FDES100(TENS AND EMS STIMULATOR, FDES102(ED402) TENS STIMULATOR
K113010 · Famidoc Technology Co., Ltd. · IPF · Dec 21, 2012 · Physical Medicine
Device Facts
Record ID
K113010
Device Name
FDES100(TENS AND EMS STIMULATOR, FDES102(ED402) TENS STIMULATOR
Applicant
Famidoc Technology Co., Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Dec 21, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For TENS mode 1. Symptomatic relief of chronic intractable pain 2. Post traumatic pain 3. post surgical pain For EMS mode 1. Relaxation of muscle spasm. 2. Increase of local blood flow circulation 3. Prevention or retardation of disuse atrophy 4. Muscle re-education 5. Maintaining or increasing range of motion 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis FDES102(ED402) TENS Stimulator 1. Symptomatic relief of chronic intractable pain 2. Post traumatic pain 3. Post surgical pain FDES103(ED403) EMS Stimulator 1. Relaxation of muscle spasm. 2. Increase of blood flow circulation 3. Prevention or retardation of disuse atrophy 4. Muscle re-education 5. Maintaining or increasing range of motion 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
Device Story
FDES101/102/103 series are portable, battery-powered TENS/EMS stimulators. Devices deliver electrical current to nerves/muscles via skin-applied electrodes. Users operate device via keypad and LCD interface to select treatment programs and adjust intensity (0-60mA). Microprocessor-controlled output channels provide bi-phase square wave stimulation. Used in physiotherapy clinics for pain management and muscle rehabilitation. Output parameters include adjustable pulse width (50-300uS) and frequency (0.5-150Hz). Safety features include automatic overload, over-current, no-load trips, and automatic shut-off. Benefits include non-invasive pain relief and muscle stimulation for post-surgical recovery and atrophy prevention.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing and compliance with voluntary safety standards including IEC 60601-1 (General safety), IEC 60601-1-2 (EMC), IEC 60601-1-4 (Programmable electrical medical systems), and IEC 60601-2-10 (Nerve/muscle stimulators). Biocompatibility of patient-contacting materials (silica gel electrodes, ABS housing) confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Portable, battery-powered (4x1.5V AAA) stimulator. ABS plastic enclosure. Two independent output channels. Bi-phase square wave output. Pulse width 50-300uS; frequency 0.5-150Hz. Microprocessor-controlled. Type BF applied part. Compliant with IEC 60601-1, 60601-1-2, 60601-1-4, 60601-2-10. Biocompatible materials (ISO 10993).
Indications for Use
Indicated for patients requiring physiotherapy for symptomatic relief of chronic intractable, post-traumatic, or post-surgical pain (TENS); or for muscle spasm relaxation, increased local blood flow, prevention/retardation of disuse atrophy, muscle re-education, maintenance/increase of range of motion, and prevention of venous thrombosis (EMS).
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
KIISOID
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
# DEC 2 1 2012
# 510(k) SUMMARY
[As required by 21CFR 807.92]
