PreXion3D Excelsior is intended to produce two dimensional digital panoramic and cephalometric images, and three dimensional digital x-ray images of the maxillofacial, and ENT (Ear, Nose and Throat) region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients. Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
Device Story
PreXion3D Excelsior is a dental cone-beam computed tomography (CBCT) system. It captures X-ray data via a rotating arm with an X-ray tube and flat panel detector (FPD). The system operates in CT, panoramic, and cephalometric modes. Data is transferred to a console computer where firmware and software reconstruct diagnostic 2D/3D images. Used in clinical settings by healthcare professionals for diagnostic support in maxillofacial and ENT regions. Output includes cross-sectional diagrams and images for printing/viewing. The device aids clinicians in treatment planning and diagnosis by providing high-resolution anatomical visualization.
Clinical Evidence
No formal clinical trial data provided. Safety and effectiveness supported by non-clinical bench testing, including MTF assessment for 3D imaging performance, and clinical considerations performed by qualified assessors in accordance with FDA guidance for solid-state X-ray imaging devices.
Technological Characteristics
Dental CBCT system; X-ray generation (60-110kV, 1-6mA); 0.3mm focal spot; FPD (TFT) detector; 16-bit (CT/Panorama) and 14-bit (Ceph) depth. Connectivity via console computer. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2/1-3/1-6/2-63, IEC 62366, IEC 62304, IEC 61223-3-4/3-5, IEC 60825-1, ISO 14971, NEMA PS 3.1-3.20, ISO 10993-1/5/10.
Indications for Use
Indicated for adult and pediatric patients requiring 2D panoramic, 2D cephalometric, or 3D maxillofacial and ENT diagnostic imaging. Cephalometric imaging includes hand/wrist carpus imaging for growth/maturity assessment.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
i-CAT FLX Cone Beam 3D and 2D Panoramic Dental Imaging System (K133620)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird design formed by three overlapping profiles of human faces. The faces are oriented to the right and are stacked on top of each other. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 3, 2016
PreXion Corporation % Mr. Shumpeita Torii Director Funai Tokyo Technology Center Building. 1-14-1 Kandasuda-cho Chiyoda-ku, Tokyo 101-0041 JAPAN
Re: K161881
Trade/Device Name: PreXion3D Excelsior Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: September 8, 2016 Received: September 9, 2016
Dear Mr. Torii:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ocks
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K161881
Device Name PreXion3D Excelsior
Indications for Use (Describe)
PreXion3D Excelsior is intended to produce two dimensional digital panoranic and cephalometric images, and three dimensional digital x-ray images of the maxillofacial, and ENT (Ear, Nose and Throat) region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
Cephalometric imaging also includes the hand wrist to obtain carpus innages for growth and maturity assessment.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
510(k) Number: K161881
# 5.1 Applicant Information
| Date Prepared: | Sep. 8th, 2016 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name and Address: | PreXion Corporation<br>Funai Tokyo technology center building,<br>1-14-1 Kandasuda-cho, Chiyoda-ku<br>Tokyo, 101-0041<br>Japan |
| Contact Person: | Mr. Shumpeita Torii<br>Director, Quality Assurance and Regulatory Division<br>Phone: +81-3-5297-7551<br>FAX: +81-3-5297-7552<br>Email: torii@prexion.co.jp |
# 5.2 Device Information
| Device Type: | Dental Cone-beam Computed Tomography |
|-------------------------|--------------------------------------|
| Regulation Description: | Computed Tomography X-Ray System |
| Review Panel: | Radiology |
| Regulation Number: | 21 CFR 892.1750 |
| Product Code: | OAS |
| Device Class: | II |
| Device Name: | Prexion3D Excelsior |
# 5.3 Predicate Devices
The legally marketed devices to which substantial equivalence is being claimed are:
| 510(k) Number: | K122199 | K103659 | K133620 |
|--------------------|-----------------------------------------|-------------------------|---------------------------------------------------------------------|
| Applicant: | THE YOSHIDA<br>DENTAL MFG. CO.,<br>LTD. | Trophy | Imaging Sciences<br>International, LLC |
| Device Name: | PREXION 3D<br>ECLIPSE | CS 9300 and CS<br>9300C | i-CAT FLX Cone Beam<br>3D and 2D Panoramic<br>Dental Imaging System |
| Regulation Number: | 21 CFR 892.1750 | 21 CFR 892.1750 | 21 CFR 892.1750 |
| Product Code: | OAS | OAS | OAS |
| Device Class: | II | II | II |
{4}------------------------------------------------
# 5.4 Device Description
PreXion3D Excelsior consists of a scanner, which is used for generating X-ray and detecting image data, and a Console, which is used for operating the scanner and managing the data. The scan data acquired by the scanner will be transferred to the Console. PreXion3D Excelsior Image Analysis System will then perform the image analysis (2D/3D) or image edition (creating cross-section diagram, etc.), and output the image to a printer.
