The DirectVision® Catheter, Uncoated is indicated for use in the drainage and/or collection and/or measurement of urine. It is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as the nephrostomy tract.
Device Story
DirectVision® Catheter, Uncoated is a flexible, 100% silicone, latex-free, 3-way Foley catheter; features straight or angled council tip with eyelet for guidewire insertion. Device includes three lumens: one for urinary drainage to collection container, one for balloon inflation/deflation, and one for bladder irrigation. Designed for use with PercuVision DirectVision™ Guide System (K091600) to provide a sheath for a microendoscope; includes irrigation channel to clear debris from microendoscope field of view. Used by clinicians in clinical settings for urinary tract drainage. Output is physical drainage of fluids; enables visualization when used with compatible microendoscope system. Benefits include facilitated catheter insertion and improved visualization during urological procedures.
Clinical Evidence
No clinical data. Substantial equivalence based on design, material, and intended use comparisons to predicate devices.
Technological Characteristics
100% silicone, latex-free construction. Triple lumen (3-way) Foley design. Available in 14-24 French sizes with 10cc or 30cc balloon. Straight or angled council tip. Supplied sterile. Non-lubricated surface.
Indications for Use
Indicated for drainage, collection, and/or measurement of urine in patients requiring urological catheterization via urethra, suprapubic, or nephrostomy placement.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
K110214
(pg 1 of 3)
#### Percu Vision DirectVision® Catheter, Uncoated Special 510(k) Premarket-Notification Submission
| Special 510(k) Summary | |
|------------------------|---------------------------------------------------------------|
| | MAY 27 2011 |
| A) Submitted by: | PercuVision |
| | 6264 S. Sunbury Rd. |
| | Westerville, Ohio 43081 |
| | Phone number is 614-891-4800 |
| | Toll Free 1-877-913-6333 |
| | Fax 614-891-3500 |
| Contact: | MEDIcept |
| | 200 Homer Ave |
| | Ashland, MA 01721 |
| | F. David Rothkopf |
| | 508-231-8842 Tel |
| | 508-231-8861 Fax |
| B) Device Name: | |
| Common Name: | Urologic Catheter |
| Proprietary Name: | DirectVision® Catheter, Uncoated |
| Device Class: | 21 CFR 876.5130 Urological Catheter and Accessories, Class II |
| Product Code | EZL (Catheter, Retention Type, Balloon) |
| C) Predicate: | K090262 PercuCath™ Urinary Catheter (with lubricious coating) |
| | K860484 Dover Foley Catheter |
D) Device Description:
్
The DirectVision® Catheter,-Uncoated is a flexible tubular device that is inserted through the urethra and used to pass fluids to and from the urinary tract. It is a 100% silicone, latex-free, triple lumen (3-Way) Foley catheter with a straight or angled council tip to facilitate insertion. The DirectVision® Catheter, Uncoated is composed of a silicone tube that trifurcates into three (3) lumens, a silicone balloon, and a two-way valve. Of the three lumens, one lumen is used for urinary drainage and can be connected to a urine collection container (drainage bag or urine meter); one lumen has a two-way valve for inflation/deflation of the Foley balloon; and the third lumen can be used for irrigation of the bladder. The council tip has an opening (eyelet) in the tip that can be used to pass a guidewire or similar device to facilitate catheter insertion. This catheter device will be available with a single or double drainage eyelet, 14 through 24 French shafts, and either a 10 cc or 30 cc balloon. The DirectVision® Catheter, Uncoated is designed to be usable with the PercuVision DirectVision™ Guide System (K091600). In the DirectVision™ Guide System, the DirectVision® Catheter, Uncoated provides a sheath around the microendoscope to facilitate its use. The catheter also incorporates a channel to be used for supplying irrigation to the microendoscope tip for clearing away debris from the field of view during use.
January 24, 2011
{1}------------------------------------------------
# PercuVision DirectVision® Catheter, Uncoated Special 510(k) Premarket-Notification Submission
### E) Intended Use:
- (
The DirectVision® Catheter, Uncoated is indicated for use in the drainage and/or collection and/or measurement of urine. It is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as the nephrostomy tract.
F) Comparison to Predicate Device(s):
| | DirectVision®<br>Catheter,<br>Uncoated | PercuCath™<br>Urinary Catheter<br>(with lubricious<br>coating) | Dover Foley<br>Catheter |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| 510K Number | | K090262 | K860484 |
| Product code | EZL | EZL | KOD |
| Indication for<br>Use | For use in the<br>drainage and/or<br>collection and/or<br>measurement of<br>urine. | For use in the<br>drainage and/or<br>collection and/or<br>measurement of<br>urine. | For use in the<br>drainage and/or<br>collection and/or<br>measurement of<br>urine. |
| Intended Use | It is intended to be<br>used as a<br>urological catheter<br>inserted through<br>the urethra for the<br>purpose of<br>draining urine and<br>other fluids from<br>the urinary tract.<br>However, drainage<br>is sometimes<br>accomplished by<br>suprapubic or<br>other placement of<br>the catheter, such<br>as the<br>nephrostomy tract. | It is intended to be<br>used as a<br>urological catheter<br>inserted through<br>the urethra for the<br>purpose of<br>draining urine and<br>other fluids from<br>the urinary tract.<br>However, drainage<br>is sometimes<br>accomplished by<br>suprapubic or<br>other placement of<br>the catheter, such<br>as the<br>nephrostomy tract. | Not available |
| Design<br>Coating | Uncoated | Coated | Uncoated |
| Sterility | Supplied sterile | Supplied sterile | Supplied sterile |
| Biocompatible | Yes | Yes | Yes |
January 24, 2011
{2}------------------------------------------------
#### PercuVision DirectVision® Catheter, Uncoated Special 510(k) Premarket-Notification Submission
#### Substantial Equivalence Discussion
There is no change in Indication for Use or Intended use of the DirectVision® Catheter, Uncoated. The Direct Vision® Catheter, Uncoated is identical in all aspects to the Percu Vision PercuCath™ Catheter Urinary Catheter with lubricious coating, except for being uncoated. The DirectVision® Catheter, Uncoated is similar to the Dover Foley catheter, which also does not have a lubricious coating. Percu Vision does not believe that removal of the coating raises new issues of safety or effectiveness, and that the DirectVision® Catheter, Uncoated is substantially equivalent to the predicates in terms of materials, intended use, basic design concept and biomechanical properties.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is depicted with three lines forming a wing-like shape above a staff with a snake coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60! Silver Spring, MD 20993-0002
PercuVision, LLC c/o F. David Rothkopf President MEDIcept, Inc. 200 Homer Avenue ASHLAND MA 01721
# MAY 2 7 2011
Re: K110214
Trade/Device Name: DirectVision® Catheter, Uncoated Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: May 3, 2011 Received: May 4, 2011
Dear Mr. Rothkopf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable; the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR.Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lehman MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## PercuVision DirectVision® Catheter; Uncoated Special 510(k) Premarket-Notification Submission
# Indications for Use
14110214 510(k) Number (if known):
Device Name: DirectVision® Catheter, Uncoated
Indications for Use:
The DirectVision® Catheter, Uncoated is indicated for use in the drainage and/or collection and/or measurement of urine. :
Prescription Use _ 21CFR 801, Subpart D OR Over-the-Counter Use _ 21 CFR 876.1500
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NECESSARY)
Hula Rume
ivision /Silan-Off n of Reproductive, Gastro-Renal, and logical Devices 10(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.