K121410 · Sedecal., Sa. · IZL · Aug 2, 2012 · Radiology
Device Facts
Record ID
K121410
Device Name
DRAGON LW
Applicant
Sedecal., Sa.
Product Code
IZL · Radiology
Decision Date
Aug 2, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
Device Story
Portable diagnostic X-ray system (Dragon LW models SPSLW4/SPSLW8) for clinical use by physicians or technicians. System comprises X-ray generator, power module, HV tank, collimator, handswitch, and portable column with articulated arm. Integrates with previously cleared Canon digital X-ray receptor panels (CXDI-501, 70C, or 80C). Operates via AC line power; features touchscreen control interface for radiographic operations. Produces diagnostic radiographic images for clinical evaluation. Facilitates bedside or point-of-care imaging, aiding in rapid diagnosis and clinical decision-making for various body parts.
Clinical Evidence
Bench testing only. Performance validated through electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance standards (IEC 60601-1-3, IEC 60601-2-7). Clinical images were acquired and evaluated in accordance with FDA guidance for solid-state X-ray imaging devices to demonstrate diagnostic quality.
Technological Characteristics
Mobile X-ray system; 4kW or 8kW generator; AC line operated. Includes touchscreen control, HV tank, and collimator. Integrates with Canon digital X-ray receptor panels. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-7. Design and materials are identical to predicate devices.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic radiographic imaging of skull, spine, chest, abdomen, and extremities. Patient may be sitting, standing, or recumbent. Contraindicated for mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
DIGITAL PORTABLE X-RAY UNITS, MODEL SP-HF-4.0 D (K042361)
Submission Summary (Full Text)
{0}------------------------------------------------
## Sedecal 510(k) Summary 510(k) Number K121410 Page 1 of 2
l. Submitter: SEDECAL SA C/ Pelaya, 9 -- 13, Pol. Ind. Río de Janciro 28110 Algete, Madrid, España (Spain) Tel .- +34 91 6280544, Fax .- +34 91 6280574 Date Prepared: June 12, 2012 Contact: Mª Luisa Gómez de Agüero, Quality and Regulatory Manager
- 2. Identification of the Device: Proprietary-Trade Name: Sedecal Digital Portable X-ray Units: SPSLW4 and SPSLW8 ("Dragon LW")
Classification Name: Mobile x-ray system, Product Code IZL and Digital X-Ray Receptor Panel Product Code MQB
Common/Usual Name: Portable Diagnostic X-Ray System with Digital X-Ray Receptor
- 3. Equivalent legally marketed device: K103522 SPL-HF-4.0 Portable (Sedecal) AND K042361 DIGITAL PORTABLE X-RAY UNITS, MODEL SP-HF-4.0 D (Sedecal)
- 4. Description of the Device: The Sedecal Dragon LW portable x-ray generator consists of: X-ray Unit with: Control Panel with controls and displays for radiographic operations; Power Module containing control and power components; HV Tank that comprises the High Voltage Transformer, the Filament Transformer and the X-ray Tube; and a Collimator with controls to limit the X-ray beam; a Handswitch and a portable column with an Articulated Arm, a Cassette Basket, and any of the following (previously cleared) Canon Digital X-Ray Receptor Panels: CXDI-501, CXDI-70C, or CXDI-80C. Each of these panels has their own 510(k) clearance. The two generator units are capable of either 4 or 8 kW depending on the model. They are AC line operated. The user interface is the same touchscreen control that was introduced in the K042361 SP-HF-4.0 D Portable.
- 5. Indications for Use (intended use) Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
- 6. Technological Characteristics: This device has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate devices. Specifications are nearly identical. This submission represents the combination of the same generator cleared in K103533 with Canon cleared panels. The Touch Screen Control present in the K042361 Portable has been carried forward into the updated design.
{1}------------------------------------------------
### Sedecal 510(k) Summary K121410 Page 2 of 2
- 7. Discussion of the nonclinical and clinical tests in the premarket notification submission for a determination of substantial equivalence: We performed clectrical safety (IEC 60601-1), electromagnetic compatibility testing, (IEC 60601-1-2), software validation testing, and testing to IEC 60601-1-3 and IEC 60601-2-7. Clinical images were acquired and evaluated based on the FDA Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices."
- 8. Conclusion. Based on the results of the nonclinical tests (that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed devices) we conclude that this modified x-ray system is safe and effective as the predicate identified in paragraph (3). Furthermore, the materials and construction methods are nearly identical to the predicate.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
> 2 2012 AUG
SEDECAL SA % Daniel Kamm, P.E. Submission Correspondent Kamm & Associates 8870 Ravello Court NAPLES FL 34114
Re: K121410
Trade/Device Name: Sedecal Digital Portable X-ray Units: SPSLW4 and SPSLW8
("Dragon LW") Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL and MOB Dated: July 13, 2012 Received: July 30, 2012
#### Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r earlie, ontrols provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (24 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. ,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
# Indications for Use
#### ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Number (if known): K12
Device Name: Sedecal Digital Portable X-ray Units: SPSLW4 and SPSLW8 ("Dragon LW")
#### Indications For Use:
These Portable Diagnostic Radiographic Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. (Not for mammography)
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-vitro Diagnostics (OIVD)
(Division Sign-Off)
(Division of Radiological Devices)
Division of Radiological Devices
510k K12498
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.