STRYKER UNIVERSAL SMARTLOCK HYBRID MMF SYSTEM

K122313 · Stryker · JEY · Oct 31, 2012 · Dental

Device Facts

Record IDK122313
Device NameSTRYKER UNIVERSAL SMARTLOCK HYBRID MMF SYSTEM
ApplicantStryker
Product CodeJEY · Dental
Decision DateOct 31, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4760
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing.

Device Story

Maxillomandibular fixation (MMF) system for temporary stabilization of jaw fractures. Components include pure titanium fixation plate, self-drilling titanium alloy locking screws (6mm/8mm), bending instrument, and spacing instrument. Plate features in-plane bendable loops for screw fixation to maxilla or mandible; spacer prevents gingival pressure. Used by clinicians to maintain occlusion during fracture healing. System provides mechanical fixation via screws and ligature wires. Benefits include stable fracture alignment during healing process.

Clinical Evidence

No clinical testing was performed. Non-clinical testing included biocompatibility, cleaning, screw insertion, corrosion resistance, mechanical stability of construct, locking tests, system handling, and cadaver-based end-user testing.

Technological Characteristics

Materials: Commercially pure titanium and titanium alloy. Components: Fixation plate, self-drilling locking screws, bending/spacing instruments. Fixation: Inter-radicular screw or circumferential wiring. Delivery: Non-sterile. Mechanical principle: Locking screw fixation with spacer to prevent gingival pressure.

Indications for Use

Indicated for treatment of mandibular and maxillary fractures in adults and adolescents (age 12+) with erupted permanent teeth.

Regulatory Classification

Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Stryker Universal SMARTLock Hybrid MMF System Image /page/0/Picture/1 description: The image shows the word "stryker" in black font. The font is bold and sans-serif. There is a small "c" in superscript to the right of the word. The word is likely a logo or brand name. Craniomaxillofacial | 1<br>ﮨﮯ<br>t | ( C<br>C<br>10<br>S | (<br>SI | |--------------|---------------------|---------| | | | | # 510(k) Summary | Date prepared: | July 31, 2012 | OCT 31 20 | |----------------------------|-------------------------------------------------------------------------------------------|-----------| | Submitter: | Stryker Craniomaxillofacial<br>750 Trade Center Way<br>Portage, MI 49002<br>USA | | | Contact: | Rob Yamashita<br>Phone: (269) 389-4258 / Fax: (269) 329-2235<br>rob.yamashita@stryker.com | | | Proprietary Name: | Stryker Universal SMARTLock Hybrid MMF System | | | Common Name: | Universal Hybrid MMF | | | Proposed Regulatory Class: | Class II | | | Product Codes: | DZL – Intraosseous fixation screw or wire<br>JEY – Bone Plate | | | Predicate Device: | Stryker MMF Screw – K050535<br>KLS Martin Erich Arch Bar – K061271 | | ## Intended Use: The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. #### Indication for Use: The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and ad olescents (age 12 and higher) in whom permanent teeth have erupted. ## Device Description: The Stryker Universal SMARTLock Hybrid MMF System is a maxillomandibular fixation system that consists of a pure titanium fixation Plate, 2 self-drilling locking Screws made out of titanium alloy (6mm and 8mm), a bending instrument and a spacing instrument. These devices are stored in a dedicated module and in a dedicated generic instrument tray, respectively. Both the tray and the module are stored in a g eneric storage container. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and there is a small superscript symbol to the right of the "r". The text is black against a white background. #### Craniomaxillofacial The Plate has in-plane bendable attachment loops for screw fixation on either the maxilla or mandible. The fixation loops can be bent with the Bender. The locking screw fixation together with use of the Spacer is designed to prevent pressure on the gingiva from the fixated Plate. #### Technological Characteristics: The Stryker Universal SMARTLock Hybrid MMF System is similar to its Predicate Devices in the following technological characteristics: - The Stryker Universal SMARTLock Hybrid HMMF Material: . System is made of either commercially pure titanium or titanium alloy. The Predicate Devices are made of stainless steel. The Stryker Universal SMARTLock Hybrid MMF Design: . System is a c ombination of the legally marketed MMF Screw System (K050535) and Erich Arch Bar (K061271). - Inter-radicular Screw Fixation or Circumferential Mode of Fixation: . wiring of dentures or teeth as with the Predicate Devices (K050535, K061271). - With ligature wires as with the Predicate Devices Intermaxillary . (K050535, K061271). Stabilization: - The devices of the Stryker Universal SMARTLock Packaging: . Hybrid MMF System are delivered non-sterile as are the Predicate Devices. #### Clinical Testing: No clinical testing was performed to support this submission. #### Non-Clinical Testing: The Stryker Universal SMARTLock Hybrid MMF System proved to be successful in all the tests conducted with regards to biocompatibility, cleaning, screw insertion, corrosion resistance, mechanical stability of construct, locking test, a system handling test and an end-user (cadaver) test. ## Substantial Equivalence to Predicate Devices: The Stryker Universal SMARTLock Hybrid MMF System combines the principles of operation of the Predicate Devices. It has nearly identical intended use and indications. Material and technological characteristics are similar to the legally marketed Predicate Devices. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 31 2012 Stryker® Craniomaxillofacial Mr. Rob Yamashita Associate Manager, Regulatory 750 Trade Centre Way, Suite 200 Portage, Michigan 49002 Re: K122313 Trade/Device Name: Stryker Universal SMART Lock Hybrid MMF System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY, DZL Dated: July 19, 2012 Received: August 2, 2012 Dear Mr. Yamashita: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. ph for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K122313 Device Name: Indications For Use: Stryker Universal SMARTLock Hybrid MMF System The Stryker Universal SMARTLock Hybrid MMF System is intended to be used for temporary stabilization of mandibular and maxillary fractures in order to maintain proper occlusion during fracture healing. The Stryker Universal SMARTLock Hybrid MMF System is indicated for the treatment of mandibular and maxillary fractures in adults and adolescents (age 12 and higher) in whom permanent teeth have erupted. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runon (Division Sign-Off) (Division Sign-On) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...