During scanning, X-rays are generated from the x-ray tube head mounted in the arm of the scanner and the x-rays passing through a patient are then detected by the flat panel detector of the scanner under the control of the firmware inside and the console software installed on the qualified Computer. The detected x-ray absorption data is processed by the console software to reconstruct the diagnostic images. The PreXion3D Excelsior has three operation modes, CT scan, Panoramic scan and Cephalometric exposure.
# 5.5 Intended Use/Indications for Use
PreXion3D Excelsior is intended to produce two dimensional digital panoramic and cephalometric images, and three dimensional digital x-ray images of the maxillofacial, and ENT (Ear, Nose and Throat) region at the direction of healthcare professionals as diagnostic support for adult and pediatric patients.
Cephalometric imaging also includes the hand and wrist to obtain carpus images for growth and maturity assessment.
# 5.6 Technological Characteristics
The subject device compares to the legally marketed devices as follows:
| | Subject Device | Predicate device | | |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | PreXion3D<br>Excelsior | PREXION 3D<br>ECLIPSE<br>(K122199) | CS 9300 and CS<br>9300C<br>(K103659) | i-CAT FLX Cone<br>Beam 3D and 2D<br>Panoramic<br>Dental Imaging<br>System<br>(K133620) |
| Device | | | | |
| Intended<br>Use/Indications<br>for Use | PreXion3D<br>Excelsior is<br>intended to<br>produce two<br>dimensional digital<br>panoramic and<br>cephalometric<br>images, and three<br>dimensional digital<br>x-ray images of<br>the maxillofacial,<br>and ENT (Ear,<br>Nose and Throat)<br>region at the | PREXION3D<br>ECLIPSE is<br>intended to<br>produce two-<br>dimensional digital<br>panoramic and<br>cephalometric<br>images, and three-<br>dimensional digital<br>x-ray images of<br>the dento-maxillo-<br>facial region at the<br>direction of<br>healthcare | The CS 9300 and<br>CS 9300C are<br>systems intended<br>to produce two-<br>dimensional and<br>three-dimensional<br>digital x-ray<br>images of the<br>dento-maxillo-<br>facial, and ENT<br>(Ear, Nose and<br>Throat) regions at<br>the direction of<br>healthcare | Devices of the i-<br>CAT family consist<br>of an x-ray system<br>that uses a cone<br>beam with a<br>rotational<br>sequence,<br>providing two<br>dimensional<br>images and three<br>dimensional<br>volume<br>reconstructions of<br>the head area, |
{5}------------------------------------------------
| Device | Subject Device | PreXion3D<br>Excelsior | PREXION 3D<br>ECLIPSE<br>(K122199) | Predicate device | CS 9300 and CS<br>9300C<br>(K103659) | i-CAT FLX Cone<br>Beam 3D and 2D<br>Panoramic<br>Dental Imaging<br>System<br>(K133620) |
|-------------------------------------------------|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| | | direction of<br>healthcare<br>professionals as<br>diagnostic support<br>for adult and<br>pediatric patients.<br>Cephalometric<br>imaging also<br>includes the hand<br>and wrist to obtain<br>carpus images for<br>growth and<br>maturity<br>assessment. | professionals as<br>diagnostic<br>support. | professionals as<br>diagnostic support<br>for pediatric and<br>adult patients.<br>In addition, the CS<br>9300C is intended<br>to produce<br>cephalometric<br>images. This<br>includes imaging<br>the hand and wrist<br>to obtain the<br>carpus image for<br>growth and<br>maturity<br>assessment. | which includes<br>ENT and<br>maxillofacial areas<br>(such as TM Joint<br>studies, mandible<br>& maxilla for<br>implant planning,<br>sinuses), for use<br>in planning and<br>diagnostic support<br>in adult and<br>pediatric care.<br>Devices of the i-<br>CAT family<br>comprise a<br>package of<br>software modules<br>capable of<br>handling 2D and<br>3D data. This<br>includes 3D<br>reconstruction,<br>storage, retrieval,<br>viewing, and<br>processing of 2D<br>and 3D-image<br>data. | |
