K111651 · Medical Components, Inc. · MSD · Dec 8, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K111651
Device Name
SPLIT CATH III
Applicant
Medical Components, Inc.
Product Code
MSD · Gastroenterology, Urology
Decision Date
Dec 8, 2011
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5540
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian vein. Catheters greater than 40cm are intended for femoral vein insertion. The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis in the adult patient. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian vein and inferior vena cava as required. Catheters greater than 40cm are intended for femoral vein or inferior vena cava insertion. Translumbar insertion via inferior vena cava is indicated when all other access sites are identified as non-viable.
Device Story
Split Cath® III is a 14 French, double-lumen, radiopaque polyurethane hemodialysis catheter designed for long-term vascular access. It features a cuff for anchoring and a soft pliable hub with suture wings. The device is inserted percutaneously by a clinician into the internal jugular, subclavian, femoral, or inferior vena cava. Red and blue clamps/sleeves on the extension tubes prevent air/fluid communication. The device facilitates blood flow for hemodialysis and apheresis procedures. It benefits patients requiring chronic vascular access by providing a durable, biocompatible conduit for extracorporeal blood treatment.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993), air leakage, liquid leakage, priming volume, flow vs. pressure, tensile strength, recirculation, chemical exposure, and accelerated aging. Performance met ISO 10555-1, ISO 10555-3, and ISO 594-2 standards.
Technological Characteristics
14 French double-lumen catheter; radiopaque polyurethane material; includes cuff for long-term implant; 20cm to 55cm lengths; Luer lock fittings per ISO 594-2; sterile, single-use; non-electronic/mechanical device.
Indications for Use
Indicated for long-term vascular access for hemodialysis and apheresis in adult patients. Insertion sites include internal jugular, subclavian, femoral, or inferior vena cava (including translumbar approach when other sites are non-viable).
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
K121848 — SPLIT CATH III 24CM SPLIT CATH III 24CM SPLIT CATH III 32CM CATH III 36CM SPLIT CATH III 40CM AND 55CM · Medical Components, Inc. · Sep 21, 2012
K151967 — Centros Long-Term Hemodialysis Catheter, CentrosFLO Long-Term Hemodialysis Catheter · Merit Medical Systems, Inc. · Apr 4, 2016
K020465 — MEDCOMP ASH SPLIT-CATH II · Medcomp · May 22, 2002
K051280 — MEDCOMP SPLIT CATH II · Medcomp · Nov 30, 2005
{0}------------------------------------------------
DEC - 8 2011
\$\qquad\$Kl.
page 1 of 3
## 510(k) Summary
Split Cath® III Summary of Safety and Effectiveness Prepared June 10, 2011
| Section 5 | |
|-----------------------|-------------------------------------------------------------------------------------------------------|
| General Information | |
| Submitter: | MEDCOMP®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Phone: (215) 256-4201<br>Fax: (215) 256-9191 |
| Contact: | Jean Callow<br>Regulatory Specialist |
| Device Trade Name: | Split Cath® III |
| Common Name: | Hemodialysis Catheter, Implanted |
| Classification Name: | MSD - Blood access device and accessories |
| CFR Reference: | 21 CFG 876.5540, Class III |
| Classification Panel: | Gastroenterology / Urology |
### Predicate Devices:
| Device Trade Name: | Split Cath® II |
|-------------------------|----------------------------------------------|
| Common Name: | Hemodialysis Catheter, Implanted |
| Classification Name: | MSD - Blood access device and accessories |
| CFR Reference: | 21 CFR 876.5540, Class III |
| Classification Panel: | Gastroenterology / Urology |
| Premarket Notification: | K040318, concurrence date February 3, 2011 |
| | K020465, concurrence date July 22, 2002 |
| | K091953, concurrence date September 16, 2009 |
| | K051280, concurrence date November 30, 2005 |
| | K981125, concurrence date February 26, 1999 |
Performance Standards: Performance standards have not been established by FDA under section 514 of the Federal Food, Drug, and Cosmetic Act.
### Indications for Use:
The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis.
It may be inserted percutaneously and is primarily placed in the internal jugular vein.
Alternate insertion sites include the subclavian vein.
Catheters greater than 40cm are intended for femoral vein insertion.
