The Acumed Congruent Bone Plate System provides fixation for fractures, fusions, or ostectomies for the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
Device Story
The Acumed Congruent Bone Plate System is a metallic bone fixation system. This submission adds 2.3mm Locking Variable Angle Screws to the existing Acu-Loc 2 Plate, which is used for distal radius fracture fixation. The screws are implanted by surgeons in a clinical setting to provide mechanical stability to bone segments. The device functions by securing the bone plate to the bone, facilitating healing through rigid fixation. The addition of variable angle screws allows for greater flexibility in screw trajectory during surgical placement, potentially improving fixation in complex fracture patterns.
Clinical Evidence
Bench testing only. Mechanical testing was performed to characterize the screw/plate interface, demonstrating that the new screws met all defined acceptance criteria.
Technological Characteristics
Materials: Titanium alloy conforming to ASTM F136. Components: 2.3mm Locking Variable Angle Screws for use with the Acu-Loc 2 Plate. Principle: Metallic bone fixation via screw/plate interface. No software or electronic components.
Indications for Use
Indicated for patients requiring fixation for fractures, fusions, or osteotomies of the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K181533 — EVOS Wrist Fracture Plating System · Smith & Nephew, Inc. · Aug 9, 2018
K031725 — SYNTHES TI-15 MO LOCKING DISTAL RADIUS PLATING SYSTEM · Synthes (Usa) · Jun 24, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
# k |20903
## JUL 1 3 2012
#### 5. 510(k) Summary
Device Trade Name: Acumed Congruent Bone Plate System: Acu-Loc 2 Plate Acumed, LLC Manufacturer: 5885 NW Cornelius Pass Road Hillsboro, OR 97124 Contact: Ms. Brittany Cunningham Regulatory Specialist Phone: 503.207.1467 Fax: 503.686.7102 Prepared by: Musculoskeletal Clinical Regulatory Advisers, LLC 1331 H Street NW, 12th Floor Washington, DC 20005 Phone: (202) 552-5800 Fax: (202) 552-5798 March 23, 2012 Date Prepared: 21 CFR 888.3030, Single/multiple component metallic Classifications: bone fixation appliances and accessories 21 CFR 888.3040, Smooth or threaded metallic bone fixation fastener Class: II
Product Codes:
#### Indications For Use:
The Acumed Congruent Bone Plate System provides fixation for fractures, fusions, or ostectomies for the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
HRS: HWC
#### Device Description:
The Acu-Loc 2 Plate is a volar plate which is intended to provide fixation for distal radius fractures.
The purpose of this Special 510(k) is to add 2.3mm Locking Variable Angle Screws for use with the Acu-Loc 2 Plate. All components are made of titanium alloy conforming to ASTM F136.
This addition does not affect the intended use of the device or alter the fundamental scientific technology of the device.
{1}------------------------------------------------
### Predicate Device:
The subject 2.3mm Variable Angle Locking Screws are substantially equivalent to predicate screws and pegs (K102998) with respect to indications, design, function, and materials.
### Preclinical Testing:
The new screws were subjected to mechanical testing to characterize the screw/plate interface. The results demonstrate that the acceptance criteria defined in the Design Control Activities Summary were met.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Acumed, LLC % Ms. Brittany Cunningham Regulatory Specialist 5885 Northwest Cornelius Pass Road Hillsboro, Oregon 97124-9432
JUL 13 2012
Re: K120903
Trade/Device Name: Acumed Congruent Bone Plate System: 2.3 mm Variable Angle Locking Screw for Use with Acu-Loc2 Plate
Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and
accessories
Regulatory Class: Class II Product Code: HRS, HWC Dated: June 14, 2012 Received: June 15, 2012
Dear Ms. Cunningham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r read be action and a determination that your device complies with other requirements of the Act that I Dri has interes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
### Page 2 - Ms. Brittany Cunningham
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson VIT Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use 4.
510(k) Number (if known): _ V 12 09 03
Device Name: Acumed Congruent Bone Plate System
The Acumed Congruent Bone Plate System provides fixation for fractures, fusions, or osteotomies for the clavicle, humerus, radius, ulna, metacarpal, metatarsal, malleolus, tibia, and fibula.
ど Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) :
Concurrence of CDRH, Office of Device Evaluation (ODE)
Asin
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K120903
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.