Care back pain relief stimulator is intended for temporary relief of pain associated with sore and aching muscles in the lower back due to strain form exercise or normal household and working activities.
Device Story
Care back pain relief stimulator (models TENS 7000, LT3074) is a non-invasive transcutaneous electrical nerve stimulator (TENS) for over-the-counter use. The device consists of a stimulation unit, support belt, electrodes, and snap cables. Electrodes are sewn into the support belt, which is worn on the lower back. The device delivers electrical pulses to the treatment area to provide temporary pain relief. Users operate the device to manage muscle soreness and aching resulting from exercise or daily activities. The device is powered by 9V or 4.5V batteries. It provides adjustable pulse width, frequency, and timer settings. Healthcare providers do not operate the device; it is intended for patient self-use. The device benefits patients by providing a non-pharmacological option for managing lower back pain.
Clinical Evidence
Bench testing only. No clinical data was conducted or relied upon. Conformity to voluntary standards (IEC60601-1, IEC60601-1-2, IEC60601-2-10, ISO 10993, ISO 14971) and software verification per FDA guidance were provided.
Indicated for temporary relief of pain associated with sore and aching muscles in the lower back due to strain from exercise or normal household and working activities.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
K110390
File No: WMI-04-OTCT -FDA-05 Version: 1.3
JUL 2 9 2011
)
# 510(k) SUMMARY
Care back pain relief stimulator, K (
07/27/2011 Date of Submission: Shenzhen Dongdixin Technology Co., Ltd. 510(k) Submitter's Name: No.3bldg. xiliyangguang industrial Estae xilixiaobaimang Nanshan district Shenzhen, China Address: 518108 +86(755) 27652471 Telephone: Fax: +86(755) 27652674 510(k) Correspondent Kang Jian Ping contact: microcong@gmail.com E-mail:
{1}------------------------------------------------
#### 1. Proposed Device:
| Trade Name: | Care back pain relief stimulator |
|----------------------|------------------------------------|
| Model: | TENS 7000,LT3074 |
| Classification Name: | Stimulator, Nerve, Transcutaneous. |
| | Over-The-Counter |
| Regulation Number: | 21CRF 882.5890 |
| Product Code: | NUH |
| Device Class: | II |
#### 2. Predicate Device:
| Predicate Device: | Low Back Pain Relief System |
|-------------------|-----------------------------|
| 510(k) Number: | K060222 |
| Manufacturer: | Gemore Technology Co.,Ltd |
#### 3. Description of Proposed Device:
Care back pain relief stimulator, which includes models TENS 7000 and LT3074 are non-invasive devices which are tented for over the counter use in temporary relief of pain associated with sore and aching muscles in the lower back due to strain form exercise or normal household and working activities.
The devices contain the following main parts: TENS stimulation unit, Support Belt, electrode, and snap cable.
The electrode is permanently sewn up in the support belt. Each belt is flexible, and is available in range of sizes to ensure good patient contact. A male snap connector is placed within the electrode and is connected via the female snap connector to a short lead wire. The lead wire has a female pin connection at the distal end which accepts the lead wire connection from the stimulator.
The device can be worn on the low back part of user so as to place the electrode on the treatment location of low back.
#### 4. Proposed Device Intended Use Statement:
Care back pain relief stimulator is intended for temporary relief of pain associated with sore and aching muscles in the lower back due to strain form exercise or
{2}------------------------------------------------
normal household and working activities.
