The Opto Hyalus Green Laser is indicated for retinal photocoagulation. Used for the following ophthalmic medical condition and treatment: - Retinal photocoagulation, panretinal photocoagulation and intravitreal endophotocoagulation of vascular and structural abnormalities of the retina and choroids, including: - Proliferative and nonproliferative diabetic retinopathy - Choroidal neovascularization - Branch retinal vein occlusion - The treatment of choroidal neovascularization associated with wet age related macular degeneration Retinal tears and detachments - Iridotomy, iridectomy, and trabeculoplasty in angle closure glaucoma and open angle glaucoma
Device Story
Solid-state, frequency-doubled, 532nm green Nd:YAG surgical laser for photothermolysis of soft tissue. Physician-operated via slit lamp, laser indirect ophthalmoscope (LIO), or probe. System features three encoders for pulse duration (10-1000ms), power (50-2000mW), and pulse interval (50-3000ms). Includes red aiming beam for positioning. User interface consists of a 240x324 pixel TFT touch-screen display for navigation, setting preferences (loudness, brightness, laser sighting), and pulse configuration (single or pulse trains). Used in clinical settings to treat retinal/choroidal vascular abnormalities and glaucoma. Output allows precise thermal coagulation of ocular tissue, reducing risk of infection and mechanical trauma compared to contact-lens-based procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and verification of design specifications against the predicate device.
Technological Characteristics
Solid-state, optically pumped semiconductor laser; 532nm emission wavelength. Includes red aiming beam. Compatible with slit lamp adapters, LIO, and probes. User interface: 240x324 TFT touch-screen. Complies with IEC 60825-1 (laser safety), IEC 60601-1 (general safety), IEC 60601-2-22 (therapeutic laser safety), and IEC 62304 (software life cycle).
Indications for Use
Indicated for patients requiring retinal photocoagulation, panretinal photocoagulation, or intravitreal endophotocoagulation for vascular/structural retinal and choroidal abnormalities, including proliferative/nonproliferative diabetic retinopathy, choroidal neovascularization (including wet age-related macular degeneration), branch retinal vein occlusion, and retinal tears/detachments. Also indicated for iridotomy, iridectomy, and trabeculoplasty in angle-closure and open-angle glaucoma.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Predicate Devices
IRIS Medical OcuLight GL/GLx Laser Systems (K031665)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "opto" in a stylized font. Below the word "opto" is the phrase "Science in Sight" in a smaller font. The word "opto" is in all lowercase letters and has a unique design. The phrase "Science in Sight" is in a serif font and is slightly italicized.
AUG 2002012
K111460
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
### Submitter Information 1.
Submitter: Opto Eletrônica S/A
Address: NO. 1071 - Joaquim Augusto Ribeiro de Souza street- Jardim Santa Felícia - São Carlos - SP/ Brazil
Phone: + 55 (16) 2106-7029
Contact name: Paulo Aneas Lichti
Date Prepared: march 10, 2011
Registration number: 9613205
- 2. U.S. Agent to respond to FDA requests:
Establishment: Opto Usa Corporation
Contact Person: Paulo Aneas Lichti
Address: 12550 BISCAYNE BLVD, 605, MIAMI - FL - 33181
Phone: 1 305 981 2979 — Fax: 1 305 981 2980 —
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "opto" in a stylized, bold font. Each letter is filled with a textured pattern, giving them a slightly distressed or vintage look. The letters are closely spaced, creating a compact and unified design.
- 3. Device Classification Name
Device Name: Opto Hyalus Green Laser
Common name: Ophthalmic Laser, Surgical Laser
Classification name:
HQF, Laser, Ophthalmic
## Regulation Number:
21 CFR 886.4390, Ophthalmic Laser
Regulatory Class: II
Performance standards: 21 CFR 1040.10
#### Predicate Device Name: 4.
IRIS Medical OcuLight GL/GLx Laser Systems, Iridex Corporation, K031665
#### 5. Product Description
The Opto Hyalus Green Laser solid state, frequency-doubled, green Nd;YAG surgical laser system is an instrument used in the photothermolysis (photocoagulation) of soft tissue at en emission wavelength of 532nm.
