GIBRALT OCCIPITAL PLATE SCREWS, SET SCREW, GIBRALT OCCIPITAL SYSTEM ARTICULATING ROD, GIBRALT OCCIPITAL PLATES
K121877 · Exactech, Inc. · KWP · Feb 12, 2013 · Orthopedic
Device Facts
Record ID
K121877
Device Name
GIBRALT OCCIPITAL PLATE SCREWS, SET SCREW, GIBRALT OCCIPITAL SYSTEM ARTICULATING ROD, GIBRALT OCCIPITAL PLATES
Applicant
Exactech, Inc.
Product Code
KWP · Orthopedic
Decision Date
Feb 12, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used with the Gibralt System, the Gibralt Occipital Spine System is intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput -T3). The Gibralt Occipital Spine System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/dislocation, occipital-cervical dislocation, atlanto-axial fracture with instability, failed previous fusion and/or tumors. The Occipital Bone Screws are limited to occipital fixation only. The use of polyaxial screws is limited to placement in TI-T3 in treating thoracic conditions only. Polyaxial screws are not intended to be placed in the cervical spine.
Device Story
Posterior spinal fixation system providing immobilization and stabilization as an adjunct to fusion. Components include occipital plates, occipital bone screws, articulating rods, and set screws; used with Gibralt Spine System (hooks, polyaxial screws). Implanted via traditional open surgical procedure by surgeons. System provides mechanical stability to spinal segments. Benefits include stabilization of the occipital-cervical-thoracic region to promote fusion in patients with degenerative or traumatic spinal conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included static compression bend, dynamic compression bend, static yield torsion, dynamic torsion, and axial pullout testing per ASTM F2706 and ASTM F543 standards.
Technological Characteristics
Components manufactured from titanium alloy per ASTM F136. System includes occipital plates, bone screws, articulating rods, and set screws. Non-sterile, single-use, steam-sterilized by hospital. Mechanical performance validated per ASTM F2706 and ASTM F543. No software or electronic components.
Indications for Use
Indicated for patients requiring spinal fusion in the cervical spine and occipito-cervico-thoracic junction (occiput-T3) due to DDD, spondylolisthesis, spinal stenosis, fracture/dislocation, occipital-cervical dislocation, atlanto-axial fracture with instability, failed fusion, or tumors. Occipital bone screws restricted to occipital fixation; polyaxial screws restricted to T1-T3 thoracic conditions.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Zimmer (Abbott Spine) Nex-Link OCT Cervical Plating System (K090060)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(k) Summary
### FEB 1 2 2013
| Company: | Exactech®, Inc<br>2320 NW 66th Court<br>Gainesville, FL 32653 |
|----------------------|---------------------------------------------------------------|
| Date: | January 14, 2013 |
| Contact Person: | Vladislava Zaitseva<br>Sr. Regulatory Affairs Specialist |
| | Phone: (352) 377-1140<br>Fax: (352) 378-2617 |
| Proprietary Name: | Exactech® Gibralt® Occipital Spine System |
| Common Name: | Occipital Plate System |
| Classification Name: | Appliance, fixation, spinal interlaminal |
| Product Code: | KWP |
#### Legally Marketed Devices to Which Substantial Equivalence Is Claimed
- DePuy Mountaineer OCT Spinal System (K110353, K042508, and K041203) .
- t Medtronic Vertex Select Reconstruction System (K110522, K091365, K090714, and K082728)
- Zimmer (Abbott Spine) Nex-Link OCT Cervical Plating System (K090060) .
#### Device Description
The Gibralt Occipital Spine System is a new posterior system intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the occipital-cervical-thoracic region of the spine.
The system consists of a variety of sizes of the occipital plates and occipital bone screws, and articulating rod and set screws. All components of the system are manufactured from titanium allov per ASTM F136.
The Gibralt Occipital system should be used in conjunction with the Gibralt Spine System components such as hooks, polyaxial screws and connecting components. The Gibralt Spine System has been cleared by FDA through 510(k) # K 110197 on July 15, 2011.
The Gibralt Occipital Spine System is provided with a complete instrumentation system to assist the surgeon in the implantation of the components according to a traditional open surgical procedure.
{1}------------------------------------------------
#### Indications for Use
When used with the Gibralt Spine System, the Gibralt Occipital Spine System is intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput -T3). The Gibralt Occipital Spine System is indicated for the following: DDD (neck pain of discogenic origin with deveneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/dislocation, occipital-cervical dislocation, atlanto-axial fracture with instability, failed previous fusion and/or tumors.
The Occipital Bone Screws are limited to occipital fixation only.
The use of polyaxial screws is limited to placement in T1-T3 in treating thoracic conditions only. Polyaxial screws are not intended to be placed in the cervical spine.
#### Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use/Indications for Use . The proposed Gibralt Occipital Spine System and predicates have similar indications for use statements.
- . Materials
The proposed Gibralt Occipital Spine System components and predicate devices are composed of titanium allov, a biocompatible material conforming to a recognized industry standard for permanent implants.
#### . Design Features
The proposed Gibralt Occipital Spine System components and predicate devices have similar design features.
- • Dimensions
The proposed Gibralt Occipital Spine System components and predicate devices are dimensionally comparable.
#### . Packaging and Sterilization
The proposed Gibralt Occipital Spine System components and predicate devices are provided non-sterile for single use only, and will be steam sterilized by the hospital prior to use in the operating room using the same sterilization method.
- . Device Shelf Life
Neither the Gibralt Occipital Spine System components nor cited predicate devices have shelf life expiration dating.
- . Performance specifications The Gibralt Occipital Spine System and cited predicates all withstand clinically relevant biomechanical loads.
{2}------------------------------------------------
#### Substantial Equivalence Conclusion
The following mechanical testing and engineering analyses were conducted to demonstrate substantial equivalence of the proposed Gibralt Occipital Spine System to cited predicates:
- Static Compression Bend, Dynamic Compression Bend, Static Yield Torsion, . Dynamic Torsion, Axial Pullout and Static Torque testing per ASTM F2706 and ASTM F543.
- A biomechanical assessment comparing Gibralt Occipital Spine System . mechanical performance and design features to cited predicate devices.
The results of mechanical testing and analysis demonstrate the proposed device is substantially equivalent to cited predicates.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings outstretched, symbolizing protection and care.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### Letter dated: February 12, 2013
Exactech®, Incorporated % Travis Arola, MS, RAC, CTBS Director, Regulatory, Spine & Biologics 2320 Northwest 66th Court Gainesville, Florida 32653
Re: K121877
Trade/Device Name: Exactech® Gibralt® Occipital Spine System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP Dated: January 14, 2013 Received: January 16, 2013
Dear Mr. Arola:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Travis Arola, MS, RAC, CTBS
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
## Mark Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
K121877 510(k) Number:
Device Name: Exactech® Gibralt® Occipital Spine System
#### INDICATIONS FOR USE:
When used with the Gibralt System, the Gibralt Occipital Spine System is intended to promote fusion of the cervical spine and occipito-cervico-thoracic junction (occiput -T3). The Gibralt Occipital Spine System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture/dislocation, occipital-cervical dislocation, atlanto-axial fracture with instability, failed previous fusion and/or tumors.
The Occipital Bone Screws are limited to occipital fixation only.
The use of polyaxial screws is limited to placement in TI-T3 in treating thoracic conditions only. Polyaxial screws are not intended to be placed in the cervical spine.
Over-The-Counter Use Prescription Use X and/or (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C)
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Ronald PaJean -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K121877
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.