| X-ray<br>Gene<br>ration<br>Device | Tube<br>Voltage | 60-110KV | 50-90 kV | 60 - 90 kV | 84-120 kV<br>94 kV (Panoramic)<br>84 kV (Panoramic<br>small) | |
| | Pulse<br>Exposu<br>re<br>function | Yes | N/A | Yes | Yes | |
| | Tube<br>Current | 1-6mA | 2.6-4 mA | 2 - 15 mA | 3-7mA<br>5mA (Panoramic<br>and Panoramic<br>small) | |
| | Focal<br>Spot<br>Size | 0.3mm | 0.2 mm | 0.7 mm | 0.5mm | |
| X-ray<br>Imag<br>e | Detector | FPD (TFT) | FPD (CMOS) | FPD (TFT) | FPD (TFT) | |
| | Pixel<br>Size | 125 μm | 200 μm (CT) | 127 μm | 127um x 127um | |
| Device | Subject Device | PreXion3D Excelsior | PREXION 3D ECLIPSE (K122199) | Predicate device CS 9300 and CS 9300C (K103659) | i-CAT FLX Cone Beam 3D and 2D Panoramic Dental Imaging System (K133620) | |
| uring Devic e | | x250μm(CT)<br>125 μm x 125μm(Panoramic )<br>140 μm x 140μm (Ceph) | 100 μm (Panoramic)<br>54 μm (Ceph) | | | |
| | Pixel Number | 1280x1024 (CT)<br>100x1280 (Panoramic)<br>2112x1754 (Cephalometric) | 640x656 (CT)<br>80x1312 (Panoramic)<br>128x4080 (Ceph) | 64 x 1536 (Panoramic) | 1536 x 1920<br>768 x 960 | |
| | Size of Area Receiving X-ray | 160mm x 128mm (CT)<br>160mm x 12.5mm (Panoramic)<br>295.68 x 245.56mm (Ceph) | 128.1 mm x 131.3 mm (CT)<br>8 mm x 131.3 mm (Panoramic)<br>6.9 mm x 312 mm (Ceph) | 5 x 149 mm max (Panoramic) | 242mm x 192mm (CT)<br>10mm x 100mm (Panoramic) | |
| | Number of Bits | 16bits (CT, Panorama)<br>14bits (Ceph) | 14 bits (CT, Panoramic)<br>16 bits (Ceph) | 14 Bits | 16 bits | |
| | SID/SOD | 700mm/ 470mm (CT, Panoramic)<br>1650mm / 1500mm (Ceph) | 620 mm / 400 mm (CT, Panoramic)<br>1650 mm / 1500 mm (Ceph) | 615 mm (SID) | 714mm / 495mm | |
| Scan ner | Dimension (WxDxH) | 930 mm x 1230 mm x 2220 mm (CT, Panoramic)<br>1747 mm x 1230 mm x 2220 mm (with Ceph) | 1245 mm x 1288 mm x 2045 mm (CT, Panoramic)<br>1805 mm x 1288 mm x 2045 mm (Ceph) | 1158 mm x 1595 mm x 2378 mm | 1222mm x 1340mm x 1820mm | |
| | Device | Subject Device<br>PreXion3D<br>Excelsior | PREXION 3D<br>ECLIPSE<br>(K122199) | Predicate device<br>CS 9300 and CS<br>9300C<br>(K103659) | i-CAT FLX Cone<br>Beam 3D and 2D<br>Panoramic<br>Dental Imaging<br>System<br>(K133620) | |
| | Weight | 165 kg (CT,<br>Panoramic)<br>200kg (Ceph) | 260 kg (CT,<br>Panoramic)<br>300 kg (Ceph) | 160 kg (CT,<br>Panoramic)<br>190 kg (Ceph) | 231kg | |
| Imaging Mode | | CT scan,<br>Panoramic scan,<br>Cephalometric<br>radiography | CT scan,<br>Panoramic scan,<br>Cephalometric<br>radiography | CT scan,<br>Panoramic scan,<br>Cephalometric<br>radiography | CT scan,<br>Panoramic scan | |
| Panoramic<br>Scan<br>Performance<br>(Scan Time) | | 4-16sec | Standard mode:<br>16 sec | 4 - 16 sec | 18.3 -20.0 sec | |
| Cephalometric<br>Radiography<br>(Scan Time) | | 0.5-0.8 sec | LA, PA, Carpus:<br>8, 10, 12, 15 sec | 0.1 - 3.2 sec | N/A | |
| CT<br>Scan<br>Perfo<br>rman<br>ce | Scan<br>Time | 6.9-25.6sec | Light mode: 8.7<br>sec<br>High Definition<br>mode: 8.7 sec<br>Ultra High<br>Definition mode:<br>17.4 sec<br>Wide mode: 9.1<br>sec x 2 | 12 - 20 sec<br>28 sec | 4.8 - 26.9 sec | |
