FUJINON G5 BRONCHOSCOPES, MODELS EB-470S, EB-250S, AND EB-270P
K050907 · Fujinon, Inc. · EOQ · May 19, 2005 · Ear, Nose, Throat
Device Facts
Record ID
K050907
Device Name
FUJINON G5 BRONCHOSCOPES, MODELS EB-470S, EB-250S, AND EB-270P
Applicant
Fujinon, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
May 19, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
Device Story
Fujinon G5 Bronchoscopes (EB-470S, EB-250S, EB-270P) are flexible endoscopes used by physicians for visualization and treatment of the trachea and bronchial tree. The device captures optical images of the airway, transmitting them to a compatible processor and monitor for real-time clinician viewing. The G5 series features ergonomic improvements, including a repositioned jet water wash port (J-Port), removal of the lens wash port (L-Port), and modified forceps inlet ports with rubber caps to enhance suction and reduce leakage. Internal check valves were replaced with external, removable valves to facilitate cleaning and prevent clogging. The device is used in clinical settings alongside light sources, processors, and electrosurgical instruments. By providing clear visualization and access to the airway, the device assists clinicians in diagnosing and treating pulmonary conditions, potentially improving patient outcomes through minimally invasive intervention.
Clinical Evidence
Bench testing only. The device underwent electrical safety, thermal, and EMC testing. No clinical data was required or provided as the device is substantially equivalent to the predicate with only minor design modifications.
Technological Characteristics
Flexible endoscope with forward viewing direction. Distal end diameters range from 3.8mm to 5.3mm; working length 600mm. Features include 120-degree field of view, 180-degree up/130-degree down bending. Materials are identical to the predicate. Connectivity is via standard Fujinon processor/peripherals. Reprocessing facilitated by external, removable check valves and CA-500 cleaning adaptor. Non-software based device.
Indications for Use
Indicated for observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree in patients requiring bronchoscopy.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
510(k) Notification Fujinon G5 Bronchoscopes MAY 1 9 2005
õ50907
# 510 (k) Summary
# Date Prepared [21 CFR 807.92(a)(1)]
March 31, 2005
## Submitter's Information [21 CFR 807.92(a)(1)]
Joseph M. Azary C/o Fujinon Inc. 543 Long Hill Avenue Shelton, CT. 06484
Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor Fujinon Inc. 10 High Point Drive, Wayne, NJ 07470, Establishment Registration# 2431293.
## Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
Trade Names: Fujinon G5 Bronchoscopes Models: EB-470S, EB-250S, and EB-270P
Common Name: Bronchoscope Classification: Class II, 21 CFR 874.4680, EOQ
# Predicate Device [21 CFR 807.92(a)(3)]
- Fujinon EB-310S (K954707) .
The G5 Technology has received FDA marketing clearance in the following 510(k) Premarket Notifications:
- K042076 Fujinon G5 Duodenoscopes .
- K042043 Fujinon G5 Gastroscopes .
- K041903 Fujinon G5 Colonoscopes .
The subject device have the same indications for use, material composition, viewing direction, image size, bending, reprocessing/sterilization method, and working dimensions as the predicate. The subject device uses the same processor and peripherals as the predicate device.
The main differences between the subject device and predicate device are as follows:
- Minor differences with observation range, field of view, diameter, and length. .
- The subject device includes the G5 upgrade, which is characterized by the following minor differ-. ences:
- The L-Port has been eliminated. The L-port functioned as a lens wash port . Doctors had o the option to take a syringe to inject a fluid to use it as a high pressure wash for the lens.
{1}------------------------------------------------
This function was eliminated because demand was low and it was rarely used by the surgeons.
- o The J-Port was repositioned. The J-Port is used as a jet water wash port. The J-Port was repositioned based on doctor feedback. The port was moved from the bottom part to the top (end) of the scope. There was also a desire to eliminate check valves to facilitate reprocessing and cleaning, as well as prevent clogging.
- A G5 forceps inlet port was modified. The new port is smaller and comes with a rubber O cap. The smaller port and rubber cap help increase suction and reduce leakage.
- The jet wash line check valve was removed. Internal check valves were removed to elimio nate the potential for clogging and to facilitate cleaning, disinfection, and sterilization. The valves are now external and removable.
- The suction and air/water cylinders and valves were upgraded. They were updated to ac-0 commodate the new valves. The function of the valves is the same.
- Addition of the FOV, which is the rubber forceps inlet valve cover. This helps create a wao tertight seal when the endoscope is used.
- Upgrade to CA-500 cleaning adaptor. The cleaning adaptor allows the scope to be cono nected to tubes for cleaning.
A chart comparing the subject device to the predicate device can be found in Annex 8. Additional information regarding the predicate device can be found in Annex 4.
# Description of the Device [21 CFR 807.92(a)(4)}
The Fujinon G5 Bronchoscopes are medical endoscopes used for the observation, diagnosis, and endoscopic treatment of the trachea and broncho.
- The G5 Bronchoscopes are offered in a 200 and 400 series. The EB-250S represents the G5 . Bronchoscope 200 series and the EB-470S represents the G5 Bronchoscope 400 series.
The G5 Bronchoscopes include minor changes that improve the useability, ergonomics, and cleaning of the devices. The G5 scopes do not have an L-Port. The L-port functioned as a lens wash port. Doctors had the option to take a syringe to inject a fluid to use it as a high pressure wash for the lens. This function was eliminated because demand was low and it was rarely used by the surgeons.
