The ORTHOLOC™ 3Di Midfoot/Flatfoot System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the feet. Specific examples include: Flatfoot Osteotomies - Lateral column Lengthening (Evans Osteotomy) - Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy) - Medial Displacement Calcaneal Osteotomy (MDCO) Mid / Hindfoot Fusions - LisFranc Arthrodesis and/or Stabilization - 1st (Lapidus), 2nd, 3rd, 4th, and 5th Tarsometatarsal (TMT) Fusions - Intercuneiform Fusions - Navicular-Cuneiform (NC) Fusion - Talo-Navicular (TN) Fusion - Calcaneo-Cuboid (CC) Fusion - Medial Column Fusion
Device Story
System comprises 23 titanium alloy bone plates in 8 styles; designed for stabilization of fresh fractures, revision procedures, joint fusion, and reconstruction of small foot bones. Plates feature polyaxial locking screw holes (up to 15° off-axis) and optional non-locking compression slots/k-wire holes. Used by orthopedic surgeons in clinical settings. Plates accept 2.7 mm and 3.5 mm ORTHOLOC 3Di locking and non-locking screws. Device provides mechanical stabilization of bone segments to facilitate healing and reconstruction. No software or algorithmic components.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, specifically mechanically validated Finite Element Analysis (FEA) and dimensional engineering analysis to compare bending performance against predicate systems.
Technological Characteristics
Titanium alloy (Ti6Al4V) per ASTM F136/ISO 5832-3. Polyaxial locking mechanism (up to 15° off-axis). 23 plate variants across 8 styles. Compatible with 2.7 mm and 3.5 mm locking/non-locking screws. Mechanical fixation device.
Indications for Use
Indicated for stabilization of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the feet, including specific flatfoot osteotomies and mid/hindfoot fusions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K121651 1/3
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
OCT 1 2 2012
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the URTHOLOCTM 3Di Midfoot/Flatfoot System.
(a)(1). Submitted By:
Date:
Contact Person:
(a)(2). Proprietary Name:
Common Name:
Classification Name and Reference:
Device Product Code, Device Panel:
(a)(3). Predicate Device:
Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002
June 1, 2012
Leslie Fitch Regulatory Affairs Specialist (901) 867-4120
ORTHOLOC™ 3Di Midfoot/Flatfoot System
Bone Plate System
21 CFR 888.3030 - Class II
HRS: Orthopedic
K061808 - DARCO Locking Bone Plate System
K113014 - CLAW II Polyaxial Compression System
K100618 - Orthohelix Variable Angle Plates
K 100502 - Ascension Plating System
K11663 - DePuy A.L.P.S. Total Foot System
#### (a)(4). Device Description
Wright Medical's ORTHOLOC™ 3Di Midfoot/Flatfoot System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of bones of the feet and toes. The subject plates are modified from the DARCO Locking Bone Plate System (K061808) or the CLAW II Polyaxial Compression System (K113014). The system contains 23 plates belonging to 1 of 8 plate styles with various sizes and options, each contoured for specific anatomy and designed for specific procedures. All plates feature polyaxial locking screw holes, and some plates have non-locking compression slots and kwire holes. The plates are made from titanium alloy conforming to ASTM F136 or ISO 5832-3 and accept 2.7 mm and 3.5 mm ORTHOLOC™ 3Di locking and ORTHOLOC™ non-locking screws.
# (a)(5). Intended Use
The ORTHOLOC™ 3Di Midfoot/Flatfoot System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the feet. Specific examples include:
{1}------------------------------------------------
K121651\$\frac{2}{3}\$
Flatfoot Osteotomies
- · Lateral column Lengthening (Evans Osteotomy)
- · Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy)
- · Medial Displacement Calcaneal Osteotomy (MDCO)
Mid / Hindfoot Fusions
- · LisFranc Arthrodesis and/or Stabilization
- · 1ª(Lapidus), 2nd, 3d, 4th, and 5th Tarsometatarsal (TMT) Fusions
- · Intercuneiform Fusions
- · Navicular-Cuneiform (NC) Fusion
- · Talo-Navicular (TN) Fusion
- · Calcaneo-Cuboid (CC) Fusion
- · Medial Column Fusion
The indications statement is similar to the DARCO predicate system's indications for use statement and has been tailored to the intended use of the subject device, which has fewer plates than the DARCO system. The bones of the hand, wrist, ankles, and fingers, and toes included in the predicate indications statement are omitted from the indications for this system, because this system includes only plates designed for midfoot and flatfoot procedures. The indications statement has also been modified to include specific procedures for which the system is designed. The addition of these specific examples of osteotomies and arthrodeses does not alter the intended therapeutic effect of the subject device. Some of these procedures are part of a predicate indications statement, and the others are supported by technologically similar predicates designed and marketed for these procedures.
# (a)(6). Technological Characteristics Comparison
While many of the technological characteristics are the same for the subject device system and the DARCO predicate system, some design changes have been made. The subject plates have a new polyaxial locking feature that offers locking up to 15° off-axis. The subject's locking feature is similar to the locking feature of the predicate CLAW II and identical to the locking feature for ORTHOLOC™ 3Di Hallux (K120359). The subject plates are made from titanium alloy (Ti6AlaV), while some of the corresponding predicate plates are commercially pure titanium or stainless steel.
#### (b)(1). Substantial Equivalence - Non-Clinical Evidence
Performance testing shows that no new worst-case plates (in bending) are introduced in this system. Through mechanically validated FEA analysis, the worst-case subject ORTHOLOC™ 3Di Midfoot/Flatfoot plates were found not to represent a new worst-case in bending for any of the eight plate styles evaluated when compared to the most similar DARCO or CLAW II plates. A dimensional engineering analysis was also provided in order to help establish mechanical equivalence.
## (b)(2). Substantial Equivalence - Clinical Evidence
N/A
#### (b)(3). Substantial Equivalence - Conclusions
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains a series of alphanumeric characters. The characters 'K121651' are written in a clear, legible font. To the right of these characters, there is a fraction '3/3' written in a smaller font size.
The new design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. Performance testing shows that no new worst-case plates (in bending) are introduced in this system. From the evidence submitted in this 510(k) the subject device system can be expected to perform at least as well as the predicate systems. . •
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
OCT 12 2012
Wright Medical Technology, Inc. % Ms. Leslie H. Fitch 5677 Airline Road Arlington, TN 38002
. Re: K121651
Trade/Device Name: ORTHOLOC 3Di Midfoot/Flatfoot System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: September 19, 2012 Received: September 20, 2012
Dear Ms. Fitch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may . publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Leslie H. Fitch
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): Not yet assigned.
Device Name: ORTHOLOC™ 3Di Midfoot/Flatfoot System
Indications for Use:
The ORTHOLOC™ 3Di Midfoot/Flatfoot System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the feet. Specific examples include:
Flatfoot Osteotomies
- Lateral column Lengthening (Evans Osteotomy) ●
- Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton ● Osteotomy)
- Medial Displacement Calcaneal Osteotomy (MDCO) ●
Mid / Hindfoot Fusions
- LisFranc Arthrodesis and/or Stabilization .
- 15(Lapidus), 2nd, 3rd, 4th, and 5th Tarsometatarsal (TMT) Fusions ●
- Intercuneiform Fusions . ●
- . Navicular-Cuneiform (NC) Fusion
- Talo-Navicular (TN) Fusion .
- Calcaneo-Cuboid (CC) Fusion ●
- Medial Column Fusion
Asundi
(Division Sign-Off) ﻟﻠﻘ Division of Surgical, Orthopedic, and Restorative Devices
K121651
510(k) Number
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.