Medcomp Split Cath® III (carbothane) 510(k) Summary Page 1 of 2
{1}------------------------------------------------
K111651 Page 2 of 3
## Indications for Use: {translumbar placement}
The Medcomp® Split Cath® III is indicated for use in attaining tong term vascular access for Hemodialysis and Apheresis in the adult patient.
It may be inserted percutaneously and is primarily placed in the internal jugular vein.
Alternate insertion sites include the subclavian vein and inferior vena cava as required.
Catheters greater than 40cm are intended for femoral vein or inferior vena cava insertion.
Translumbar insertion via inferior vena cava is indicated when all other access sites are identified as non-viable.
#### Device Description:
- 14 French, double "D" Jumen design with cuff for long-term implant. .
- Variety of lumen lengths from 20cm to 55cm. .
- . Soft radiopaque polyurethane material
- Lumen is connected to the extension via a soft pliable hub with a suture wing .
- Red and blue clamps and red and blue sleeves are provided on the extension tube � to prevent air/fluid communications
- The hub contains the device name and French size, clamp I.D. Rings are printed . with the priming volume.
#### Safety and Performance Tests
Biocompatibility requirements of ISO 10993 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing for externally communicating, blood contacting, long-term devices were met. All materials used in the manufacture of the Split Cath® III were previously cleared for similar applications by Medcomp, Inc.
Performance testing of the Split Cath® III was conducted in accordance with the following international standards:
- ISO 10555-1: 1997, Sterile Single Use-Intravascular Catheters, General . Requirements
- . ISO 10555-3: 1997, Sterile Single Use-Intravascular Catheters, Central Venous Catheters
- ISO 594-2: Conical Fittings with a 6% (Luer) Taper for Syringes, Needles, and . Certain Other Medical Equipment - Part 2: Lock Fittings
Subject product testing has yielded acceptable safety and performance outcomes.
The results of these tests, in conjunction with the substantial equivalence claims effectively demonstrate that the Split Cath® III is substantially equivalent to the cited predicate devices.
Testing performed: Air Leakage Liquid Leakage Priming Volume Flow verse Pressure Force at Break / Tensile Strength Medcomp Split Cath® III (carbothane) 510(k) Summary
Page 2 of 2
{2}------------------------------------------------
K11 1651 Page 3/3
Recirculation Chemical Exposure Accelerated Aging
# Summary of Substantial Equivalence
Based on the indications for use and safety and performance testing, the Split Cath® III meets the requirements that are considered for its intended use and is substance equivalent in design materials, sterilization, and indications for use to the predicated device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC - 8 2011
Ms. Jean Callow Regulatory Specialist Medcomp 1499 Delp Drive HARLEYSVILLE PA 19438
Re: K111651
Trade/Device Name: Medcomp® Split Cath® III Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: November 23, 2011 Received: November 28, 2011
Dear Ms. Callow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Jean Callow
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.goy/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
homi M. Dhany
So Flerbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): KI /65/
Device Name: Medcomp® Split Cath® III (translumbar insertion)
Indications for Use:
The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis in the adult patient.
It may be inserted percutaneously and is primarily placed in the internal jugular vein.
Alternate insertion sites include the subclavian vein and inferior vena cava as required,
Catheters greater than 40cm are intended for femoral vein or inferior vena cava insertion.
Translumbar insertion via inferior vena cava is indicated when all other access sites are identified as non-viable.
Prescription Use X (Part 21 CFR 801 Subpart AND/OR D) ·
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
hymn M.M.
Page 1 of 1
tive. Gastro-Renal, and
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K1/ 165 |
Device Name: Medcomp® Split Cath® III
Indications for Use:
The Medcomp® Split Cath® III is indicated for use in attaining long term vascular access for Hemodialysis and Apheresis.
It may be inserted percutaneously and is primarily placed in the internal jugular vein.
Alternate insertion sites include the subclavian vein.
Catheters greater than 40cm are intended for femoral vein insertion.
Prescription Use X (Part 21 CFR 801 Subpart AND/OR D)
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aomkthting
Page 1 of 1
| (Division Sign-Off) | |
|---------------------------------------------|---------|
| Division of Reproductive, Gastro-Renal, and | |
| Urological Devices | |
| 510(k) Number | K111651 |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.