SKAFFOLD CIS
K111538 · Skeletal Kinetics, LLC · MTJ · Oct 25, 2011 · SU
Device Facts
| Record ID | K111538 |
| Device Name | SKAFFOLD CIS |
| Applicant | Skeletal Kinetics, LLC |
| Product Code | MTJ · SU |
| Decision Date | Oct 25, 2011 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
CAAP Bone Wax is indicated to control the bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.
Device Story
CAAP Bone Wax is a sterile, single-use kit comprising calcium phosphate powder, dilute sodium silicate liquid, and a mixing system (bowl, pestle, spatula). Components are mixed immediately prior to use to form a paste. Applied directly to cut or damaged bone sites, the paste hardens into a hydroxyapatite material, mimicking native bone mineral. It functions as a mechanical barrier or tamponade to achieve hemostasis. Used in clinical settings by healthcare providers to manage bleeding during surgical procedures involving bone.
Clinical Evidence
No clinical data. Evidence consists of bench testing and animal studies. Bench testing confirmed physiological pH and temperature stability. Animal study demonstrated adequate hemostasis as a mechanical tamponade compared to predicate devices.
Technological Characteristics
Composition: Calcium phosphate and sodium silicate solution; forms hydroxyapatite upon hardening. Sterilization: Radiation (ANSI/AAMI/ISO 11137-2:2006). Biocompatibility: ISO 10993-1. Form factor: Kit with mixing system.
Indications for Use
Indicated for control of bleeding from cut or damaged bone in patients requiring hemostasis at bony sites.
Predicate Devices
- Auto Suture Bone Wax (K971680)
- Ostene (K062280)
- Ethicon Bone Wax (Preamendment)
- Lukens Bone Wax (K791495)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text including "K111538" and "p1/2". Below this, the date "OCT 25 2011" is written. The text appears to be written in black ink on a white background. The handwriting is somewhat stylized, with distinct letter formations.
Image /page/0/Picture/1 description: The image shows a logo with a circular graphic on the left and the words "Skeletal Kinetics" on the right. The circular graphic resembles a globe with intersecting lines. The text "Skeletal Kinetics" is stacked, with "Skeletal" on top of "Kinetics", and both words are in a bold, sans-serif font.
A COLSON ASSOCIATE
Product Code:
### 510(k) Summary
## General Information as required by 21 CFR 807.92 (a) (1)
Christine Kuo, Contact Person: Director, Regulatory Affairs and Quality Assurance Phone: (408) 350-5842, Fax: (408) 366-1077
September 22, 2011 Date Prepared:
### Device Name as required by 21 CFR 807.92 (a) (2)
| Trade Names: | CAAP (Calcium Apatite) Bone Wax |
|-----------------|---------------------------------|
| Common Name: | Calcium phosphate bone wax |
| Classification: | Unclassified |
MTJ
# Predicate Devices as required by 21 CFR 807.92 (a) (3)
The subject device is substantially equivalent in safety and effectiveness to the following legally marketed devices (predicates):
| I. Auto Suture Bone Wax | K971680 |
|-------------------------|--------------|
| II. Ostene | K062280 |
| III. Ethicon Bone Wax | Preamendment |
| IV. Lukens Bone Wax | K791495 |
## Device Description as required by 21 CFR 807.92 (a) (4)
The single use CAAP (Calcium Apatite) Bone Wax sterile kit contains: Calcium Phosphate Powder, Dilute Sodium Silicate Liquid, and a Mixing System (Mixing Bowl, Pestle and Spatula). CAAP Bone Wax is to be mixed immediately prior to use.
### Intended Use as required by 21 CFR 807.92 (a) (5)
CAAP Bone Wax is indicated to control the bleeding from cut or damaged bone by
{1}------------------------------------------------
acting as a mechanical barrier or tamponade.
..............................................................................................................................................................................
## Summary of Technological Characteristics as required by 21 CFR 807.92 (a) (6)
CAAP Bone Wax is composed of calcium phosphate and sodium silicate solution, when mixed together it forms a paste that can be applied directly to sites of bleeding bone, the resulting hardening material from the paste is composed of hydroxyapatite similar to the mineral phase of native bone tissue.
CAAP Bone Wax is substantially equivalent to previously cleared bone wax devices believed to control the bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.
Predicate device I, K971680, Auto Suture Bone Wax (U.S. Surgical) is indicated in the control of bleeding from bone surfaces.
Predicate device II, K062280, Ostene (Ceremed) is indicated for use in the control of bleeding from bone surfaces.
Predicate device III, preamendment, Bone Wax (Ethicon Surgical) is to control bleeding from bone.
Predicate device IV, K791495, Lukens Bone Wax (Lukens) is acts as a mechanical barrier in achieving hemostasis in bleeding bone.
### Summary of Non-clinical Tests as required by 21 CFR 807.92 (b) (1)
The biocompatibility of CAAP Bone Wax is in accordance with the standard set forth in ISO 10993-1, Biological Evaluation of Medical Devices and the radiation sterilization validation is in compliance to ANSVAAMI/ISO 11137-2:2006, Sterilization of Health Care Products - Radiation - Establishing the Sterilization Dose.
The results of bench testing demonstrated that CAAP Bone Wax remains within the physiological pH and temperature range and would be expected to show no adverse biological consequence when applied as a physical hemostatic material.
The animal study provided substantial data that the use of CAAP Bone Wax as a mechanical tamponade to control bleeding in a bony site results in adequate hemostasis as compared with the predicate devices.
### Summary of Clinical Tests as required by 21 CFR 807.92 (b) (2)
CAAP Bone Wax does not require clinical test.
### Conclusion as required by 21 CFR 807.92 (3)
The summary demonstrates that CAAP Bone Wax is substantially equivalent to the predicate devices and is as safe, as effective and performs as well as the legally market devices identified.
{2}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
OCT 25 2011
Skeletal Kinetics LLC % Ms. Christine Kuo Director, Regulatory Affairs and Quality Assurance 10201 Bubb Road Cupertino, California 95014
Re: K111538
Trade/Device Name: CAAP (Calcium Apatite) Bone Wax Regulatory Class: Unclassified Product Code: MTJ Dated: September 22, 2011 Received: September 23, 2011
Dear Ms. Kuo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your devire (ran be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act or any Federal statutes and regulations administered by other Federal agencies. Your must
{3}------------------------------------------------
Page 2 - Ms. Christine Kuo
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regalations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Emil Keith
Fo Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Section 4: Indications for Use
510(K) Number (if Known): K11538
Device Name:
CAAP (Calcium Apatite) Bone Wax
Indications for Use:
CAAP (Calcium Apatite) Bone Wax is indicated to control bleeding from cut or damaged bone by acting as a mechanical barrier or tamponade.
Prescription Use x (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21CFR 801 Subpart C
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
E. Keith
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111538