The Image-Arena software platform is intended to import, export, store, retrieve and report digital studies. The Image-Arena software is based on a SQL database and is intended as an image management system. The Image-Arena software can import certain digital 2D or 3D image file formats of different modalities. Image-Arena offers a Generic Clinical Application Package interface in order to connect TomTec applications as well as commercially available analysis and quantification tools to the Image-Arena platform. The software is suited for stand-alone workstations as well as for networked multisystem installations and therefore is an image management system for physician practices and hospitals. It is intended as a general purpose digital medical image processing tool. Image-Arena is not intended to be used for reading of mammography images. Image-Com software is intended for reviewing and measuring of digital medical data of different modalities. It can be driven by Image-Arena or other third party platforms and is intended to launch other commercially available analysis and quantification tools. Echo-Com software is intended to serve as a versatile solution for Stress echo examinations in patients who may not be receiving enough oxygen because of blocked arteries. Echo-Com software is intended for reviewing, wall motion scoring and reporting of stress echo studies.
Device Story
Image-Arena is an SQL-based image management platform for importing, storing, retrieving, and reporting digital medical studies (excluding mammography). It functions as a workstation or networked server/client system. The platform utilizes a Generic Clinical Application Package (CAP) interface to integrate TomTec and third-party analysis/quantification tools. Image-Com and Echo-Com are specific application packages for reviewing, measuring, and reporting medical data, including stress echo wall motion scoring. The system runs on Intel Pentium-based hardware using Windows. It does not generate original medical images; it processes and displays data imported from various modalities. Physicians and clinical staff use the system to review studies and generate reports, facilitating clinical decision-making by providing tools for quantitative analysis and standardized reporting.
Clinical Evidence
No clinical trial data provided. The document states the product concept was clinically accepted and performance was validated through internal software testing protocols at the module and system level.
Technological Characteristics
Software-based image management system; runs on Intel Pentium high-performance computers; Windows OS; SQL database; supports DICOM, HL7, and TCP/IP protocols; modular architecture with Generic CAP interface; standalone or networked client-server deployment.
Indications for Use
Indicated for use in physician practices and hospitals as a general-purpose digital medical image management and processing system. Echo-Com module is indicated for patients undergoing stress echo examinations, specifically those with potential myocardial ischemia due to blocked arteries.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for TOMTEC Imaging Systems. The logo is in black and white. The word "TOMTEC" is in large, bold letters, with the "M" stylized to look like a mountain. Below the word "TOMTEC" are the words "IMAGING SYSTEMS" in smaller letters.
K110667
February 10th, 2011
# Traditional 510(k) Summary
APR - 8 2011
Image-Arena and Image-Arena Applications Image-Arena 4.5 Echo-Com 4.5 Image-Com 4.5
Additional Marketing Name: IQ Workstation
### Owner's Name and Address
TomTec Imaging Systems GmbH Edisonstrasse 6 D-85716 Unterschleissheim
### Contact Person
Inge Scheidt QM & RA Officer Phone ++49-89-32175-515 Fax ++49-89-32175-750
## Common, Classification & Proprietary Names
Various Image Analysis System Common Name: Software Classification Name: Picture archiving and communications system Proprietary Name(s): Image-Arena and Image-Arena Applications
Image-Arena 4.5 Image-Com 4.5 Echo-Com 4.5
summary
page 1 of 6
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for TOMTEC Imaging Systems. The logo is in black and white and features a stylized font. The word "TOMTEC" is in large, bold letters, with a line above the "TEC" portion. Below the main word, in smaller letters, it says "IMAGING SYSTEMS".
## Predicate Devices
| Predicate Device 1 | K083348 | Image-Arena Platform 4.0, Server<br>Manager 4.0, Echo-Com 4.0,Image-Com<br>4.0, TomTec Imaging Systems GmbH |
|--------------------|---------|-------------------------------------------------------------------------------------------------------------|
| Predicate Device 2 | K061995 | Xcelera, Philips Medical Systems North<br>America CO. |
## Device Description
Image-Arena is an SQL database based image management system that provides the capability to import, export, store, retrieve and report digital studies.
lmage Arena is developed as a common interface platform for TomTec - and commercially available analysis and quantification tools (= clinical application packages) that can be connected to Image-Arena through the Generic Clinical Application Package interface (= Generic CAP Interface)
lmage-Arena manages different digital medical data from different modalities except digital mammography.
