AFT PROXIMAL HUMERUS FRACTURE PLATE

K121672 · Shoulder Options, Inc. · HRS · Oct 1, 2012 · Orthopedic

Device Facts

Record IDK121672
Device NameAFT PROXIMAL HUMERUS FRACTURE PLATE
ApplicantShoulder Options, Inc.
Product CodeHRS · Orthopedic
Decision DateOct 1, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AFT™ Proximal Humerus Fracture Plate is intended for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone.

Device Story

AFT™ Proximal Humerus Fracture Plate is a low-profile, anatomically contoured metallic bone fixation device. Used by orthopedic surgeons in clinical settings to stabilize proximal humerus fractures, dislocations, osteotomies, or non-unions. The plate features holes for 3.5mm locking/nonlocking screws and 4.5mm partially threaded screws. Provided non-sterile for single-use. The device provides mechanical stabilization of bone segments to facilitate healing, particularly in osteopenic bone.

Clinical Evidence

Bench testing only. Performance demonstrated via static and dynamic bending testing per ASTM F382-99.

Technological Characteristics

Material: Ti-6Al-4V (ASTM F136). Design: Low-profile, anatomically shaped plate with holes for 3.5mm and 4.5mm screws. Energy source: None (mechanical). Sterilization: Non-sterile (intended for single-use).

Indications for Use

Indicated for patients with fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus, including those with osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K121672 Page 1/2 OCT # 8. 510(k) Summary Shoulder Options, Inc. 100 E. South Main St. P.O. Box 1458 Waxhaw, NC 28173 June 5, 2012 Date Prepared: John Kapitan, CEO Contact Person: Tel: (704)512-0000 Fax: (704)831-5316 Email: kapitan@shoulderoptions.com #### 2. Device Name | Common/Usual Name: | Plate, Fixation, Bone | |----------------------|-----------------------------------------------------------------------------| | Classification Name: | Single/multiple component metallic bone fixation appliances and accessories | | Regulation Number: | 888.3030 | | Product Code: | HRS, HWC | | Classification: | II | | Panel: | Orthopedic | ## 3. Predicate Devices The Shoulder Options AFT™ Proximal Humerus Fracture Plate is substantially equivalent to the following devices: | 510(k) Number | Device | Manufacturer | |---------------|--------------------------------------------------|--------------| | K041860 | Synthes LCP Proximal Humerus Plate | Synthes | | K011815 | Synthes LCP Proximal Humerus Plate | Synthes | | K963172 | Synthes 4.5mm Non-Locking Cannulated<br>Screw | Synthes | | K042695 | 4.5mm NCB Proximal Humerus Non-<br>Locking Screw | Zimmer | {2}------------------------------------------------ K121672 page 2/2 #### 4. Description of the Device The AFT™ Proximal Humerus Fracture Plate is a low-profile, anatomically shaped plate which matches the natural contour of the proximal humerus. The plate is available in 'long' and 'short' configurations with left and right options. The plate contains holes for 3.5mm locking and nonlocking screws and 4.5mm partially threaded screws. The screws are available in various lengths. Both the plates and the screws are manufactured from Ti-6Al-4V (ASTM F136). The components are provided non-sterile for single-use. #### 5. Indications for Use The AFT™ Proximal Humerus Fracture Plate is intended for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. #### 6. Summary of Performance Data Testing of the AFT™ Proximal Humerus Fracture Plate to demonstrate substantial equivalence included static and dynamic bending testing per ASTM F382-99. #### 7. Safety & Effectiveness The AFT™ Proximal Humerus Fracture Plate is substantially equivalent to the Synthes LCP Proximal Humerus Fracture Plate (K041860, K011815), Synthes 4.5mm Non-Locking Cannulated Screw ( K963172), and Zimmer NCB Proximal Humerus Non-Locking Screw (K042695). The devices have the same "Indications for Use," are available by prescription only, and are provided non-sterile for single-use only. Based on this and the design similarities, it can be concluded that the AFT™ Proximal Humerus Fracture Plate is both a safe and effective device and is substantially equivalent to the predicate devices. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows a seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT קומא א Shoulder Options, Incorporated % Mr. John Kapitan CEO 100 East South Main Street P.O. Box 1458 Waxhaw, North Carolina 28173 Re: K121672 Trade/Device Name: AFT" Proximal Humerus Fracture Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: August 15, 2012 Received: August 20, 2012 #### Dear Mr. Kapitan, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ ### Page 2 - Mr. John Kapitan comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark A. Milliken Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## 7. Indications for Use Statement 510(k) Number (if known): K121672 Device Name: AFT™ Proximal Humerus Fracture Plate Indications for Use: The AFT™ Proximal Humerus Fracture Plate is intended for fractures and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. Over-The-Counter Use Prescription Use × AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Aht (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices k121672 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...