| | FOV<br>(Voxel<br>Size) | Diameter 100mm<br>x H76mm<br>(0.100 - 0.200mm)<br>Diameter 150mm<br>x H75 (0.200mm)<br>Diameter 100mm<br>x H50mm<br>(0.100 - 0.200mm)<br>Diameter 50mm x<br>H50mm<br>(0.100 - 0.200mm) | Light mode, High<br>Definition mode,<br>Ultra High<br>Definition mode:<br>Diameter 81 mm,<br>H 75 mm<br>Wide Mode:<br>Diameter 113 mm,<br>H 72 mm | Diameter 170 mm/<br>H135 mm (0.090<br>mm - 0.500 mm)<br>Diameter 170 mm/<br>H110 mm (0.090<br>mm - 0.500 mm)<br>Diameter 170 mm/<br>H60 mm (0.090<br>mm - 0.500 mm)<br>Diameter 100 mm/<br>H100 mm (0.090<br>mm - 0.500 mm) | Diameter x Height<br>8 cm x 8 cm<br>16 cm x 4 cm<br>16 cm x 6 cm<br>16 cm x 8 cm<br>16 cm x 10 cm<br>16 cm x 11 cm<br>16 cm x 13 cm<br>23 cm x 17 cm<br>(3D ceph)<br>Voxel size: 0.125,<br>0.2, 0.25, 0.3,<br>0.4mm | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| | Subject Device | Predicate device | | | |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Device | PreXion3D<br>Excelsior | PREXION 3D<br>ECLIPSE<br>(K122199) | CS 9300 and CS<br>9300C<br>(K103659) | i-CAT FLX Cone<br>Beam 3D and 2D<br>Panoramic<br>Dental Imaging<br>System<br>(K133620) | |
| Volume Size | 512x512x512<br>1024x1024x800 | 512x512x512 | 250 x 250 x 250<br>267 x 267 x 267<br>339 x 339 x 220<br>567 x 567 x 367-<br>450<br>850 x 850 x 300 | 320x320x288<br>536x536x443<br>768x768x288 | |
| Dose Level<br>(CTDIw) | Standard (110KV,<br>3.3mA, 12.8s)<br>10cm: 2.9mGy<br>5cm: 1.9mGy<br><br>Rapid (110KV,<br>2.1mA, 6.9s)<br>10cm: 0.9mGy<br>5cm: 0.8mGy<br><br>High Resolution<br>(110KV, 3.6mA,<br>25.6s)<br>10cm: 4.9mGy<br>5cm: 3.2mGy<br><br>High Contrast<br>(110KV, 2.1mA,<br>25.6s)<br>10cm: 8.7mGy<br>5cm: 5.7mGy<br><br>Standard Child<br>(95KV, 2.2mA,<br>12.8s)<br>10cm: 1.4mGy<br>5cm: 1.0mGy<br><br>Rapid Child | High Definition<br>(90KV, 4mA, 8.7s)<br>8cm: 4.57mGy<br><br>Ultra High<br>Definition (90KV,<br>4mA, 17.4s)<br>8cm: 9.32mGy<br><br>Wide (90KV, 4mA,<br>9.1s x 2)<br>11cm: 6.67mGy<br><br>Light (90KV,<br>2.6mA, 8.7s)<br>8cm: 3.01mGy | 90KV, 6.3mA,<br>12sec, 8cm x 8cm:<br>4.31mGy<br><br>90KV, 6.3mA,<br>12sec, 5cm x 5cm:<br>3.70mGy<br><br>90KV, 6.3mA,<br>20sec, 5cm x 5cm:<br>6.46mGy | 3D Ceph (120KV,<br>5mA, 8.9s)<br>13cm: 1.9mGy*,<br>11cm: 1.9mGy*,<br>10cm: 1.9mGy*,<br>8cm: 2.0mGy*,<br>6cm: 2.0mGy*,<br>4cm: 2.1mGy*<br><br>Quick Scan<br>(120KV, 5mA,<br>4.8s)<br>13cm: 1.0mGy*,<br>11cm: 1.0mGy*,<br>10cm: 1.0mGy*,<br>8cm: 1.0mGy*,<br>6cm: 1.0mGy*,<br>4cm 1.6mGy*:<br><br>Quick Scan+<br>(90KV, 3mA, 4.8s)<br>13cm: 0.3mGy*,<br>11cm: 0.3mGy*,<br>10cm: 0.3mGy*,<br>8cm: 0.3mGy*,<br>6cm: 0.3mGy*,<br>4cm: 0.3mGy*<br><br>HD (120KV, 5mA, | |
| | Subject Device | | Predicate device | | |
| Device | PreXion3D<br>Excelsior | PREXION 3D<br>ECLIPSE<br>(K122199) | CS 9300 and CS<br>9300C<br>(K103659) | i-CAT FLX Cone<br>Beam 3D and 2D<br>Panoramic<br>Dental Imaging<br>System<br>(K133620) | |
| | (95KV, 1.6mA,<br>6.9s)<br>10cm: 0.5mGy<br>5cm: 0.4mGy | | | 26.9s)<br>13cm: 3.8mGy*,<br>11cm: 3.9mGy*,<br>10cm: 3.6mGy*,<br>8cm: 4.0mGy*,<br>6cm: 4.0mGy*,<br>4cm: 4.2mGy*<br>Quick Scan HD<br>(120KV, 5mA,<br>14.7s)<br>13cm: 2.1mGy*,<br>11cm: 2.1mGy*,<br>10cm: 2.0mGy*,<br>8cm: 2.1mGy*,<br>6cm: 2.1mGy*,<br>4cm: 2.2mGy*<br>* the sum of<br>1/3 of the phantom<br>central dose and<br>2/3 of the<br>peripheral dose. | |