The equipment consists of three encoders which enable the physician to adjust the pulse duration, the laser power and the interval between successive pulses. All information is displayed on a color display 240x324 pixels TFT technology. On this display there is a touch-screen film responsible for the navigation software and by setting preferences, such as loudness, intensity laser sighting and display brightness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "opto" in a bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance. The letters are evenly spaced and aligned horizontally.
Physicians have the option of applying single pulses or pulse trains previously adjusted. Power levels ranging from 50mW up to 2000mW, with resolution of 10mW. The duration can be set between 10ms and 1000 ms with a resolution of 10ms. The interval between pulses can be set between 50ms and 3000ms with 10ms resolution or can be set to single pulse (in this case, the range is set to zero).
#### Intended Use 6.
The Opto Hyalus Green Laser is indicated for retinal photocoagulation. Used for the following ophthalmic medical condition and treatment:
- Retinal photocoagulation, panretinal photocoagulation and intravitreal ● endophotocoagulation of vascular and structural abnormalities of the . retina and choroids, including:
- イ Proliferative and nonproliferative diabetic retinopathy
- V Choroidal neovascularization
- イ Branch retinal vein occlusion
- イ The treatment of choroidal neovascularization associated with wet age related macular degeneration Retinal tears and detachments
- Iridotomy, iridectomy, and trabeculoplasty in angle closure glaucoma and . open angle glaucoma
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "opto" in a stylized font. The letters are bold and have a textured fill. The letters are connected to each other, forming a single, cohesive design.
## Advantages of the LIO (Laser Indirect Ophthalmoscope)
- 1. Allow binocular visualization of the peripheral retina:
- 2. Used to evaluate patients who must be examined in a supine position;
- 3. The region of the vitreous band can be inspected and treated;
- 4. Allow easier treatment of superior quadrant than does the endolaser;
- 5. Allow better visibility through the phakic gas-filled eyes since there are fewer reflections than present with biconcave lenses;
- Minimizes the risk of lens damage by the endolaser when treating the far periphery; 6.
- 7. The peripheral edges of a giant tear can be completely inspected and treated with the Laser Indirect Ophthalmoscope;
- 8. Avoids discomfort induced by pressure from a contact lens in the postoperative patient;
- 9. Reduces the stress to recently sutured wounds;
- 10. Minimizes the risk of infection.
### Technical Characteristics 7. ·
The Opto Hyalus Green Laser beam has a wavelength of 532nm, which is in the visible spectrum and is a green light. A red aiming beam is used to position the treatment green beam prior to delivery.
The Opto Hyalus Green Laser is a solid state, Optical Pumped Semiconductor surgical laser.
It's an instrument used in the photothermolysis (photocoagulation) of soft tissue at en emission wavelength of 532nm.
Compatible delivery devices include: slit lamp adapters/, laser indirect ophthalmoscopes (LIO) and Probe. All these accessories are specific for Opto Hyalus Green Laser.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "opto" in a bold, sans-serif font. The letters are filled with a textured pattern, giving them a slightly rough appearance. The letters are arranged horizontally, with each letter clearly distinct and legible.
#### 8. Accessories
- Eye Safety Filter .
- Pedal Multifunctional
- Slit Lamp Adapter .
### Compatible Delivery Devices 9.
- Ophthalmoscope
- Probe
### 10. Performance Standard
Opto Hyalus Green Laser is designed, tested and wilt be manufactured in accordance with both mandatory and voluntary standards, including:
- 60825-1:2007, Safety of laser products Part I : Equipment . IEC classification, requirements and user's guide.
- IEC 60601-2-22: Ed 1995, Medical electrical Equipment Part 2 : . Particular requirements for the safety of diagnostic and therapeutic laser equipment.
- IEC 60601-1: 1988 + AI: 1991 + A2:1995, Medical Electrical equipment . -Part I : general requirement for safety.
- IEC 60601-1-2: 200 1, Medical Electrical equipment -Part I : General . requirements for safety-2, Collateral Standard : electromagnetic compatibility - requirements and tests.
- IEC 6060 1-1-4: 2000, Medical electrical equipment Part I : General � requirements for safety -4 - Collateral standard : Programmable electrical medical systems, edition 1.1.