# FDES101(ED401) Series Electro-Stimulator, K (
# 1. Submitter's Information [21CFR 807.92(a)(1)]
| Company Name: | Famidoc Technology Co., Ltd | |
|------------------|-----------------------------------------------------|--|
| Street Address: | East 2/F Zhixiang Building, 71 Block Baoan District | |
| City: | Shenzhen | |
| State/ Province: | Guangdong | |
| Country: | China | |
| Telephone: | +86(755) 27864342 | |
| Fax: | +86(755) 27864151 | |
| Contact Person: | Cao Liang | |
| Contact Title: | General Manager | |
| Contact Email: | leon@famidoc.com | |
#### Trade Name, Common Name, Classification [21CFR 807.92(a)(2)] 2.
| a) Trade Name: | FDES101(ED401) TENS and EMS Stimulator |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Electro-Stimulator or Electrical Stimulator |
| Classification Name: | Stimulator, Muscle, Powered<br>per 21 CFR § 890.5850;<br>Stimulator, Nerve, Transcutaneous. For pain relief<br>per 21 CFR § 882.5890 |
| Device Class: | Class II |
| Product Code: | IPF, GZJ |
| a) Trade Name: | FDES102(ED402) TENS Stimulator |
| Common Name: | TENS or TENS Device |
| Classification Name: | Stimulator, Nerve, Transcutaneous. For pain relief |
Section 05-Page 1 of 9
{1}------------------------------------------------
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
| Device Class: | per 21 CFR § 882.5890 |
|----------------------|--------------------------------|
| | Class II |
| Product Code: | GZJ |
| | |
| b) Trade Name: | FDES103(ED403) TENS Stimulator |
| Common Name: | EMS or EMS Device |
| Classification Name: | Stimulator, Muscle, Powered |
| | per 21 CFR § 890.5850; |
| Device Class: | Class II |
| Product Code: | IPF |
#### Identification of Predicate Device(s)[21 CFR 807.92(a)(3)] a)
| PREDICATE DEVICES | |
|----------------------------|----------------------------------------|
| Manufacturer | Shenzhen Dongdixin Technology Co., Ltd |
| Legally<br>Marketed Device | MT9000 Series Electro-Stimulator |
| 510 (K) Number | K093138 |
There are basically the same between the FDES101(ED401) Series Electro-Stimulator and the predicate device which would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended . use. The detailed differences between them please see below:
{2}------------------------------------------------
・
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
・
· ·
·
| No. | Comparison item | New device | Predicate device | Judgment |
|------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| 1 | 510K# | K113010 | K093138 | -- |
| 2 | Device Name | FDES101(ED401) TENS<br>and EMS Stimulator,<br>FDES102(ED402) TENS<br>Stimulator,<br>FDES103(ED403) EMS<br>Stimulator | MT9000 Combo<br>TENS/EMS/IF/MIC<br>Stimulator<br>MT9001 TENS<br>Stimulator<br>MT9002 EMS Stimulator<br>MT9003 IF Stimulator<br>MT9004<br>MICROCURRENT<br>Stimulator | -- |
| 3 | Manufacturer | Famidoc Technology Co.,<br>Ltd. | Shenzhen Dongdixin<br>Technology Co., Ltd. | -- |
| 4 | Power Source | DC 6V, 4×1.5V AAA<br>batteries | 9V Battery | SE |
| 4.1 | -Method of Line<br>current isolation | Battery Supply N/A | Battery Supply N/A | -- |
| 4.2 | - Patient Leakage<br>Current<br>-Normal condition<br>-Single fault<br>condition | 3.0uA<br>5.8uA | 0.61uA<br>0.68uA | SE, meet<br>the<br>requiremen<br>ts of IEC<br>60601-1 |
| 5 | Classification | | | -- |
| 5.1 | Type of protection<br>against electric<br>shock | Internally powered<br>equipment | Internally powered<br>equipment | SE |
| 5.5 | Degree of<br>protection against<br>electric shock | Type BF applied part | Type BF applied part | SE |
| 5.3 | Device Class | Class II | Class II | SE |
| 6 | Number of treatment<br>programs | FDES101(ED401)<br>30<br>FDES102(ED402)<br>15<br>FDES103(ED403)<br>15 | MT9000<br>4<br>MT9001<br>1<br>MT9002<br>1<br>MT9003<br>1<br>MT9004<br>1 | |
| No. | Comparison item | New device | Predicate device | Judgment |
| 7 | Number of Output<br>Channels | 2 | 2 | SE |
| 7.1 | - Synchronous or<br>Alternating? | Synchronous and<br>Alternating | Synchronous and<br>Alternating | SE |
