FUJINON/FUJIFILM EB-530 SERIES BRONCHOSCOPES (EB-530S, EB-530T AND EB-530H
K122535 · Fujifilm Medical System U.S.A., Inc. · EOQ · Dec 13, 2012 · Ear, Nose, Throat
Device Facts
Record ID
K122535
Device Name
FUJINON/FUJIFILM EB-530 SERIES BRONCHOSCOPES (EB-530S, EB-530T AND EB-530H
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Dec 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Fujinon/Fujifilm EB-530 Series Bronchoscopes are intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
Device Story
Fujinon/Fujifilm EB-530 Series Bronchoscopes are flexible endoscopes used for visualization and treatment of the trachea and bronchial tree. The system includes three models: EB-530S (Standard), EB-530T (Treatment, with 2.8mm forceps channel), and EB-530H (High resolution). Devices connect to a video processor, light source, monitor, and peripheral equipment. Operated by physicians in clinical settings, the bronchoscope transmits visual data to the processor for display on a monitor, enabling real-time observation and guidance for endoscopic instruments. The device facilitates diagnostic and therapeutic interventions, potentially improving patient outcomes by allowing minimally invasive access to the airway.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1, IEC 60601-1-1, IEC 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-1).
Technological Characteristics
Flexible bronchoscope with fiber-optic/video imaging. Models: EB-530S, EB-530T, EB-530H. Distal end diameters: 4.9mm to 5.8mm. Field of view: 120-140 degrees. Bending: 180 degrees up, 130 degrees down. Forceps channel: 2.0mm or 2.8mm. Connectivity: Video processor, light source, monitor. Sterilization: Reprocessing instructions validated for reusable components; pre-sterilized suction button (SB-602) used.
Indications for Use
Indicated for observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree in patients requiring bronchoscopic procedures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
# FUJIFILM
FUJIFILM Medical Systems U.S.A., Inc. 10 High Paint Drive Vayne: NJ 07470 Tel. (973) 533-5600
## 510(k) Summary
DEC 1 3 2012
,
Date: November 1, 2012
K Number: K122535
### Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293
### Contact Person:
Name: Gina Walljasper Title: Director, Quality and Regulatory Compliance Telephone: (973) 633-5600 Ext. 484 Facsimile: (973) 633-8818 E-Mail: gwalljasper@fujifilm.com
### Identification of the Proposed Device:
Fuiinon/Fuiifilm EB-530 Series Bronchoscopes Proprietary/Trade Name: Bronchoscopes Common Name: Device Class: Class 2 Review Panel: Ear Nose & Throat Classification Information:
| AND REALERS LE & LEASE & LEA BER AND<br>Company Court Call College Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Com<br><br>paties B . F<br>Classification Name | Section | <br>Product<br>Codes<br>Canadian Career Station Company of Children Company of Children Company of Children |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Separation Street Art A.R. P.A. P.O. Box C.C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C.<br>NEW YORK OF CANDRIGURANCE OF ST<br>Bronchoscope (flexible or rigid) and accessories | ילס ח"מת -<br>4 4680<br>Call of the country of the first and the comments of the comments of the first of the first of the comments of the comments of the comments of the comments of the comments of<br>and the first from the became of the many would be | and any and |
#### l. INDICATIONS FOR USE
The Fujinon/Fujifilm EB-530 Series Bronchoscopes are intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
{1}------------------------------------------------
K122535'
# FUJKFILM
FUJIFILM Medical Systems U.S.A., Inc. 10 High Pors Ony ame NJ 07470 8 1973 6325800
#### DEVICE DESCRIPTION II.
Fulinon/Fuilfilm EB-530 Series Bronchoscopes are intended for the observation, diagnosis and endoscopic treatment of the trachea and bronchial tree. The Fujinon/Fujifilm EB-530 Series Bronchoscopes are used in combination with the Fujinon/Fujifilm's video endoscope processors, light source, monitor, cart, foot switch, endoscopic accessories, other peripheral devices (e.g. printer, keyboard, etc.), and the legally marketed electrosurgical instruments.
There are three models included in this submission:
- EB-530S Standard Type •
- EB-530T Treatment Type (includes 2.8mm forceps channel) .