Image-Arena is suited as stand-alone workstation as well as networked multisystem server / client installations.
Image-Arena runs on an integrated Intel Pentium high performance computer system based on Microsoft™ Windows standards. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols.
Image-Arena provides the possibility to create user defined medical reports.
The system does not produce any original medical images.
Image-Com is a clinical application package software for reviewing and measuring of digital medical data. Image-Com is either embedded in Image-Arena platform or can be integrated into Third Party platforms, such as PACS or CVIS.
Echo-Com is a clinical application package software for reviewing and reporting of digital stress echo data. Echo-Com is either embedded in Image-Arena Platform or can be integrated into Third Party platforms, such as PACS or CVIS.
Image /page/1/Picture/13 description: The image shows a pattern of small, black, right-angle shapes arranged in a slightly curved line. The shapes are oriented in different directions, creating a dynamic and somewhat chaotic visual effect. The overall impression is one of movement and complexity within a simple geometric form.
summary
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for TOMTEC Imaging Systems. The logo is in black and white and features the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" is the phrase "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern.
### Indications for use and Intended use
The Image-Arena software platform is intended to import, export, store, retrieve and report digital studies. The Image-Arena software is based on a SQL database and is intended as an image management system. The Image-Arena software can import certain digital 2D or 3D image file formats of different modalities.
Image-Arena offers a Generic Clinical Application Package interface in order to connect TomTec applications as well as commercially available analysis and quantification tools to the Image-Arena platform.
The software is suited for stand-alone workstations as well as for networked multisystem installations and therefore is an image management system for physician practices and hospitals. It is intended as a general purpose digital medical image processing tool.
Image-Arena is not intended to be used for reading of mammography images.
Image-Com software is intended for reviewing and measuring of digital medical data of different modalities. It can be driven by Image-Arena or other third party platforms and is intended to launch other commercially available analysis and quantification tools.
Echo-Com software is intended to serve as a versatile solution for Stress echo examinations in patients who may not be receiving enough oxygen because of blocked arteries. Echo-Com software is intended for reviewing, wall motion scoring and reporting of stress echo studies.
### Technological Characteristics Comparison
The proposed and the predicate devices are software devices that run on high performance computer system to receive, store, display and digital process of medical images.
The Image-Arena Application software tool package is modular structured and consists of different software modules, combining the advantages of the previously FDA cleared software products:
summary
page 3 of 6
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Tomtec Imaging Systems. The logo is in black and white and features the word "Tomtec" in large, bold letters. Below the word "Tomtec" is the phrase "Imaging Systems" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
| Predicate Device 1 | K083348 | Image-Arena Platform 4.0, Server<br>Manager 4.0, Echo-Com 4.0,Image-Com<br>4.0, TomTec Imaging Systems GmbH |
|--------------------|---------|-------------------------------------------------------------------------------------------------------------|
| Predicate Device 2 | K061995 | Xcelera, Philips Medical Systems North<br>America CO. |
.
.
:
summary
.
:
:
५
. . . . .
page 4 of 6
.
ﻠ
:
. .
ਦੇ 8
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for TOMTEC Imaging Systems. The logo is in black and white, with the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" is the phrase "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern, and it is likely used to represent the company's brand.
#### Predicate Devices:
#### TomTec Image-Arena and Image-Arena
Applications
Image /page/4/Figure/4 description: The image shows a diagram of the Image-Arena platform 4.0. The diagram shows the different components of the platform, including Image-Arena Applications, Echo-Com 4.0, Image-Com 4.0, and Third Party Applications. The diagram also shows the Generic Interface, Database and Data Management System, Dicom Services, Import and Export Functions, Report and Archive Functions, and Basic Image Viewer.
・・ IA 3.0
applications
New Device: TomTec Image-Arena 4,5 and Image-Arena
Image /page/4/Figure/7 description: The image contains the word "Applications" in bold, black font. The word is presented in a simple, straightforward manner. The text appears to be isolated without any surrounding context or additional elements.
