K120948 · The Yoshida Dental Mfg. Co., Ltd. · OAS · Dec 21, 2012 · Radiology
Device Facts
Record ID
K120948
Device Name
PREXION 3D
Applicant
The Yoshida Dental Mfg. Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
PreXion 3D is intended to produce two-dimensional and three-dimensional digital X-ray images of the dento-maxillo-facial regions at the direction of healthcare professionals as diagnostic support.
Device Story
PreXion 3D is a dental computed tomography (CT) and panoramic X-ray system. It uses an X-ray tube head and flat panel detector to capture X-ray absorption data from multiple directions. The system includes a scanner, a console computer, and software for image processing and analysis. It generates 2D panoramic images and 2D/3D CT images of hard tissues (bones, teeth) in the jaw and mouth cavity. Operated by healthcare professionals in a clinical setting, the device provides diagnostic support through cross-section views and image editing. Clinicians use these images to visualize anatomical structures, aiding in diagnosis and treatment planning. The system benefits patients by providing high-resolution imaging for dental and maxillofacial assessment.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included Modulation-Transfer Function (MTF) per IEC 61223-3-5 to evaluate spatial resolution, metal artifact analysis compared to the FINECUBE predicate, and electrical/radiation safety testing per IEC 60601 series standards.
Indicated for healthcare professionals to produce 2D and 3D digital X-ray images of the dento-maxillo-facial region for diagnostic support.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
# DEC 2 1 2012
## 510(k) Summary
- a. Owner/Company name, address THE YOSHIDA DENTAL MFG CO., LTD. 1-3-6, Kotobashi, Sumida-ku Tokyo 130-8516, Japan
Michizo Yamanaka President
· Contact person Hidenori Watanabe Regulatory Affairs Phone: 011-81-3-3631-2165 Fax: 011-81- 3-3633-9420 Email: hi-watanabe@yoshida-net.co.jp
## b. Contact/Application Correspondent
Fumiaki Kanai, Ph.D. President and CEO MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
011-81-3-3818-8577 Phone: 011-81-3-3818-8573 Fax: Email: kanaif(@mici.co.jp
- c. Date prepared March 28, 2012
## d. Name of device
PREXION 3D Trade Name: Computed tomography x-ray system Common Name: Classification Name: X-ray, tomography, computed, dental Classification Regulation: 21 CFR 892.1750
YOSHIDA DENTAL MFG CO., LTD. NOVEMBER 9, 2012
ATTACHMENT 1 A 1-12 OF A 1-17
{1}------------------------------------------------
## e. Predicate devices
The PREXION 3D is substantially equivalent to the following legally marketed device: .
| 510(k) Number | Trade name | Product code |
|---------------|------------|--------------|
| K063622 | FINECUBE | OAS |
| K103659 | CS 9300 | OAS |
The predicate devices are hereinafter called "the FINECUBE (k063622)" or "CS 9300 (k103659)" in this application.
## f. Description of the device
The PREXION 3D consists of scanner and two software including Console software, and Viewer software used for the Image Analysis System and Data processing. A qualified computer named Console computer is distributed with the PREXION 3D.
The PREXION 3D uses the Image Analysis System and the processed data acquired by the scanner to analyze 2D and 3D images, perform image edition, such as creating cross-section views, and output results to a printer or other output device.
During scanning, X-rays are generated from the X-ray tube head mounted in the arm of the scanner and the X-rays passing through a patient are then detected by the flat panel detector of the scanner under the control of the firmware and the Console software installed on the qualified computer. The detected X-ray absorption data is processed by the Console software and viewer software on a computer to reconstruct images. Scanning is performed using X-ray penetration signals of a patient taken from multiple directions for the diagnosis of hard tissue including bones and teeth of the jaw and mouth cavity. The PREXION 3D performs two types of scanning including CT scan generating two (2) and three (3) dimensional images and Panoramic scan generating two (2) dimensional images.
