K110090 · Endologix, Inc. · DYB · May 26, 2011 · Cardiovascular
Device Facts
Record ID
K110090
Device Name
AFX INTRODUCER SYSTEM
Applicant
Endologix, Inc.
Product Code
DYB · Cardiovascular
Decision Date
May 26, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The AFX Introducer System facilitates the introduction of guidewires, and other medical devices into the vasculature and minimizes blood loss associated with such introduction.
Device Story
Endologix AFX Introducer System is a vascular access device consisting of a 45 cm hydrophilically-coated Pebax introducer sheath and two dilators (one single lumen, one dual lumen). The sheath features a PTFE liner, stainless steel braid, Vectran strands for strength, and a Platinum/Iridium radiopaque marker. The system includes an ABS hemostasis valve housing with a butyl rubber valve and silicone gasket to minimize blood loss. Used by clinicians in clinical settings to provide a conduit for interventional devices; the dilators provide stability and support during insertion over a 0.035" guidewire. The dual lumen dilator supports two-guidewire procedures. The device is provided sterile for single use. By facilitating safe vascular access and reducing hemorrhage, the system enables successful delivery of endovascular therapies.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included sterilization validation (ISO 11135-1), shelf-life packaging validation (ASTM F2096-04, F1929-98, F88M-09), biocompatibility (ISO 10993 series), and performance testing per ISO 10555-1 and ISO 25539-1. Performance metrics included successful guidewire passage, crossing profile verification, leak testing at 3.0 PSI, and kink resistance during 90-degree bends.
Indicated for use in patients requiring vascular access to facilitate the introduction of guidewires, catheters, and other interventional medical devices into the vasculature while minimizing blood loss. The dual lumen dilator configuration is indicated for use in two-guidewire procedures.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
Gore Introducer Sheath with Silicone Pinch Valve (K082356 and K032073)
Cook Medical Extra Large Check-Flo Introducers (K902469)
{1}------------------------------------------------
### MAY 2 6 2011
# APPENDIX 1
### 510(K) SUMMARY OR 510(K) STATEMENT
### Endologix AFX Introducer System 510(k) Summary - K110090 May 6, 2011
| 1. Company: | Endologix Incorporated<br>11 Studebaker<br>Irvine, CA 92618 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Janet M. Fauls<br>Vice President, Regulatory and Clinical Affairs<br>(949) 595-7203<br>(949) 595-7313 (fax)<br>jfauls@endologix.com |
| 2. Proprietary Trade Name: | Endologix AFX Introducer System |
| 3. Classification Name: | Catheter, Introducer (21 CFR 870.1340) |
| 4. Product Code: | DYB |
### Product Description: 5.
The Endologix AFX Introducer System consists of an Introducer Sheath (45 cm working length) and two Dilators. The system is enclosed in a sterile package. During clinical use, the system will allow introduction of a standard 0.035" guidewire within its inner lumen.
- · Introducer Sheath: The Introducer Sheath provides a conduit for introducing other interventional devices, including guidewires and interventional catheters, into the vasculature. The main components of this assembly are a hydrophilically-coated, PTFE lined, 55D Pebax introducer sheath, acrylobutylstyrene (ABS) hemostasis valve housing with butyl rubber valve, and silicone gasket assembly, 3 way stopcock/valve, and sideport diethylhexyl phthalate (DEHP)-free polyvinylchloride (PVC) tubing. The Pebax introducer sheath also includes an embedded stainless steel braid for increased strength and kink resistance, and embedded Vectran® strands for increased longitudinal strength. The distal tip of the introducer sheath includes an embedded radiopaque marker composed of 90% Platinum/10% Iridium.
- · Dilators: The dilators for the system are used to provide support and stability to the Introducer Sheath during deployment into the vasculature. The dilators are designed specifically for the Endologix AFX Introducer System and can accommodate a 0.035" guidewire through the central lumen. Two dilators are included with the AFX Introducer System. Both dilators are included within the package. One dilator is a single lumen device and one dilator is a dual lumen device. Both dilators are composed of 63D Pebax with 20% Barium Sulfate (BaSO4) for radiopacity and include a
{2}------------------------------------------------
hydrophilic coating on the outer surface. These dilators are designed in diameter and length to be placed within the AFX Introducer Sheath component. The dilators have a tapered, atraumatic distal tip. The proximal end of the dilators includes a luer port. The dilators provide stiffness to the assembly so that the introducer sheath can be placed within the vasculature during a standard vascular access.
### 6. Indications:
The AFX Introducer System facilitates the introduction of guidewires, and other medical devices into the vasculature and minimizes blood loss associated with such introduction.
