The A200 Knee System is intended for use in total knee arthroplasty for the following indications: 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved. 2. Correction of varus, valgus, or posttraumatic deformity. 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure. This device is intended for cemented use only.
Device Story
A200 Knee System is a semi-constrained, cemented total knee prosthesis. Components include femoral components (Co-Cr-Mo), tibial trays (Co-Cr-Mo or Ti-6Al-4V), tibial inserts (highly crosslinked UHMWPE, with/without vitamin E), and patellar components (UHMWPE). System includes instrumentation for trial and implantation. Used by orthopedic surgeons in clinical settings for total knee arthroplasty. Implants replace damaged joint surfaces to restore function and reduce pain. Design principles and materials are equivalent to existing knee systems; performance verified through mechanical testing of fatigue, contact stress, constraint, and wear rates.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing: component fatigue, contact area/stress, constraint, and tibial insert-tray interlock strength. UHMWPE material characterization included density, oxidation index, vitamin E content, and free-radical content. Mechanical testing included Izod impact, tensile/compressive properties, and fatigue crack propagation. Wear testing performed per ISO 14243-1 and ISO 14243-2.
Indicated for patients requiring total knee arthroplasty due to painful/disabled knee joints from osteoarthritis, rheumatoid arthritis, or traumatic arthritis; correction of varus, valgus, or posttraumatic deformity; or revision of failed osteotomy, arthrodesis, or previous joint replacement. Intended for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
510(k) Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
# 510(k) Summary Renovis Surgical Technologies, LLC A200 Knee System K120038
January 29, 2013
### ADMINISTRATIVE INFORMATION
Manufacturer Name: Renovis Surgical Technologies, LLC 1901 W. Lugonia Avenue, Suite 340 Redlands, CA 92374 +1 (909) 557-2360 Telephone: Fax: +1 (909) 307-8571 Official Contact: Anthony DeBenedictis Vice President of Quality Assurance Representative/Consultant: Kevin A. Thomas, PhD Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone: +1 (858) 792-1235 Fax: +1 (858) 792-1236
Email:
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name:<br>Common Name:<br>Classification Names: | A200 Knee System<br>Total Knee Prosthesis System<br>Prosthesis, knee, patellofemorotibial, semi-constrained, cemented,<br>polymer/metal/polymer;<br>Prosthesis, knee, patellofemorotibial, semi-constrained, cemented,<br>polymer + additive/metal/polymer + additive |
|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Regulations: | 21 CFR 888.3560, Class II |
| Product Code: | JWH, OIY |
| Classification Panel: | Orthopedic Products Panel |
| Reviewing Branch: | Orthopedic Devices Branch |
kthomas@paxmed.com flarson@paxmed.com
# FEB 1 4 2013
{1}------------------------------------------------
#### INTENDED USE
The A200 Knee System is intended for use in total knee arthroplasty for the following indications:
- 1. Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
This device is intended for cemented use only.
#### DEVICE DESCRIPTION
The A200 Knee System consists of the following components: Femoral Components - nine (9) sizes in a cruciate-retained (CR) design, each for left and right sides, manufactured from Co-Cr-Mo allov: Tibial Tray Components - nine (9) sizes, each for the left and right sides, and each in two stem designs, a conical design manufactured from Co-Cr-Mo alloy, and a keeled design manufactured from Ti 6Al 4V alloy: Tibial Insert Components - three (3) footprint sizes, each for the left and right sides, with articulating thicknesses ranging from 10 mm to 20 mm, all in CR designs, in standard and deep dish configurations, manufactured from highly crosslinked ultra-high molecular weight polyethylene with and without vitamin E: Patellar Components five (5) sizes, in dome design, manufactured from highly crosslinked ultra-high molecular weight polvethylene with and without vitamin E: and A200 Knee Instrumentation including instruments for use with the system components, trial implant components, and general instruments.
The A200 Knee System implants are manufactured from Co-Cr-Mo alloy conforming to ASTM F75, Ti-6A1-4V alloy conforming to ASTM F136, and ultra-high molecular weight polyethylene conforming to ASTM F648 and ASTM F2695. All UHMWPE is crosslinked by gamma irradiation. The metallic implant components (femoral components and tibial tray components) are sterilized by Co6 gamma irradiation, and all UHMWPE components are sterilized by ethylene oxide.
