FRESENIUS 2008 HEMODIAYSIS MACHINE WITH BIBAG SYSTEM
K120017 · Fresenius Medical Care North America · KDI · Feb 2, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K120017
Device Name
FRESENIUS 2008 HEMODIAYSIS MACHINE WITH BIBAG SYSTEM
Applicant
Fresenius Medical Care North America
Product Code
KDI · Gastroenterology, Urology
Decision Date
Feb 2, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fresenius bibag system is used with Fresenius three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.
Device Story
The Fresenius 2008T Hemodialysis Machine with bibag System is a high-permeability hemodialysis delivery system for acute and chronic renal failure. It performs hemodialysis, hemofiltration, and hemoconcentration. The bibag component is a single-use, sealed bag containing USP-grade dry sodium bicarbonate powder; the machine draws water into the bag to generate a saturated bicarbonate solution online for 1:44 dilution. Modifications include an optional 900g bibag (replacing/supplementing the 650g version) and a keyboard update featuring a 'Fn Lock' key and high-tear-resistant elastomer material. The device is operated by clinicians in clinical settings. The system uses a three-stream proportioning method. The output is the delivery of dialysate to the dialyzer, which facilitates the removal of toxins and fluid from the patient's blood. The system includes safety alarms and monitoring to ensure treatment parameters are maintained according to physician prescription.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included functional testing of the 900g bibag (bioburden, endotoxin, shelf life, stability), software verification/validation for the keyboard update (regression, safety systems, simulated dialysis), EMC testing, and physical testing of the new elastomer material (life cycle/use life).
Technological Characteristics
High permeability hemodialysis system; three-stream proportioning. Materials: USP grade dry sodium bicarbonate powder; high-tear-resistant elastomer for keyboard. Connectivity: Standalone/integrated module. Software: Embedded firmware for machine control and keyboard interface. Sterilization: Not specified (disposable bibag).
Indications for Use
Indicated for patients with acute or chronic renal failure requiring bicarbonate hemodialysis. Used with Fresenius three-stream proportioning hemodialysis systems equipped with a bibag module.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
Predicate Devices
Fresenius 2008T Hemodialysis Machine with bibag System (K101715)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K120017 page 1 of 6". The text appears to be handwritten. The text "K120017" is on the first line, and the text "page 1 of 6" is on the second line.
Image /page/0/Picture/2 description: The image shows the Fresenius Medical Care logo. The logo consists of a geometric shape resembling an inverted pyramid or funnel with horizontal lines inside, positioned above the text "Fresenius Medical Care". The text is in a simple, sans-serif font and is aligned horizontally.
#### Section 5 510(k) Summary
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.
#### 5.1 Submitter's Information
| Name: | Fresenius Medical Care North America |
|----------------------|--------------------------------------------------------------------------------------------|
| Address: | 920 Winter Street<br>Waltham, MA 02451-1457 |
| Phone: | (781) 699-4479 |
| Fax: | (781) 699-9635 |
| Contact Person: | Denise Oppermann, Senior Director<br>Regulatory Affairs - Devices<br>Renal Therapies Group |
| Date of Preparation: | 30 December 2011 |
#### 5.2 Device Name
| Trade Name: | Fresenius 2008T Hemodialysis Machine with<br>bibag System |
|------------------------------------|-----------------------------------------------------------|
| Common Name: | Hemodialysis Delivery System |
| Classification Name | High Permeability Hemodialysis System |
| Classification Number | Class II per § 876.5860 |
| Product Code/Classification Panel: | 78KDI/Gastroenterology/Urology Panel |
#### 5.3 Legally Marketed Predicate Device (unmodified device)
Fresenius 2008T Hemodialysis Machine with bibag System (K101715).
{1}------------------------------------------------
K120017
page 2 of 6
Image /page/1/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a triangle shape with horizontal lines inside, positioned above the text "Fresenius Medical Care". The text is in a simple, sans-serif font and is left-aligned.
#### 5.4 Device Description
The Fresenius 2008T Hemodialysis Machine with bibag System is indicated for use in bicarbonate hemodialysis for acute and chronic renal failure.
