PaX-Flex3D is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy on a real time basis by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the anatomic structures by acquiring 360° rotational image sequences of oral and maxillofacial area for a precise treatment planning in adult and pediatric dentistry. The device is operated and used by physicians, dentists, and x-ray technicians.
Device Story
PaX-Flex3D (PHT-7000) is a dental radiographic imaging system providing panoramic, cephalometric, and cone beam computed tomography (CBCT) imaging. The device utilizes an X-ray tube, generator, and CMOS digital X-ray detector to capture 360° rotational image sequences of the oral and maxillofacial area. It is operated by physicians, dentists, and X-ray technicians in clinical settings. The system reconstructs X-ray data from different angles into 3D images or 2D panoramic/cephalometric views. These images are used by healthcare providers for diagnostic assessment and treatment planning, such as oral surgery or implant procedures, potentially improving clinical outcomes through precise anatomical visualization.
Clinical Evidence
No clinical trials were performed. Substantial equivalence was supported by non-clinical performance data, including bench testing for electrical, mechanical, and environmental safety (IEC 60601 series), and image quality assessment of the new sensors compared to the predicate device.
Technological Characteristics
System utilizes CMOS digital X-ray detectors (Xmaru series). Energy source: X-ray generator (50-90 kV, 2-10 mA). Connectivity: DICOM 3.0 compatible. Standards: IEC 60601-1, IEC 60601-1-2 (EMC), IEC 61223-3-4/5 (acceptance testing), NEMA PS 3.1-3.18. Form factor: Compact design. Software: Reconstruction algorithm for 3D CBCT and 2D panoramic/cephalometric imaging.
Indications for Use
Indicated for adult and pediatric patients requiring panoramic, cephalometric, or cross-sectional imaging of the oral and maxillofacial anatomy for treatment planning.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
K121412
# Special 510(k) Summary
. This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 872.1800.
Date
5/2/2012
### Manufacturer
Vatech Co., Ltd.
23-4, Seogu-Dong, Hwaseong-Si, Gyeonggi-Do, 445-170, Korea Republic
Tel: +82-31-379-9585 Fax: +82-31-379-9638 Contact person: Mr. Sonny Park
### Official Correspondent (U.S. Designated agent)
Mtech Group
12946 Kimberley Ln, Houston, TX 77079
Tel: +713-467-2607
Fax: +713-464-8880
Contact person: Mr. Dave Kim (davekim@mtech-inc.net)
#### Trade/Proprietary Name:
PaX-Flex3D (PHT-7000)
#### Common Name:
Dental Computed Tomography X-ray System
### Classification Name:
X-ray, Tomography, Computed, Dental (21CFR 892.1750, Product code OAS, Class2)
{2}------------------------------------------------
Description:
PaX-Flex3D (PHT-7000), a dental radiographic imaging system, consists of dual image acquisition modes; panoramic, cephalometirc and cone beam computed tomography. Specifically designed for dental radiography of the teeth or jaws, PaX-Flex3D (PHT-7000) is a complete dental X-ray system equipped with x-ray tube, generator and dedicated SSXI detector for dental panoramic, cephalometric and cone beam computed tomographic radiography. The dental CBCT system is based on CMOS digital X-ray detector. CMOS CT detector is used to capture radiographic diagnostic images of oral anatomy in 3D for dental treatment such as oral surgery or implant. The device can also be operated as the panoramic and cephalometric dental x-ray system based on CMOS X-ray detector.
#### Indication for use:
PaX-Flex3D is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy on a real time basis by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the anatomic structures by acquiring 360° rotational image sequences of oral and maxillofacial area for a precise treatment planning in adult and pediatric dentistry. The device is operated and used by physicians, dentists, and x-ray technicians.
