DISPOSABLE HIGH PRESSURE STOPCOCK/MANIFOLD MODEL HP STOPCOCK, MP STOPCOCK, HP MANIFOLD, PHP MANIFOLD
K123084 · Elcam Medical Acal · FMG · May 23, 2013 · General Hospital
Device Facts
Record ID
K123084
Device Name
DISPOSABLE HIGH PRESSURE STOPCOCK/MANIFOLD MODEL HP STOPCOCK, MP STOPCOCK, HP MANIFOLD, PHP MANIFOLD
Applicant
Elcam Medical Acal
Product Code
FMG · General Hospital
Decision Date
May 23, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
Elcam's High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are intended to serve as flow control and delivery devices for I.V fluids injection to the patient's vascular system. The devices are indicated for medium and high pressure injection of fluids such as, but not limited to, fluids used during angiography and angioplasty and during interventional radiology and cardiology procedures. Devices indication for medium and high pressure does not preclude its use for low pressure procedures. The High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are intended for single use only.
Device Story
Stopcocks and manifolds function as flow control and delivery interfaces for I.V. fluids; used in cardiac angiography, angioplasty, interventional radiology, and cardiology. Devices connect to I.V. sets to manage fluid administration; units can be used individually or ganged together as manifolds. Designed to withstand high-pressure injections (up to 1200 psi) compared to standard low-pressure stopcocks. Operated by clinicians in clinical settings. Provides a secure, pressure-resistant pathway for contrast media or other fluids; ensures integrity during high-pressure procedures; prevents leakage or failure under stress. Single-use, disposable design.
Clinical Evidence
Bench testing only. Performance validated via high-pressure testing per ISO 8536-10, biocompatibility per ISO 10993, and sterilization/packaging integrity per ISO 11135-1 and ISO 11607. No clinical data required for this device type.
Technological Characteristics
Materials: Polycarbonate LEXAN (body/rotator), ACETAL-ULTRAFORM (handle), Liquid Silicone Rubber (O-ring). Pressure ratings: 500-1200 psi depending on model. Standards: ISO 8536-10 (pressure), ISO 594-1/2 (connectors), ISO 10993 (biocompatibility), ISO 11135-1 (sterilization), ISO 11607-1/2 (packaging). Sterilization: EtO. Form factor: Stopcock and manifold configurations.
Indications for Use
Indicated for patients undergoing angiography, angioplasty, interventional radiology, or cardiology procedures requiring medium or high-pressure fluid injection into the vascular system. Suitable for use in low-pressure procedures as well. Single-use only.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
# K123084
Image /page/0/Picture/1 description: The image contains a logo for Elcam Medical. The logo consists of a stylized letter "E" on the left, followed by the text "Elcam" in a bold, sans-serif font. Below "Elcam" is the word "MEDICAL" in a smaller, sans-serif font. Underneath the logo is the text "Where everything connects" in a smaller font.
Dec. 25, 2012
## 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92
## Submitter Information
Elcam Medical A.C.A.L. Kibbuts BarAm, M.P. Merom HaGalil, 13860, Israel Tel: (972) 4 6988120/1/2, Fax: (972) 4 6980777
## Submission contact person:
Aharon Cohen Kibuz Bar Am 30889, Israel TEL: 972-4-6988324
## Device Classification
| Proprietary Device Name: | Elcam High and Medium Pressure Stopcocks and High and<br>Premium High Pressure Manifolds. |
|----------------------------|-------------------------------------------------------------------------------------------|
| Common name: | Stopcocks and Manifolds for I.V. |
| Product Code: | FMG |
| Classification Name: | Stopcock, I.V. Set |
| Classification Regulation: | 21 CFR § 880.5440 |
| Regulatory Class: | II |
## Identification of the legally Marketed Predicate Devices
Elcam's legally cleared Stopcock and Manifold devices (K11016) are indicated for flow control and delivery of I.V drugs and fluids used in I.V procedures. Devices are intended for normal pressures of gravity feed, sampling and bolus injection.
Scientific Device Manufacturer, LLC's cleared SDM Angiographic Manifold devices (K96031) are intended for I.V medium and high pressure fluids injection, used in angiography procedures.
