The 3M™ Unitek™ Lava™ Chairside Oral Scanner is an optical impression system (CAD/CAM) used to record the topographical characteristics of teeth. Data generated from the 3M™ Unitek ™ Lava™ Chairside Oral Scanner may be used in conjunction with the production of orthodontic appliances, retainers and accessories.
Device Story
3M™ Unitek™ Lava™ Chairside Oral Scanner is an optical impression system for dental CAD/CAM applications. System comprises a computer, proprietary software, and a handheld dental wand. Operator (dental professional) uses wand in close proximity to teeth to capture 3D topographical data. Software processes input to create and store 3D models alongside patient information. Used in clinical settings to facilitate production of orthodontic appliances, retainers, and accessories. Output allows clinicians to design and manufacture dental devices, potentially improving workflow efficiency and patient care compared to traditional impression methods.
Clinical Evidence
Bench testing only. In-vitro test data provided to demonstrate safety and effectiveness of software modifications for the expanded indications.
Technological Characteristics
Optical impression system (CAD/CAM). Components: computer, proprietary software, handheld dental wand. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-4 (programmable electrical medical systems), and ISO 10993 (biocompatibility).
Indications for Use
Indicated for recording topographical characteristics of teeth for the production of orthodontic appliances, retainers, and accessories. No specific age or gender restrictions noted. Contraindications not specified.
Regulatory Classification
Identification
An optical impression system for computer assisted design and manufacturing (CAD/CAM) is a device used to record the topographical characteristics of teeth, dental impressions, or stone models by analog or digital methods for use in the computer-assisted design and manufacturing of dental restorative prosthetic devices. Such systems may consist of a camera, scanner, or equivalent type of sensor and a computer with software.
Special Controls
*Classification.* Class II (Special Controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of the chapter subject to the limitations in § 872.9. The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations; Guidance for Industry and FDA.” For the availability of this guidance document, see § 872.1(e).
{0}------------------------------------------------
Page 16 of 174
SEP - 5 2008
#### 5.0 510K Summary
Applicant: Brontes Technologies, Inc. (A 3M Company) 10 Maguire Road, Suite 310 Lexington, MA 02421 Tel: (781) 541-5200
## Contact: Mark Olsen
# 510(k) Numbers and Product Codes of equivalent devices:
Brontes Technologies, Inc. (A 3M Company), Lexington, MA 3M™ ESPE Lava™ Chairside Oral Scanner Code: NOF; K073199 1
#### Indications for Use
The 3M™ Unitek™ Lava™ Chairside Oral Scanner is an optical impression system (CAD/CAM) used to record the topographical characteristics of teeth. Data generated from the 3M™ / Unitek ™ Lava™ Chairside Oral Scanner may be used in conjunction with the production of orthodontic appliances, retainers and accessories.
#### Device Description
The 3M™ Unitek™ Lava™ Chairside Oral Scanner is a system intended to obtain 3D images of the topographical characteristics of Teeth. The System consists of a computer, proprietary software and a hand held dental 'wand' for scanning the surface of the teeth.
Using the wand in close proximity to the teeth, the image of the teeth is communicated to the software and a 3D model is created by the software. The 3D model is then stored on disk along with pertinent patient information.
In this 510K pre-market notification, the software has been modified to adapt to the revised intended use statement, which adds a new device with the application of generating models for orthodontic appliances, retainers and accessories. An analysis of the changes to the device along with In-Vitro test data demonstrates that no new issues of safety or effectiveness are raised by the changes to the device discussed in this premarket notification.
The 3M™ Unitek™ Lava™ Chairside Oral Scanner will be manufactured by Brontes Technologies, Inc., a 3M Company, and distributed by 3M Unitek, another 3M Company,
#### Performance Standards
There are no changes to the performance standards with The 3M™ Unitek™ Lava™ Chairside Oral Scanner as compared to the predicate device. The 3M™ ESPE Lava™ Chairside Oral Scanner meets the following Performance Standards:
- . Tripartite Guidance - 1987 (G87-1)
- . Special Controls Guidance Document: Optical Impression Systems for Computer Assisted Design and Manufacturing (CAD/CAM) of Dental Restorations; Guidance for Industry and FDA
{1}------------------------------------------------
- . ISO/EN 10993-1; 1997 Biological Evaluation of Medical Devices, Part I: Evaluation and Testing
- ISO/EN 10993-5; 1999 Tests for In-Vitro Cytotoxicity, 20d Ed �
- ISO/EN 10993-10; 2002 Tests for Irritation and Delayed Type . Hypersensitivity, 2nd Ed
- IEC 60601-1 Medical Electrical Equipment Part 1: General . Requirements for Safety
- IEC 60601-1-2 Medical Electrical Equipment Part 1 General . Requirements for Safety; Electromagnetic Compatibility -Requirements and Tests
- IEC 60601-1-4 medical Electrical Equipment Part 1 Medical . Electrical Equipment part 1-4: General Requirements for Collateral Standard: Programmable Electrical Medical Systems
- FDA Guidance on Dental Handpieces (Draft) July 1995 .
- . ISO 7405 - Pre-Clinical evaluation of Biocompatibility of medical devices used in dentistry, test methods for dental materials.
- FDA; Guidance for the Content of Pre-market Submissions for . Software Contained in Medical Devices - May 2005
- FDA Guidance; Cyber Security for Networked Medical Devices . Containing Off the Shelf Software
- ISO 14971 Medical devices Risk management Part 1: Application . of risk analysis
### Conclusion:
There are more similarities than differences between the predicate devices and the 3M™ Unitek™ Lava™ Chairside Oral Scanner. The predicate device and the 3M™ Unitek™ Lava™ Chairside Oral Scanner have similar intended uses, warnings and contraindications. Both the 3M™ Unitek™ Lava™ Chairside Oral Scanner and predicate device may be used with a powder and powdering delivery system. When used in accordance with the instructions for use, by qualified personnel, the 3M™ Unitek™ Lava™ Chairside Oral Scanner is safe and effective, as indicated, for the intended use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, serif font. The text is centered and appears to be a header or title. The words are all capitalized and evenly spaced.
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Brontes Technologies, Incorporated C/o Mr. John Greenbaum President Generic Devices Consulting, Incorporated 20310 SW 48th Street Southwest Ranches, Florida 33332
Re: K081961
Trade/Device Name: 3M™ Unitek™ Lava™ Chairside Oral Scanner Regulation Number: 872.3661 Regulation Name: Optical Impression Systems for CAD/CAM Regulatory Class: II Product Code: NOF Dated: July 8, 2008 Received: July 9, 2008
SEP - 5 2008
Dear Mr. Greenbaum:
We have reviewed your Scotion 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{3}------------------------------------------------
Page 2 - Mr. Greenbaum
Please be advised that FDA's issuance of a substantial equivalcnce determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
J.W.amelis-Reed, m.y.
Chiu S. Lin. Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K081961
Device Name: 3M™ Unitek™ Lava™ Chairside Oral Scanner
Indications For Use:
The 3M™ Unitek™ Lava™ Chairside Oral Scanner is an optical impression system (CAD/CAM) used to record the topographical characteristics of teeth. Data generated from the 3M™ Unitek ™ Lava™ Chairside Oral Scanner may be used in conjunction with the production of orthodontic appliances, retainers and accessories.
Prescription Use X (Per 21 CRF 801.109) OR
Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Turner
(Optional Format 1-2-96)
(Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dantal Devices
510(k) Number: K081961
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.