TENS stands for Transcutaneous Electrical Nerve Stimulation. The Ultima NEO TENS system is used to provide symptomatic pain relief for chronic, acute or post-operative pain. EMS stands for Electrical Neuromuscular Stimulator. The Ultima NEO EMS system is indicated for: - Relaxation of muscle spasm; - Increasing local blood circulation and muscle re-education; - Prevention or retardation of disuse atrophy; - Prevention of venous thrombosis of the calf muscles immediately after surgery; - Maintaining or increasing range of motion. MIC stands for Microcurrent Stimulation. The Ultima NEO MIC system is used to provide symptomatic pain relief for chronic intractable pain, post traumatic pain or post surgical pain. IF stands for Interferential Stimulation. The Ultima NEO IF system is indicated for: Symptomatic relief of chronic intractable pain.
Device Story
Ultima NEO is a portable, multi-modality electrotherapy device combining TENS, EMS, MIC, and IF stimulation in one enclosure. Device features dual isolated output channels, LCD display, and rechargeable lithium battery. Operation is restricted to one modality at a time via mechanical/electronic interlocks. User adjusts frequency, pulse width, and timing parameters; includes doctor lock/unlock feature. Device provides electrical stimulation to nerves or muscles via electrodes. Healthcare providers or patients use the device to manage pain or perform muscle rehabilitation. Output affects clinical decision-making by providing non-pharmacological pain management and physical therapy support. Benefits include symptomatic relief and improved muscle function.
Clinical Evidence
Bench testing only. Device performance verified through compliance with recognized international consensus standards including IEC 60601 series, ISO 14971, and ISO 13485. Risk analysis and testing confirm safety and effectiveness within specified output ranges.
Technological Characteristics
Portable electrotherapy device; dual isolated output channels; rechargeable lithium battery; LCD interface. Modalities: TENS, EMS, MIC, IF. Software-controlled parameters (frequency, pulse width, timer). Safety features: open circuit detectors, doctor lock/unlock, interlocks for single-mode operation. Complies with IEC 60601-1, 60601-1-2, 60601-1-4, 60601-2-10, IEC 62304, ISO 14971, ISO 13485, ISO 9001, UL 60601-1.
Indications for Use
Indicated for patients requiring symptomatic pain relief (chronic, acute, post-operative, intractable, post-traumatic, post-surgical) or muscle re-education/therapy (spasm relaxation, circulation improvement, atrophy prevention, venous thrombosis prevention, range of motion maintenance).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
K120054
JUN - 5 2012
# 510(k) Summary
Date of Summary prepared: 31 December 2011
Submitter: Pain Management Technologies Address: 1340 Home Ave. Bldg. A Akron, OH 44310 Tel (330)-564-0820 : Fax (330) 564-0818 : E-Mail josh@paintechnology.com :
Contact person: Joshua Lefkovitz
Address of the manufacturing facility:
EasyMed Instruments Co., Ltd 5/F-6/F, Block A, Gupo Gongmao Building, Fengxin Road, Fengxiang Industrial District, Daliang, Shunde, Foshan, Guangdong, China Zip code: 528300
Submitted Device:
Generic name: Powered Muscle Stimulator (PMS) (or Electrical Muscle Stimulator (EMS)), Transcutaneous Electrical Nerve Stimulator (TENS), Microcurrent (MIC) TENS stimulator and Interferential Current Therapy Stimulator (IF)
Ultima NEO Trade name:
Common name: For the TENS/MIC system-TENS device For the EMS system-Powered Muscle Stimulators For the IF system-Interferential Stimulator
#### Classification name: For EMS system
Powered Muscle Stimulator for re-education of muscles - IPF; 21 CFR 890.5850
For TENS / MIC system
Stimulator, Nerve, Transcutaneous, for Pain Relief - GZJ; 21 CFR 882.5890.
For IF system Interferential current therapy - LIH
{1}------------------------------------------------
K120054
Device Classification: For EMS, TENS / MIC and IF systems - Class II
Predicate Devices: EasyMed EMS/TENS Model NMS-28 (K050921); MT9000 Combo TENS/EMS/IF/MIC (K093138)/(Trade name : InTENSity™ Select Combo) ; GM3 Series IF TENS Model GM3X2IF (K032719)
The class of the predicate Devices: Class II
Device Description: A portable TENS/EMS/MIC/IF combo device for pain relief or muscle re-education.
