | Detailed Comparison |
| | General Public | is intended to be used by the general public so that a user can measure his/her own blood pressure and pulse rate and his/her own weight. Additionally, the SoloHealth Station is intended to screen adults for clarity of central vision.<br>SoloHealth Station does not provide a general screening of visual function and does not provide a diagnosis of eye health or other disease. The SoloHealth Station only screens clarity of central vision. Users should consult their personal physicians if they have concerns regarding their eyesight. | available | for use under the direct supervision of an ophthalmologist or optometrist in the measurement of vision at fixation in one or both eyes, with or without optical correction. | available | for clarity of central vision is the same intent and is similar to the indications statement by Central Vision Analyzer. The SoloHealth Station does measure vision at fixation in one or both eyes with or without optical correction in central vision as does the predicate Central Vision Analyzer. |
| | General public | | | | General Public | Similar to the Optec 5000, the SoloHealth Station is intended for the general public use. The Optec 5000 is used by department of transportation to test visual |
| | | | | | General Public with supervision of ophthalmologist or optometrist | |
| Subject Device | Similar Vision Screening Devices | | | | | |
| | SoloHealth Station | Eye Chart | Vimetrics Central Vision<br>Analyzer 1000 (CVA-1000) | Optec 5000 Series<br>Vision Tester | Detailed Comparison | |
| Near Distance<br>Screening | Yes; Screening<br>simulated at 17 inches | Yes | Yes | Yes; Screening<br>simulated 16 in. | All devices are used for near<br>distance screening. |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.