K113533 · Baxano, Inc. · HAE · Apr 11, 2012 · Neurology
Device Facts
Record ID
K113533
Device Name
NEURO CHECK DEVICE WITH IO-FLEX WIRE
Applicant
Baxano, Inc.
Product Code
HAE · Neurology
Decision Date
Apr 11, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4840
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K113533 · Apr 11, 2012
NEURO CHECK DEVICE WITH IO-FLEX WIRE
Baxano, Inc.
Post-market complaint data
Post-market review of device complaints was used to identify the need for design modifications to the iO-Flex Wire to improve fatigue stress resistance.
Post-market review of iO-Flex Wire complaints; Retrospective review of post-market complaint data
Patients who experienced issues with the predicate iO-Flex Wire
Not applicable for this study
Identification of fatigue-related failure modes
Indications for Use
The Baxano Inc. MicroBlade Shaver and Accessories are designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.
Device Story
MicroBlade Shaver system comprises proximal handle, rigid shaft, and flexible cutting platform; used for spinal surgery. Accessories include Probe (telescoping handle, cannula, deployable catheter) for site access and iO-Flex Wire placement; Distal Handle for wire control. System allows manual control of MicroBlade Shaver and Neuro Check devices in neural foramen. Device is manual; operated by surgeon in OR. iO-Flex Wire (Nitinol) modified with heat treatment and chemical etching to improve fatigue stress resistance. Device enables access to compromised neural areas in spinal column; facilitates decompression procedures. Output is mechanical tissue/bone removal; surgeon visualizes site and controls device manually. Benefits include improved access to spinal neural areas and enhanced fatigue resistance of wire component.
Clinical Evidence
No clinical data. Substantial equivalence supported by mechanical bench top comparison testing of modified vs. non-modified iO-Flex Wire designs, including tensile strength, fatigue cycling, and repeated connectivity testing.
{1}------------------------------------------------
APR 1 1 2012
# 510(K) SUMMARY (21 CFR 807.92)
# MICROBLADE SHAVER DEVICE AND ACCESSORIES
510(k) Owner:
从113533
Baxano, Inc. 655 River Oaks Parkway San Jose, CA 95134 Tel: 408-514-2200 Fax: 408-514-2201
Contact Person:
Edward J. Sinclair Tel: 408-514-2220 E-mail: esinclair@baxano.com
Date Prepared:
02-22-2012
Trade Name:
MicroBlade Shaver and Accessories
Common Name:
Classification:
Class II (21 CFR 882.4840)
Rongeur, Manual
Product Code:
HAE
Predicate Device Information:
Baxano MicroBlade® Shaver Device and Accessories (K063231, K080494, K100958)
Device Description:
The MicroBlade Shaver device is comprised of a proximal handle, rigid shaft, and flexible cutting platform used for cutting and biting soft tissue and bone. The Probe Accessory is comprised of a telescoping proximal handle, cannula, and deployable catheter used to access the decompression site and place the accessory iO-Flex® Wire. The Distal Handle Accessory is comprised of a handle to accommodate the iO-Flex Wire using a wire locking mechanism and
Page 1 of 4
{2}------------------------------------------------
K113533
P.2 of 4
wire capture receptacle and allows manual control of the MicroBlade Shaver and Neuro Check® devices.
This 510(k) pre-market notification describes a modification solely to the iO-Flex Wire. The modifications include an additional surface finish process that employs heat treatment and a subsequent chemical etching process applied to the proximal wire section intended to improve the fatigue stress resistance. There were no changes in the wire material and a small change to one nominal dimension.
The modified iO-Flex Wire is manufactured from nickel-titanium ("Nitinol") and is designed to connect to the distal end of the Neuro Check Device and MicroBlade® Shaver device for placement and allows manual control of the MicroBlade Shaver and Neuro Check devices in the neural foramen. The iO-Flex Wire is packaged with the Neuro Check Device and available as a separately packaged accessory device.
The Baxano Inc. MicroBlade Shaver and Accessories are designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.