Image /page/4/Figure/8 description: The image shows a diagram of a software architecture. The architecture consists of several layers, including Image-Arena Applications, Echo-Com 4.5, Image-Com 4.5, TomTec Clinical Application Packages, and Third Party Applications. These applications are built on top of a Generic Interface, which in turn is built on top of a Database and Data Management System Imtproof. The database system provides Dicom Services, Import and Export Functions, Report and Archive Functions, and a Basic Image Viewer.
Predicate
Devices:
Image /page/4/Figure/10 description: The image shows the text "K083348" in a simple, sans-serif font. The text is black against a white background, providing a clear contrast. The numbers and letters are evenly spaced and easily readable.
TomTec ImageArena Platfrom 4.0,
Server Manager
Server Manager Server Manager
4.0, EchoCom 4.0, Image-Com 4.0
K061995
Xcelera
New Device:
TomTec Image-
Arena 4.5 and
Image-Arena
Applications
summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for TOMTEC Imaging Systems. The logo is black and white and features the word "TOMTEC" in large, bold letters. Below the word "TOMTEC" are the words "IMAGING SYSTEMS" in smaller letters. The logo is simple and modern.
## Discussion according non-clinical performance data testing
Testing was performed according to internal company procedures. Software testing and validation were done at the module and system level according to written test protocols established before testing was conducted. Test results were reviewed by designated technical professionals before software proceeded to release.
## Discussion according clinical performance data testing
The overall product concept was clinically accepted and the clinical test results support the conclusion that the device is as safe as effective, and performs as well as or better than the predicate device.
## Test Conclusions of non-clinical and clinical performance data
Test results support the conclusion, that the device is as safe as effective, and performs as well as or better than the predicate device.
Munich, February 10th, 2011
J. Heald
Inge Scheidt OM & RA Officer
Image /page/5/Picture/10 description: The image shows a series of black, right-angle shapes arranged in a curved line. The shapes are oriented in different directions, creating a dynamic and somewhat chaotic pattern. The background is plain white, which helps to emphasize the black shapes.
summary
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a symbol on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The symbol is a stylized representation of a human figure, and the text is in a bold, sans-serif font.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room – WO66-G609 Silver Spring, MD 20993-0002
Mr. Inge Scheidt QM & RA Officer TomTec Imaging Systems GmbH Edisonstrasse 6 Unterschleissheim, Bavaria, D-85716 GERMANY
APR - 8 2011
Re: K110667
Trade/Device Name: Imag-Arena Image.Arena Applications: Image-Arena 4.5; Echo-Com 4.5; Image-Com 4.5 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II
Product Code: LLZ Dated: February 10, 2011
Received: March 9, 2011
Dear Mr. Scheidt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{7}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary S Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510{k) Number (if known): کای 0 (6 6 7
Device Name:
## lmage-Arena and Image-.Arena Applications Image-Arena 4.5 Echo-Com 4.5 Image-Com 4.5
## Additional Marketing Name: IQ Workstation
Indications for Use:
The Image-Arena software platform is intended to import, export, store; retrieve and report digital studies. The Image-Arena software is based on a SQL - database and is intended as an image management system. The Image-Arena software can import certain digital 2D or 3D image file formats of different modalities.
Image-Arena offers a Generic Clinical Application Package interface in order to connect TomTec applications as well as commercially available analysis and quantification tools to the Image-Arena platform.
The software is suited for stand-alone workstations as well as for networked multisystem installations and therefore is an image management system for physician practices and hospitals. It is intended as a general purpose digital medical image processing tool.
Image-Arena is not intended to be used for reading of mammography images.
Image-Com software is intended for reviewing and measuring of digital medical data of different modalities. It can be driven by Image-Arena or other third party platforms and is intended to launch other commercially available analysis and quantification tools. .
Echo-Com software is intended to serve as a versatile solution for Stress echo examinations in patients who may not be receiving enough oxygen because of blocked arteries. Echo-Com software is intended for reviewing, wall motion scoring and reporting of stress echo studies.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--|--------------------------------------------------------|
| | <img alt="Signature" src="signature.png"/> |
Division Sign Off
Division of Technologycal Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
દત
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.