K111016 · Elcam Medical Acal · FPA · Jun 9, 2011 · General Hospital
Device Facts
Record ID
K111016
Device Name
SAFEPORT MANIFOLD (TM) (OR STOPCOCK)
Applicant
Elcam Medical Acal
Product Code
FPA · General Hospital
Decision Date
Jun 9, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Safeport Manifold is indicated to serve as a flow control and a conduit device for I.V fluids delivery to the patient's vascular system. The product is intended for delivering 1.V drugs or fluids, allowing gravity feed, sampling, bolus injection and elimination of reflux of fluids during operation
Device Story
SafePort Manifold is a flow control/conduit device for I.V. fluid delivery; used in clinical settings for drug/fluid administration, gravity feed, sampling, and bolus injection. Device consists of a one-piece injected body with side female ports, handles, elastomers, and optional rotors/check valves. Available in standard or swabable valve configurations; swabable valves function as closed luer-activated valves to prevent leakage/contamination and eliminate need for caps. Operated by clinicians to manage fluid flow and vascular access. Benefits include improved convenience, robustness, and reduced risk of contamination via needleless injection sites.
Clinical Evidence
No clinical trials performed. Substantial equivalence supported by non-clinical bench testing.
Technological Characteristics
One-piece injected body; side female ports; handles; elastomers; optional rotors and check valves. Available with or without swabable valves (closed luer-activated). Materials include unspecified polymers and elastomers. Sterilization method and shelf life are identical to predicate.
Indications for Use
Indicated for patients requiring I.V. fluid or drug delivery, gravity feed, sampling, or bolus injection via a vascular access system.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for Elcam Medical. The logo is black and white and features the words "Elcam Medical" stacked on top of each other. To the right of the text is a stylized image that appears to be a medical device or component. The logo is simple and modern, and it is likely used to represent the company's brand.
Special 510(k) for SafePort Manifold™ (or Stopcock)
Applicant: Elcam Medical ACAL
KIIIOIb
## 510(k) SUMMARY [As Required by 21 CFR 807.92(c)]
JUN - 9 2011
Date Prepared: April 3, 2011
Applicant Name: Elcam Medical ACAL
Kibbutz BarAm, Merom Ha Galil 13860, Israel
Contact Person Name: Ms. Natasha Evronyan – R.A. and Q.A. Engineer
Telephone: 972-4-6988821
Fax: 972-4-6980777
E-mail: Natasha.ev@elcam.co.il
Elcam Medical's U.S. Agent: Mr. Lloyd Fishman – President
Elcam Medical, Inc.
2 University Plaza, Suite 620,
Hackensack, NJ 07601, USA
Phone: 201- 457- 1120
Fax: 201- 457- 1125
E-mail: Lloyd.fishman@elcam-medical.com
#### Predicate Device:
Elcam's Induction & Sampling Manifold (or Stopcock) cleared under 510(k) # K032393.
Device Name
Common/ Usual Name: SafePort Manifold™ (or Stopcock)
Proprietary/ Trade Name: SafePort Manifold™ (or Stopcock)
April 3, 2011
- Confidential-
nave 43 of 45
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for Elcam Medical. The logo consists of a black square with the words "Elcam" and "Medical" stacked on top of each other in white font. The text is positioned in the lower-left corner of the square. The logo is simple and modern, with a focus on the company name.
Classification: Elcam's SafePort Manifold has been classified as Class II device under the following classification names:
| Classification Name | Product Code | 21 CFR Ref. | Panel |
|------------------------------------------|--------------|-------------|-------------------|
| Intravascular Administration<br>I.V. Set | FMG | 880.5440 | General Hospitals |
## Device Description:
Modified SafePort Manifold (or Stopcock) is a redesign of Induction & Sampling Manifold (or Stopcock). It has the same appearance and intended use while several design modifications were introduced in order to contribute to the device's convenience of use and robustness. The device will be available in two (2) versions swabable or standard ports configurations. The first version, standard port configuration is similar to the predicate device but with more robust design. The second version, besides of the design changes have an additional feature. The feature is the swabable valves which can be incorporated into the female side ports and function as a closed luer activate valve. The swabable valve enables the female port (luer) to be closed when it is not in use and saves the need to close it with a cap in order to avoid leakage and/ or port's contamination and serves as a needleless injection site integrated in the stopcock. Devices having the swabable valves have been demonstrated in a 510(k) submission for Elcam's Closed Stopcock.