#### EQUIVALENCE TO MARKETED DEVICES
Renovis Surgical Technologies, LLC, submits the following information in this Premarket Notification to demonstrate that for the purposes of FDA's regulation of medical devices, the A200 Knee System is substantially equivalent in indications and design principles to the following legally marketed predicate devices, each of which has been determined by FDA to be substantially equivalent to a legally marketed predicate device:
DePuy Orthopaedics, Inc., DePuy Sigma Knee Femoral Adapter, K060515;
Encore Medical, L.P., Highly Cross-Linked Vitamin E UHMWPE Tibial Insert, K103223;
Biomet Manufacturing Corp., E1™ Tibial Bearings, K100048;
Encore Orthopedics, Inc., Foundation Knee System, K923277;
Zimmer, Inc., MG II Porous Total Knee System, K892800;
U.S. Medical Products, Inc., Consensus Knee System - Primary Knee, K932837; and
Hospital for Special Surgery, Insall-Burstein II, K935080.
{2}------------------------------------------------
## 510(k) Summary K120038
The A200 Knee System and the predicate device K060515 are patellofemorotibial knee prosthesis systems, intended for cemented total knee arthroplasty, having the same intended use and similar indications for use.
The A200 Knee System and the predicate device K060515 are made of the same materials, Co-Cr-Mo alloy (femoral components and tibial tray components) and highly crosslinked UHMWPE (tibial insert components and patellar components), and all are provided sterilized by gamma irradiation or ethylene oxide.
The subject and predicate device K060515 have femoral components designed specifically for the right and left knees, in cruciate retaining (CR) designs, and provided in a range of sizes. Similarly, the subject device and the predicate device K060515 also include tibial trays that incorporate a similar stabilizing keel design, and both devices also include UHMWPE tibial inserts in CR designs for use with the corresponding femoral components.
The A200 Knee System and the predicate device K060515 components are similar in overall shape and design. The subject device and the predicate devices encompass a similar range of physical dimensions, including the AP and ML dimensions of the femoral and tibial tray components, the thickness of the tibial insert components, and the thickness and diameter of the patellar components. Constraint, contact area, and contact stress testing demonstrated the A200 Knee System to be substantially equivalent to K923277, K892800, K932837, and K935080.
The A200 Knee System and the predicate devices K103223 and K100048 include tibial inserts manufactured from UHMWPE containing alpha-tocopherol (vitamin E). Performance testing demonstrated that the wear rate of the A200 Knee System UHMWPE with vitamin E was substantially equivalent to the predicate devices K103223 and K100048.
Any differences in the technological characteristics between the subject and predicate devices do not raise new issues of safety or efficacy.
Non-clinical testing data that was submitted. referenced, or relied upon to demonstrate substantial equivalence included: component testing (tibial tray fatigue, contact area. contact stress, constraint, and tibial insert-tray interlock strength); UHMWPE material characterization (density, swell ratio, oxidation index, trans-vinylene index, melting point, crystallinity, vitamin E content, free-radical content, assessment of consolidation by scanning electron microscopy, and analysis of extraction residues); UHMWPE mechanical testing (Izod impact resistance, tensile properties, compressive properties, small punch testing, fatigue crack propagation, and environmental stress cracking); and force-control wear testing according to the methods of ISO 14243-1 and ISO 14243-2.
The data included in this submission demonstrates substantial equivalence to the predicate devices listed above.
Overall, the A200 Knee System has the following similarities to the predicate devices:
- . has the same intended use,
- . uses the same operating principles.
- . incorporates the same basic designs,
- incorporates the same or very similar materials, and .
- has similar packaging and is sterilized using the same materials and processes. .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 14, 2013
Renovis Surgical Technologies, LLC % Paxmed International, LLC Kevin Thomas, Ph.D. 11234 El Camino Real, Suite 200 San Diego, California 92130
Re: K120038
Trade/Device Name: A200 Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: OIY, JWH Dated: January 29, 2013 Received: January 30, 2013
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
### Page 2 - Kevin Thomas, Ph.D.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Image /page/4/Picture/7 description: The image shows the name "Erin DKeith" in a stylized font. The letters "Erin" and "eith" are in a standard font, while the letters "D" and "K" are stylized with a layered, outlined design. The dot in the "I" is a solid black circle.
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Summary K120038
# Indications for Use
510(k) Number: K120038
Device Name: A200 Knee System
The A200 Knee System is intended for use in total knee arthroplasty for the following indications:
- 1. Painful and disabled knee joint resulting from osteoarthritis, theumatoid arthritis, traumatic arthritis where one or more compartments are involved.
- 2. Correction of varus, valgus, or posttraumatic deformity.
- 3. Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.
This device is intended for cemented use only.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
013.02.08 14:36:18 Anton E. Dmitriev, PhD Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.