| 7.2 | - Method of<br>Channel<br>Isolation? | By electrical circuit and<br>software | By electrical circuit and<br>software | SE |
| 8 | Constant Current?<br>Constant Voltage? | Yes<br>No | Yes<br>No | SE |
| 9 | Software/Firmware/<br>Microprocessor<br>Control? | Yes | Yes | SE |
| 10 | Automatic Overload<br>Trip?<br>Automatic Over<br>Current Trip? | Yes<br>Yes | Yes<br>Yes | SE |
| 11 | Automatic No Load<br>Trip? | Yes | Yes | SE |
| 12 | Automatic Shut off? | Yes | Yes | SE |
| 13 | Patient Override<br>Control? | No | No | SE |
| 14 | Indication function | | | -- |
| 14.1 | -On/Off Status? | Yes | Yes | SE |
| 14.2 | -Voltage/Current<br>Level? | Yes | Yes | SE |
| 14.3 | -Low Battery? | Yes | Yes | SE |
| 15 | Timer Range<br>(minutes) | 0-60 minutes | 0-60 minutes | SE |
| 16 | Housing materials | Plastic (ABS) enclosure | Plastic (ABS) enclosure | SE |
| 17 | Target population | Patients who need<br>physiotherapy treatment | Patients who need<br>physiotherapy treatment | SE |
| 18 | Performance | Use friendly interface,<br>easy to operate | Use friendly interface,<br>easy to operate | SE |
| 19 | Treatment area | Any area (Except those<br>treatment area which<br>been described in the<br>user manual can not<br>use ) ), such as Hand,<br>Arm, Chest, Waist,<br>Buttock, Thigh, Calf,<br>back and low back etc. | Any area (Except those<br>treatment area which<br>been described in the<br>user manual can not<br>use ) ), such as Hand,<br>Arm, Chest, Waist,<br>Buttock, Thigh, Calf,<br>back and low back etc. | SE |
| No. | Comparison item | New device | Predicate device | Judgment |
| 20 | List of patient<br>contacting<br>material(s) | Electrode - Transparent<br>silica gel (same supplier<br>TOP-RANK)<br>Electrode cord - PVC<br>Enclosure - ABS<br>(AG15A1)<br>Belt clip -- ABS(AG15A1) | Electrode - Transparent<br>silica gel (same<br>supplier TOP-RANK)<br>Electrode cord - PVC .<br>Enclosure - ABS<br>(AG15A1)<br>Belt clip - ABS<br>(AG15A1) | SE |
| 20.1 | Electrode lead wires | Compliance with 21CFR<br>898,<br>PVC plug:<br>L=1.15m, please see the<br>page 16 clause 7<br>Electrode cords of<br>Section-08-01 Device<br>Description for detailed. | Compliance with 21CFR<br>898,<br>PVC plug: φ<br>2mm,L=1.5m | SE |
| 20.2 | Electrode pads | Compliance with ISO<br>10993-1, provided by<br>TOP-RANK. 50*50mm,<br>Transparent silica gel | Compliance with ISO<br>10993-1, provided by<br>TOP-RANK. 40*40mm,<br>Transparent silica gel | SE |
| 21 | Standards | | | -- |
| 21.1 | Biocompatibility | Compliant with<br>requirements of ISO<br>10993-5 and ISO<br>10993-10 standards | Compliant with<br>requirements of ISO<br>10993-5 and ISO<br>10993-10 standards | SE |
| 21.2 | Mechanical Safety | Compliant with<br>requirements of IEC<br>60601-1, IEC 60601-2-10<br>safety standards | Compliant with<br>requirements of IEC<br>60601-1, IEC<br>60601-2-10 safety<br>standards | SE |
| 21.3 | Electrical Safety | Compliant with<br>requirements of IEC<br>60601-1, IEC<br>60601-2-10, IEC<br>60601-1-2 safety<br>standards | Compliant with<br>requirements of IEC<br>60601-1, IEC 60601-2<br>-10, IEC 60601-1-2<br>safety standards | SE |
| 21.4 | Energy delivered | The delivered energy is<br>limited according to<br>requirements of collateral<br>IEC 60601-2-10 safety<br>standards | The delivered energy is<br>limited according to<br>requirements of<br>collateral IEC<br>60601-2-10 safety<br>standards | SE |
| 22 | Used at (hospital,<br>home, ambulances) | Physiotherapy clinics | Physiotherapy clinics | SE |
| No. | Comparison item | New device | Predicate device | Judgment |
| 23 | Operating<br>temperature and<br>humidity | 5-40°C<br>30%-75% | 5-40°C<br>30%-75% | SE |
| 24 | Storage<br>temperature and<br>humidity | -10-50°C<br>10%-90% | -10-50°C<br>10%-90% | SE |
| 25 | Weight (Ibs.) | 0.35 | 0.28 | SE |
Section 05-Page 3 of 9
·
{3}------------------------------------------------
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
Section 05-Page 4 of 9
. . . . . .
·
{4}------------------------------------------------
・
·
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
Section 05-Page 5 of 9
:
' .
