BIODENTA DENTAL IMPLANT SYSTEM- BONE LEVEL D 3.0 TO 6.0 MM
K123512 · Biodenta Swiss AG · DZE · Mar 28, 2013 · Dental
Device Facts
Record ID
K123512
Device Name
BIODENTA DENTAL IMPLANT SYSTEM- BONE LEVEL D 3.0 TO 6.0 MM
Applicant
Biodenta Swiss AG
Product Code
DZE · Dental
Decision Date
Mar 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Biodenta dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures.
Device Story
System of endosseous root-form dental implants; includes implants, prosthetic parts, and surgical instruments. Implants surgically placed in mandibles or maxillae to support prosthetic devices (artificial teeth) for partially or fully edentulous patients. Device provides foundation for single/multiple tooth restorations, bridgework, or overdentures. Operates via internal hexagon connection (or narrower version for 3.0 mm diameter). Clinician performs surgical placement. Mechanical stability supports prosthetic function. Benefits include restoration of chewing function and esthetics following natural tooth loss.
Clinical Evidence
No clinical data. Safety and effectiveness supported by non-clinical bench testing, including fatigue testing per FDA guidance for root-form endosseous dental implants and comparative implant surface area analysis.
Indicated for patients requiring single or multiple tooth restorations, fixed or removable bridgework, or overdenture retention in the mandible or maxilla.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Biodenta Dental Implant System - Bone Level (K111003)
{0}------------------------------------------------
Premarket Notification / 510(k) Submission Biodenta Dental Implant System – Bone Level D 3.0 to 6.0 mm 5 - 510(k) Summary
ﺴ ﺴ ﺴ
・
.
Image /page/0/Picture/1 description: The image shows the logo for "biodenta". The logo consists of the word "biodenta" in a bold, sans-serif font, with a stylized graphic above the word. The graphic appears to be three curved lines stacked on top of each other, resembling a leaf or a wave. There is a small registered trademark symbol to the right of the word "biodenta".
K123512
## MAR 2 8 2013
# 5
## 510(k) Summary
| Owner's name | Biodenta Swiss AG |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Address: | Tramstrasse 16<br>9442 Berneck<br>Switzerland |
| Phone: | +41 71 747 11 11 |
| Fax number: | + 41 71 747 11 12 |
| Contact person: | Mr. David Eiler, Regulatory Manager |
| Date summary prepared: | February 18, 2013 |
| Trade / proprietary name: | Biodenta Dental Implant System - Bone Level D 3.0 to 6.0<br>mm |
| Common name: | Endosseous dental implant |
| Device classification name: | implant, endosseous, root-form |
| Product code: | DZE |
| Regulation number: | 21 CFR 872.3640 |
| Subsequent Product Code: | NHA (Abutment, Implant, Dental, Endosseous;<br>Regulation #: 21 CFR 872.3630) |
| Legally marketed device to which equivalence is claimed (predicate device): | |
| 1. Company: | Biodenta Swiss AG |
| Device name: | Biodenta Dental Implant System - Bone Level |
| 510(k) number: | K111003 |
| 2. Company: | Nobel Biocare AB |
| Device name: | Nobel Active 3.0 |
| 510(k) number: | K102436 |
2.0
10f4 Revision
Section Section page
{1}------------------------------------------------
#### Premarket Notification / 510(k) Submission Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm 5 - 510(k) Summary
Image /page/1/Picture/1 description: The image shows the word "biodenta" in a bold, sans-serif font. To the right of the word is a small, superscripted "R" in a circle, indicating a registered trademark. Above the word is a stylized graphic that resembles a crescent shape with textured shading, possibly representing a leaf or a dental form.
