K061995 · Philips Medical Systems North America Co. · LLZ · Sep 6, 2006 · Radiology
Device Facts
Record ID
K061995
Device Name
XCELERA
Applicant
Philips Medical Systems North America Co.
Product Code
LLZ · Radiology
Decision Date
Sep 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Philips Xcelera software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital medical images. Philips Xcelera software is an integrated multimodality image and information system, designed to perform the necessary functions required for import /export/ storage / archiving / review / analysis/ quantification (for example: area, circumference, volume, velocity, length, percent, time, ejection fraction, pressure gradient, LV volumes) / reporting and database management of digital cardiovascular images, waveforms and data related to cardiology. Xcelera offers support for third party plug-ins in order to enable the use of commercially available tools and for analysis, quantification and reporting Xcelera offers support to launch specified 3th party programs from the user interface (Desktop Integration). It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams. Philips Xcelera software runs on standard information technology hardware and software. The Xcelera proprietary software product utilizes the standard Microsoft Windows Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols. Philips Xcelera will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components. The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.
Device Story
Philips Xcelera is an integrated multimodality image and information system (PACS) for cardiology. It ingests digital cardiovascular images, waveforms, and data; performs storage, archival, review, analysis, and quantification (e.g., area, volume, ejection fraction, pressure gradient); and generates reports. The system runs on standard IT hardware using the Microsoft Windows OS. It supports third-party plug-ins and desktop integration to launch external analysis tools. Used in clinical environments, it allows multiple users to access and exchange cardiology exams. The system utilizes standard TCP/IP, DICOM, and HL7 protocols for data exchange. It is configurable to support various modalities and viewing sites, aiding clinicians in diagnostic decision-making and workflow management.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics, intended use, and compliance with established standards (DICOM/HL7).
Technological Characteristics
Integrated multimodality PACS software. Runs on standard IT hardware with Microsoft Windows OS. Connectivity via TCP/IP, DICOM, and HL7 protocols. Modular design supports third-party plug-ins and desktop integration. No specific materials or energy sources defined as it is a software-based system.
Indications for Use
Indicated for the management, review, analysis, and quantification of digital cardiovascular images, waveforms, and data in a clinical setting. Intended for use by healthcare professionals to support cardiology diagnostics and reporting.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Siemens Medical Systems Inc., LEONARDO syngo Cardiology Workstation (K042203)
Submission Summary (Full Text)
{0}------------------------------------------------
Koli995
Image /page/0/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The word is presented in all capital letters and is horizontally oriented.
## Philips Medical Systems
SEP - 6 2006
# 510(k) Summary
Philips Xcelera
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
I General Information
| Company Name: | Philips Medical Systems North America Company | |
|----------------------|---------------------------------------------------------------------------------------------------------------|--|
| Address: | 22100 Bothell Everett Highway<br>Bothell Washington 98021-8431<br>USA | |
| Contact Person | Lynn T. Harmer | |
| Telephone Number: | 425-478-7312 | |
| Prepared (date): | 2006 May 12 | |
| Manufacturing Site: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best, The Netherlands | |
| Device Name: | Philips Xcelera | |
| Classification Name: | Picture Archiving and Communication System | |
| Regulation number | 892.2050 | |
| Classification: | Class: II | |
| ProCode: | 90 LLZ | |
| Common/Usual Name: | Workstation | |
| Predicate Devices: | Philips Medical Systems,<br>Harmony<br>Siemens Medical Systems Inc.,<br>LEONARDO syngo Cardiology Workstation | |
{1}------------------------------------------------
#### I I Information Supporting Substantial Equivalence Determination
#### System Description:
Philips Xcelera software is an integrated multimodality image and information system, designed to perform the necessary functions required for import /export/ storage / archiving / review / analysis/ quantification / reporting and database management of digital cardiovascular images, waveforms and data related to cardiology,
Xcelera offers support for third party plug-ins in order to enable the use of commercially available tools and for analysis, quantification and reporting Xcelera offers support to launch specified 3th party programs from the user interface (Desktop Integration). It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams.
Philips Xcelera software runs on standard information technology hardware and software. The Xcelera proprietary software product utilizes the standard Microsoft Windows Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols.
Philips Xcelera will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components
The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.
### Intended Use:
Philips Xcelera software product is an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital medical images.
### General Safety and Effectiveness:
The Xcelera complies with ACR/NEMA DICOM digital imaging communication standard.
### Conclusion:
The Philips Xcelera does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers the Xcelera to be substantially equivalent to the Philips Medical Systems, Harmony (K022788) and the Siemens Medical Systems Inc., LEONARDO syngo Cardiology Workstation (K042203).
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the eagle symbol.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## SEP - 6 2006
Philips Medical Systems North America c/o Mr. Marc M. Mouser Program Reviewer Underwriters Laboratories, Inc. 2600 N. W. Lake Road CAMAS WA 98607-8542
K061995 Re:
Trade/Device Name: Philips Xcelera Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 4, 2006 Received: August 21, 2006
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Booton's ro(t) predantially equivalent (for the indications for relerenced above and nave decemined anceted predicate devices marketed in interstate commerce use stated in the encrosury manations p Medical Device Amendments, or to devices that prob io May 26, 1770, the chaomistic date of the Federal Food, Drug, and Cosmetic
have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosm have been recrassmed in accordance with and portication (PMA). You may, ACt (Act) that do flor require approval or a peneral controls provisions of the Act. The general therefore, mailier the device, subjoct to the general for annual registration, listing of devices, good controls provisions of also ing, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket II your device is classifica (see above) into entrols. Existing major regulations affecting your Approval), it thay be subject to such adamons. Title 21, Parts 800 to 898. In addition, FDA active can be round in the cements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular logo with the text "FDA Centennial" and the years "1906-2006". The logo also features three stars at the bottom. The text is arranged around the circle's perimeter, and the FDA acronym is prominently displayed in the center.
d Promoting Public He
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r toase of acressionation that your device complies with other requirements of the Act that I Drederal statutes and regulations administered by other Federal agencies. You must or any I bathar all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IT Fart 6077, adoling (21 CFR Part 820); and if applicable, the electronic form in the quant) Bystems (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) rins letter wif anon you to ough finding of substantial equivalence of your device to a legally premaired predicated on: "The Dr. In a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rf you desire spoorned at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" / tion, prease note and regain other general information on your responsibilities under (21 CF R F art 60 : 37). I coof Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincercly yours,
Nancy C. Brogdon
Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(K) Number (if known):
Device Name: Philips Xcelera
Philips Xcelera software is an integrated multimodality image and information system, designed to perform the necessary functions required for import /export/ storage / archiving / review / analysis/ quantification (for example: area, circumference, volume, velocity, length, percent, time, ejection fraction, pressure gradient, LV volumes) / reporting and database management of digital cardiovascular images, waveforms and data related to cardiology.
Xcelera offers support for third party plug-ins in order to enable the use of commercially available tools and for analysis, quantification and reporting Xcelera offers support to launch specified 3th party programs from the user interface (Desktop Integration). It allows multiple users fast access to, and exchange of specific and/or multiple cardiology exams.
Philips Xcelera software runs on standard information technology hardware and software. The Xcelera proprietary software product utilizes the standard Microsoft Windows Operating System and user interface. Communication and data exchange are done using standard TCP/IP, DICOM and HL7 protocols.
Philips Xcelera will also be made available for use on specified Cardiovascular Monitoring Systems, which use suitable hardware components.
The modular design allows configurability to tailor the image import, archive and communications solution to one's particular budgetary and performance needs. The number of modalities and reporting and/or viewing sites can be configured per system.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Hepson
(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.