NOVATION INTEGRIP ACETABULAR AUGMENTS, GROUPS 4 AND 5

K113609 · Exactech, Inc. · LPH · Jan 4, 2012 · Orthopedic

Device Facts

Record IDK113609
Device NameNOVATION INTEGRIP ACETABULAR AUGMENTS, GROUPS 4 AND 5
ApplicantExactech, Inc.
Product CodeLPH · Orthopedic
Decision DateJan 4, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Exactech Novation InteGrip Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. The Exactech Novation InteGrip Acetabular Augment is affixed to the mating acetabular shell using PMMA bone cement. Therefore, acetabular shells with HA coating must not be used with InteGrip Acetabular Augments. The assembled construct is intended for press-fit fixation.

Device Story

Titanium alloy acetabular augments designed to address segmental acetabular deficiencies during primary or revision hip replacement. Augments are affixed to compatible acetabular shells using PMMA bone cement; the final construct is intended for press-fit fixation. Used by orthopedic surgeons in clinical settings. Modification expands size range to accommodate Novation Crown Cup acetabular shells (60-68mm). Provides structural support as an alternative to allograft.

Clinical Evidence

Bench testing only. Engineering evaluations were conducted to verify that the geometric features of the modified devices correspond to compatible acetabular shells and are dimensionally and functionally similar to the predicate device.

Technological Characteristics

Material: Titanium alloy. Design: Acetabular augments with screw holes and specific height configurations. Fixation: PMMA bone cement to shell; press-fit to bone. Sterilization: Gamma radiation (SAL 10^-6). Compatibility: Compatible with Novation Crown Cup acetabular shells (60-68mm); incompatible with HA-coated shells.

Indications for Use

Indicated for skeletally mature patients undergoing primary or revision hip replacement surgery requiring prosthetic alternatives to structural allograft for segmental acetabular deficiencies. Contraindicated for use with HA-coated acetabular shells.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113609 Page 1/2 ### Exactech® Novation® InteGrip™ Acetabular Augments Special 510(k) - 510(k) Summary of Safety and Effectiveness Exactech® Inc. Sponsor: 2320 N.W. 66th Court Gainesville, FL 32653 > Phone: (352) 377-1140 Fax: (352) 378-2617 FDA Establishment Number 1038671 Vladislava Zaitseva Contact: Regulatory Affairs Specialist Date: January 4, 2012 Trade or Proprietary or Model Name(s): Exactech® Novation® InteGrip™ Acetabular Augments Common Name: Acetabular Augment #### Classification Name: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented (21 CFR 888.3358, Device Class: II, Product Code: LPH) #### Information on devices to which substantial equivalence is claimed: 510/k) Number Trade or Proprietary or Model Name | 510(K) Number | Trade or Proprietary or Model Name | Manufacturer | |---------------|------------------------------------------------|---------------| | K101761 | Exactech Novation InteGrip Acetabular Augments | Exactech, Inc | ### Indications for Use: The Exactech Novation InteGrip Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. The Exactech Novation InteGrip Acetabular Augment is affixed to the mating acetabular shell using PMMA bone cement. Therefore, acetabular shells with HA coating must not be used with InteGrip Acetabular Augments. The assembled construct is intended for press-fit fixation. {1}------------------------------------------------ K113609 Page 2/2 # Exactech® Novation® InteGrip™ Acetabular Augments Special 510(k) - 510(k) Summary of Safety and Effectiveness ### Device Description: The proposed Exactech Novation InteGrip Acetabular Augments are a modification of the Exactech InteGrip Acetabular Augments cleared through premarket notification #K101761. The predicate and proposed devices have the same intended use and the same basic fundamental scientific technology. The modified devices share the following similarities with predicate devices: - . Indications for Use - Design Features (e.g. screw holes, height configurations) � - Material (titanium alloy) . - . Shelf Life (5 years) - Packaging and sterilization materials and processes (gamma radiation sterilization to a sterility assurance level of 10°) The proposed devices include the following design change: - Expand the current scope of augments to include sizes that are compatible with . the Novation Crown Cup acetabular shell sizes 60-68mm (both plasma spray and InteGrip). ### Substantial Equivalence Conclusion: The following engineering analyses were conducted to demonstrate substantial equivalence of the proposed device to the predicate Exactech Novation InteGrip Acetabular Augments: - An engineering evaluation to determine that the geometric features of the . proposed devices correspond to the compatible acetabular shells and are dimensionally and functionally similar to the predicate devices. In addition to the design similarities, the results of engineering analyses demonstrate the proposed device is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 JAN - 4 2012 Exactech® Inc. % Ms. Vladislava Zaitseva 2320 N.W. 66th Court Gainesville, FL 32653 Re: K113609 Trade/Device Name: Exactech® Novation® InteGrip™ Acetabular Augments Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH Dated: December 5th, 2011 Received: December 6ta, 2011 Dear Ms. Zaitseva: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Vladislava Zaitseva Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Einol Keith Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Exactech® InteGrip™ Acetabular Augments Special 510(k) - Indications for Use 510(k) Number: K113609 # Device Name: Exactech® InteGrip™ Acetabular Augments #### INDICATIONS The Exactech Novation InteGrip Acetabular Augments are indicated for use in skeletally mature individuals undergoing primary or revision surgery for hip replacement and whose orthopedic surgeon desires a prosthetic alternative to structural allograft in cases of segmental acetabular deficiencies. The Exactech Novation InteGrip Acetabular Augment is affixed to the mating acetabular shell using PMMA bone cement. Therefore, acetabular shells with HA coating must not be used with InteGrip Acetabular Augments. The assembled construct is intended for press-fit fixation. | Prescription Use | X | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | | and/or | | | Over-The-Counter Use | | | (21 CFR 807 Subpart C) | | Please do not write below this line - use another page if needed. Concurrence of CDRH, Office of Device Evaluation (ODE) BBRS Lr (Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113609
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%