Above legally Marketed Devices used as substantial equivalence to Elcam proposed High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds (modified version of the legally cleared Stopcock and Manifold devices - K111016)
## Device Description
Stopcocks and Manifolds are generally used for administration of fluids into the human. body. There is a wide use of Stopcocks as flow controls and sampling sites in I.V sets, critical care applications and monitoring kits. Stopcocks are used individually or in the form of Manifolds, units comprised of several Stopcocks joined together ("gang").
{1}------------------------------------------------
High pressure Stopcocks and Manifolds are used in cardiac angiography and angioplasty procedures in which fluids are administered under pressure. These procedures require stopcocks and manifolds which are robust under high pressures.
The line of pressure resistant Stopcocks and Manifolds presented in this file includes four (4) products:
- i. High Pressure (HP) Stopcocks
- ii. Medium Pressure (MP) Stopcocks
- High Pressure (HP) Manifolds iii.
- Premium High Pressure (PHP) Manifolds iv.
## Intended Use of Device
Elcam's High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are intended to serve as flow control and delivery devices for I.V fluids injection to the patient's vascular system. The devices are indicated for medium and high pressure injection of fluids such as, but not limited to, fluids used during angiography and angioplasty and during interventional radiology and cardiology procedures. Devices indication for medium and high pressure does not preclude its use for low pressure procedures. The High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are intended for single use only.
#### Safety & Effectiveness
The proposed and predicate devices are similar in design, materials of construction, components, intended use and labeling.
Based on the performance results provided (including test results and clinical data) and the analysis of similarities and differences presented above, Elcam Medical believes that the proposed device safe & effectiveness is substantially equivalent to the predicate device without raising new safety and/or effectiveness issues.
### Rational for Substantial Equivalency
Substantial equivalency between the proposed devices, to its predicate devices was demonstrated by comparison in a tabular way of the intended use, indications, effectiveness, safety, performance and basic technological characteristics.
| # | Comparison<br>parameter | Proposed device:<br>Elcam Disposable High<br>Pressure Stopcock/Manifold | Elcam Disposable Stopcock/<br>Manifold | Scientific Device<br>Manufacturer, LLC-<br>SDM Angiographic<br>Manifold |
|---|-------------------------|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| 1 | K No. | K123084 | K111016 | K960431 |
| 2 | Owner | Elcam Medical A.C.A.L. | Elcam Medical A.C.A.L. | Scientific Device<br>Manufacturer, LLC |
| 3 | components<br>materials | The device is composed of the<br>following materials:<br>Body: Polycarbonate LEXAN<br>Handle: ACETAL- | The product is made of the<br>following materials:<br>Body: Former Lexan<br>Handle: High Density | The product is made of the<br>following materials:<br>- Handle - Acetal |
#### SE discussion:
{2}------------------------------------------------
| # | Comparison<br>parameter | Proposed device:<br>Elcam Disposable High<br>Pressure Stopcock/Manifold | Elcam Disposable Stopcock/<br>Manifold | Scientific Device<br>Manufacturer, LLC-<br>SDM Angiographic<br>Manifold |
|---|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | ULTRAFORM<br>High Pressure Rotator:<br>Polycarbonate LEXAN<br>O-Ring: Liquid Silicone<br>Rubber (Apple Rubber) | Polyethylene | O-Ring- Ethylene<br>Propylene Terpolymer<br>Rubber (EPDM)<br>Body-Polycarbonate<br>Handle plugs -<br>Polyethylene |
| 2 | Pressure<br>Rating | HP Stopcocks - 1200 psi<br>(82 Bar)<br>MP Stopcocks - 500 psi )35<br>Bar)<br>HP Manifold - 600 psi (41.3<br>bar)<br>PHP Manifolds - 800 psi<br>(54.6 bar) | Pressure is up to 3 bar (44 psi) | Pressure Ranges 500 -1200psi |
| 3 | Performance<br>parameters | Designed to withstand high<br>pressure injections | Not designed to withstand<br>high pressure injections | Designed to withstand high<br>pressure injections |