- EB-530H Super Image Type (High resolution) .
#### lll. SUMMARY OF STUDIES
The subject device has been subjected to and passed electrical safety and EMC test requirements.
Fujinon/Fujifilm EB-530 Series Bronchoscopes were evaluated in accordance with the following safety and performance requirements in addition the applicable quality requlations:
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for<br>safety |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC60601-1-1 | Medical electrical equipment - Part 1-1: General requirements<br>for<br>safety - Collateral standard: Safety requirements for medical<br>electrical systems |
| IEC60601-1-2 | Medical electrical equipment - Part 1-2: General Requirements<br>for Safety - Collateral standard: Electromagnetic Compatibility -<br>Requirements and tests |
| IEC60601-2-18 | Medical electrical equipment - Part 2-18: Particular<br>requirements for the safety of endoscopic equipment |
| ISO10993-1 | Biological evaluation of medical devices |
New pre-sterilized suction button SB-602 conforms to the applicable internal and international ISO testing requirements for Sterility.
The reprocessing instructions were updated and validated. No clinical testing was conducted.
{2}------------------------------------------------
# FUJIFILM
FUJIFILM Medical Systems U.S.A., Inc. 0 High Point Drive Wavne, NJ 07470 1. (973) 833-5600
#### SUBSTANTIAL EQUIVALENCE IV.
Fuilnon/Fujifilm EB-530 Series Bronchoscopes are substantially equivalent to the following device:
| Legally Marketed Device(s) | 510(k) # |
|----------------------------|----------|
| Fujinon G5 Bronchoscopes | K050907 |
Comparison is outlined in the table below.
| | Legally marketed<br>Fujinon G5<br>Bronchoscope<br>EB-470S (K050907) | Proposed Device<br>Model EB-530S | Proposed Device<br>Model EB-530T | Proposed Device<br>Model EB-530H |
|--------------------------------|---------------------------------------------------------------------|----------------------------------|----------------------------------|----------------------------------|
| Viewing<br>direction | Forward/0degree | Forward/0degree | Forward/0degree | Forward/0degree |
| Observation<br>range | 3-100mm | 3-100mm | 3-100mm | 3-100mm |
| Field of view | 120 degrees | 120 degrees | 120 degrees | 140 degrees |
| Distal end<br>diameter | 4.9mm | 4.9mm | 5.8mm | 5.4mm |
| Flexible portion<br>diameter | 4.9mm | 4.9mm | 5.9mm | 4.9mm |
| Bending | Up: 180 degrees | Up: 180 degrees | Up: 180 degrees | Up: 180 degrees |
| capability | Down: 130 degrees | Down: 130<br>degrees | Down: 130<br>degrees | Down: 130<br>degrees |
| Forceps<br>channel<br>diameter | 2.0mm | 2.0mm | 2.8mm | 2.0mm |
| Working length | 600 mm | 600 mm | 600 mm | 600 mm |
| Total length | 870mm | 870mm | 870mm | 870mm |
All patient contact materials used in the proposed EB-530 Series Bronchoscopes remain the same as the legally marketed device.
#### V. CONCLUSION
Fujinon/Fujifilm EB-530 Series Bronchoscopes are substantially equivalent to the legally marketed device and conform to applicable medical device safety and performance: standards.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of three overlapping human figures.
December 13, 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
FUJIFILM Medical Systems U.S.A., Inc. % Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW. Buffalo, MN 55313
Re: K122535
Trade/Device Name: Fujinon/Fujifilm EB-530 Series Bronchoscopes (EB-530S, EB-530T and EB-530H) Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOQ Dated: November 1, 2012 Received: November 2, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Eric A. Mann
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K122535
## Indications for Use
## 510(k) Number (if known): K122535
Device Name: Fujinon/Fujifilm EB-530 Series Bronchoscopes (EB-530S, EB-530T AND EB-230H)
Indications for Use: This device is intended for the observation, diagnosis, and endoscopic treatment of the frachea and bronchial tree.
X Prescription Use _ . (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number __
**Prescription Use**
(Per 21 CFR 801.109)
K122535; Fujinon/Fujifilm EB-530 Series Bronchoscopes
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.