PS X SERIES (BASIC COMPONENTS AND CATHETER), PD X SERIES (CATHETER WITH ALL COMPONENTS)
K121383 · Medical Components, Inc. · FJS · Sep 5, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K121383
Device Name
PS X SERIES (BASIC COMPONENTS AND CATHETER), PD X SERIES (CATHETER WITH ALL COMPONENTS)
Applicant
Medical Components, Inc.
Product Code
FJS · Gastroenterology, Urology
Decision Date
Sep 5, 2012
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The "X" Series PD Catheter is indicated for peritoneal dialysis in adults. It is intended for use in patients who are poor candidates for conventional peritoneal dialysis catheters and are candidates for presternal insertion.
Device Story
The "X" Series PD Catheter is a peritoneal dialysis access device consisting of an abdominal silicone rubber catheter with a coiled tip and single polyester cuff, connected via a titanium double-barbed connector to a subcutaneous silicone rubber extension catheter with a double polyester cuff and preformed tubing arc. The device is implanted by a physician to provide peritoneal access while allowing the skin exit site to be located remotely from the lower abdomen. It is supplied in kits containing insertion tools (dilator/sheath, guidewire, scalpel, tunneling tool). The device facilitates chronic peritoneal dialysis in patients who are poor candidates for conventional catheters. It functions as a conduit for dialysis fluid exchange.
Clinical Evidence
Bench testing only. Performance testing included air/liquid leakage, gravity flow, force at break, pull force, priming volume, site care agent effects, cuff shear, and tip separation force. Biocompatibility testing met ISO 10993 requirements.
Technological Characteristics
Materials: Silicone rubber, polyester cuffs, titanium connector. Design: Coiled tip abdominal segment, subcutaneous extension with preformed arc. Sterilization: Not specified. Connectivity: None. Software: None.
Indications for Use
Indicated for chronic peritoneal dialysis in adults. Applicable for patients requiring remote skin exit sites, including those with obesity, floppy abdominal skin folds, urinary/fecal incontinence, chronic yeast intertrigo, intestinal stomas, or those desiring deep tub baths.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
CT Power Injectable implantable Infusion Port (K110424)
Submission Summary (Full Text)
{1}------------------------------------------------
K121383
Page 12
5 2012 SEP
ﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
#### Section 5 510(k) SUMMARY Traditional 510K Submitter Information: Submitter: MEDCOMP® 1499 Delp Drive Harleysville, PA 19438 (215) 256-4201 Telephone (215) 256-9191 Fax Contact: Rosanna Severini, Compliance Director Date Prepared: August 10, 2012 "X" Series PD Catheter Device Name: Common Name: Peritoneal Dialysis Catheter Classification Name: Peritoneal Dialysis Catheter & Accessories C.F.R. Section: 876.5630 Classification Panel: Gastroenterology and Urology Class: II. 78 FJS Predicate Devices: K071167, Flex-Neck® ExxTended™ PD Catheter & Accessories, concurrence date Primary:
August 31, 2007. Class II CFR §876.5630 K950042, Swan Neck Presternal Peritoneal Dialysis Catheter, concurrence date December 15, 1995. Class II CFR \$876.5630 K915490, Peritoneal Dialysis Catheter, concurrence date February 2, 1994. Class II CFR §876.5630 Secondary: K113487: Hemo-Cath K110424: CT Power Injectable implantable Infusion Port
# Device Description:
The "X" Series PD Catheter is comprised of a silicone rubber, coiled tip, single polyester cuff abdominal catheter segment that is joined with a titanium double bared connector to a silicone rubber, double polyester cuff subcutaneous extension catheter possessing a preformed tubing arc bend between the two cuffs. The abdominal catheter segment is implanted into the peritoneal cavity. The attached subcutaneous extension piece allows remote location of the catheter skin exit site away from the lower abdominal region.
The catheter will be sold in two kit configurations, as the catheter only with a female luer, clamp, and male luer end cap and titanium connector. The catheter set will also include the addition of a dilator/sheath introducer, guidewire, scalpel, introducer needle, tunneling tool, gauze pads and syringe.
#### Intended Used:
The "X" Series PD Catheter is indicated for peritoneal dialysis in adults. It is intended for use in patients who are poor candidates for conventional peritoneal dialysis catheters and are candidates for presternal insertion.
#### Indications for Use:
The "X" Series Catheter is indicated for chronic peritoneal dialysis.
{2}------------------------------------------------
K12 1383
The extended length of the “X” Series Catheter makes it especially applicable for peritoneal dialysis
patients when it is necessary to locate the skin exit site remote from the usual lower abdominal
region. The catheter may be particularly indicated in patients with obesity, floppy abdominal skin
folds, urinary or fecal incontinence, chronic yeast intertrigo, intestinal stomas, or in patients who
desire to take deep tub baths.
# Comparison to Predicate Devices:
The "X" PD Series Catheter is substantially equivalent to the predicate devices in terms of intended use, materials, anatomical location, basic design, performance, labeling, manufacturing process and method of sterilization.
### Performance Standards:
Performance standards have not been established by the FDA under section 514 of the Federal Food, Drug and Cosmetic Act.
## Performance Testing:
- Air Leakage Liquid Leakage Gravity Flow Force at Break Pull Force Testing Priming Volume Effects of Site Care Agents Cuff Shear Testing Tip Separation Force
#### Biocompatibility:
Results for all biocompatibility testing demonstrate the materials used meet the requirements of ISO 10993. All materials have been cleared under past approved 510K's.
#### Technological Characteristics:
The principles of operation are the same as the predicate devices. There are no new questions raised regarding the safety or effectiveness of the device.
#### Summary of Substantial Equivalence:
The proposed device meets the performance criteria of design verification as specified by ISO standards, guidance documents and test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed device is substantially equivalent to the legally marketed predicate devices.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three stripes representing the department's mission. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP -5 2012
Ms. Rosanna Severini Compliance Director Medcomp (Medical Components) 1499 Delp Drive HARLEYSVILLE PA 19438
Re: K121383
Trade/Device Name: "X" Series PD Catheter Regulation Number: 21 CFR§ 876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: FJS Dated: August 24, 2012 Received: August 24, 2012
Dear Ms. Severini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hulda Reimer
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K/2/383
"X" Series PD Catheter __ Device Name:
Indications for Use:
The "X" Series Catheter is indicated for chronic peritoneal dialysis.
The extended length of the "X" Series Catheter makes it especially applicable for peritoneal dialysis patients when it is necessary to locate the skin exit site remote from the usual lower abdominal region. The catheter may be particularly indicated in patients with obesity, floppy abdominal skin folds, urinary or fecal incontinence, chronic yeast intertrigo, intestinal stomas, or in patients who desire to take deep tub baths.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hubert Lehman
vision Sign-Off ision of Reproductive, Gastro-Renal, and rological Devices 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.