K112841 · Calypso Medical Technologies, Inc. · IYE · Feb 16, 2012 · Radiology
Device Facts
Record ID
K112841
Device Name
SURFACE BEACON TRANSPONDERS
Applicant
Calypso Medical Technologies, Inc.
Product Code
IYE · Radiology
Decision Date
Feb 16, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Intended Use
The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders. The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces. Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures. Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy. Surface Beacon® transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy.
Device Story
Calypso System uses non-ionizing electromagnetic and optical technology for real-time patient localization during radiation therapy. System inputs include signals from passive Beacon transponders (implanted or surface-mounted) and optical data from cameras/infrared targets. System processes these inputs to provide 3D localization of treatment targets or surface motion. Dynamic Edge™ Gating feature interfaces with external radiation therapy systems to trigger beam-holds when motion exceeds defined tracking limits. Used in clinical radiation therapy environments by trained operators. Output provides objective, continuous position data to clinicians for patient alignment and motion monitoring. Benefits include increased precision in radiation delivery and potential for reduced dose to healthy tissue via gating.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed that Surface Beacon transponders satisfy intended use for real-time monitoring of respiratory and patient motion.
Technological Characteristics
Sensing: Cameras, infrared targets, and electromagnetic Beacon transponders. Connectivity: Interface to external radiation therapy systems for gating. Safety standards: IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2. System components: Console and tracking station.
Indications for Use
Indicated for patients undergoing radiation therapy requiring alignment and monitoring of treatment targets. Implanted transponders for soft tissue marking (prostate/peri-prostatic). Surface transponders for temporary skin placement to monitor respiratory and patient motion during radiotherapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
K123137 — CALYPSO SYSTEM · Varian Medical Systems · Dec 5, 2012
K112768 — CALYPSO SYSTEMS WITH DYNAMIC EDGE GATING · Calypso Medical Technologies, Inc. · Nov 18, 2011
K102373 — CALYPSO SYSTEM WITH DYNAMIC EDGE GATING · Calypso Medical Technologies, Inc. · Sep 30, 2010
K060906 — CALYPSO 4D LOCALIZATION SYSTEM · Calypso Medical Technologies, Inc. · Jul 28, 2006
K080726 — CALYPSO 4D LOCALIZATION SYSTEM · Calypso Medical Technologies, Inc. · May 14, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
# (1128-
### 510(K) SUMMARY
# FEB 1 6 2012
This 510(k) Summary is provided in accordance with 21 CFR 807.92.
Date of preparation: 28 September 2011
Submitter information:
Calypso Medical Technologies, Inc. 2101 Fourth Avenue, Suite 500 Seattle, WA 98121
Phone: 206-254-0600 Fax: 206-254-0606
| Contact: | Marcia A Page Vice President Quality Assurance and Regulatory Affairs |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Device trade name: | Calypso® System Surface Beacon® transponders |
| Common name: | Patient localization system |
| Classification name: | Medical charged-particle radiation therapy system |
| Classification: | CFR 892.5050 Class II Product code – LHN, IYE |
| Predicate devices: | Calypso System (K060906, K080726, K102373) RPM Respiratory Gating System (K102024) Respisens (K092845) |
#### Device Description:
The Calypso System utilizes non-ionizing electromagnetic and optical technology to provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy.
Use of the Calypso System for target localization is based on the system's detection of non-ionizing electromagnetic signals from passive markers, called Beacon® transponders
{1}------------------------------------------------
which are implanted in or near the treatment target or placed externally on the surface of a patient.
Implanted Beacon transponders when used with the Calypso System enable objective measurement of the location of the target in 3 dimensions. The system operator uses this information to align the patient's treatment target to the isocenter of the therapy system prior to treatment. This information can also be used to monitor (track) the position of the treatment target during radiation therapy treatment.
When using Surface Beacon transponders this information can be used to monitor respiratory motion and other patient motion in real time during radiotherapy. When correlated to the patient's treatment target motion the system provides objective measurement of the location of the target in 3 dimensions.
The Dynamic Edge™ Gating feature enables the Calypso System to interface with radiation therapy systems configured with gating capabilities via an interface to external systems. With this optional feature added to the Calypso System, a beam hold is signaled to the radiation therapy system when the treatment target position has moved outside the defined tracking limits. The radiation therapy system is signaled to remove the beam-hold upon the target's return to a position inside the defined tracking limits.
#### Indications for Use:
The Calypso System is intended for use as an adjunct in treatment planning and radiation therapy, to align and monitor the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
Implanted Beacon transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed).
Surface Beacon transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy.
{2}------------------------------------------------
Technological Characteristics - See device comparison table below
| Feature and/or Specification<br>of new/modified device | Calypso System with implanted<br>Beacon Transponders | Calypso System with Surface<br>Beacon Transponders |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Sensing Technology | Cameras and infrared targets on<br>array and Beacon Transponders | Cameras and infrared targets on<br>array and Surface Beacon<br>Transponders |
| Monitoring | Real Time Monitor patient<br>motion | No change |
| Tracking | Real Time Tracking Patient<br>motion | Real Time Tracking Patient and<br>Respiratory motion |
| Gating | Interface to external systems | No change |
| Records | Display and Record motion | No change |
| Computer hardware and<br>software | Computer systems (console and<br>tracking station) to control<br>device functions and provide for<br>user interface | No change |
| Electromagnetic technology<br>(array) | Used to track motion during<br>therapeutic procedures | No change |
| Compatibility with the<br>environment and other<br>devices | Operates in a radiation therapy<br>system environment | No change |
| Electrical safety mechanical<br>safety | IEC 60601-1; IEC 60601-1-1 | No change |
| EMC Safety | IEC 60601-1-2 | No change |
## Summary of Performance Testing:
Results of verification and validation testing demonstrate that the Surface Beacon transponder satisfies the intended use as described above.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling human figures or waves, stacked on top of each other.
Public Health Service
FEB 1 6 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Marcia Page Vice President Quality Assurance and Regulatory Affairs Calypso Medical Technologies, Inc. 2101 Fourth Ave South, Suite 500 SEATTLE WA 98121
Re: K112841
Trade/Device Name: Calypso® System Surface Beacon transponders Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: January 3, 2012 Received: January 4, 2012
Dear Ms. Page:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary S. Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): K112841 ----------
Device Name: Calypso@ System Surface Beacon transponders
The Calypso System is intended for use as an adjunct in treatment planning and radiation the our poor ijo align and/or the patient's position relative to the isocenter of a radiation therapy system. The Calypso System provides accurate, precise and continuous localization of a treatment isocenter by using two or more Beacon transponders.
The optional Dynamic Edge™ Gating component may signal a radiation therapy system to impose a beam-hold when the treatment target position has moved outside the defined tracking limits and to signal the radiation therapy system to remove the beam-hold upon the target's return to a position inside the defined tracking limits. The Dynamic Edge Gating feature has been shown to be compatible with Varian and Siemens radiation therapy treatment systems with external gating interfaces.
Implanted Beacon® transponders are indicated for use to radiographically and electromagnetically mark soft tissue for future therapeutic procedures.
Permanent Beacon transponders are indicated for implantation in the prostate and the peri-prostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy.
Surface Beacon® transponders are indicated for temporary external placement on the skin, to monitor respiratory motion and other patient motion in real time during radiation therapy.
x Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Pastel
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.