Biodenta bone level dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures.
Device Story
Biodenta Dental Implant System - Bone Level consists of threaded, root-form endosseous dental implants, abutments, and surgical instruments. Designed for surgical placement in mandibles or maxillae by dental professionals to support prosthetic restorations (single/multiple teeth, bridges, overdentures) in partially or fully edentulous patients. Implants feature internal hexagon connections and are fabricated from titanium; abutments are titanium alloy. System provides mechanical foundation for prosthetic devices, restoring patient chewing function and esthetics. Clinical benefit derived from stable osseointegration and prosthetic support.
Clinical Evidence
Bench testing only. Fatigue testing conducted per FDA guidance and ISO 14801: Dentistry-Implants-Dynamic fatigue test for endosseous dental implants. Worst-case scenario testing demonstrated sufficient mechanical strength for intended clinical application.
Technological Characteristics
Threaded, root-form endosseous dental implants; internal hexagon connection; titanium implant material; titanium alloy abutment material; diameters 3.5mm, 4.1mm, 4.8mm; angled abutments (0°, 15°). Fatigue testing performed per ISO 14801.
Indications for Use
Indicated for partially or fully edentulous patients requiring surgical placement of endosseous dental implants in the mandible or maxilla to support single/multiple tooth restorations, fixed/removable bridgework, or overdentures.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "biodenta" in bold black font with a registered trademark symbol next to it. Above the word is a curved graphic. Below the word is the text "JC111003" in a handwritten style.
SEP 1 5 2011
510(k) Summary
Biodenta Swiss AG Owner's name: Tramstrasse 16 9442 Berneck Address: Switzerland Phone: +41 71 747 11 11 + 41 71 747 11 12 Fax number: Contact person: Mr. David Eiler, Regulatory Manager Date summary prepared: April 8, 2011 Trade / proprietary name: Biodenta Dental Implant System - Bone Level Endosseous dental implant Common name: Device classification name: implant, endosseous, root-form Product code: DZE Regulation number : 21 CFR 872.3640 NHA (Regulation number: 21 CFR 872.3630) Subsequent Product Code: Legally marketed device to which equivalence is claimed (predicate device): Company: Biodenta Swiss AG Biodenta Dental Implant System Device name: K093630 510(k) number: Nobel Biocare Ab Company: Device name: Nobelactive Internal Connection Implant K071370 510(k) number:
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "biodenta" in bold, black font, with a registered trademark symbol to the right of the word. Above the word is a curved, abstract design. Below the word "biodenta" is the handwritten number "K111003".
### Indications for Use:
Biodenta bone level dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or intermediate abutment support for fixed or removable bridgework and to retain overdentures.
# Device Description:
The Biodenta Dental Implant System - Bone Level is an integrated system of endosseous dental implants, which are designed to support prosthetic devices for partially or fully edentulous patients. The system consists of a variety of dental implants, abutments and prosthetic parts and related surgical instruments.
# Non-clinical Testing Data:
Fatigue testing was conducted according to FDA Guide: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments -Guidance for Industry and FDA Staff and ISO 14801 Dentistry-Implants-Dynamic fatigue test for endosseous dental implants. The worst case scenario for the Biodenta Dental Implant System - Bone Level implants and abutments was tested. These results show that Biodenta Dental Implant System - Bone Level have sufficient mechanical strength for their intended clinical application.
#### Equivalence to marketed device:
Biodenta Swiss AG demonstrated that, for the purposes of FDA's regulation of medical devices, the Biodenta Dental Implant System - Bone Level is substantially equivalent to the predicate devices in intended use, material composition, fundamental scientific technology, principles of operation, and basic design. When compared with the predicate devices, no new questions of safety or effectiveness have been raised for the Biodenta Dental Implant System - Bone Level.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "biodenta" in bold, black font, with a registered trademark symbol to the right of the word. Above the word is a logo that appears to be three curved lines stacked on top of each other. Below the word "biodenta" is the text "K111003" in a handwritten-style font.
# Substantial Equivalence Comparison to predicate device:
| | Subject Device | Predicate Devices | |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Biodenta Dental Implant<br>System - Bone Level | Biodenta Dental Implant<br>System (K093630) | Nobelactive Internal Connection<br>Implant (K071370) |
| Intended use | Biodenta bone level dental<br>implants are intended for<br>surgical placement in<br>mandibles or maxillae to<br>support single or multiple<br>tooth restorations or<br>terminal or intermediate<br>abutment support for fixed<br>or removable bridgework<br>and to retain overdentures. | Biodenta dental implants<br>are intended for<br>surgical placement in<br>mandibles or maxillae to<br>support single or multiple<br>tooth restorations or<br>terminal or intermediate<br>abutment support for fixed<br>or removable bridgework<br>and to retain overdentures.<br>Specific indications for small<br>diameter (Ø 3.5 mm)<br>implants: It is recommended<br>not to place small implants<br>with "NP" platform in the<br>molar or premolar region. | Nobel Biocare's NobelActive<br>implants are endosseous implant<br>intended to be surgically placed in<br>the bone of the upper or lower jaw<br>arches to provide support for<br>prosthetic devices, such as an<br>artificial tooth, in order to restore<br>patient esthetics and chewing<br>function. Nobel Biocare's<br>NobelActive implants are indicated<br>for single or multiple unit<br>restorations in splinted or non-<br>splinted applications. Nobel<br>Biocare's NobelActive implants may<br>be placed immediately and put into<br>immediate function provided that<br>initial stability requirements detailed<br>in the manual are satisfied. |
| System | threaded, root-form dental<br>implant, bone level | threaded, root-form dental<br>implant, tissue level | threaded, root-form dental<br>implant, bone level |
| Diameter | 3.5 mm, 4.1 mm, 4.8 mm | 3.5 mm, 4.1 mm, 4.8 mm | 3.5 mm, 4.3 mm, 5.0 mm |
| Connection | Internal Hexagon | Internal Octagon | Internal Hexagon |
| Implant<br>Material | Titanium | Titanium | Titanium |
| Abutment<br>System | Angled Abutment | Angled Abutment | Angled Abutment |
| Abutment<br>Angle | 0°, 15° | 0°, 15°, 20° | 0°, 15° |
| Abutmet<br>Material | Titanium Alloy | Titanium Alloy | Titanium Alloy |
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MID 20993-0002
Mr. David Eiler Regulatorv Manager Biodenta Swiss AG Tramstrasse 16 9442 Berneck Switzerland
SEP 15 2011
Re: K111003
Trade/Device Name: Biodenta Dental Implant System - Bone Level Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: September 9, 2011 Received: September 12, 2011
Dear Mr. Eiler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind vou, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{4}------------------------------------------------
Page 2 - Mr. Eiler
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for "biodenta". The logo consists of the word "biodenta" in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Above the word is a stylized graphic that appears to be a curved shape with lines inside.
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Biodenta Dental Implant System - Bone Level
Indications for Use:
Biodenta bone level dental implants are intended for surgical placement in mandibles or maxillae to support single or multiple tooth restorations or terminal or intermediate abutment support for fixed or removable bridgework and to retain overdentures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Rumm
(Division Sign-Ott) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111003
Page 1 of 1
Revision 1.1 Section な 2 01 4 Section page
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.