| Performance<br>Standards | ANSI/AAMI<br>ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 62366<br>IEC 62304<br>IEC 60601-2-63<br>IEC 61223-3-4<br>IEC 61223-3-5<br>IEC 60825-1<br>ISO 14971<br>NEMA PS 3.1 -<br>3.20<br>ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10 | UL 60601-1<br>IEC 60601-1-1<br>IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-2-7<br>IEC 60601-2-28<br>IEC 60601-2-32<br>IEC 60825-1<br>IEC 61233-3-4<br>IEC 61223-3-5<br>IEC 62304<br>ISO 14971<br>NEMA PS 3.1 -<br>3.20 | Unknown | IEC 60601-1:<br>2005<br>IEC 60601-1-<br>6:2010<br>IEC 62366:2007<br>IEC 60601-2-<br>63:2012<br>IEC 60601-1-2:<br>2007<br>ISO 10993-5<br>ISO 10993-10 | |
{9}------------------------------------------------
{10}------------------------------------------------
## 5.7 Non-Clinical Performance Data
The subject device has demonstrated conformance to non-clinical performance requirements through evaluation and testing in accordance with the following harmonized standards:
ANSI/AAMI ES60601-1 IEC 60601-1-2 IEC 60601-1-3 IEC 60601-1-6 IEC 62366 IEC 62304 IEC 60601-2-63 IEC 61223-3-4 IEC 61223-3-5 IEC 60825-1 ISO 14971 NEMA PS 3.1 - 3.20 ISO 10993-1 ISO 10993-5 ISO 10993-10
In addition to the conformance with the above recognized standards, the following testing and nonclinical considerations were preformed:
- . Testing for 3D imaging performance to assess MTF for three image orientations (x, y, z) for the applicable 3D modes of device operation.
- . Non-clinical considerations according to FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices"
Results of all non-clinical testing and non-clinical considerations support the safety and effectiveness of the subject device.
# 5.8 Clinical Performance Data
Clinical considerations according to FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" were performed by the qualified clinical assessors. Results of the clinical consideration support the safety and effectiveness of the subject device.
{11}------------------------------------------------
## 5.9 Conclusions
The subject device has the same intended use/Indications for use as the predicate devices. The subject and predicate devices also share similar technological characteristics such as specifications of X-ray Generation Device. X-ray Image Capturing Device and Scanner. Imaging Mode (i-CAT does not have a Cephalometric imaging mode), Scan Mode and conformance with Performance Standards. The subject and predicate devices also have differences in technological characteristics, such as Tube Voltage, Pulse Exposure, Tube Current, Focal Spot Size, Pixel Number, Size of Area Receiving X-ray, Detector, Number of Bits, Scanner SID/SOD, Panoramic Scan Performance (Scan Time), Cephalometric Radiography (Scan Time), CT Scan Time, FOV (Voxel Size), Volume Size, Dose Level (CTDIw) and Performance Standards.
We conducted non-clinical and clinical performance testing as follows:
- Conformance with the harmonized standards .
- Testing for 3D imaging performance to assess MTF for the three image orientations .
- . Non-clinical and Clinical considerations according to FDA Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices"
Based on the above information and all data provided in this submission, the comparison of intended uses, technological characteristics, and non-clinical performance data and clinical performance data demonstrates that the subject device is substantially equivalent to the legally marketed devices identified in this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.