SPECTRA OPTIA APHERESIS SYSTEM

K113480 · Caridianbct, Inc.. · LKN · Jan 20, 2012 · GU

Device Facts

Record IDK113480
Device NameSPECTRA OPTIA APHERESIS SYSTEM
ApplicantCaridianbct, Inc..
Product CodeLKN · GU
Decision DateJan 20, 2012
DecisionSESE
Submission TypeSpecial
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Spectra Optia® Apheresis System, a blood component separator, is intended for use in therapeutic plasma exchange.

Device Story

Spectra Optia Apheresis System is a centrifugal blood component separator used for therapeutic plasma exchange. The system consists of a machine and a disposable Exchange Set. It processes whole blood to separate cellular components from plasma, facilitating the removal and replacement of patient plasma with donor plasma. The device is operated by healthcare professionals in clinical settings. Embedded software monitors the procedure; recent updates enhance the system's ability to detect potential platelet loss during repeated RBC spillover events, providing earlier user notifications to prevent excessive platelet removal. The system provides real-time alerts to the clinician, who then takes remedial action to ensure patient safety and optimize therapeutic outcomes.

Clinical Evidence

No clinical data were necessary. The modifications did not impact the performance specifications of the system's therapeutic apheresis protocols. Evidence consists of non-clinical bench testing, including reliability, usability, robustness, and regression testing.

Technological Characteristics

Centrifugal blood component separator. System includes a machine and a disposable Exchange Set. Software-controlled separation process. Connectivity/interoperability details not specified. Sterilization method not specified.

Indications for Use

Indicated for patients requiring therapeutic plasma exchange to remove and replace plasma for the treatment of adverse hematologic and other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 2 0 2012 # 510(k) Summary | Owner/Manufacturer: | CaridianBCT, Inc.<br>10811 W. Collins Avenue<br>Lakewood, Colorado 80215 | |----------------------|--------------------------------------------------------------------------------------| | Contact Person: | Tina O'Brien<br>Senior Regulatory Affairs Specialist<br>Office Phone: (303) 239-2082 | | Summary Date: | November 18, 2011 | | Trade Name: | Spectra Optia® Apheresis System | | Common Name: | Therapeutic Apheresis System | | Classification Name: | Separator, Automated, Blood Cell and Plasma<br>Therapeutic | | Product Code: | LKN | | Predicate Device: | Spectra Optia Apheresis System | #### Device Description: The Spectra Optia® Apheresis System is a centrifugal system that separates whole blood into its cellular and plasma components. The system is used therapeutically to remove and replace the removed plasma with healthy donor plasma (i.e. plasma exchange) from patients who suffer from adverse hematologic and other conditions. The system includes a disposable Exchange Set through which the patient's blood passes, and a machine that controls the separation of blood and the removal and replacement of plasma. The Spectra Optia system's embedded software has been enhanced to reduce the likelihood of excessive platelet loss/removal during Therapeutic Plasma Exchange (TPE) procedures when there are repeated RBC spillover alerts and no remedial action is taken. #### Intended Use: The Spectra Optia " Apheresis System, a blood component separator, is intended for use in therapeutic plasma exchange. #### Technological Comparison: The base technology of the Spectra Optia system is not affected by this software change to minimize the potential for spillover conditions and to provide earlier detection and notification to the user of a potential platelet loss if appropriate action is not taken. There is no change to the disposable Exchange Set. #### Discussion of Non-clinical Data: The correct functioning of the software modifications was verified through the following activities: ez 1/2 {1}------------------------------------------------ - Reliability: Reliability was assessed using tests to simulate expected usage . profiles. - Usability was verified with simulated use runs on the affected . Usability: protocol. - Boundary condition testing was conducted to explore the Robustness: . software for expected behavior. Boundary conditions address critical values, maximum and minimum values, as well as values just inside/outside established boundaries. - Regression: Specific regression testing was conducted related to the software . updates for the platelet loss mitigation. ### Discussion of Clinical Data: Clinical validation data were not necessary, as the modifications did not impact the performance specifications of the system's therapeutic apheresis protocols. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human services. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Tina O'Brien Senior Regulatory Affairs Specialist Caridian BCT, Inc. 10810 West Collins Avenue LAKEWOOD CO 80215 JAN 2 0 2012 Re: K113480 Trade/Device Name: Spectra Optia® Apheresis System Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKN Dated: December 21, 2011 Received: December 23, 2011 Dear Ms. O'Brien: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass battle in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ # Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Benjamin K. Evanko Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for CaridianBCT. The logo consists of a stylized graphic to the left of the text "CaridianBCT". The graphic is made up of five curved lines that converge at the left and then diverge as they move to the right. The text is in a sans-serif font and is black. Spectra Optia®Apheresis System Software Version 6.1 Special 510(k) Submission #### Intended Use Statement 4 510(k) Number: _ Device Name: Spectra Optia® Apheresis System # Intended Use The Spectra Optia® Apheresis System, a blood component separator, is intended for use in therapeutic plasma exchange. Prescription Use: YES (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use: NO (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) n Slanluctive, Gastro-Renal, and
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...