| 4 | Integrity of<br>materials and<br>functionality<br>after EtO<br>sterilization | The materials and the product<br>functionality are not affected<br>by EtO sterilization process or<br>as a result of aging over rated<br>life time specification. | The materials and the product<br>functionality are not affected<br>by EtO sterilization process or<br>as a result of aging over rated<br>life time specification. | The materials and the product<br>functionality are not affected<br>by EtO sterilization process or<br>as a result of aging over rated<br>life time specification. |
| 5 | Biocompatibil<br>ity standards | Tests performed according to<br>ISO 10993 | Test performed according to<br>ISO 10993 | Test performed according to<br>ISO 10993 |
| 6 | Performance<br>Standards | Tests performed according to:<br>ISO 8536-10<br>ISO 594-1<br>ISO 594-2 | Tests performed according to:<br>ISO 594-1<br>ISO 594-2 | Unknown |
| 7 | Sterilization<br>and sterile<br>package<br>integrity | Tests performed according to:<br>ISO 11135-1<br>ISO 11607-1 and ISO 11607-2 | Tests performed according to:<br>ISO 11135-1<br>ISO 11607-1 and ISO 11607-2 | Unknown |
The above table compares similarities and differences between Elcam's High Pressure Stopcock/Manifold to legally cleared Elcam Stopcock and Manifold devices (K111016) and to Scientific Device Manufacturer, LLC-SDM Angiographic Manifold (K960431). The following summarizes the main tests performed to demonstrate substantial equivalence resulting from the use of different materials required for sustaining high pressure injection:
- High pressure performance tests for conformity to ISO 8536-10 standard .
- Biocompatibility .
- Integrity of materials and functionality after EtO sterilization .
- Sterilization and sterile package integrity tests for conformity to ISO 11135-1, ISO .
- l 1607-1 and ISO 11607-2
#### Conclusion
The mentioned nonclinical tests and clinical use experience of Elcam's predicate legally marketed Disposable Stopcock/ Manifold demonstrate that the Disposable High Pressure Stopcock/Manifold is as safe, as effective, and performs as well as or better than the legally marketed devices identified in the K123084 submission.
{3}------------------------------------------------
#### Substantial Equivalence Statement
Based on the above, it is Elcam Medical's opinion that the proposed Elcam High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are substantially equivalent in terms design principles, performance features and of safety & effectiveness to the legally cleared predicate devices : Elcam's Disposable Stopcocks/ Manifolds (K111016) and to Scientific Device Manufacturer, LLC's- SDM Angiographic Manifold (K960431), referred to in chapter 4 of this 510(K) submission executive summary document.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### May 23, 2013
Mr. Aharon Cohen Regulatory Affairs Manager Elcam Medical A.C.A.L. Kibbutz Bar-Am D.N. Merom Hagalil Israel 13860
Re: K123084
Trade/Device Name: Elcam High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FMG Dated: April 3, 2013 Received: April 25, 2013
Dear Mr. Cohen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Cohen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runner, DOSIA
Digitally signed by Mary S. Runner -S
ON: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Mary S. Runner
0.9.2342.19200300.100.1.1=1300087950
Date: 2013.05.23 14:13:45 -04'00'
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K.........................
Device Name: Elcam High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds
#### Indications for Use:
Elcam's High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are intended to serve as flow control and delivery devices for I.V fluids injection to the patient's vascular system. The devices are indicated for medium and high pressure injection of fluids such as, but not limited to, fluids used during angiography and angioplasty and during interventional radiology and cardiology procedures. Devices indication for medium and high pressure does not preclude its use for low pressure procedures. The High and Medium Pressure Stopcocks and High and Premium High Pressure Manifolds are intended for single use only.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/6/Picture/10 description: The image shows the FDA logo. The logo is made up of the letters FDA in a stylized font. The letters are outlined in black and the inside of the letters are white. The logo is simple and recognizable.
Richard C. Chapman 2013.05.23 13:56:27 -04'00'
(Division Sign-Off) (Division Sign-On)
Division of Anesthesiology, General Hospital Division Control, Dental Devices
Page 1 of 1
**510(k) Number:** K12084
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.