K063231 · Baxano, Inc. · HAE · Apr 16, 2007 · Neurology
Device Facts
Record ID
K063231
Device Name
MICROBLADE SHAVER AND ACCESSORIES
Applicant
Baxano, Inc.
Product Code
HAE · Neurology
Decision Date
Apr 16, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Baxano, Inc. Microblade Shaver and Accessories are designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.
Device Story
Microblade Shaver and Accessories are manual surgical instruments used to access neural foramen and decompress targeted areas of the spinal column. The system includes a flexible, low-profile stainless steel shaver, a probe, a guide, a needle wire, and a distal handle. During surgery, the surgeon pulls the shaver into the foramen using the distal handle; tissue and bone are removed by pulling up on the handle to engage the cutting mechanism. The device is intended for use by surgeons in open decompression procedures. It provides a method for accessing compromised neural areas where exposure is difficult, allowing for tissue removal without damaging neural structures.
Clinical Evidence
No clinical trials were performed. Evidence consists of mechanical performance testing to verify design specifications and cadaver studies conducted by physicians. Cadaver study results demonstrated that the device provided tissue removal comparable to commercially available rongeurs with no visible damage to neural structures and ease of use comparable to the standard of care.
Indicated for accessing, cutting, and biting soft tissue and bone during spinal column surgery.
Regulatory Classification
Identification
A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.
Predicate Devices
Codman Laminectomy Shaver
Integra Ruggles Instrument
Zeppelin Laminectomy Rongeur
Ellman Disc FX System (accessories)
Submission Summary (Full Text)
{1}------------------------------------------------
K063231
APR 16 2007
Image /page/0/Picture/2 description: The image shows the word "BAXANO" in a bold, sans-serif font, enclosed in a rectangular box. A stylized, curved line bisects the "X" in "BAXANO", adding a unique visual element to the design. The overall design is simple, bold, and modern, with a focus on the word "BAXANO" as the central element.
## 510(K) SUMMARY (21 CFR 807.92)
## MICROBLADE SHAVER AND ACCESSORIES
| 510(k) Owner: | Baxano, Inc.<br>2660 Marine Way, Suite B<br>Mountain View, CA 94043<br>Tel: 650-937-1400<br>Fax: 650-937-1410 |
|---------------|---------------------------------------------------------------------------------------------------------------|
|---------------|---------------------------------------------------------------------------------------------------------------|
- Contact Person: Jefferey Bleam Tel: 650-937-1400 ext. 408 E-mail: jbleam@baxano.com
- Date Prepared: October, 2006
Trade Name: Microblade Shaver and Accessories
- Manual rongeur for cutting or biting bone in the skull or spinal Common Name: column
Classification: Manual rongeur (per 21 CFR section 882.4840)
Predicate Devices: Codman Laminectomy Shaver, Integra Ruggles Instrument, Zeppelin Laminectomy Rongeur Ellman Disc FX System (accessories)
- Device Description: The Microblade Shaver and Accessories are used to access the neural foramen and decompress targeted areas. The Microblade Shaver is pulled into the foramen with the Distal Handle and tissue is removed by pulling up on the distal handle to cut the targeted soft tissue and bone. The Accessories include a Probe, Guide, Needle Wire and Distal Handle.
- Intended Use: The Baxano, Inc. Microblade Shaver and Accessories are designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.
{2}------------------------------------------------
| Substantial | |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Equivalence: | The Microblade Shaver and Accessories have the same indications<br>for use as the predicates; to access, cut, and bite soft tissue and<br>bone involving the spinal column. In addition, the Accessories<br>have the same intended uses; to access, guide, and probe in<br>neurosurgical applications. Any differences between the<br>Microblade Shaver and Accessories and the predicates are specific<br>in design and do not raise new questions of safety or effectiveness.<br>The device and the predicates are labeled for open decompression<br>procedures. |
| Technological | |
| Characteristics: | The Microblade Shaver is made of stainless steel and is flexible,<br>with a low profile, to allow access to compromised neural areas<br>when exposure is otherwise difficult to obtain. These are the same<br>materials and technological characteristics as the predicate devices.<br>The Accessories include a probe, guide and wire substantially<br>equivalent to probes, guides and wires commercially available for<br>discectomy procedures in the lumbar spine. |
| Non-Clinical | |
| Performance Data: | Mechanical performance tests were conducted to verify that the<br>device meets design specifications and intended performance<br>characteristics, based on the application for removing bone in<br>compromised neural areas. |
| | The Microblade Shaver and Accessories were used in multiple<br>cadaver studies by physicians skilled in procedures for<br>decompressing neural foramen and lateral recess. Results<br>demonstrated that the Microblade Shaver and Accessories<br>provided comparable tissue removal with no visible damage to<br>neural structures, and an ease of use comparable to the standard of<br>care, when compared to commercially available rongeurs. |
| Conclusions: | Baxano has determined, based on the performance testing and<br>cadaver studies, that the Microblade Shaver and Accessories<br>conform to the design specifications and are substantially<br>equivalent to the predicate devices for accessing and removing soft<br>tissue and bone in the spinal column. |
Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not interpreted as an admission or any other type of evidence in pay patent in any patent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The procent ubbission and statements therein therefore should not be construed as affecting or relating to the scope of any patent of passent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three wavy lines, which represent the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Baxano, Inc. % Rockwell & Associates Ms. Sharon Rockwell VP, Regulatory & Clinical Affairs 5582 Chalon Road Yorba Linda, California 92886
APR 1 6 2007
Re: K063231
Trade/Device Name: Microblade Shaver and Accessories Regulation Number: 21 CFR 882.4840 Regulation Name: Manual rongeur Regulatory Class: II Product Codc: HAE Dated: March 15, 2007 Received: March 30, 2007
Dear Ms. Rockwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Sharon Rockwell
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
For
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Microblade Shaver and Accessories
Indications for Use: The Baxano, Inc. Microblade Shaver and Accessories are designed for accessing, cutting, and biting soft tissue and bone during surgery involving the spinal column.

(Division Sign-Off) Division of General, Restorative, and Neurological Devices 11063231
Prescription Use
(Part 21 CFR 801 Subpart D)
X
AND/OR
510(k) Number_
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.