K110434 · BEIJING SYNTECH LASER CO., LTD. · GEX · Apr 19, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110434
Device Name
TRIXEL CO2 LASER
Applicant
BEIJING SYNTECH LASER CO., LTD.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 19, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The equipment is used for human tissue vaporization. coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology,
Device Story
Trixel CO2 Laser (CFL-10) and Trixel II CO2 Laser (UFL-60) are carbon dioxide laser systems for medical and aesthetic use. The device emits a high-energy laser beam that is highly absorbed by water in human tissue, causing vaporization and coagulation. It is used for treating skin conditions including fine and coarse wrinkles, scars, uneven pigmentation, and dilated pores. The system includes a mainframe, footswitch, and various handpieces, including a bitmap scanner handpiece and an ordinary treatment handpiece with an aiming beam. Operated by clinicians in medical settings, the laser facilitates photothermolysis to stimulate deep cell regeneration for skin improvement. The device is used for tissue vaporization and coagulation across multiple surgical specialties.
Clinical Evidence
Bench testing only. Compliance with IEC 60825-1:2007, IEC 60601-2-22:2007, IEC 60601-1:1988+A1:1991+A2:1995, and IEC 60601-1-2:2001+A1:2004 was verified to demonstrate substantial equivalence.
Technological Characteristics
Carbon dioxide laser system; wavelength-based tissue interaction; air or water cooling systems; aiming beam <5mW; power input 120V. Complies with IEC 60825-1, IEC 60601-2-22, IEC 60601-1, and IEC 60601-1-2 standards.
Indications for Use
Indicated for human tissue vaporization and coagulation in dermatology, plastic surgery, general surgery, gynecology, podiatry, dental, and otorhinolaryngology procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Page (1) of (4)
Section III 510(k) Summary
Project #: MIDI INK0382010Bd
1 9 2011
### Section III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number:
1. Date of Submission: 2010.12.21 2. Sponsor Beijing Syntech Laser Co., Ltd. 5F, No.1 Building, No.65 Jiancaicheng West Road, Xisanqi, Haidian District, Beijing, 100096, P.R.China Contact Person: Jin Jinghua Position: Vice general manager Tel: +86-10-82939866-234 Fax: +86-10-82935528 Email: sales@syntechlaser.com
3. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
4. Proposed Device Identification
Proposed Device Name: Trixel CO2 Laser/ Trixel II CO2 Laser Proposed Device Model: CFL-10/ UFL-60
Classification: 2 Product Code: GEX Regulation Number: 21 CFR 878.4810 Review Panel: 878 General and Plastic Surgery
## Intended Use Statement:
The equipment is used for human tissue vaporization. coagulation in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology,
{1}------------------------------------------------
K110434 Page (2) of
Section III 510(k) Summary
5. US agent
U.S. Agent Contact Name WEI HUANG
U.S. Agent Business Name MID-LINK INTERNATIONAL, INC.
U.S. Agent Full Address 307 EAGLE HEIGHTS APT E, MADISON, WISCONSIN, 53705, UNITED STATES
U.S. Agent E-mail info@mid-link.net
U.S. Agent Phone Number 608 -3355979
U.S. Agent Fax Number 760 -4665084
6. Predicate Device Identification
510(k) Number: K101555 Product Name: CO2 LASER SYSTEM Manufacturer: ADVANCED TECHNOLOGY LASER CO., LTD.
## 7. Device Description
Trixel CO2 Laser is a carbon dioxide laser used in medical and aesthetic industry for treatment of such skin conditions as fine and coarse wrinkles, scars of various origin, uneven pigmentation and dilated pores. Due to the CO2 laser's high absorption of water, its high-energy beam of laser light interacts with the skin's surface causing the upper layer to peel off and use photothermolysis to stimulate deep cell regeneration and then achieve the target of skin improvement.
The equipment is mainly used for human tissue vaporization, coagulation and exposure to achieve the purpose of treatment.
CFL-10 Trixel CO2 Laser:
The equipment includes the following:
One mainframe;
One footswitch; One data cable;
One power wire;
One ventilation tube;
Six Wire clips;
Two keys:
One trixel (bitmap) scanner handpiece;
Two trixel laser position heads
One ordinary treatment handpiece with aimed light
A laser goggles;
Three 3.15A fuses;
One user manual.
UFL-60 Trixel II CO2 Laser:
The equipment includes the following:
One mainframe;
{2}------------------------------------------------
Premarket Notification 510(k) Submission
Section III 510(k) Summary
One set of footswitch; One data cable; One power wire: One ventilation tube; Six Wire clip; Two keys; One trixel (bitmap) scanner handpiece; Three trixel laser position heads; One ordinary treatment handpiece with aimed light A laser goggles; Three SA fuses; One user manual.
#### 8. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- IEC 60825-1: 2007, Safety of laser products Part 1: Equipment classification, requirements. .
- . IEC 60601-2-22: 2007, Medical Electrical Equipment - Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment.
- . IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment - Partl: General requirements for safety.
- . IEC60601-1-2:2001+A1:2004, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
Substantially Equivalent Conclusion 9.
The proposed device, Trixel CO2 Laser/Trixel H CO2 Laser, has same Product Code, Regulation Number. Class. Maximum Power. Work Mode. Wavelength. Mode Structure, Spot Size, Pulse Setting, etc., but only different in five aspects as following:
- … Intended Use
- 2. Light Arm
- 3. Aiming Beam
- 4. Cooling System
- న్న Power Input
For these different, we have performed analysis as following:
Analysis 1 the intended use of predicate device includes the one of proposed device. They both have the function of vaporization and coagulation. The above clearance for intended use does not introduce any problems regarding the safety and effectiveness of the device.
{3}------------------------------------------------
Avalysis 2: though the length of light arm is different. CFL-10: 0.97m. UFL-60: 1.17m, predicate device: 1.3m. it doesn't affect the treatment itself. This difference does not introduce any problems regarding the safety and effectiveness of the device.
See analysis 3: Both the output range of aiming beam is <SmW. which complies with the requirement of IEC 60601-2-22:2007. This difference does not introduce any problems regarding the safety and effectiveness of the device.
Analysis 4: air cooling system and water cooling system can both cool down the system. This difference does not introduce any problems regarding the safety and effectiveness of the device, they both comply with the requirement of IEC 60601-1.
Analysis 3: they have the similar power input, CF1,-10 Trixel CO2 Laser: - 120 V. UFL-60 Trixel II CO2 Laser: 120 V. predicate device: 115V. This difference does not introduce any problems regarding the safety and effectiveness of the device.
Basic on analysis above, proposed device has determined to be Substantially Equivalent (SE) to the predicate device, eBeam CQ2Laser System (K101555), in respect of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular text element and a stylized graphic of an eagle. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion. To the right of the text is a stylized graphic of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
APR 1 9 2011
Beijing Syntech Laser Co., Ltd % Underwriters Laboratories, Inc. Mr. Marc M. Mouser 2600 NW Lake Road Camas, Washington 98607-9526
Re: K110434
Trade/Device Name: Trixel CO2 Laser/Trixel II CO2 Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 28, 2011 Received: April 4, 2011
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
Page 2 - Mr. Dennis Hahn, RAC
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.html 15809.html fast the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR, Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free no (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
AS B- nh
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Section II Indications for Use
510(k) Number: Device Name: Trixel CO2 Laser/Trixel II CO2 Laser
## Indications for Use:
The equipment is used for human tissue vaporization in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology,
�PRESCRIPTION USE (Part 21 CFR 801 Subpart D) [ ] OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Neitke Ogiden for mxm
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110434
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.