ETS BLUE POWDER-FREE PATIENT NITRILE EXAMINATION GLOVES
Applicant
Northstar Healthcare Holdings
Product Code
LZA · General Hospital
Decision Date
Sep 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
ETS Blue Powder Free Nitrile Patient Exam Glove is a non-sterile, single-use, ambidextrous nitrile glove. Worn by healthcare providers on hands or fingers during patient examinations to serve as a protective barrier against contamination. Device is manufactured from nitrile; provided in dispenser boxes. Performance verified against ASTM standards for dimensions, physical properties, pinhole resistance, and residual powder. Biocompatibility confirmed via skin irritation and sensitization testing. No clinical data required; device functions as a passive physical barrier.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through testing against ASTM D6319-10 (dimensions, physical properties), ASTM D6124-06 (residual powder), and ASTM D5151-06 (freedom from pinholes). Biocompatibility assessed per ISO 10993-10, confirming the device is non-irritating and does not display potential for sensitization.
Technological Characteristics
Material: Nitrile. Form factor: Ambidextrous, powder-free, non-sterile, textured. Standards: ASTM D6319-10 (dimensions, physical properties), ASTM D6124-06 (residual powder), ASTM D5151-06 (pinholes), ISO 10993-10 (biocompatibility). No energy source or software.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
# 510(k) SUMMARY
DATE:
OWNER:
February 17, 2012
SEP 10 2012
Northstar Healthcare Holdings 70 Sir John Rogerson's Quay Dublin 2, Ireland
#### OFFICIAL
#### CORRESPONDENT:
Alex Nagy Manager, Quality Systems and Regulatory Affairs Cypress Medical Products LLC 1200 South Route 31 McHenry, IL 60050 Telephone: 815-385-0100 Fax: 815-385-0114
#### DEVICE NAME:
| Trade Name: | ETS Blue Powder Free Nitrile Patient Exam Glove |
|-----------------|-------------------------------------------------|
| Common Name: | Patient Examination Gloves |
| Classification: | Patient Examination Gloves |
| Class: | Class I |
| Product Code: | LZA |
## PREDICATE DEVICE(S):
| Predicate<br>510(k) | Device Name | Indication | Clearance<br>Date | Company |
|---------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--------------------------------------------------------------------------------------|
| K062764 | Nitrile Powder Free<br>Examination Gloves<br>(White, Blue) | The patient examination glove is a<br>disposable device intended for medical<br>purposes that is worn on the examiner's hand<br>to prevent contamination between patient and<br>examiner. | 27 Feb 07 | Kossan<br>Latex<br>Industries<br>SDN BHN.<br>Selangor<br>Darul<br>Ehsan,<br>Malaysia |
DEVICE DESCRIPTION: Powder free blue nitrile patient exam glove, meets all the requirements of ASTM standard D6319, except for sterility requirements.
Northstar Healthcare Holdings
{1}------------------------------------------------
## STATEMENT OF INTENDED USE:
The nitrile examination glove is a disposable device intended for medical purposes to be worn on the examiner's hand or finger to prevent contamination between patient and examiner.
TECHNOLOGICAL CHARACTERISTICS: The ETS Blue Powder Free Nitrile Patient Exam Glove is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with nitrile using similar manufacturing processes.
| Feature | Nitrile Powder Free Examination Gloves Tested<br>for Use with Chemotherapy Drugs (Blue,<br>White)<br>K062764<br>Predicate | ETS Blue Powder Free Nitrile<br>Patient Exam Gloves<br>Proposed |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Intended for medical purposes that is worn on<br>the examiner's hand to prevent contamination<br>between patient and examiner. Tested for Use<br>with Chemotherapy Drugs. | Intended for medical purposes that<br>is worn on the examiner's hand to<br>prevent contamination between<br>patient and examiner. |
| Indications for Use<br>Statement | The patient examination glove is a disposable<br>device intended for medical purposes that is<br>worn on the examiner's hand to prevent<br>contamination between patient and examiner.<br>This device is not intended to be used as a<br>chemical barrier. | The examination glove is a<br>disposable device intended for<br>medical purposes that is worn on the<br>examiner's hand or finger to prevent<br>contamination between patient and<br>examiner. |
| Description | Non-sterile, powder free, examination gloves<br>made of nitrile and colored white. The textured<br>gloves are provided in sizes extra small, small,<br>medium, large and extra large. The smooth<br>gloves are provided in sizes small, medium,<br>large and extra large. | Same |
| Presentation | Non-sterile gloves are provided in dispenser<br>boxes. | Same |
| Material | Nitrile | Same |
| Sterilization | Non-sterile | Same |
| Single Use | Yes | Same |
| Ambidextrous | Yes | Same |
| Dimensions | Meets ASTM D6319-10 | Same |
| Tensile Strength | Meets ASTM D6319-10 | Same |
| Ultimate Elongation | Meets ASTM D6319-10 | Same |
| Freedom from<br>Pinholes | Meets ASTM D5151-06 and ASTM D6319-10 | Same |
| Residual Powder | Meets ASTM D6124-06 | Same |
| Biocompatibility | Passes Primary Skin Irritation in Rabbits | Same |
| Tests | Passes Closed Patch Sensitization in Guinea Pigs | Same |
Northstar Healthcare Holdings
{2}------------------------------------------------
# ASSESSMENT OF NONCLINICAL DATA:
| Characteristic | Standard | Device Performance |
|-----------------------|------------------------------------|-----------------------------|
| Dimension | ASTM D6319 | Meets |
| Physical Properties | ASTM D6319 | Meets |
| Freedom from Pinholes | ASTM D6319 & ASTM D5151 | Meets |
| Powder Residual | ASTM D6124 | Meets |
| | | Results generated values |
| | | below 2mg of residual |
| | | powder |
| Biocompatibility | Primary Skin Irritation in rabbits | Gloves are non-irritating |
| | (ISO 10993-10) | |
| | Dermal Sensitization in the guinea | Gloves do not display any |
| | pig (ISO 10993-10) | potential for sensitization |
#### CONCLUSIONS:
The ETS Blue Powder Free Nitrile Patient Exam Gloves meet the requirements of established standards ASTM D6319-10, ASTM D6124-06, ASTM D5151-06, and ISO 10993-10.
Based on the comparison of intended use, design, materials and performance, the ETS Blue Powder Free Nitrile Patient Exam Gloves are substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Northstar Healthcare Holdings C/O Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
SEP 10 2012
Re: K121947
Trade/Device Name: ETS Blue Powder Free Nitrile Patient Exam Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LZA Dated: August 28, 2012 Received: August 28, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2- Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
For Q. Shem.D.
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(k) Number (if known): _ K 12 1947
#### ETS Blue Powder Free Nitrile Patient Exam Glove Device Name:
Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) |
|--------------------------------------------------------------------------------|
| Division of Anesthesiology, General Hospital Infection Control, Dental Devices |
Page 1 of _510(k) Number: