MEDCOMP SPLIT CATH II

K051280 · Medcomp · MSD · Nov 30, 2005 · Gastroenterology, Urology

Device Facts

Record IDK051280
Device NameMEDCOMP SPLIT CATH II
ApplicantMedcomp
Product CodeMSD · Gastroenterology, Urology
Decision DateNov 30, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K051280 · Nov 30, 2005MEDCOMP SPLIT CATH IIMedcompPublished clinical literatureThe sponsor utilized published clinical papers to support the use of the translumbar insertion technique for the device in patients where all other traditional vascular access sites have been exhausted.Translumbar insertion; Hemodialysis; Clinical literature

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Published clinical papers on translumbar technique; Literature review of clinical practiceAdult patients requiring long-term hemodialysis with no other viable access sitesNot applicable for this studyCatheter function and clinical outcome

Indications for Use

The Medcomp Split Cath® II is indicated for use in attaining long-term vascular access for hemodialysis and apheresis in the adult patient. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion site is the subclavian vein or inferior vena cava as required. Catheters greater than 40cm are intended for femoral insertion or inferior vena cava insertion. Translumbar insertion via inferior vena cava is indicated only when all other access sites are identified as non-viable.

Device Story

Polyurethane, double-lumen catheter for hemodialysis and apheresis; features two segregated 'D'-shaped lumens with side holes for 360-degree blood flow; distal venous lumen extends beyond arterial lumen to minimize recirculation; includes fixed polyester cuff for tissue ingrowth. Inserted percutaneously by physicians in clinical settings; primary placement in internal jugular vein, with alternate sites including subclavian, femoral, or inferior vena cava (translumbar). Lumens are split/peeled apart prior to insertion for free-floating placement within the vessel. Output is vascular access for blood removal and return. Benefits include reliable long-term access for patients with exhausted traditional sites.

Clinical Evidence

No new clinical trials conducted. Evidence relies on published clinical literature regarding the translumbar technique for catheter insertion. Literature confirms that translumbar placement is a reliable alternative for patients with exhausted traditional access sites and does not negatively impact catheter function.

Technological Characteristics

Polyurethane double-lumen catheter; 'D'-shaped lumen geometry; polyester cuff for tissue ingrowth; luer connectors; 360-degree side hole orientation. Biocompatibility per ISO 10993. Sterilization method not specified. No electronic or software components.

Indications for Use

Indicated for long-term vascular access for hemodialysis and apheresis in adult patients. Translumbar insertion via inferior vena cava is indicated only when all other access sites are non-viable.

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K051280 NOV 3 0 2005 # 510(k) SUMMARY ### Submitter Information: A. Submitter: MEDCOMP® 1499 Delp Drive Harleysville, PA 19438 (215) 256-4201 Telephone (215) 256-9191 Fax Jean Callow May 11, 2005 Contact: Date Prepared: Predicate Device: - B. Trade Name: Common Name: Classification: C.F.R. Section: Medcomp Split Cath® II Hemodialysis Catheter, Implanted MSD 876.5540 K020465 Medcomp® Ash Split Cath® II K884375 Argon Medical Corp. Translumbar Aortography Needle Catheter. #### D. Device Description: The Medcomp Split Cath® II is a polyurethane, double lumen catheter used to remove and return blood through two-segregated lumen passages. Both lumens are "D" shaped, tapered at the distal tip, with three side holes on each tip. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed polyester cuff allows for tissue ingrowth for long term placement. The arterial and venous lumens are designed to be split, or peeled apart, prior to insertion to provide two free-floating lumens within the vessel. The side holes are orientated to allow 360-degree arterial uptake and venous return. The lumens are connected to the extensions via a soft pliable hub with suture wing. Red and blue luer connectors and clamps identify the arterial and venous extensions. Priming volume information is printed an identification ring housed within the extension line clamp. #### intended Use: E. The Medcomp Split Cath® II is indicated for use in attaining fong-term vascular access for hemodialysis and apheresis in the adult patient. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion site is the subclavian vein or inferior vena cava as required. ﺎ ﻓﻲ ﺍﻟﻤﺴﺘﻘ Medcomp Split Cath II Translumbar 510(k) Summary Page 1 of 2 C. {1}------------------------------------------------ K051280 Page 2 of 2 Catheters greater than 40cm are intended for femoral insertion or inferior vena cava insertion. Translumbar insertion via inferior vena cava is indicated only when all other access sites are identified as non-viable. ## Comparison to Predicate Device: F. The technological characteristics of the Split Cath® II are identical to the predicate device in terms of design, design specifications, performance, manufacturing process and method of sterilization. The difference between the proposed device and the predicate is the inclusion of inferior vena cava as insertion site. Labeling indicating translumbar placement instructions and indicating that translumbar placement of the catheter is a useful and reliable alternative for patients that require long-term hemodialysis but all other access sites are identified as non-viable. ### Performance Data: G. In Vitro performance data for the legally cleared Medcomp Ash Split Cath® II, included tensile strength, joint strength, leakage, recirculation, flow performance, and flexural apply to the Split Cath® II since there are no significant material or design changes. Performance data is submitted for peel testing and force @ break only due to a change in solvent from the predicate device (solvent was cleared under K972207). Chemical testing is submitted for recent ointment testing performed by Medcomp. Biocompatibility testing on the Ash Split Cath® II demonstrates the lumen materials meet the requirements of ISO 10993 for a permanent contact device. Biocompatibility data on the Split Cath® II was not deemed necessary since substantial equivalence is addressed by way of comparison to a legally marketed device. Clinical data is based on numerous published clinical papers on the need for and presently used translumbar technique for catheter insertion. Due to criteria stressed in the published clinical papers this technique is for a very small patient population as an absolute last resort when all other traditional access sites have been exhausted. Properly inserted catheters via the translumbar technique do not have a detrimental affect on the clinical outcome as to catheter function. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. NOV 3 0 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Ms. Jean Callow Regulatory Specialist MedComp® 1499 Delp Drive HARLEYSVILLE PA 19438 Re: K051280 KUS I 280 Trade/Device Name: MedComp 14F Split Cath II for Translumbar Insertion Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: November 9, 2005 Received: November 28, 2005 Dear Ms. Callow: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section > rotty premained is substantially equivalent (for the indications felerenced above and nave decembed by marketed predicate devices marketed in interstate for use stated in the enciosale) to regars atment date of the Medical Device Amendments or to commerce prior to May 20, 1976, the encordance with the provisions of the Federal Food, Drug, devices that have been reclusined in accesses in a creat the device, subject to the general controls and Cosment Act (Fict). Totalina), are responsible to determine that the medical devices you provisions of the Act. "However, you de reap determined as substantially equivalent under the use as components in the Kir nave etater 500k) of the act), or were legally on the market prior to premarket notifeation process (Section of Secure in endments. Please note: If you May 20, 1970, the chacinent and 07 thel (i.e., unfinished) and further process (e.g., sterilize) purchase your device components in luding these components in your kit. The general you must submit a new 9 regis oread equirements for annual registration, listing of devices, controls provisions of the or use labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 - Ms. Jean Callow Please be advised that FDA's issuance of a substantial equivalence determination does not mean Frease of advised hat I Dri 3 issualite or our device complies with other requirements of the Act that I DA has made a actoring administered by other federal agencies. You must of any it cach sundles and regulanents, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will anow you to oegn mainer of substantial equivalence of your device to a legally premative lieutification " - e results in a classification for your device and thus, permits your marketed proceed to the market. If you desire specific advice for your device on the labeling device to proceed to the mance in Joa Compliance at (240) 276-0115. Also, please note the regulation, prease contact anding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may oounn other generational and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Edmund Ch. Huyckson for Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number: KO5 1280 Device Name: SPLIT CATH® II CATH® II CATHETER Indications for use: THE MEDCOMP SPLIT CATH® II CATHETER IS INDICATED FOR USE IN ATTAINING LONG TERM THE MEDICOME OF EFF ORTHO'!! SKULTION SPHERESIS IN THE ADULT PATIENT. IT MAY BE INSERTED PERCUTANEOUSLY AND IS PRIMARILY PLACED IN THE INTERNAL JUGULAR VEIN. ALTERNATE INSERTION SITES INCLUDE THE SUBCLAVIAN VEIN AND INFERIOR VENA CAVA AS REQUIRED. CATHETERS GREATER THAN 40CM ARE INTENDED FOR FEMORAL VEIN OR INFERIOR VENA CAVA INSERTION. TRANSLUMBAR INSERTION VIA INFERIOR VENA CAVA IS INDICATED WHEN ALL OTHER ACCESS SITES ARE IDENTIFIED AS NON-VIABLE. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | | |------------------------------------------|-------------------------------| | | OR Over-The-Counter__________ | | (Division Sign-Off) | | |---------------------------------------------------------------|---------| | Division of Reproductive, Abdominal, and Radiological Devices | | | 510(k) Number | K051280 | (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...