The Digital Impression System for Orthodontic Use is an optical impression system intended for use by dental professionals to record the topographical characteristics of teeth, gingiva and/or palate. The Digital Impression System is intended for use in conjunction with the production of orthodontic appliances.
Device Story
Handheld intraoral 3D imager; captures 3D topography of teeth, gingiva, and palate. Operates via blue laser diode (405 nm) using interferometry to measure surface geometry. Video camera in wand captures images; proprietary software processes data into 3D models in real-time. Used in orthodontic offices by orthodontists or assistants; replaces traditional PVS cast impressions. Output is digital 3D file used by manufacturers to design/produce customized orthodontic appliances. System includes handheld wand, computer base unit, and touch screen monitor. Disposable wand tips used for cross-contamination control. Benefits include elimination of traditional impression materials and powdering requirements (unlike predicate), and streamlined digital workflow for appliance fabrication.
Clinical Evidence
No clinical data. Bench testing included biocompatibility studies (cytotoxicity, irritation, sensitization) on disposable wand tips and validation of software performance. 3D topographical accuracy verified by comparing digital files generated by the system against digitized PVS traditional impressions.
Technological Characteristics
Handheld intraoral 3D imager. Light source: 405 nm blue laser diode. Measurement principle: interferometry. Components: handheld wand, computer base unit, touch screen monitor. Connectivity: cable-connected wand to base unit. Disposable single-patient use wand tips. Software: proprietary 3D image acquisition and processing. No tooth coating required.
Indications for Use
Indicated for dental professionals to record topographical characteristics of teeth, gingiva, and/or palate for the production of orthodontic appliances.
Regulatory Classification
Identification
An optical impression system for computer assisted design and manufacturing (CAD/CAM) is a device used to record the topographical characteristics of teeth, dental impressions, or stone models by analog or digital methods for use in the computer-assisted design and manufacturing of dental restorative prosthetic devices. Such systems may consist of a camera, scanner, or equivalent type of sensor and a computer with software.
Special Controls
*Classification.* Class II (Special Controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of the chapter subject to the limitations in § 872.9. The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations; Guidance for Industry and FDA.” For the availability of this guidance document, see § 872.1(e).
{0}------------------------------------------------
K122065
510(k) Summary
Image /page/0/Picture/2 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a swooping line above them. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are printed in a smaller, sans-serif font.
FEB 8 2013
## Submitter:
Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7472 - Facsimile Wendy Garman - Contact Person
Date Summary Prepared:
February 2013
### Device Name:
- Trade Name Digital Impression System for Orthodontic Use .
- Common Name Intraoral Imaging System .
- Classification Name Optical Impression Systems for CAD/CAM, per 21 CFR § 872.3661 .
- Product Codes Optical Impression System, Computer Assisted Design and Manufacturing . (CAD/CAM) of Dental Restorations (NOF)
### Devices for Which Substantial Equivalence is Claimed:
- 3M Unitek Lava Chairside Oral Scanner, Brontes Technologies, Inc. K081961 .
### Device Description:
The Digital Impression System for Orthodontic Use (DIS) is a handheld intraoral 3D imager intended for use inside the human oral cavity to digitally capture the three-dimensional topography of teeth, gingiva, and/or palate in clinical settings. It is intended to replace traditional cast impressions which are the current standard for recording dental and orthodontic features of patients undergoing orthodontic treatment. The system is intended to be used by orthodontists and orthodontic assistants in an orthodontic office environment. The system consists of a handheld wand connected to a computer which is housed in a base unit. The computer contains proprietary software to acquire, process, and store the digital 3D image data. Patient information is entered into the software using a touch screen monitor connected to the computer. To capture a 3D image of the patient's dental arch and/or bite, the operator
> 1717 West Collins Avenue, Orange, CA 92867 800-537-7824. 714-516-7400.
{1}------------------------------------------------
moves the wand over the surface of the teeth to be scanned. A video camera inside the wand captures images of the teeth surfaces. Algorithms in the sòftware process these images into a 3D image and display the 3D image on the computer monitor in real time. The software also saves the 3D image data and identifying patient information to be used by orthodontic manufacturers to design and manufacture customized orthodontic appliances.
### Indications for Use:
The Digital Impression System for Orthodontic Use is an optical impression system intended for use by dental professionals to record the topographical characteristics of teeth, gingiva, and/or palate. The Digital Impression System is intended for use in conjunction with the production of orthodontic appliances.