## g. Indications for Use
The PREXION 3D is intended to produce two-dimensional and three-dimensional digital X-ray images of the dento-maxillo-facial region at the direction of healthcare professionals as diagnostic support.
### h. Statement of substantial equivalence
The PREXION 3D was developed from the FINECUBE (k063622) by adding panoramic scan function. Accordingly, the characteristics of the PREXION 3D are identical or similar to those of the FINECUBE (k063622) regarding X-ray Generation Device characteristics including tube voltage, tube current, and Focal Spot Size, X-ray image capturing device including type of detector, Pixel size, pixel number. (See Table 6-1 Comparison Table).
The only difference of the PREXION 3D from the FINECUBE (k063622) is addition of panoramic scanning function. The panoramic scanning function of the PREXION 3D is similar to CS 9300 (k103659). Both the PREXION 3D and CS 9300 (k103659) have the panoramic scanning function in addition to CT scan function. The PREXION 3D has the similar characteristics regarding intended use, X-ray Generation Device characteristics
YOSHIDA DENTAL MFG CO., LTD. NOVEMBER 9, 2012
ATTACHMENT 1 A1-13 OF A1-17
{2}------------------------------------------------
including tube voltage, tube current, and Focal Spot Size, X-ray image capturing device including type of detector, pixel number as CS 9300 (k103659).
By considering above device characteristics, the intended use of the PREXION 3D is identical to part of intended use of the CS 9300 (k103659), and similar to the intended use of the FINECUBE (k063622).
In order to ensure same performance characteristics as predicate devices, software verification/validation, performance testing, and risk analysis were performed. Such test results and risk analysis indicate that the PREXION 3D meets the requirements of the recognized consensus or voluntary standard. Based on the information presented above we conclude that the PREXION 3D is substantially equivalent to the predicate devices and does not raise any new questions regarding safety or effectiveness.
#### Comparison table i.
Table 6-1 compares the characteristics between the PREXION 3D and the predicates.
{3}------------------------------------------------
ON 3D (K120948)
ONSE TO FDA LETTER DATED MAY
x 149 mm max (Panoram x 1536 pixels (Panorar PD (TFT ર0 - 90k પ 2 - 15m/ 0.7mm l 4 bits .6mm x 123.2m 608 x 610 0.2mm 200μm l 2bits 90k V 4mA FPD Xion sional pisal managalad to provinsi
Adimensional manakal manakaran para na mana mana mana marka marin
sa of the villaga marka mana mana mana mana mana mana mana mana man 123.2mm x 121.6mm (CT)
mm x 131.6mm (Panoram 616 x 608 (CT)
216 x 72 (Panorami 200μm (CT)
0μm (Panorami 0.2mm 90k V 4bits 4mA FPD Tube Current
Tocal Spot Size Number of B ube Voltag ixel numbe Size of Area
receiving receiving
X-ray le 6-1. Comparison Ts ixel size Detector ications for l X ray Generatic
Device K ray image
capturing devi
SHIDA DENTAL MFG CO.,
VEMBER 9, 2012
ATTACHMENT
A I-15 OF A I-1
{4}------------------------------------------------
XION 3D (K120948)
K... SPONSE TO FDA LETTER DATED MAY 16, 2012
| | PREXION 3D | FINECUBE (k063622) | CS 9300 (k103659) | |
|-------------------------------|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Scanner | SID/SOD | | | |
| | 700mm/468mm<br>(Hi-Def/Ultra Hi-Def/Rapid Scan)<br>700mm/322mm<br>(Hi-Res) | 700mm/322mm(Magnification<br>imaging)<br>700mm/468mm(Wide area<br>imaging) | 615mm (SID) | |
| | Dimension<br>(W×D×H) | 1170mm×1530mm×1930mm | 1158mm×1595mm×2378mm | |
| | Weight | 390kg | 390kg | 160kg |
| Scan mode | | CT Scan and Panoramic Scan | CT Scan | CT Scan and Panoramic Scan |
| Panoramic scan<br>Performance | Scan time | 16.2 sec | | 4 - 16 sec |