The MicroBlade Shaver is designed to be flexible, with a low profile to allow access to compromised neural areas in the spinal column. The accessories include a Probe, Distal Handle and the iO-Flex Wire. The fundamental scientific technology and operating controls remain unchanged from the predicate iO-Flex Wire design.
Predicate Device Comparison
Intended Use:
Technological
Characteristics:
Modifications were made to the iO-Flex Wire as a result of post market review of iO-Flex Wire complaints. Specific changes and the reason for change are described in the table below:
Page 2 of 4
{3}------------------------------------------------
にいるちろ �-3 弯-4
:
| | Predicate<br>iO - Flex Wire<br>Specifications | Subject<br>iO - Flex Wire<br>Specifications | Reason for<br>Modification |
|--------------------------------------|--------------------------------------------------------------------------|----------------------------------------------|----------------------------------------------------------------------|
| Dimensional | | | |
| Overall iO-<br>Flex Wire<br>Length | 26.0" | Same as<br>Predicate | n/a |
| Barrel Length | 0.030" | Same as<br>Predicate | n/a |
| Proximal<br>Barrel | Ø 0.0350" | Ø 0.0345" | Manufacturing<br>Tolerance |
| Distal Tip<br>Angulation | 11 Degree | Same as<br>Predicate | n/a |
| Tapered<br>Section | Ø 0.023" | Same as<br>Predicate | n/a |
| Surface Treatment | | | |
| Surface<br>Treatment | Finish to be<br>within 8 microns<br>or better | Heat<br>Treatment and<br>Chemical<br>Etching | Surface<br>Treatment to<br>increase fatigue<br>cycling<br>resistance |
| Tensile Strength and Fatigue Cycling | | | |
| Tensile<br>Strength | ≥ 38 lbs. | Same as<br>Predicate | n/a |
| Fatigue Cycling | 12 cycles @ 38<br>lbs. of tensile<br>load without<br>gross failures | Same as<br>Predicate | n/a |
| | 180 cycles @ 6<br>lbs. of axial<br>load of<br>simulated<br>decompression | Same as<br>Predicate | n/a |
| Repeated<br>Connectivity | 12.<br>attach/detach<br>cycles without<br>failure | Same as<br>Predicate | n/a |
{4}------------------------------------------------
Performance Data:
Substantial Equivalence: Mechanical bench top comparison testing was conducted on the modified and non-modified iO-Flex Wire designs to verify that the modified iO-Flex Wire meets current design specifications and performance characteristics, based upon the intended use.
The MicroBlade Shaver Device and Accessories are substantially equivalent to the MicroBlade Shaver and Accessories (K100958, cleared on July, 23, 2010). The MicroBlade Shaver and Accessories have the same indications for use and fundamental scientific technology as their predicate.
The indications for use, technological characteristics and operating controls remain the same. Non-clinical performance test results of the modified iO-Flex Wire do not raise new questions of safety or effectiveness.
Conclusions:
Baxano has determined, based on the same intended use, technological characteristics, operating controls and non-clinical performance testing that the modified iO-Flex Wire is substantially equivalent to the predicate device.
Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relotes to whether the product addressed in this submission may be lawfully market approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any potent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The present submission and statements therefore should not be construed as affecting or relating to the scope of any patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird symbol, with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
APR 1 1 2012
Baxano, Inc. c/o Mr. Edward J. Sinclair Vice President, Clinical, Regulatory and Quality Affairs 655 River Oaks Parkway San Jose, CA 95134
Re: K113533
Trade/Device Name: MicroBlade Shaver Device and Accessories Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: March 9, 2012 Received: March 12, 2012
Dear Mr. Sinclair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal-Register.
{6}------------------------------------------------
## Page 2 - Mr. Edward J. Sinclair
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Signature
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health -
Enclosure
{7}------------------------------------------------
K//3533
Indications for Use
510(k) Number (if known):
Device Name: Baxano MicroBlade® Shaver Device and Accessories
Indications For Use:
The Baxano Inc. MicroBlade Shaver and Accessories are designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.
Prescription Use __ x (Part 21 CFR 801 Subpart D)
510(k) Number.
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
John Doucett
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
KII3533
**Prescription Use**
**(Per 21 CFR 801.109)**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.