{5}------------------------------------------------
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
#### Description of Device[21 CFR 807.92(a)(4)] b)
The FDES101(ED401) Series Stimulator, which includes models FDES101(ED401), FDES102(ED402) and FDES103(ED403), are Transcutaneous Electrical Nerve Stimulator and muscle stimulator for pain relief and/or Electrical Muscle Stimulator. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The parameters of units are controlled by the press buttons. Its intensity level is adjustable according to the needs of patients.
- The three models FDES101(ED401), FDES102(ED402) and FDES103(ED403) have the same housing in a molded portable plastic case with a viewable LCD display, an accessible keypad, and accessible battery storage compartment. The case shape is rectangular. The LCD is located on the upper half of the rectangular face of the device, above the keypad. The LCD is used to display system information to the user.
- The device is equipped with a keypad composed of push buttons which are located below the LCD. The function for each button is defined by a symbol on the LCD corresponding to the button immediately below it. The process to set the parameter and attach lead wires to the five models is also the same. Yet, they have different liquid crystal display and parameters for patients to create their own settings.
The FDES101(ED401) TENS and EMS Stimulator is the combination unit with the TENS. EMS functions, the function can be selected by press buttons. The range of settings is identical to those of FDES102(ED402) and FDES103(ED403). The difference on the five units can be identified by the LCD display.
#### Intended Use[21 CFR 807.92(a)(5)] c) FDES101(ED401) TENS and EMS Stimulator
For TENS mode
Section 05-Page 6 of 9
{6}------------------------------------------------
File No: XW-Stimulator A-FDA-05 Version: 1.2 . Data: Aug.20, 2012
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. post surgical pain
## For EMS mode
- 1. Relaxation of muscle spasm.
- 2. Increase of local blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- 4. Muscle re-education
- 5. Maintaining or increasing range of motion
- 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
## FDES102(ED402) TENS Stimulator
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
## FDES103(ED403) EMS Stimulator
- 1. Relaxation of muscle spasm.
- 2. Increase of blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- 4. Muscle re-education
- ട. Maintaining or increasing range of motion.
- Immediate post-surgical stimulation of calf muscles to prevent venous റ. thrombosis
#### Technological Characteristics[21 CFR 807.92(a)(6)] d)
### a) For TENS and EMS Stimulator FDES101(ED401)
| Model: | FDES101(ED401) |
|----------------------|-------------------------------------|
| Use Output Channels: | Two independent output channels |
| Intensity: | 0-60 levels adjustable |
| Output Amplitude: | 0-60mA adjustable (at 1000ohm load) |
| Treatment Mode: | TENS, EMS |
Section 05-Page 7 of 9
{7}------------------------------------------------
.
File No: XW-Stimulator A-FDA-05 Version: 1.2 . Data: Aug.20, 2012 ,
Pulse Width: 50-300uS : ・
| Pulse Frequency: | 0.5-150Hz |
|--------------------|----------------------------------------------|
| Treatment Time: | 1-60minutes adjustable and default 30minutes |
| Output Plus Wave: | Bi-phase square wave |
| Power Supply: | 4×1.5V AAA batteries, DC 6V |
| Weight: | 0.35 Ibs (With batteries) . |
| Device Dimensions: | 129.7×76×35.1mm(L×W×H) |
# b) For TENS Stimulator FDES102(ED402)
| Model: | FDES102(ED402) |
|----------------------|----------------------------------------------|
| Use Output Channels: | Two independent output channels |
| Intensity: | 0-60 levels adjustable |
| Output Amplitude: | 0-60mA adjustable (at 1000ohm load) |
| Treatment Mode: | TENS |
| Pulse Width: | 50-300uS |
| Pulse Frequency: | 0.5-150Hz |
| Treatment Time: | 1-60minutes adjustable and default 30minutes |
| Output Plus Wave: | Bi-phase square wave |
| Power Supply: | 4 × 1.5V AAA batteries, DC 6V |
| Weight: | 0.35 Ibs (With batteries) |
| Device Dimensions: | 129.7×76×35.1mm(L×W×H) |
# c) For EMS Stimulator FDES103(ED403)
| Model: | FDES103(ED403) |
|----------------------|----------------------------------------------|
| Use Output Channels: | Two independent output channels |
| Intensity: | 0-60 levels adjustable |
| Output Amplitude: | 0-60mA adjustable (at 1000ohm load) |
| Treatment Mode: | EMS |
| Pulse Width: | 50-300uS |
| Pulse Frequency: | 1-150Hz |
| Treatment Time: | 1-60minutes adjustable and default 30minutes |
| Output Plus Wave: | Bi-phase square wave |
| Power Supply: | 4 × 1.5V AAA batteries, DC 6V |
| Weight: | 0.35 lbs (With batteries) |
| Device Dimensions: | 129.7×76×35.1mm(L×W×H) |
Section 05-Page 8 of 9
{8}------------------------------------------------
File No: XW-Stimulator A-FDA-05 Version: 1.2 Data: Aug.20, 2012
### e) Biocompatibility Certification
Electrodes to be provided with this device are from the manufacturer Top-Rank Health Care Equipment Co., Ltd (K070612) who submitted in 2007.