SafePort Manifold or Stopcock is composed of the following components:
- -One piece injected body having a side female ports. (with or without swabable valves according to the end user request)
- Handles assembled into the female side ports. ー
- -Elastomer placed between each handle and side female port body, which is placed to function as a pressure activated valve activated valve.
- Rotor assembled to the male port (connected toward the patient ) for . " connection locking.
- -Check valve assembled to the male port which is connected toward the patient in order to enable connection locking. (this part will not be included, unless otherwise requested by the end user)
- Port covers and colored buttons. (if requested by an end user) -
April 3, 2011
#### - Confidential-
{2}------------------------------------------------
新闻 - 文章
Image /page/2/Picture/1 description: The image shows a logo for Elcam Medical. The logo is a black square with the words "Elcam Medical" in white text. The word "Elcam" is on top of the word "Medical". The logo is simple and modern.
## Indication for Use:
New modified SafePort Manifold or Stopcock is a one or multiple ports product, which is indicated to serve as a flow control and a conduit device for I.V fluid delivery to the patient's vascular system. The product is intended for delivering of I.V. drugs or fluids, allowing gravity feed, sampling bolus injection and elimination or reflux of fluid during operation.
#### Technological Characterizes and Substantial Equivalence:
Elcam's modified SafePort Manifold (or Stopcock )without luer activated/ swabable valves has the same indication for use, principle of operation, shape, sterilization method and shelf life as its main predicate Induction and Sampling Manifold (or Stopcock).
The modified SafePort Manifold with luer activated/ swabable valves, a new feature cleared in the 510 (k) submissions for Closed Stopcock (MRVLS) is also the same as the predicate from the aspects mentioned above except for principles of operation requiring the specific treatment for the valves. The same principles have been introduced in the 510(k) submission for Closed Stopcocks (MRVLS).
There were no questions regarding the new product safety and effectiveness that were raised due to the non- identical technological characteristics.
#### Performance Data:
Testing relating functionality of the SafePort Manifold's different variation has been conducted in a series of non-clinical tests presented in Section 7 of the submission. It was shown that the modified device is as safe and as effective as its predicates.
#### Clinical Data:
No clinical trials have been preformed for the modified SafePort Manifold.
#### Conclusion:
lt was demonstrated that the modified SafePort Manifold is substantially equivalent to its predicate, Induction & Sampling Manifold (or Stopcock), and do not raise any questions of safety and effectiveness.
- Confidential-
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of an abstract symbol that resembles a person embracing an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Natasha Evronyan O.A and R.A. Engineer Elcam Medical A.C.A.L. Kibbutz Bar-Am MP Merom HaGalil, Israel 13860
JUN - 9 2011
Re: K111016
Trade/Device Name: Safeport ManifoldTM Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: May 24, 2011 Received: May 31, 2011
Dear Ms. Evronyan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Ms. Evronyan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Elcam Medical. The logo consists of the letters 'lc' stacked vertically, with the word 'Elcam' above the word 'Medical' to the left of the letters. A solid black circle is positioned above the 'c' in 'lc'.
# Indication for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Safeport Manifold
Indications for Use: Safeport Manifold is indicated to serve as a flow control and a conduit device for I.V fluids delivery to the patient's vascular system. The product is intended for delivering 1.V drugs or fluids, allowing gravity feed, sampling, bolus injection and elimination of reflux of fluids during operation
Prescription Use _ X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
www.elcam-medical.com
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nuhul the
Acting for
Russell Chapman
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K111016
Elcam Medical
Tel: 972-4-698-8120/1/2, Fax: 972-4-698-0777 sales@elcam.co.il
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.