The Medical Frequency Doubled YAG Laser is intended for use in otolaryngological, dermatological and ophthalmic surgical procedures. A complete list is contained in the Indications for Use Statement.
Device Story
Solid-state, frequency-doubled, green Nd:YAG surgical laser; emits 532nm continuous wave energy for photothermolysis/photocoagulation of soft tissue. Used in clinical settings by physicians. Delivery via slit lamps, laser indirect ophthalmoscopes (LIO), or endoprobes; includes LaserLink adapter for slit lamp coupling. System monitors treatment parameters (wavelength, pulse rate, power, spot size) and safety systems (door interlock) to allow user intervention. Provides precise thermal energy to target tissues; benefits include controlled coagulation and tissue ablation for ophthalmic, dermatological, and otolaryngological conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing, including safety, performance, and functional testing.
Technological Characteristics
Solid-state, frequency-doubled Nd:YAG laser; 532nm emission wavelength. Continuous wave energy delivery. Compatible with slit lamps, LIO, and endoprobes. Includes safety interlocks and monitoring displays. Complies with 21 CFR 1040.10 and 1040.11 performance standards.
Indications for Use
Indicated for ophthalmology (retinal/pan-retinal photocoagulation, endophotocoagulation, macular treatments, laser trabeculoplasty), otolaryngology (stapedectomy, stapedotomy, myringotomy, lysis of adhesions, bleeding control, acoustic neuroma removal, soft tissue adhesion in otologic procedures), and dermatology (vascular and pigmented lesions). Prescription use only.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
# 510(K) Summary
# SEP 1 8 2009
This is a 510(K) summary in accordance with CFR807.82(c).
A. Submitter Information:
Lightmed Corporation Submitter:
Address: NO.1-1, Lane1, Pao-An St. Sec. 3, Shulin City, Taipei Hsien 23861, Taiwan
Owner/Operator Number: Mr. Gary Lee, President / 9040850
Contact person: Anita Chen, Regulatory Affair
TEL: +886-2-2688-1726
FAX: +886-2-2676-4920
- B. Device Name:
Product Name: Medical Frequency Doubled YAG Laser Trade Name: LightLas 532
Trade Name: Digoxin 552
Common name: Ophthalmic Laser, Surgical Laser
Classification name:
86 HQF, Laser, Ophthalmic
79 GEX, Laser Powered Surgical Instrument
Regulation Number:
21 CFR 886.4390, Ophthalmic Laser
21 CFR 878-4810, Laser surgical instrument for use in general and plastic
surgery and in dermatology
Regulatory Class: II
Performance standards: 21 CFR 1040.10 & 1040.11
C. Predicate Device Names:
(1) Device Names: Novus Spectra Laser System (K022327)
(2) Device Names: Viridis Derma(K020071)
(3) Device Nambe: VariLite Laser System (K041930)
{1}------------------------------------------------
#### D. Device Description:
The Medical Frequency Doubled YAG Laser is a solid state, frequency-doubled, green Nd: YAG surgical laser. It's an instrument used in the photothermolysis (photocoagulation) of soft tissue at en emission wavelength of 532nm. LightLas LaserLink : Laser delivery adapter that may be coupled to each of the above Selecta models and connected to a currently cleared LightLas 532 retinal photocoagulator to allow use of the slit lamp to deliver 532 nm continuous wave laser energy for retinal photocoagulation.
Compatible delivery devices include: slit lamps, slit lamp adapters/ attachments, laser indirect ophthalmoscopes(LIO) and endoprobe.
The intended use has not changed from the predicate devices.
E. Intended Use:
The Medical Frequency Doubled YAG Laser is intended for use in otolaryngological, dermatological and ophthalmic surgical procedures. A complete list is contained in the Indications for Use Statement.
The intended use has not changed from the predicate devices.
Technological Characteristics summary & Substantial Equivalence F. The Medical Frequency Doubled YAG Laser has the same indications for use as the Irides-OcuLight TX (K062369). They have similar functional elements such as treatment wavelengths, pulse rates, treatment power, spot size and cooling system. Control systems such as the door interlock, and the safety systems and displays are constantly monitored in these systems for user intervention during a procedure or maintenance.
- G. Performance Data Summary:
The appropriate testing including safety, performance and functional testing to determine substantial equivalence of the Medical Frequency Doubled Y AG Laser System.
- H. Conclusion
The LightLas 532 is substantially equivalent to predicate devices currently legally marketed for the indication of retinal photocoagulation. laser trabeculoplasty. the treatment of vascular and pigmented skin lesions, and other laser treatments.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
SEP 1 8 2009
Lightmed Corporation % Ms. Anita Chen Regulatory Affairs NO, 1-1, Lane 1, Pao-An St. Sec. 3 Shulin City, Tapei Hsien 23861 Taiwan
Re: K091534
Trade/Device Name: Medical Frequency Doubled YAG Laser Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic laser Regulatory Class: Class II Product Code: HQF Dated: August 24, 2009 Received: August 26, 2009
Dear Ms. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket avproval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registfation and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Ànita Chen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Milliman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1114162011 101 บอง มเลเบแนน เมเนเนเนเน บร มหมุนบอนบน บร
Kogi 534 510(K) Number (if Known): ____________________________________________________________________________________________________________________________________________________
Product Name: Medical Frequency Doubled YAG Laser Trade Name: LightLas 532
#### Indications for Use:
Ophthalmology:
- Retinal Photocoagulation .
- Pan Retinal Photocoagulation
- Endophotocoagulation
- Macular Treatments
- Laser Trabeculoplasty
## Otolaryngology:
- Stapedectomy
- · Stapedotomy
- · Myringotomy
- · Lysis of adhesions
- Control of bleeding
- · Removal of acoustic neuromas ·
- · Soft tissue adhesion in micro/macro otologic procedures.
#### Dermatology:
- · Vascular lesion
- · Pigmented lesion
The intended use has not changed from the predicate devices (K022327, K020071, K062369)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDR, Office of Device Evaluation (ODE)
| Prescription Use: | <span style="text-decoration: overline;">X</span> OR Over-The-Counter Use: |
|----------------------|----------------------------------------------------------------------------|
| (Per 21 CFR 801.109) | |
Med RE Ol for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
| 510K_ENT_A4.doc | 66 |
|-----------------|----|
|-----------------|----|
| 510(k) Number | <span style="text-decoration: overline;">K091534</span> |
|---------------|---------------------------------------------------------|
|---------------|---------------------------------------------------------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.