| 3. Company: | MIS - Implant Technologies Ltd. | Company | Device Name | 10(k) Number | Intended use | Implant Type | Implant to Abut. Connection | Implant Diameter (Length) | Abutment Angle | Implant Material | Abutment Material | Surface Treatment | Sterilization |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|---------------------------------|-------------------------------------------------------------|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|-----------------------------|-----------------------------------------------------------------------------------------------------|-------------------|----------------------|----------------------|-----------------------------------------|-------------------------------------|
| Device name:<br>ﻤﻴﺮﺍ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ | Seven Implants; Biocom Implants; Lance Implants | Biodenta Swiss AG | Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm | New device | Biodenta dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures. | Root-form endosseous dental implant Bone Level Implant | Internal Hexagon | D 3.0 mm (L 10 - 14 mm) D 4.1 mm (L 6.5 mm) D 4.8 mm (L. 6.5 mm) D 6.0 mm (L. 6.5-12 mm) | 0°, 15° | Titanium Grade 4 | Ti alloy (Ti 6AI 4V) | Spark Anodization | Delivered Sterile Gamma Irradiation |
| t · 510(k) number: | K103089 | Biodenta Swiss AG | Biodenta Dental Implant System - Bone Level | K111003 | Biodenta bone level. dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures. | Root-form endosseous dental implant Bone Level Implant | Internal Hexagon | D 3.5 mm (L 8 - 14 mm) D 4.1 mm (L 8 - 14 mm) D 4.8 mm (L. 8 - 14 mm) | 0°, 15° | Titanium Grade 4 | Ti alloy (Ti 6AI 4V) | Spark Anodization | Delivered Sterile Gamma Irradiation |
| 4. Company: | Keystone Dental | Nobel Biocare AB | Nobel Active 3.0 | K102436 | The NobelActive 3.0mm implant is indicated for the use in the treatment of missing maxillary lateral incisors of the mandibular central and lateral incisors to support prosthetic devices, such as artificial teeth, in order to restore chewing function in partially edentulous patents. The NobelActive 3.0 implants may be put into immediate function provided that stability requirements detailed in the manual are satisfied | Root-form endosseous dental implant Bone Level Implant | Internal Hexagon | D 3.0 mm (L 10 - 15 mm) | 0°, 15° | Titanium Grade 4 | Ti alloy (Ti 6AI 4V) | Spark Anodization | Delivered Sterile Gamma Irradiation |
| age and the country of the consider which while and<br>! Device name: | Genesis Implant System | MIS - Implant Technologies Ltd. | Seven Implants; Biocom Implants; Lance Implants | K103089 | MIS dental implants are intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function. When a one stage surgical procedure is applied, the implant may be immediately loaded when good primary stability is achieved and the occlusal load is appropriate. MIS short implants are to be used only with straight abutment | Root-form endosseous dental implant Bone Level Implant | Internal Hexagon | D 4.0 (L 6 mm) D 5.0 (L 6 mm) D 6.0 (L 6 mm) | 0°, 15°, 20°, 25° | Ti alloy (Ti 6AI 4V) | Ti alloy (Ti 6AI 4V) | Sand blasted and acid etched | Delivered Sterile Gamma Irradiation |
| 510(k) number: | K101545 | Keystone Dental | Genesis Implant System | K101545 | The Genesis Implant System is intended for use in single-stage or two-stage surgical procedures in all types of bone in partially or fully edentulous mandibles and maxillae. The Genesis Implant System supports single or multiple-unit restorations to re-establish patient chewing function and esthetics. Genesis implants are intended for placement following natural tooth loss or for immediate placement into an extraction socket. Immediate function may be achieved when good primary stability is established and appropriate occlusal loading is applied. | Root-form endosseous dental implant Bone Level Implant | Internal 6 lobe | D 3.8 mm (L 8.5 - 18 mm) D 4.5 mm (L 8.5 - 18 mm) D 5.5 mm (L 8.5 - 16 mm) D 6.5 mm (L 8.5 - 13 mm) | 0°, 15° | Ti alloy (Ti 6AI 4V) | Ti alloy (Ti 6AI 4V) | Double-acit-etching & Spark Anodization | Delivered Sterile Gamma Irradiation |
#### Indications for Use:
Biodenta dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures.
