Air Techniques High Frequency Intra-Oral X-Ray is to be used as an extraoral source of X-Rays in Dental radiography.
Device Story
High-frequency intra-oral X-ray system for dental diagnostic radiography; utilizes extraoral X-ray source. System components include base unit, tube housing with beam-limiting device, control console with 9.84 ft coiled cord, rotating yoke, and scissor arm. Operates via two microprocessors: one for control/supervisory functions, one for user interface. Feedback circuits ensure X-ray output accuracy and reproducibility. Power supply provides selectable 50-70 kVp (1 kV steps) and 4, 6, 7, or 8 mA tube current. Exposure times range 0.04-4.00 seconds. Predefined parameters stored and selected via operator control panel. Used in dental clinics by dental professionals to acquire diagnostic images; output assists clinicians in identifying dental conditions. Benefits include precise, reproducible X-ray exposure for diagnostic purposes.
Clinical Evidence
Bench testing only. Performance demonstrated through compliance with UL/IEC 60601-1, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, and EMC standard IEC 60601-1-2. Performance testing conducted per FDA 21 CFR 1020.30 and 1020.31. All test results were satisfactory.
Indicated for use as an extraoral X-ray source in dental radiography for patients requiring dental diagnostic imaging.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
MAY 2 7 2011
K 111330.
Image /page/0/Picture/2 description: The image shows the logo for SKANRAY TECHNOLOGIES. The logo features a cluster of black circles on the left side, resembling a stylized atom or a group of particles. To the right of the circles, the word "SKANRAY" is written in bold, black capital letters. Below "SKANRAY", the word "TECHNOLOGIES" is written in smaller, black capital letters.
PRE-MARKET NOTIFICATION 510 (K) SUBMISSIONS-Section 5: Summary for
Air Techniques High Frequency Intra-Oral X-Ray Document #: TDS-000004-2-20, Rev#:1
### 510(K) SUMMARY
As required by CFR section 807.92(c)
Sponsor 5.1
| Date: | April, 06, 2011 |
|-----------------|--------------------------------------|
| Applicant: | Skanray Technologies Private Limited |
| | Plot # 15-17, Hebbal Industrial Area |
| | Mysore-570016, India |
| Contact Person: | Parasuramappa Belur |
| Telephone: | +91-821-2415559 |
| Fax: | +91-821-2403344 |
# 5.2 Initial Distributor/ Importer:
| Address: | Air Techniques Inc,<br>1295 Walt Whitman Road,<br>Melville, NY 11747-3062 |
|------------|---------------------------------------------------------------------------|
| Telephone: | 001-516-214-5541 |
| Fax: | 001-516-433-7683 |
## 5.3 Establishment Registration Number
The firm will be registered and listed prior to distribution of medical device.
### 5.4 Device Name
| Trade Name | Air Techniques High Frequency Intra-Oral X-Ray |
|---------------------|----------------------------------------------------------------------------------------------|
| Variants | Wall, Chair Mount & Floor Stand Identified by Part number<br>303-000134-X where, X= 6 to 11. |
| Common Name | High Frequency Intra Oral X-Ray System, |
| Classification Name | 76 EHD - Unit, X-Ray, Extra oral with Timer |
#### 5.5 Predicate Device
| INTRA SKAN DC | K103579 |
|------------------|----------|
| Progeny PREVA | K043092 |
| Dentsply Gx-770: | K935046. |
| Gendex 765DC: | K992610. |
## 5.6 Product Description
INTRA SKAN DC Variant Part number 303-000134-X is same as private labelled under Air Techniques High Frequency Intra-Oral X-Ray.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Skanray Technologies. The logo features a cluster of dots arranged in a semi-circular shape on the left side. To the right of the dots is the word "SKANRAY" in bold, stylized letters, and below that is the word "TECHNOLOGIES" in smaller, less bold letters.
PRE-MARKET NOTIFICATION 510 (K) SUBMISSIONS-Section 5: Summary for
Air Techniques High Frequency Intra-Oral X-Ray Document #: TDS-000004-2-20, Rev#:1
Air Techniques High Frequency Intra-Oral X-Ray is an X-Ray System with an extraoral X-Ray source for dental diagnostic radiography. The system houses two microprocessors, one for control / supervisory functions and another for man-machine/user interface. The technology incorporates feedback circuits to ensure accuracy & reproducibility of X-Ray output.
Air Techniques High Frequency Intra-Oral X-Ray consists of the following main components:
Base Unit
Tube Housing
Beam Limiting Device-inbuilt with tube Housing
Control Console, 9.84 Ft (3 m) coiled cord.
Rotating yoke for tube housing mounting
Extension arm
Scissor arm
Optional Components:
Long Cone 11.8in (300mm)
9.84 Ft (3m) coiled cord with exposure switch
The Power supply is regulated to provide a selectable 50 to 70 kVp in step of 1kV at a selectable tube current of 4, 6, 7, or 8 mA. The range of exposure times is 0.04 to 4.00 seconds with 1:15 duty. Predefined exposure parameters kV, mA & exposure time values may be stored in, selected & operated via the operator control panel.
# 5.7 Indications for Use
Air Techniques High Frequency Intra-Oral X-Ray is to be used as an extraoral source of X-Rays in Dental radiography
### 5.8 Safety, EMC and Performance Data
Safety and effectiveness is demonstrated by:
- Electrical, mechanical, environmental safety and performance testing according to . standard UL/IEC 60601-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32 was performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. All test results were satisfactory. Refer EMC Test Summary & Safety Test Report Summary in section "Electromagnetic Compatibility and Electrical Safety" of this submission document.
- . Performance testing performed according to FDA 21 CFR 1020.30, 21CFR1020.31 standards, Design Requirement specification, verification and validation plans. All test results were satisfactory. Refer summary of performance in section 18 "Performance Testing-Bench" of this submission document.
- Same indications for use as predicate devices. .
All of the above details combine to demonstrate that the Air Techniques High Frequency Intra-Oral X-Ray is safe and effective when the device is used as labelled.
#### END OF DOCUMENT
Section 5-2 of 2
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a caduceus-like symbol with three wavy lines representing the branches of government, and the words "USA" are included in the circular text.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Skanray Technologies c/o Mr. Morten Christensen Reviewer/Staff Engineer Underwriters Laboratores Inc. 455 East Trimble Road SAN JOSE CA 95131
MAY 2 7 2011
Re: K111330
Trade/Device Name: Air Techniques High Frequency Intra-Oral X-Ray Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: May 6, 2011 Received: May 11, 2011
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the ' Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): __
Device Name: Air Techniques High Frequency Intra-Oral X-Ray
Indications For Use:
Air Techniques High Frequency Intra-Oral X-Ray is to be used as an extraoral source of X-Rays in Dental radiography.
Prescription Use __ V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary Slsted
(Division Sign Off)
:fice of In Vitro D
K 111330
510K
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.