The bibag is a specialized, single-use, sealed bag filled with USP grade dry sodium bicarbonate powder. The bibag attaches to a special connector incorporated into the front of the 2008T Hemodialysis Machine. The hemodialysis machine draws dialysis grade water into the bibag to produce a saturated solution of sodium bicarbonate online. This online generation of sodium bicarbonate can only be prepared using a specially modified Fresenius 2008T Hemodialysis Machine with bibag System and can only be used with 45x (1:44) dilution. The bibag cannot be used with other non-Fresenius hemodialysis machines capable of using cartridge type dry sodium bicarbonate because of the unique connection between the bibag disposable, the bibag connector, and the hemodialysis machine.
Modifications of the previously cleared 2008T Hemodialysis Machine with bibag system includes:
#### 5.4.1 900g bibag Disposable:
Addition of an optional disposable 900g bibag which is equivalent to 650g bibag but increased in length to accommodate an additional 250g dry sodium bicarbonate powder. The length of a treatment session is dependent upon pre-treatment set-up time (including potential delays) and physician-prescribed dialysate flow rate and sodium bicarbonate concentration. (For example: The larger bag will provide sufficient USP grade dry sodium bicarbonate powder for 8 hours when running at a dialysate flow rate (Qd) of 500 mL/min and sodium bicarbonate concentration setting of 32 mEq/L).
There were no hardware or software changes made to the 2008T Hemodialysis Machine with bibag System to accommodate the 900g bag option. The existing bibag module attached to the machine will be used to connect the new 900g bibag with the same connector used for the 650g bibag.
Additionally, this submission includes a minor software maintenance change made to the 2008T Hemodialysis Machine with bibag System since the last clearance (K101715).
{2}------------------------------------------------
K120017
page 3 of 6
Image /page/2/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of an inverted triangle with horizontal lines inside. Below the triangle is the text "Fresenius Medical Care".
The following modifications were implemented following a regulatory assessment that the changes did not affect the fundamental scientific technology or intended use of the device. Based on FDA guidance "Deciding When to Submit a 510(k) for a Change to an Existing Device", FMCNA determined that these modifications did not necessitate a 510(k) submission:
# 5.4.2 Function (Fn) Lock Keyboard
The existing fold-down compact keyboard used on the 2008T Hemodialysis Machine (K101715) is located directly below the display screen. The keyboard folds down to allow the operator to enter treatment parameter values, chart with CDX or make selections inside the treatment screen and folds up again to prevent unintentional changes.
For added user convenience while navigating their MIS systems, FMCNA added function lock capability to the current 2008T's keyboard. The modification includes repurposing the "Fn" key to a "Fn Lock" key and adding a function lock indicator light to the immediate right of the Caps Lock indicator light. Software and hardware modifications were made to implement this feature is only available in 2008T Hemodialysis Machines with activation of the CDX option. In dialysis mode, the Fn Lock LED will be off and the function lock feature will be off. In addition, the keyboard is now equipped with a new elastomer material with higher tear resistance properties as compared to the standard silicon elastomer of the existing keyboard in the unmodified device (K101715).
Treatment modalities for the modified Fresenius 2008T Hemodialysis Machine remain identical to those for the unmodified 2008T (K101715):
The 2008T is a high permeability hemodialysis system used for the treatment of patients with acute or chronic kidney failure, fluid overload or toxemic conditions. Therapies include hemodialysis, hemofiltration and hemoconcentration. The 2008T will accommodate the use of both low flux and high flux dialyzers.
#### 5.5 Indications for Use
The modified Fresenius 2008T Hemodialysis Machine with bibag System has the same indications for use as the unmodified device. The Fresenius Medical Care bibag System is used with Fresenius Medical Care three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended
{3}------------------------------------------------
K120017
page 4 of 6
Image /page/3/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a triangle shape with horizontal lines inside, above the text "Fresenius Medical Care". The text is in a simple, sans-serif font and is left-aligned.
for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag system is extracorporeal bicarbonate hemodialysis according to a physician's intended for prescription.