### Summary of Technological Characteristics:
The Digital Impression System for Orthodontic Use is substantially equivalent to another legally marketed device in the United States. The Digital Impression System for Orthodontic Use functions in a manner similar to and is intended for the same use as 3M Unitek Lavo Chairside Oral Scanner that is currently marketed by Brontes Technologies, Inc.
| Features | Digital Impression System for<br>Orthodontic Use | 3M Unitek Lava Chairside Oral<br>Scanner |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Digital Impression System for<br>Orthodontic Use is an optical<br>impression system intended for<br>use by dental professionals to<br>record the topographical<br>characteristics of teeth, gingiva,<br>and/or palate. The Digital<br>Impression System is intended for<br>use in conjunction with the<br>production of orthodontic<br>appliances. | The 3M Unitek Lava Chairside Oral<br>Scanner is an optical impression<br>system (CAD/CAM) used to record<br>the topographical characteristics<br>of teeth. Data generated from the<br>3M Unitek Lava Chairside Oral<br>Scanner may be used in<br>conjunction with the production<br>of orthodontic appliances,<br>retainers and accessories. |
| Target users | Dental Professionals trained in<br>orthodontics | Dental Professionals trained in<br>orthodontics |
| Where used | Dental offices, used at dental chair | Dental offices, used at dental chair |
| Anatomical Sites | Upper and lower arches of teeth,<br>left and right bite | Upper and lower arches of teeth,<br>left and right bite |
| Features | Digital Impression System for<br>Orthodontic Use | 3M Unitek Lava Chairside Oral<br>Scanner |
| Technique to | Interferometry measurement of | Stereoscopic measurement of |
{2}------------------------------------------------
| Produce 3D Images | video camera images | video camera images |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Light used for illumination | Blue (405 nm) | Blue (405 nm) |
| Light Source | Diode (laser) | Diode (light emitting) |
| Tooth Coating | None | Titanium dioxide powder |
| Components | Handheld imaging wand containing a high accuracy video camera<br>Computer mounted in housing that can be moved from room to room<br>Cable connecting wand with computer<br>Touch screen monitor | Handheld imaging wand containing a high accuracy video camera<br>Computer mounted in housing that can be moved from room to room<br>Cable connecting wand with computer<br>Touch screen monitor |
| Key Characteristics | Handheld wand is moved over the teeth to acquire 3D images<br>Real-time display of 3D images while patient is being imaged | Handheld wand is moved over the teeth to acquire 3D images<br>Real-time display of 3D images while patient is being imaged |
| Device Output | Output is a software file that can be used as input to CAD/CAM dental processes | Output is a software file that can be used as input to CAD/CAM dental processes |
| Patient Contact Areas | Cheek, tongue, lips, teeth, palate | Cheek, tongue, lips, teeth, palate |
| Patient Contact Areas; Cross-contamination Control | Wand tip is single patient/single use disposable | Wand tip must be sterilized between patients |
| Checks of software and hardware function | Scan of verification target | Scan of verification target |
## Non-Clinical Test Data:
Biocompatibility studies (Cytotoxicity, Irritation and Sensitization) have been completed on the disposable wand tip included in the Digital Impression System for Orthodontic Use product.
3D topographical files were created by the Digital Impression System for Orthodontic Use and compared to files derived from, digitized polyvinylsiloxane (PVS) traditional impressions. Test results confirmed that digital impressions generated by the Digital Impression System for Orthodontic Use are substantially equivalent to PVS impressions made by traditional means.
{3}------------------------------------------------
Additionally, the Digital Impression System for Orthodontic Use software has been successfully validated to confirm the performance of the device.
### Clinical Test Data:
Clinical testing has not been conducted on this product.
### Conclusion:
Based upon the biocompatibility studies, similar technological characteristics to the predicate device, comparison between 3D topographical files created by the Digital Impression System for Orthodontic Use and files created from digitized PVS impressions and successful validation of the Digital Impression System for Orthodontic Use software, the performance of the Digital Impression System for Orthodontic Use is deemed to be substantially equivalent to 3M Unitek Lava Chairside Oral Scanner.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings outstretched, symbolizing the department's mission to protect and promote the health and well-being of the nation.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### February 8, 2013
Ormco Corporation C/O Ms. Wendy Garman Director, Regulatory Affairs Sybron Dental Specialties. Incorporated 1717 West Collins Avenue ORANGE CA 92867
Re: K122065
Trade/Device Name: Digital Impression System for Orthodontic Use Regulation Number: 21 CFR 872.3661 Regulation Name: Optical Impression Systems for CAD/CAM Regulatory Class: II Product Code: NOF Dated: February 1, 2013 Received: February 5, 2013
### Dear Ms. Garman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Garman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony De m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
KI22065
# Indications for Use
510{k} Number (if known):
Device Name: Digital Impression System for Orthodontic Use
Indications For Use:
The Digital Impression System for Orthodontic Use is an optical impression system intended for use by dental professionals to record the topographical characteristics of teeth, gingiva and/or palate. The Digital Impression System is intended for use in conjunction with the production of orthodontic appliances.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Andrew I. Steen 2013.02.07 15:52:22=05'00'
Division Sianlology, General Hospital
**510(k) Number:** k122065
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.