Features:
- Innovative design
- · Large LCD display
- · Dual output isolated channels
- · One rechargeable lithium battery
- · Adjustable frequency, pulse width, and timing parameters
- 18 different modes
- · Timer option
- · Doctor lock/unlock facility
- · Open circuit detectors
- Non-volatile
The intended use of the device:
TENS stands for Transcutaneous Electrical Nerve Stimulation. The Ultima NEO TENS system is used to provide symptomatic pain relief for chronic, acute or post-operative pain.
EMS stands for Electrical Neuromuscular Stimulator. The Ultima NEO EMS system is indicated for:
- . Relaxation of muscle spasm;
- Increasing local blood circulation and muscle re-education; .
- Prevention or retardation of disuse atrophy; .
- Prevention of venous thrombosis of the calf muscles immediately . after surgery;
- . Maintaining or increasing range of motion.
MIC stands for Microcurrent Stimulation. The Ultima NEO MIC system is used to provide symptomatic pain relief for chronic intractable pain, post traumatic pain or post surgical pain.
{2}------------------------------------------------
IF stands for Interferential Stimulation. The Ultima NEO IF system is indicated for:
- Symptomatic relief of chronic intractable pain. .
The intended use and indications for use of the new device are very similar to that of the predicate devices.
#### Technological Comparison:
- 1. The new device Ultima NEO is a portable electrotherapy device which combines four systems : Transcutaneous Electrical Nerve Stimulation (TENS), Electrical Muscle Stimulation (EMS), Microcurrent (MIC) and Interferential (IF), into one enclosure. These four systems work separately. When one system is working, the others do not work and thus do not affected each other. This feature has been fully guaranteed by the design.
- 2. The marketed device EasyMed EMS/TENS Model NMS-28 (K050921) combines the systems of a TENS and an EMS (Electrical Muscle Stimulator or Powered Muscle Stimulator), into one package. It is a digital TENS/EMS combo unit which has been well exploited the digital technology, supplying the user preset programs and manual selectable programs with full ranges of parameters.
- 3. The new device Ultima NEO also combines the systems of TENS and EMS into one package. It is also substantial equivalent to the marketed device EasyMed EMS/TENS Model NMS-28 (K050921), their features and performances in muscle controlling as well as pain relief therapy are almost the same substantially.
- 4. The designed circuitry of TENS and EMS modes in the new device Ultima NEO is almost the same as the marketed device EasyMed EMS/TENS Model NMS-28 (K050921). For the TENS and EMS systems, they are of almost the same circuit diagrams and same working principle; The main difference between the two devices is that the new device Ultima NEO is also combined the Microcurrent (MIC) and Interferential (IF) systems into one package. But these systems are worked separately.
- 5. The marketed device MT9000 Combo TENS/EMS/IF/MIC (K093138) is a combo device which includes the Microcurrent stimulation system. The Microcurrent stimulation can be generated by a very low and relatively long period current. The specification of
{3}------------------------------------------------
K120054
microcurrent system in MT9000 Combo TENS/EMS/IF/MIC K093138 is very similar to that of Ultima NEO MIC system.
- 6. The marketed device GM3 Series IF TENS Model GM3X2IF (K032719) is an interferential stimulator which generates the real-sine waves of electrical current. The specification of GM3X2IF (K032719) is very similar to that of Ultima NEO IF system.
- 7. The new device Ultima NEO is designed and combined the Interferential (IF) and Microcurrent (MIC) systems based on the design of EasyMed EMS/TENS Model NMS-28 (K050921). The introduction of marketed device GM3 Series IF TENS Model GM3X2IF (K032719) and MT9000 Combo TENS/EMS/IF/MIC (K093138) are for the purpose of comparison of the feature of Microcurrent (MIC) and Interferential (IF).
- 8. The new device Ultima NEO is of the same ranges of parameters as those of marketed devices.