#### 5.6 Technological Characteristics
There are no changes in the technological characteristics of the previously cleared Fresenius 2008T Hemodialysis Machine with bibag System (K101715). The Fresenius 2008T Hemodialysis Machine with bibag system incorporates changes pertaining only to the size of the bibag disposable with 900g USP grade dry bicarbonate powder as an additional option to the previously cleared 650g bibag. There are no hardware or software changes made in the machine to accommodate this modification. The 900g bibag is designed and developed using dimensionally and functionally equivalent product specifications, construction and manufacturing methods as in the previously cleared 650g bibag (K101715). All water requirements, module options, functional options, performance limits, control parameters, compatible bloodlines, and language options remain unchanged from the predicate device. The following technical specifications of the modified device remain the same as the unmodified device:
- Safety system
- . System performance
- Environmental Requirements .
- Transportation and Storage condition
- . User Interface (except for the Fn Lock feature and touchscreen)
- Hardware and therapy settings
- Accessories .
- Environmental Design
- Alarms
- Accuracy and Controls .
- Protection against Mechanical Hazard
- . Protection against Electrical Hazard
- Protection against excessive temperature or other hazards ●
- Manufacturing Location and manufacturing processes (assembly, fabrication, ● testing, shipping, installation and service).
A risk analysis has been completed and potential hazards associated with the modifications are identified and mitigated. Mitigations are verified wherever applicable. All potential risks
{4}------------------------------------------------
K120017
page 5 of 6
Image /page/4/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a triangle pointing downwards with three horizontal lines inside. Below the triangle is the text "Fresenius Medical Care".
were deemed acceptable after mitigation. Performance and safety tests were conducted to ensure the safety and effectiveness of the device after the proposed modifications.
#### 5.7 Performance Data
The performance of the modified device (900g bibag disposable and all other modifications discussed in this submission) was evaluated according to existing FMCNA procedures, protocols, declared performance standards and guidelines of the quality system regulation (21 CFR 820). Design verification and validation tests were conducted to ensure that the modifications described in this submission did not affect the essential performance of the device and the device functions as intended.
The following tests were conducted for the 900g bibag disposable:
- Functional Verification
- . Bioburden and Endotoxin
- Shelf life
- . Stability
- . Unstructured testing
The following tests were conducted for the Fn Lock Keyboard:
#### 1. Software Verification and Validation Testing:
- Software Verification (Functional Tests) .
- . Regression
- . Safety Systems Verification
- Simulated Dialysis Treatment .
- Production Test Procedure
- . Unstructured and Static Code Verification
- 2. Safety Testing
- EMC Testing (ESD immunity)
- 3. Unit Testing (Fn Lock Keyboard)
- 4. Physical Testing (Fn Lock Keyboard)
- . High Tear Elastomer: Use Life (Shelf Life)
- . High Tear Elastomer: Life cycle
{5}------------------------------------------------
K120017
page 6 of 6
Image /page/5/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a geometric shape resembling an inverted pyramid or funnel with horizontal lines inside. Below the shape, the text "Fresenius Medical Care" is written in a simple, sans-serif font.
#### 5.8 Conclusion
Test results demonstrated that the modified 2008T Hemodialysis Machine with bibag System functions as intended and met pre-determined acceptance criteria. Results of functional validation, performance testing, biorburden assessment, risk analysis, safety testing, stability, shelf-life tests, and usability evaluation indicate that the modified Fresenius 2008T Hemodialysis Machine with bibag System is substantially equivalent to the named predicate device and remains safe and effective for its intended use.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three talons, representing service to the nation. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Denise Oppermann Senior Director, Regulatory Affairs Devices Fresenius Medical Care, North America 920 Winter Street WALTHAM MA 02451
FEB - 2 2012
### Re: K120017
Trade/Device Name: 2008T Hemodialysis Machine with bibag™ System Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI, KPO Dated: December 30, 2011 Received: January 3, 2012
Dear Ms. Oppermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the.Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
# Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin R. Fisher
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Fresenius Medical Care
2008T Hemodialysis Machine with bibag™ System Special 510(k) Notification
## Indications for Use Statement
510(k) Number (if known): KIQ()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()()(
Device Name:
2008T Hemodialysis Machine with bibag™ System
Indications for Use:
,
The Fresenius bibag system is used with Fresenius three stream proportioning hemodialysis systems equipped with the bibag module such as the 2008T Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.
> Prescription Use (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
forn m2ha
Division Sign-Off
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.