The PercuCath™ Catheter is indicated for use in the drainage and/or collection and/or measurement of urine. It is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as the nephrostomy tract.
Device Story
PercuCath™ Urinary Catheter is a flexible, 100% silicone, latex-free, triple-lumen (3-way) Foley catheter. Features include a lubricious hydrophilic coating, a straight or angled council tip with an eyelet for guidewire insertion, a silicone balloon, and a two-way valve for balloon inflation/deflation. The three lumens facilitate urinary drainage, balloon control, and bladder irrigation. Available in 14-24 Fr sizes with 10 cc or 30 cc balloons. Used by clinicians to manage urinary drainage and fluid collection. The device is inserted into the urinary tract; the council tip allows for guidewire-assisted placement. Output is the passive drainage of urine into a collection container. Benefits include effective fluid management and facilitated catheterization via the council tip design.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
100% silicone construction; latex-free; triple-lumen (3-way) design; lubricious hydrophilic coating; straight or angled council tip with eyelet; 14-24 Fr shaft sizes; 10 cc or 30 cc balloon; two-way valve for balloon inflation/deflation.
Indications for Use
Indicated for drainage, collection, and measurement of urine and other fluids from the urinary tract via urethral, suprapubic, or nephrostomy placement.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Dover® 100% Silicone Foley Council Tip Catheter (K860484)
Submission Summary (Full Text)
{0}------------------------------------------------
510(k), [Percuvision]
Page 18 of 50
K090262
MAY 28 2009
# 5. 510(k) Summary as required by 21 CFR 807.92(c) -
510(k) Owner: Percuvision LLC 765 North Hamilton Road. Suite 260A Gahanna, Ohio 43230 (614) 337-8700 Telephone: Facsimile: (614) 337-2274
Barbara S. Fant, Pharm.D. Contact person: Clinical Research Consultants, Inc. 310 Terrace Avenue Suite 201 Cincinnati, OH 45220 (513) 961-8200 Phone: Facsimile: (513) 961-2858
November 04, 2008 Date:
Trade Name: PercuCath™ Urinary Catheter
Urological Foley catheter Common name:
Classification Name: Class II; 21 CFR 876.5130, Urological Catheter and Accessories
EZL: Catheter, Retention Type, Balloon Product Code:
#### Identification of a Legally Marketed Predicate Device
The PercuCath™ Urinary Catheter is substantially equivalent to the Bardex® Lubri-Sil® 3-Way Folev Catheter marketed by C.R. Bard. Inc., 510(k) Premarket Notification Number: K002868 FDA Product Code EZL. Secondly, it is substantially equivalent to the Dover® 100% Silicone Foley Council Tip Catheter with manufactured by Tyco Healthcare, 510(k) Premarket Notification Number: K860484, FDA Product Code KOD.
#### General Description
The PercuCath™ Urinary Catheter is a flexible tubular device that is inserted through the urethra and used to pass fluids to and from the urinary tract. The PercuCath™ Urinary Catheter is a 100% silicone, latex-free, triple lumen (3-Way) Foley catheter with a lubricious hydrophilic coating and a straight or angled council tip to facilitate insertion. It is composed of a silicone tube that trifurcates into three lumens, a silicone balloon, and a two-way valve. When activated, the valve permits flow in either of two directions, i.e., for inflation or deflation of the balloon. Of the three lumens, one lumen is used for urinary drainage and can
#### Percuvision LLC
Proprietary and Confidential Information
11/04/2008
{1}------------------------------------------------
510(k), [Percuvision]
Page 19 of 50
be connected to a urine collection container (drainage bag or urine meter); one lumen has a two-way valve for inflation/deflation of the Foley balloon; and, the third lumen can be used for irrigation of the bladder. The council tip has an opening (eyelet) in the tip that can be used to pass a guidewire or similar device to facilitate catheter insertion. The PercuCath™ Urinary Catheter will be available with a single eyelet or double eyelet, 14 through 24 Fr. Shafts (i.e.,14, 16, 18, 20, 22, and 24), and either a 10 cc or 30 cc balloon.
#### Indications for Use and Intended Use
The PercuCath™ Catheter is indicated for use in the drainage and/or collection and/or measurement of urine. It is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as the nephrostomy tract.
#### Performance Data
Mechanical testing was performed on the PercuCath™ Urinary Catheter and the results demonstrate that the PercuCath™ Urinary Catheter met the design specifications for the device and were substantially equivalent to the predicate devices.
#### Basis of Substantial Equivalence
The PercuCath™ Urinary Catheter is substantially equivalent to the Bardex® Lubri-Sil® 3-Way Foley Catheter manufactured by C.R. Bard, Inc. in material, intended use, basic design concept, and biomechanical properties. Secondly, it is substantially equivalent to the Dover® 100% Silicone Foley Council Tip Catheter in its intended use and its council tip with an eyelet.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name arranged in a circular pattern around a stylized emblem. The emblem consists of a stylized eagle with three lines representing its wings.
#### Public Health Service
MAY 28 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Percuvision LLC % Mr. Jeff D. Rongero Senior Project Engineer UL Health Sciences -Underwriters Laboratories, Inc. 12 Laboratory Drive RESEARCH TRIANGLE PARK NC 27709-3995
Re: K090262
Trade/Device Name: PercuCath Urinary Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: May 12, 2009 Received: May 13, 2009
Dear Mr. Rongero: .
We have reviewed your Section 510(k) premarket notification of intent to market the device w o have and have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
(Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 876.xxx (240) 276-0115 (Obstetrics/Gynecology) 21 CFR 884.xxx (240) 276-0120 21 CFR 892.xxx (Radiology) Other (240) 276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 4. Indications for Use Statement
KO90262 510(k) Number (if known):
Device Name: PercuCath™ Urinary Catheter
#### Indications for Use:
The PercuCath™ Catheter is indicated for use in the drainage and/or collection and/or measurement of urine. It is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract. However, drainage is sometimes accomplished by suprapubic or other placement of the catheter, such as the nephrostomy tract.
Prescription Use X
OR
Over-The-Counter Use
### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Huld Reuner
(Division Sign Off) Division of Reproductive, Abdominal, and Radiological Devices K090262
510(k) Number_
Percuvision LLC Proprietary and Confidential Information
11/04/2008
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.