#### Predicate Device:
| Manufacturer | : Vatech Co., Ltd |
|---------------|---------------------------------------|
| Device | : PaX-Flex-3D |
| 510(k) Number | : K102259 (Decision Date – 2/18/2011) |
#### Substantial Equivalence:
PaX-Flex3D (PHT-7000) described in this 510(k) has the similar intended use and technical characteristics as PaX-Flex3D of Vatech Co., Ltd.
| Characteristic | Proposed<br>Vatech Co., Ltd.<br>PaX-Flex3D (PHT-7000) | Predicate<br>Vatech Co., Ltd.<br>PaX-Flex3D |
|----------------|-------------------------------------------------------|---------------------------------------------|
|----------------|-------------------------------------------------------|---------------------------------------------|
{3}------------------------------------------------
## 510(k) Submission – PaX-Flex3D (PHT-7000)
.
..
| 510(k) number | K121412 | K102259 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | PaX-Flex3D is a computed<br>tomography x-ray system intended<br>to produce panoramic,<br>cephalometric or cross-sectional<br>images of the oral anatomy on a real<br>time basis by computer<br>reconstruction of x-ray image data<br>from the same axial plane taken at<br>different angles. It provides<br>diagnostic details of the anatomic<br>structures by acquiring 360°<br>rotational image sequences of oral<br>and maxillofacial area for a precise<br>treatment planning in adult and<br>pediatric dentistry. The device is<br>operated and used by physicians,<br>dentists, and x-ray technicians. | PaX-Flex3D is a computed<br>tomography x-ray system<br>intended to produce panoramic,<br>cephalometric or cross-sectional<br>images of the oral anatomy on a<br>real time basis by computer<br>reconstruction of x-ray image<br>data from the same axial plane<br>taken at different angles. It<br>provides diagnostic details of the<br>anatomic structures by acquiring<br>360° rotational image sequences<br>of oral and maxillofacial area for<br>a precise treatment planning in<br>adult and pediatric dentistry. The<br>device is operated and used by<br>physicians, dentists, and x-ray<br>technicians. |
| Performance<br>Specification | Panoramic, cephalometric and computed<br>tomography | Panoramic, cephalometric and<br>computed tomography |
| Input Voltage | AC 100-120/200-240 V | AC 110/230 V |
| Tube Voltage | 50-90 kV | 50-90 kV |
| Tube Current | $2 \sim 10$ mA | $4 \sim 10$ mA |
| Focal Spot Size | 0.5 mm | 0.5 mm |
| Exposure Time | $1.9 - 24$ s | $9 - 24$ s |
| Slice Width | 0.1 mm min. | 0.1 mm min. |
| Total Filtration | 2.8 mmAl | 2.8 mmAl |
| Performance<br>Specification | Computed tomography | Computed tomography |
| Mechanical | Compact design | Compact design |
| Electrical | LDCP logic circuit | LDCP logic circuit |
| Software | DICOM 3.0 Format compatible | DICOM 3.0 Format compatible |
| Anatomical<br>Sites | Maxillofacial | Maxillofacial |
| Image Receptor | Computed<br>Tomography<br>(Flat Panel Detector)<br>Xmaru0712CF<br>Xmaru1215CF Plus | Xmaru0808CF |
:
·
Vatech Co., Ltd.
:
·
.
.
.