#### Device Description:
The Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, prosthetic parts and related surgical instruments. The Bone Level D 3.0 to 6.0 mm implants use the same platforms and abutment connections like the Bone Level implants (K111003) for the diameter 4.1. 4.8. and 6.0 mm implants, and therefore the abutments and prosthetic parts of the Bone Level implants are used. The diameter 3.0 mm implants use the same connection concept in a narrower version. The submission includes:
- Diameter 3.0 mm Implants with Length of: 10, 12, and 14 mm; Platform B0
- Diameter 4.1 and 4.8 mm Implants with Length of: 6.5 mm; Platform B2
- Diameter 6.0 mm Implants with Length of: 6.5, 8, 10, and 12 mm; Platform B2
- Straight abutments (Straight, Temporary, Ball, Locator); B0 platform, Length 10.4 14.9 mm, Diameter 3.6 - 4.0 mm
- Healing abutments, closure screws; B0 platform, Height 0.5 7.1 mm, Diameter 2.8 3.9bmm
#### Non-clinical Testing Data:
Fatique testing was conducted according to FDA Guide: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments -Guidance for Industry and FDA Staff. The worst case scenario for the Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm implants and abutments was tested. The results show that the Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm has sufficient mechanical strength for their intended clinical application.
To compare the implant surface area of the Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm with the predicate devices an implant surface area analysis has been carried out. All Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm implant's surface areas are almost identical or higher than the predicate device surface area. Therefore the surface area is considered to be sufficient.
#### Clinical Testing:
Non-clinical test data was used to support the decision of safety and effectiveness.
#### Equivalence to marketed device:
Biodenta Swiss AG demonstrated that, for the purposes of FDA's regulation of medical devices, the Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm is substantially equivalent to the predicate devices in intended use, material composition, fundamental scientific technology, principles of operation, and basic design.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for "biodenta". The logo consists of the word "biodenta" in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Above the word is a curved, stylized graphic element that appears to be made of three curved lines.
F
### Summary Substantial Equivalence Comparison to predicate devices:
Revision 2.0 3 of 4
{3}------------------------------------------------
Premarket Notification / 510(k) Submission Biodenta Dental Implant System – Bone Level D 3.0 to 6.0 mm 5 - 510(k) Summary
:
Image /page/3/Picture/1 description: The image shows the logo for "biodenta". The logo consists of the word "biodenta" in a sans-serif font, with a stylized graphic above it. The graphic appears to be three curved lines stacked on top of each other, creating a semi-circular shape. A registered trademark symbol is present to the right of the word "biodenta".
(Blank Page)
.
.
べん・・
Revision 2.0
Section page
:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 28, 2013
Mr. David Eiler Regulatory Manager Biodenta Swiss AG Tramstrasse 16 Berneck, St. Gallen Switzerland 9442
Re: K123512
Trade/Device Name: Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: February 19, 2013 Received: February 21, 2013
Dear Mr. Eiler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/5/Picture/6 description: The image shows the name "Kwame O. Ulmer" in a stylized font. The letters are bold and black, and the name is arranged in two lines. The letters are partially obscured by a geometric pattern that resembles a maze or a circuit board. To the right of the name is a triangular shape made of three concentric triangles.
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Premarket Notification / 510(k) Submission Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm 4 - Indications for Use Statement
Image /page/6/Picture/1 description: The image shows the logo for Biodenta. The logo consists of the word "biodenta" in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Above the word is a graphic of what appears to be a stylized leaf or feather design.
## Indications for Use
K 123512 510(k) Number (if known):
Device Name: Biodenta Dental Implant System - Bone Level D 3.0 to 6.0 mm
Indications for Use:
Biodenta dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
| Mary S. Runner -S |
|---------------------|
| Susan Runner, DOSMA |
| 2013:03.21 |
| 12:14:28 -04'00' |
Page _ 1_ of _ 1_
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
1 () Revision Section નું 2 of 4 Section page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.