#### Technological Characteristics: 5.
Both the stimulator and the Predicate device stimulator have the same intended use and fundamental technology. A side-by-side comparison of the technological characteristics of the care back pain relief and the cited predicate devices is included in the 510(k) submission.
| Parameter | Care back pain relief stimulator<br>Model: TENS7000,LT3074 | Low Back Pain Relief<br>System, model<br>GM320PP |
|------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Care back pain relief stimulator is<br>intended for temporary relief of<br>pain associated with sore and<br>aching muscles in the lower back<br>due to strain form exercise or<br>normal household and working<br>activities. | The model GM320PP is<br>intended for temporary<br>relief of pain associated<br>with sore and aching<br>muscles in the lower back<br>due to strain from exercise<br>or normal household and<br>work activities. |
| Power Source | 9V Battery for TENS7000<br>4.5V Battery for LT3074 | 9V Battery |
| Max Output Voltage (V)<br>±20% | 40.0V@500Ω for TENS 7000 | 40.0V@500Ω |
| | 45.0V@500Ω for LT3074 | |
| | 100.0V@2kΩ for TENS 7000 | 120.0V@2kΩ |
| | 50.0V@2kΩ for LT3074 | |
| | 110.0V@10kΩ for TENS 7000 | 150.0V@10kΩ |
| | 60.0V@10kΩ for LT3074 | |
| Max Output Current (mA)<br>±20% | 80.0mA @500Ω for TENS 7000 | 80.0 mA @500Ω |
| | 90.0mA @500Ω for LT3074 | |
| | 50.0mA @2kΩ for TENS 7000 | 60.0 mA @2kΩ |
| | 25.0mA @2kΩ for LT3074 | |
| | 11.0mA @10kΩfor TENS 7000 | 15.0 mA @10kΩ |
| | 6.0mA @2kΩ for LT3074 | |
| Pulse Width Range | 50-300us for TENS 7000<br>50-250us for LT3074 | 150-260 us |
| Frequency | 2-150Hz for TENS 7000<br>2-110Hz for LT3074 | 2-120Hz |
| Timer Range (minutes) | 5 - 60 minutes or Continuous fo<br>TENS7000<br>30 minutes for LT3074 | 15min, 30min, 60min<br>and continuous |
| Maximum Average Current<br>(average absolute value),<br>mA (refer to the remark 1) | 3.6mA,500Ω, for TENS 7000<br>2.5mA,500Ω, for LT3074 | 2.5mA,500Ω |
| RMS current (refer to the remark 1) | 17.0 mA.r.m.s for TENS 7000 | 20.0 mA.r.m.s |
| | 14.2 mA.r.m.s for LT3074 | |
| | | |
| Maximum Current Density<br>(refer to the remark 1) | $0.56mArms/cm^2,500Ω$ for TENS<br>7000<br>$0.14mArms/cm^2,500Ω$ for LT3074 | $0.56mArms/cm^2,500Ω$ |
| Maximum Power Density<br>(refer to the remark 1) | $4.8mWrms/cm^2,500Ω$ for TENS<br>7000<br>$1.0mWrms/cm2,500Ω$ for LT3074 | $3.3mWrms/cm^2,500Ω$ |
| Compliance with Voluntar<br>Standards? | IEC60601-1, IEC60601-1-2<br>IEC60601-2-10 | IEC60601-1, IEC60601-1-2<br>IEC60601-2-10 |
| Compliance with 21 CFR<br>898? | Yes | Yes |
| Weight (grams.) | 150 grams with battery for<br>TENS7000<br>80 grams with battery for LT3074 | 140 grams with battery |
| Dimensions (mm.) H x W x T | 101*61*24.5cm for TENS7000<br>85*72*30cm for LT3074 | 108*61.5*25 |
| Housing Materials &<br>Construction | Enclosure: ABS,94V-1,80°C,UL<br>Approved | Enclosure:<br>ABS,94V-1,80°C,UL<br>Approved |
| Materials of electrodes: | Rubber or fabric | Gel |
{3}------------------------------------------------
#### File No: WMI-04-OTCT -FDA-05 Version: 1.3
#### Non-Clinical Tests Performed: 6.
Care back pain relief stimulator did not conduct, nor rely upon, clinical tests to determine substantial equivalence. Care back pain relief stimulator has received test reports certifying that the Care back pain relief stimulator was tested and found to be in conformity with voluntary standards includes IEC60601, IEC60601-1-2, IEC60601-2-10, ISO 10993 and ISO14971. The software verification has been carried out according to the FDA Guidance of the content of Premarket submissions for the software contained in Medical Devices. This test reports have provided as part of this premarket notification.