{4}------------------------------------------------
510(k) Submission – PaX-Flex3D (PHT-7000)
| | | Panoramic<br>(CMOS photodiode<br>array) | <i>Xmaru1501CF</i> | <i>Xmaru1501CF</i> |
|---------------------------|------|-------------------------------------------------|-----------------------------------------------------|---------------------|
| | | Cephalo<br>Metric<br>(CMOS photodiode<br>array) | <i>Xmaru2301CF</i> | <i>Xmaru2301CF</i> |
| Size of Imaging<br>Volume | | <i>Xmaru0712CF</i> | 5 x 5 cm / 8 x 5 cm / 8 x 8 cm | - |
| | | <i>Xmaru1215CF Plus</i> | 5 x 5 cm / 8 x 5 cm /<br>8.5 x 8.5 cm / 12 x 8.5 cm | - |
| | | <i>Xmaru0808CF</i> | - | 5 x 5 cm / 8 x 5 cm |
| Pixel<br>Resolution | CT | <i>Xmaru0712CF</i> | 3.5 lp/mm | - |
| | CT | <i>Xmaru1215CF Plus</i> | 3.5 lp/mm | - |
| | | <i>Xmaru0808CF</i> | - | 3 lp/mm |
| | Pano | <i>Xmaru1501CF</i> | 5 lp/mm | 5 lp/mm |
| | Ceph | <i>Xmaru2301CF</i> | 5 lp/mm | 5 lp/mm |
| Pixel Size | | <i>Xmaru<br/>0712CF</i> | 140 x 140 μm | - |
| | CT | <i>Xmaru<br/>1215CF<br/>Plus</i> | 140 x 140 μm | - |
| | | <i>Xmaru<br/>0808CF</i> | - | 150 x 150 μm |
| | Pano | <i>Xmaru<br/>1501CF</i> | 100 x 100 μm | 100 x 100 μm |
| | Ceph | <i>Xmaru<br/>2301CF</i> | 100 x 100 μm | 100 x 100 μm |
Indications for use, safety characteristics, and non-clinical performance for panoramic, cephalometric and CBCT sensors of PaX-Flex3D (PHT-7000) and PaX-Flex3D are similar. The primary differences are as follows: PaX-Flex3D (PHT-7000) introduces two new cone beam CT sensors:
Vatech Co., Ltd.
{5}------------------------------------------------
#### 510(k) Submission - PaX-Flex3D (PHT-7000)
Xmaru0712CF and Xmaru1215CF Plus. The non-clinical performance and clinical consideration report for the new SSXI CBCT sensors are provided separately in this submission. Based on the non-clinical and clinical consideration and the outcome of an expert review of image comparisons for both devices, new PaX-Flex3D (PHT-7000) is substantially equivalent, in terms of safety and effectiveness, with PaX-Flex3D, the predicate device.
## Safety, EMC and Performance Data:
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1(A1+A2, 1995), IEC 60601-1-1 (2001), IEC 60601-1-3 (Ed. 1, 1994), IEC 60601-2-7 (1998), IEC 60601-2-28 (Ed. 1, 1993), IEC 60601-2-32 (Ed. 1, 1994) and IEC 60601-2-44 (Ed. 2, 2002) were performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2.
PaX-Flex3D (PHT-7000) meets the provisions of NEMA PS 3.1-3.18, Digital Imaging and Communications in Medicine (DICOM) Set.
Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" were performed.
Acceptance test according to IEC 61223-3-4 and IEC 61223-3-5 was performed.
All test results were satisfactory.
#### Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Vatech Co., Ltd. concludes that PaX-Flex3D (PHT-7000) is safe and effective and substantially equivalent to predicate device as described herein.
END
Vatech Co., Ltd.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The seal is black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Vatech Co., Ltd. % Mr. Dave Kim Official correspondent (U.S. designated agent) Mtech group 12946 Kimberly Lane HOUSTON TX 77079
JUN - 8 2012
Re: K121412
Trade/Device Name: PaX-Flex3D (PHT-7000) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: May 11, 2012 Received: May 11, 2012
### Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{7}------------------------------------------------
Enclosure
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{8}------------------------------------------------
# Indications for Use
510(K) Number (if known):
Device Name: PaX-Flex3D (PHT-7000)
Classification: Computed tomography X-ray system
Indications for Use:
PaX-Flex3D is a computed tomography x-ray system intended to produce panoramic, cephalometric or cross-sectional images of the oral anatomy on a real time basis by computer reconstruction of x-ray image data from the same axial plane taken at different angles. It provides diagnostic details of the anatomic structures by acquiring 360° rotational image sequences of oral and maxillofacial area for a precise treatment planning in adult and pediatric dentistry. The device is operated and used by physicians, dentists, and x-ray technicians.
Prescription Use_ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K
Concurrence of CDRH, Office of Device Evaluation(ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.