- The device also complies with European Medical Directive 93/42/EEC 2007/47/EEC and the US Federal Performance Amendment н Standards 21 CFR 1002. 10 Requirements (21 CFR 1040. 10 and 21 CFR 1040.11 for Class IV Laser). Part 1010.2 and 1010.3,Part 820 -Quality System Regulation, and have passed ISO 9001 and 13485 System Certification.
- IEC 62304:2006, medical device software Software life cycle processes
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "opto" in a stylized font. Each letter is bold and has a textured fill, giving it a slightly rough or vintage appearance. The letters are arranged horizontally, with the "o" on either end appearing as perfect circles and the "p" and "t" having a more angular, block-like design.
### Discussion of similarities and Differences with the Predicate 11. Product
Opto Hyalus Green Laser share the same indications for use and safety compliance, similar design features, functional features, and therefore are substantially equivalent to the predicate device , the IRIS Medical OcuLight GL Laser Systems (K031665). In addition a review of the predicate device demonstrate that the Opto Hyalus Green Laser is safe and effective as the predicate device as they share equivalent specifications / characteristics and are used to perform the same indicated surgical procedures.
There is no relevant difference in the specifications/characteristics of the Opto Hyalus Green Laser and its predicate IRIS Medical OcuLight GL Laser Systems (K031665).
### Conclusion:
Opto Hyalus Green Laser use the same fundamental technology features as the IRIS Medical OcuLight GL Laser Systems (K031665) and delivers the same level of effectiveness. Therefore, it is concluded that there is no significant difference in the basic function, safety and effectiveness between the IRIS Medical OcuLight GL Laser Systems (Predicate Device) and the Opto Hyalus Green Laser.
The Opto Hyalus Green Laser is substantially equivalent to predicate devices currently legally marketed for treatments in ophthalmology.
### Non-clinical performance data and conclusions from non-clinical 12. tests
Laboratory testing was conducted to validate and verify that the proposed device, Opto Hyalus Green Laser met all design specifications and was substantially equivalent to the predicate device.
Clinical Conclusion: No Clinical information is required.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "opto" in a stylized font. The letters are bold and have a textured fill, giving them a slightly distressed appearance. The letters are arranged horizontally, with each letter distinct and easily readable.
#### Conclusion 13.
Based on the information in this notification we concluded that Opto Hyalus Green Laser is substantially equivalent to predicate device currently legally marketed for the indication of retinal photocoagulation laser. Only applicable to uses in ophthalmology.
The Opto Hyalus Green Laser shares the same intended use, indications for use, and similar technical characteristics to the predicate device.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its body and wings. The bird is positioned within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 20 2012
Opto Electronic S/A % Mr. Paulo Aneas Lichti Joaquim A.R. De Souza Street 1071-Jardim Santa Felicia Sao Carlos, Sao Paulo Brazil 1:3630-330
Re: K111460
Trade/Device Name: Opto Hyalus Green Laser Regulation Number: 21 CFR 886.4390 Regulation Name: Opthalmic laser Regulatory Class: Class II Product Code: HQF Dated: August 09, 2012 Received: August 17, 2012
Dear Mr. Lichti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{8}------------------------------------------------
### Page 2 -- Mr. Paulo Aneas Lichti
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Milliken
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use Form
## 510(k) Number (if know): K111460
### Device Name: Opto Hyalus Green Laser
The Opto Hyalus Green Laser is indicated for retinal photocoagulation. Used for the following ophthalmic medical condition and treatment:
- Retinal photocoagulation, panretinal photocoagulation and intravitreal endophotocoagulation of vascular and structural abnormalities of the retina and choroids, including:
- Proliferative and nonproliferative diabetic retinopathy V
- V Choroidal neovascularization
- > Branch retinal vein occlusion
- The treatment of choroidal neovascularization associated with V wet age related macular degeneration Retinal tears and detachments
- Iridotomy, iridectomy, and trabeculoplasty in angle closure glaucoma and ● open angle glaucoma
Over-The-Counter Use Prescription Use Prescription Use _____________________________________________________________________________________________________________________________________________________________
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device s (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111460
510(k)-
Sec: 04_ Page 1 of 1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.