The J-Port was repositioned. The J-Port is used as a jet water wash port. The J-Port was repositioned based on doctor feedback. The port was moved from the bottom part to the top (end) of the scope. There was also a desire to eliminate check valves to facilitate reprocessing and cleaning, as well as prevent clogging.
A G5 forceps inlet port was modified. The new port is smaller and comes with a rubber cap. The new design helps increase suction and reduces leakage.
The jet wash line check valve was removed. Internal check valves were removed to eliminate the potential for clogging and to facilitate cleaning, disinfection, and sterilization. The valves are now external and removable.
The suction and air/water cylinders and valves were upgraded. They were updated to accommodate the new valves. The function of the valves is the same.
{2}------------------------------------------------
The G5 Bronchoscopes also feature the addition of the FOV, which is the rubber forceps inlet valve cover. This helps create a watertight seal when the endoscope is used. Upgrade to CA-500 cleaning adaptor. The cleaning adaptor allows the scope to be connected to tubes for cleaning.
The G5 Bronchoscopes are used with the same processors, monitors, hard carts as the predicate devices. Each Duodenoscope is packaged in a protective carrying case with lens cleaner, silicon oil, forceps valve, waterproof cap, S connector cap, protective cap, adapters, valves, and the Operation Manual.
The Fujinon G5 Bronchoscopes are used in conjunction with other peripherals specified in the Operation Manual such as:
- Light Source .
- . Processor
- Cart .
- Data Keyboard .
- Foot Switch .
- Monitor .
- Video Printer .
- Camera and Hard Copy Unit .
- VCR .
- . ElectroSurgical Instruments
Specifications Chart:
| | EB-250S | EB-470S | EB-270P |
|---------------------------|-------------|-------------|-------------|
| Viewing Direction | Forward | Forward | Forward |
| Observation Range | 3-50mm | 3-100mm | 3-100mm |
| Field of view | 120 degrees | 120 degrees | 120 degrees |
| Image size | Super Image | Super Image | Super Image |
| Distal end diameter | 5.3 mm | 4.9mm | 3.8 mm |
| Flexible portion diameter | 5.3 mm | 4.9mm | 3.5 mm |
| Bending capability | | | |
| Up | 180 degrees | 180 degrees | 180 degrees |
| Down | 130 degrees | 130 degrees | 130 degrees |
| Left | N/A | N/A | N/A |
| Right | N/A | N/A | N/A |
| Forceps channel diameter | 2.0mm | 2.0mm | 1.2 mm |
| Working length | 600mm | 600mm | 600mm |
| Total length | 870mm | 870mm | 870mm |
| Processor | 200 Series | 400 Series | 200 Series |
{3}------------------------------------------------
510(k) Notification Fujinon G5 Bronchoscopes
# Intended Use [21 CFR 807.92(a)(5)]
The device is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
# Technological Characteristics [21 CFR 807.92(a)(6)]
Fujinon, Inc. believes that the subject device is substantially equivalent to the predicate device. The subject device have the same indications for use, material composition, image size, bending, reprocessing/sterilization method, and working dimensions as the prodicate. The subject device uses the same processor and peripherals as the predicate device.
The main technological differences are the minor changes associated with the G5 upgrade.
#### Performance Data [21 CFR 807.92(b)(1)]
The subject device has been subjected to and passed electrical safety, thermal, and EMC testing requirements. The materials in the endoscope are identical to the materials used in the predicate device.
#### Conclusion [21 CFR 807.92(b)(3)]
We believe the changes are minor and conclude that the subject devices are as safe and effective as the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of a bird or abstract design, consisting of three curved lines that resemble wings or feathers.
MAY 1 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Fujinon, Inc. c/o Joseph Azary Azary Technologies, LLC 543 Long Hill Avenue Shelton, CT 06484
Re: K050907
KU50907
Trade/Device Name: Fujinon G5 Bronchoscopes, Models EB-407S, EB-250S and EB-270P Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope and Accessories Regulatory Class: Class II Product Code: EOQ Dated: March 31, 2005 Received: April 11, 2005
Dear Mr. Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device wt have reviewed four becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreate) to trg. 3 the enactment date of the Medical Device Amendments, or to commerce prior to 1125 2011-11-11 accordance with the provisions of the Federal Food, Drug, de vices that have been recussions in ausire approval of a premarket approval application (PMA). and Cosmetic Free (. rece) that be nevice, subject to the general controls provisions of the Act. The r ou may, mereleve, mains of the Act include requirements for annual registration, listing of general controls pro reseries, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is olassined (600 above) als. Existing major regulations affecting your device can may of subject to subli additions, Title 21, Parts 800 to 898. In addition, FDA may be found in the Ooscements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I least be advised that I Dri a lossants of our device complies with other requirements of the Act that I Dri has made a abinistered by other Federal agencies. You must of any I catal statutes and regaranents, including, but not limited to: registration and listing (21 comply with an the rece stequirements)01); good manufacturing practice requirements as set CFN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable the electronic forth in the quality by wills (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This lotter will and in your a FDA finding of substantial equivalence of your device to a legally premaince noticated on "The classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you a the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other golferal micronal and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
David M. Whipple
Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Notification Fujinon G5 Bronchoscopes
5 10(k) Number (if known):
Device Name: Fujinon Inc. G5 Bronchoscopes
Indications For Use: The device is intended for the observation, diagnosis, and endoscopic treatment of the trachea and bronchial tree.
KOSO907
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Anatolub
(Optional Format 1-2-96)
(Division Nose and
510(k) Number
Page 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.