### 7. Conclusions:
The Care back pain relief Stimulator, which includes models TENS 7000 and LT3074, has the same intended use and technological characteristics as the cited predicate device. Moreover, bench testing, safety report and Risk Analysis Report documentation supplied in this submission demonstrates that the difference in the submitted models could maintain the same safety and effectiveness as that of the cited predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, the care back pain relief stimulator is substantially equivalent to the predicate device.
{4}------------------------------------------------
File No: WMI-04-OTCT -FDA-05 Version: 1.3
#### Remark 1
#### TENS 7000 sample calculations:
Maximum Average Current=I max*pulse rate*pulse width=80mA*150Hz*300us*$10^{-6}$=3.6mA
300us * 80mA = 17.0mA.r.m.s Ir.m.s = 150 Hz * 1000000 17.0mArm.s Ir.m.s Maximum Current Density = = 0.56mA.r.m.s/$cm^2$ 30cm2 S 300us Vr.m.s * 40V = 8.5V r.m.s 150Hz 10000000 8.5Vr.m.s Vr.m.s Maximum Voltage Density = = 0.28V r.m.s/$cm^2$ 30cm2 S Vr.m.s * Ir.m.s 8.5Vr.m.s * 17mAr.m.s Maximum Power Density = = 4.8mW r.m.s / cm2 S 30cm2
#### LT3074 sample calculations:
Maximum Average Current=I max*pulse rate*pulse width=90mA*100Hz*250us*10-6=2.5mA
$$Ir.m.s = \sqrt{\frac{5}{5} * 100 Hz * \frac{250us}{1000000}} * 90mA = 14.2mAr.m.s$$
$$Maximum Current Density = \frac{Ir.m.s}{S} = \frac{14.2mAr.m.s}{100cm^{2}} = 0.14mAr.m.s/cm^{2}$$
$$Vr.m.s = \sqrt{\frac{5}{5} * 100Hz * \frac{250us}{1000000}} * 45V = 7.1Vr.m.s$$
$$Maximum Voltage Density = \frac{Vr.m.s}{S} = \frac{7.1Vr.m.s}{100cm^{2}} = 0.07Vr.m.s/cm^{2}$$
$$Maximum Power Density = \frac{Vr.m.s * Ir.m.s}{S} = \frac{7.1Vr.m.s * 14.2mAr.m.s}{100cm^{2}} = 1.0mWr.m.s/cm^{2}$$
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Handelhaus Dittman GMBH Shenzen Dongdixin Technology Co. LTD. c/o Mr. Kang Jiang Ping R.A. Director No. 3 Building Xiliyiangguang Industrial Estate Xillixaobaim Shenzen China 518108
JUL 2 9 2011
Re: K110390
Trade/Device Name: ETG 255 Combo Stimulator, TEN 260 Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Codes: GZJ, IPF Dated: November 30, 2011 Received: February 11, 2011
Dear Mr. Jiang Ping:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Keria Alexander
Malvina B. Eydelman, M.D. Director
Division of Ophthalmic, Neurological, Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number ( ):
Device Name:
ETG 255 Combo Stimulator, TEN 260 Combo Stimulator
## Indications for Use:
ETG 255 Combo Stimulator,
### For TENS mode
Symptomatic relief of chronic intractable pain, relief of acute post-surgical and posttraumatic pain.
#### For EMS mode
- 1. Relaxation of muscle spasm.
- Increase of local blood flow circulation 2.
- Prevention or retardation of disuse atrophy 3.
- Muscle re-education 4.
- Maintaining or increasing range of motion. 5.
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis 6.
### TEN 260 Stimulator
Symptomatic relief of chronic intractable pain, relief of acute post-surgical and posttraumatic pain.
Prescription Use V V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign Off
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K110390
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.