- 9. The software of the new device Ultima NEO can be divided into four parts: the first for TENS system, the second for EMS system, the third for MIC system, and the four one for IF system. First of all, the part of software for TENS system of the new device Ultima NEO is only a subset of that of the marketed device EasyMed EMS/TENS Model NMS-28 (K050921). Secondly, the considerations on safety, effectiveness and reliability in NMS-28 (K050921) are also introduced into the EMS system of the new device; While the part of software for EMS system of the new device Ultima NEO is almost the same as that of the marketed device EasyMed EMS/TENS Model NMS-28 (K050921), by applying almost the same control methods for muscle contraction-relaxation. In other word, the TENS and EMS systems of the new device Ultima NEO are almost the same as those of the marketed device NMS-28 (K050921). Thirdly, the part of software for Microcurrent (MIC) system of the new device Ultima NEO is very similar to that of the marketed device MT9000 Combo TENS/EMS/IF/MIC (K093138). For the last, the part of software for Interferential (IF) system of the new device Ultima NEO is very similar to that of the marketed device GM3 Series IF TENS Model GM3X2IF (K032719)
10. Both the new device Ultima NEO and the marketed devices
{4}------------------------------------------------
NMS-28 (K050921)/ GM3 Series IF TENS Model GM3X2IF (K032719)/ MT9000 Combo TENS/EMS/IF/MIC (K093138) are passed the same tests of applicable recognized international consensus standards.
- 11. The accessories of the new device Ultima NEO are similar to those of the marketed devices.
Labeling Comparison: The Labelling is substantially equivalent to that of the predicate devices.
Safety information:
Design to comply with relevant safety applicable recognized consensus standards; the output enerqy is well controlled in the safety and effectiveness ranges specified by relevant FDA guidance's. Testing has been carried out in very detailed and strictly. Test results and Risk Analysis show that the new unit Ultima NEO is safe with no any hazard.
Ultima NEO combines the systems of TENS, EMS, Micocurrent, and IF into one package, it is not possible to use any other system(s) simultaneously when one system is being used. Mechanical and electronical integrity ensures that only one system can be selected at any one time.
| Standard ORG | Standard Number | Standard Title |
|--------------|-----------------|-------------------------------------------------|
| IEC | 60601 | Medical Electrical equipment- Part I: |
| IEC | 60601 - I -2 | Medical Electrical equipment- Part I-2: |
| IEC | 60601-1-4 | Medical Electrical equipment- Part I-4: |
| IEC | 60601-2-10 | Medical electrical equipment Part2 -10: |
| ISO | 14971 | Medical devices- |
| ISO | I 3485 | Medical Devices-<br>Quality Management systems- |
| IC council | 93/42/EEC | EC Directive 93/42/EEC Annex V,<br>Article 3, |
| IEC/EN | 62304 | IEC 62304 Ed. 1.0,<br>Medical device software |
{5}------------------------------------------------
| IEC-EN | 60601-1-6 | Medical electrical<br>equipment .Part I-6: |
|--------|-----------|--------------------------------------------|
| ISO | 9001 | Quality management<br>systems- |
| UL | 60601-1 | Medical electrical<br>equipment Part I : |
K120054
Conclusions:
The new device Ultima NEO, which includes TENS, EMS, MICrocurrent, and IF systems, has the same intended use and technological characteristics as the predicate devices of NMS-28 (K050921)/ MT9000 Combo TENS/EMS/IF/MIC (K093138) /GM3 Series IF TENS Model GM3X21F (K032719). Thus, the new device Ultima NEO is substantially equivalent to the predicate devices and any differences between the devices do not pose any new questions of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Pain Management Technologies c/o Mr. Joshua Lefkovitz President 1340 Home Avenue, Building A Akron, OH 44310
JUN - 5 2012
Re: K120054
Trade/Device Name: Ultima NEO Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ, IPF, LIH Dated: May 18, 2012 Received: May 24, 2012
Dear Mr. Lefkovitz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
# Page 2 - Mr. Joshua Lefkovitz
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number . (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### 510(k) Number : K120054
Device Name
: Ultima NEO
### Indications for Use
TENS stands for Transcutaneous Electrical Nerve Stimulation. The Ultima NEO TENS system is used to provide symptomatic pain relief for chronic, acute or post-operative pain.
EMS stands for Electrical Neuromuscular Stimulator. The Ultima NEO EMS system is indicated for:
- Relaxation of muscle spasm; .
- Increasing local blood circulation and muscle re-education; ' .
- Prevention or retardation of disuse atrophy; .
- Prevention of venous thrombosis of the calf muscles immediately . after surgery;
- Maintaining or increasing range of motion. .
MIC stands for Microcurrent Stimulation. The Ultima NEO MIC system is used to provide symptomatic pain relief for chronic intractable pain, post traumatic pain or post surgical pain.
IF stands for Interferential Stimulation. The Ultima NEO IF system is indicated for:
Symptomatic relief of chronic intractable pain. .
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
7
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off)
Division of Ophthalmic, Neurologiet ABAF effice of CDRH, Office of Device Evaluation (ODE) Nose and Throat Devices
K120054
510(k) Number.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.