GEL-BLOCK EMBOLIZATION PLEDGETS

K113266 · Vascular Solutions, Inc. · KRD · May 14, 2012 · Cardiovascular

Device Facts

Record IDK113266
Device NameGEL-BLOCK EMBOLIZATION PLEDGETS
ApplicantVascular Solutions, Inc.
Product CodeKRD · Cardiovascular
Decision DateMay 14, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Gel-Block Embolization Pledgets are intended for use in embolization of hypervascularized tumors and arteriovenous malformations (AVMs).

Device Story

Gel-Block Embolization Pledgets are resorbable porcine gelatin devices designed to provide a mechanical barrier to blood flow in the vasculature. The device consists of two radially-compressed cylindrical gelatin pledgets housed in individual transparent delivery assemblies. The assembly features a luer lock for flushing and a vented cap to prevent premature pledget evacuation. The device is used by clinicians to embolize hypervascularized tumors and arteriovenous malformations. During the procedure, the clinician attaches the delivery assembly to an in-place catheter; the pledgets are delivered into the target vessel to induce occlusion. The device functions by swelling upon contact with blood, similar to predicate microspheres, to block flow. The resorbable nature of the porcine gelatin allows for temporary embolization. Clinical benefit is derived from the reduction of blood flow to target lesions, aiding in tumor management or AVM treatment.

Clinical Evidence

No human clinical data provided. Evidence consists of bench testing (swelling, digestibility, flushing, deliverability, formaldehyde residuals) and a GLP animal study. The animal study involved nine test and four control (Embosphere) sheep over 12 weeks. Angiography and histopathology confirmed embolic effect in renal arteries for at least four weeks, with renal infarcts observed in test animals similar to controls. No non-targeted embolization or systemic toxicity was observed. Biocompatibility testing was performed per ISO 10993-1.

Technological Characteristics

Resorbable porcine gelatin material; cylindrical form factor; radially-compressed for catheter delivery. Mechanical swelling mechanism. Biocompatibility compliant with ISO 10993-1. Sterilization method not specified. No electronic or software components.

Indications for Use

Indicated for embolization of hypervascularized tumors and arteriovenous malformations (AVMs) in patients requiring vascular occlusion.

Regulatory Classification

Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # 510(k) Summary Date Prepared: May 9, 2012 510(k) Number: K113266 #### Submitter's Name / Contact Person #### Manufacturer Vascular Solutions, Inc. 6464 Sycamore Court Minneapolis, MN 55369 USA Tel: 763-656-4300; Fax: 763-656-4250 Establishment Registration # 2134812 Contact Person Jennifer Ruether Sr. Regulatory Product Specialist ## General Information Trade Name Common / Usual Name Classification Name Predicate Devices Gel-Block Embolization Pledgets . Device, vascular, for promoting embolization 870.3300; KRD; vascular embolization device; Class II K991549: Embosphere Microspheres (BioSphere Medical, Inc.) K052742: QuadraSphere Microspheres (BioSphere Medical, Inc.) ### Device Description The Gel-Block Embolization Pledgets are an embolic device consisting of two radially-compressed gelatin pledgets that can be delivered through a catheter system. The pledgets are stored individually within a transparent delivery assembly that has a luer lock on the proximal end and a vented cap on the distal end. The luer lock allows for attachment to syringes for flushing, while the vented cap allows the flushing to occur without pledget evacuation from the delivery assembly. After flushing, the vented cap is removed to expose a threaded luer that connects to the hub of an in-place delivery catheter. The Gel-Block Embolization Pledgets consist of the following components: - Two gelatin pledgets, in individual delivery assemblies - One syringe for pledget delivery . {2}------------------------------------------------ # Intended Use / Indications The Gel-Block Embolization Pledgets are intended for use in embolization of hypervascularized tumors and arteriovenous malformations (A VMs). # Technological Characteristics The Gel-Block Embolization Pledgets are similar in design to the predicate devices, as all of them provide a mechanical barrier to blood flow in the vasculature and are delivered using catheters. The ability of the Gel-Block to swell is similar to that of the QuadraSpheres. The subject device is available in three consistent sizes, while the predicates are available in defined size ranges. The subject and predicate devices also differ in shape; because the Gel-Block Pledgets are larger than the predicate microspheres, they are cylindrical in shape for easy delivery through a catheter. The Gel-Block Embolization Pledgets also differ from both predicate devices in terms of materials, and therefore degradation; the Embospheres consist of a non-resorbable acrylic polymer impregnated with porcine gelatin, while the Gel-Block is made wholly from resorbable porcine gelatin. The QuadraSpheres consist of a sodium acrylate-vinyl alcohol co-polymer. The technological differences between the subject and predicate devices have been evaluated through biocompatibility, bench, and animal testing to provide evidence of safe and effective use of the Gel-Block Embolization Pledgets, thereby establishing substantial equivalence to the predicate devices. # Substantial Equivalence and Summary of Studies The Gel-Block Embolization Pledgets are substantially equivalent to the specified predicate devices based on comparisons of the device functionality, technological characteristics, and Indications for Use. The device design has been verified through the following tests: - . Swelling capability - Pepsin digestibility . - Flushing capability . - ◆ Deliverability - . Formaldehyde residuals Biocompatibility was verified through the following test strategy compliant with ISO 10993-1: - Cytotoxicity - . Sensitization - . Irritation/intracutaneous reactivity - Acute systemic toxicity ● - . Subchronic toxicity - . Genotoxicity - Hemocompatibility . - . Implantation A GLP animal study was conducted on nine test and four control (Embosphere Microspheres) adult female sheep over a period of 12 weeks. Angiography and/or histopathology confirmed that the Gel-Block Embolization Pledgets provided an embolic effect in renal arteries for at least four weeks. All test animals had renal infarct indicative of successful embolization for a duration similar to that of the control animals. There were no observed incidents of non-targeted embolization in nearby vessels. No signs of systemic toxicity were observed. Anemia was observed in four test animals. However, in vitro {3}------------------------------------------------ hemocompatibility testing did not produce evidence of significant hemolysis, thus mitigating this concern. : Results of the design verification, animal study, and biomaterial tests did not raise new safety or performance questions and support the substantial equivalence of the Gel-Block Embolization Pledgets to the predicate devices. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 WAY. 1 4 2012 Vascular Solutions, Inc. c/o Jennifer Ruether Senior Regulatory Product Specialist 6464 Sycamore Court Minneapolis. MN 55369 Re: K113266 Trade/Device Name: Gel-Block Embolization Pledgets Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular embolization device Regulatory Class: Class II Product Code: KRD Dated: March 19, 2012 Received: March 20, 2012 ## Dear Ms. Ruether: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Ms. Jennifer Ruether Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, M.S. Killilea Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use #### K113266 510(k) Number (if known): Device Name: Gel-Block Embolization Pledgets Indications for Use: The Gel-Block Embolization Pledgets are intended for use in embolization of hypervascularized tumors and arteriovenous malformations (AVMs). Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) . A. Hillel Page 1 of 1 (Posted November 13, 2003) (Division Sign-Off) Division of Cardiovascular Devices 长川多266 510(k) Number Gel-Block Embolization Pledgets Traditional 510(k) - Page 4 of 2070
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...