The shell of device is used ABS material; this material has passed Biocompatibility i testing in Guanzhou Medical Instruments Quality Surveillance and Inspection Center of State Food and Drug Administration, Report No: RZ106309.
#### Discussion of Non-Clinical Tests Performed for Determination of f) Substantial Equivalence are as follows
The FDES101(ED401) Series Electro-Stimulator did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Nonclinical testing was performed in order to validate the design according with the company's specified design requirements, and to assure conformance with the following voluntary design standards:
- > IEC 60601-1 "Medical electrical equipment Part 1: General requirements for safety".
- > IEC 60601-1-2 "Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility- Requirements and tests, Interpretation Sheet."
- > IEC 60601-1-4 "Medical electrical equipment Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems
- > IEC 60601-2-10 "Medical electrical equipment Part 2: Particular requirements for the safety of nerve and muscle stimulators"
## g) Conclusions
The FDES101(ED401) Series Stimulator, which includes models FDES102(ED402) and FDES103(ED403), has the same intended use and technological characteristics as the predicate device of MT9000 Series Electro-Stimulator, Model MT9000. Moreover, bench testing, safety report and Risk Analysis Report documentation supplied in this submission demonstrates that the difference in the submitted models could maintain the same safety and effectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, the FDES101(ED401) Series Electro-Stimulator is substantially equivalent to the predicate device.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, serif font. The text is centered on the image.
Image /page/9/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
#### Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Public Health Service
### December 21, 2012
Famidoc Technology Co., Ltd. % Cao Liang East 2/F. Zhixiang Building 71 Block Baoan District Shenzhen, Guangdong China 518101
Re: K113010
Trade/Device Name: FDES101 (ED401) TENS and EMS Stimulator, FDES102 (ED402) TENS Stimulator, FDES103 (ED403) EMS Stimulator Regulation Number: 21 CFR 890.5850
Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF, GZJ Dated: November 28, 2012 Received: November 28, 2012
Dear Cao Liang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{10}------------------------------------------------
## Page 2 - Cao Liang
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Victor Krauthamer -A
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
Famiooc
File No: XW-Stimulator A-FDA-04 Version: 1.1 Data: May.3, 2012
# Indications for Use
510(k) Number: K113010
Device Name: EDES101(ED401) TENS and EMS Stimulator, FDES102(ED402) TENS Stimulator, FDES103(ED403) EMS Stimulator
# Indications for Use:
## FDES101(ED401) TENS and EMS Stimulator
## For TENS mode
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
### For EMS mode
- · 1. Relaxation of muscle spasm.
- 2. Increase of local blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- 4. Muscle re-education
- 5. Maintaining or increasing range of motion
- 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
## FDES102(ED402) TENS Stimulator
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
## FDES103(ED403) EMS Stimulator
- 1. Relaxation of muscle spasm.
- 2. Increase of blood flow circulation
- Prevention or retardation of disuse atrophy 3.
- Muscle re-education ধ:
{12}------------------------------------------------
File No: XW-Stimulator A-FDA-04 Version: 1.1 Data: May.3, 2012
- Maintaining or increasing range of motion. ട.
- Immediate post-surgical stimulation of calf muscles to prevent 6. venous thrombosis
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Brian D. Pullin -S
Division of Neurological and Physical Medicine Devices 510(k) Number: K113010
Section 04 -Page 2 of 2
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.