| | Scan time | 8.6 sec<br>(Rapid Scan)<br>16.8 sec<br>(Hi-Def/Hi-Res)<br>33.5 sec<br>(Ultra Hi-Def/Ultra Hi-Res) | 19sec. (Standard mode)<br>37sec. (High density mode) | 12 - 20 sec<br>28 sec |
| CT scan<br>Performance | | φ81mm/H74mm<br>(Hi-Def/Ultra Hi-Def/Rapid Scan)<br><br>φ56mm/H51mm<br>(Hi-Res/Ultra Hi-Res) | Wide area imaging<br>φ82.0mm H75.1mm<br><br>Magnification imaging<br>φ56.5mm, H51.7mm | φ170mm/H135mm<br>(0.090mm - 0.500mm)<br><br>φ170mm/H110mm<br>(0.090mm-0.500mm)<br><br>φ170mm/H60mm<br>(0.090mm - 0.500mm)<br><br>φ100mm/H100mm<br>(0.090mm - 0.500mm)<br><br>φ80mm/H80mm<br>(0.090mm - 0.500mm)<br><br>φ100mm/H50mm<br>(0.090mm - 0.500mm)<br><br>φ50mm/H50mm<br>(0.090mm-0.500mm) |
| | FOV<br>(voxel size) | | | |
ATTACHMENT ]
A1-16 OF A1-17
YOSHIDA DENTAL MFG CO., LTD
NOVEMBER 9, 2012
{5}------------------------------------------------
#### j. Risk Analysis
The PREXION3D was evaluated in accordance with ISO14971:2007. The risk management of the device was deemed satisfactory.
### k. Bench Testing
THE YOSHIDA DENTAL MFG CO., LTD has performed bench tests to ensure safety and effectiveness as follows;
## 1. Laser safety
The laser system of the PREXION3D is identical to that of the PANOURA 18S (K111231). Therefore, the test report for IEC 60825-1 for the PANOURA 18S (K111231) is used as the test report for the PREXION3D.
## 2. Modulation-Transfer Function
In order to evaluate the spatial resolution of the PreXion3D, we measured the MTF in accordance with IEC 61223-3-5. The spatial resolution of all scan modes met the acceptance criteria.
## 3. Artifact Analysis
In order to evaluate the artifact of the image of the PREXION 3D, the images of all scan mode of the PREXION 3D were compared to those of the FineCube (K063622). There was no difference of pattern and strength of the metal artifact between the PREXION 3D and the FineCube(K063622).
The software of the PREXION3D has been validated according to "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
EMC, Electric safety, and X-ray radiation safety are confirmed in accordance with IEC60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32.
#### I. Conclusion
The PREXION 3D has similar intended use and technical characteristics as the predicate devices including the FINECUBE (k063622) and CS 9300 (k103659). A number of test results and risk analysis indicate that the PREXION 3D meets the requirements of the recognized consensus or voluntary standard. Based on those information, we conclude that the PREXION 3D is substantially equivalent to the predicate devices and does not raise any new questions regarding safety or effectiveness.
YOSHIDA DENTAL MFG CO., LTD. NOVEMBER 9, 2012
· ATTACHMENT 1 A1-17 OF A1-17
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in black.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 21, 2012
The Yoshida Dental MFG Co., LTD. % Dr. Fumiaki Kanai President and CEO MIC international 4-1-17 Hongo, Bunkyo-ku TOKYO, 113-0033, JAPAN
Re: K120948
Trade/Device Name:PREXION 3D Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: Class II Product Code: OAS Dated: November, 09, 2012 Received: November 13, 2012
Dear Dr. Kanai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 - Dr. Kanai
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
## Janine M. Morris -S
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K120948
Device Name: PreXion 3d
Indications For Use:
PreXion 3D is intended to produce two-dimensional and three-dimensional digital X-ray images of the dento-maxillo-facial regions at the direction of healthcare professionals as diagnostic support.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Janine M. Morris -S 2012.12.21 10:48:57 -05'00'
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.