### 7. Substantial Equivalence:
Documentation provided includes mechanical test results and detailed comparison to the predicate devices demonstrate that the Endologix AFX Introducer System is substantially equivalent to the Gore Introducer Sheath with Silicone Pinch Valve (K082356 and K032073), the Cook Medical Extra Large Check-Flo Introducers (K902469), and the Endologix IntuiTrak Introducer Sheath (P040002/S021). Reference is also made to the following Endologix products to support the clearance of the Endologix Dual Lumen Dilator: Dual Lumen Catheter (K991601) and IntuiTrak Single Lumen Dilator (P040002/S021)
Testing includes sterilization, shelf-life packaging validation, biocompatibility testing, and performance testing to ISO 10555-1 for intravascular catheters. A summary of this testing is provided in the following tables.
| Test | Test Description | Result<br>(Pass/Fail) |
|-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Sterility | Validation of the ethylene oxide sterilization cycle in<br>accordance with ISO 11135-1 to verify that the ethylene oxide<br>sterilization process provides a SAL of 10-6 | Pass |
Table 7-1: Steriliztation Validation
| | Table 7-2: Shelf-Life Packaging Validation | | |
|--|--------------------------------------------|--|--|
|--|--------------------------------------------|--|--|
| Test* | Test Description | Result<br>(Pass/Fail) |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Visual Inspection | Visual inspection of the packaging (inner and outer pouch) to<br>identify any anomalies that could impact integrity of the<br>package | Pass |
| Bubble Emission | ASTM F2096-04 Standard Test Method for Detecting Gross<br>Leaks in Medical Packaging by Internal Pressurization | Pass |
| Dye Penetration | ASTM F1929-98 (2004) Standard Test Method for Detecting<br>Seal Leaks in Porous medical Packaging by Dye Penetration | Pass |
| Peel Test | ASTM F88M-09 Standard test method for seal strength of<br>flexible barrier materials | Pass |
* Testing per ISO 11607-1: Packaging for Terminally Sterilized Medical Devices
{3}------------------------------------------------
### Table 7-3: Biocompatibility
.
| Test | Test Description | Result<br>(Pass/Fail) |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------|
| Cytotoxicity<br>ISO 10993-5 | MEM Elution Using L0929 Mouse Fibroblast Cells | Pass |
| Sensitization<br>ISO 10993-10 | Murine Local Lymph Node Assay | Pass |
| Intracutaneous Reactivity<br>ISO 10993-10 | Intracutaneous Irritation Test | Pass |
| Acute Systemic Toxicity<br>ISO 10993-11 | Acute Systemic Injection Test | Pass |
| Hemocompatibility<br>ISO 10993-4 | Complement Activation<br>Hemolysis Assay<br>Partial Thromboplastin Time<br>Platelet and Leukocyte Counts | Pass<br>Pass<br>Pass<br>Pass |
| Pyrogenicity USP<br>< 151> | Rabbit Pyrogen Test | Pass |
### Table 7-4: Performance Testing
| Test* | Test Description | Result<br>(Pass/Fail) |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------|
| Leak Testing | Connection fittings and valve seal do not leak liquid when<br>tested at 3.0 PSI | Pass |
| Bond and Component<br>Tensile Strength | Break force of each bond meets or exceeds requirements | Pass |
| Particle Size:<br>10-25μm <6,000<br>>25μm <600 | USP <788> requirements for particulate matter in injection<br>solutions of 100 mL or less in volume | Pass<br>Pass |
* Testing per ISO 10555-1 for intravascular catheters
| | Table 7-5: Compatibility with Other Devices |
|--|---------------------------------------------|
|--|---------------------------------------------|
| Test* | Test Description | Result<br>(Pass/Fail) |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Guidewire Passage | Verify the delivery system accepts 0.035" guidewire | Pass |
| Crossing Profile | Verify the catheter dimensions comply with product<br>specifications | Pass |
| Open Lumen Testing | Ensure that the guidewire passes through the system while in<br>a bent position | Pass |
| Kink Test | Ensure that the delivery system will not kink when inserted<br>through and retracted from the introducer sheath at a 90° angle | Pass |
| Introducer Sheath Transfer<br>Force Testing | Verify transfer force is in compliance with product<br>specifications | Pass |
* Testing per ISO 25539-1: Cardiovascular Implacts – Endovascular Devices
{4}------------------------------------------------
| Feature | Gore Introducer Sheath with<br>Silicone Pinch Valve<br>(K082356 and K032073) | Cook Medical Extra Large<br>Check-Flo Introducers<br>(K902469) | Endologix<br>Dual Lumen Catheter<br>(K991601) | Endologix IntuiTrak<br>Delivery System<br>- Introducer Sheath -<br>Single Lumen Dilator -<br>(P040002 / S021) | Endologix<br>AFX Introducer System<br>(K110090) |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DYB | DYB | DQY | --- | DYB |
| Intended Use/<br>Indications for<br>Use | Intended for use to facilitate the<br>introduction of guidewires,<br>catheters and other interventional<br>medical devices into the vascular<br>system, and to minimize blood loss<br>associated with such introduction | Intended for use to facilitate the<br>introduction of guidewires,<br>catheters and other interventional<br>medical devices into the vascular<br>system, and to minimize blood<br>loss associated with such<br>introduction | For use in a two guidewire<br>procedure | Integrated into the IntuiTrak Delivery<br>System; intended for use to facilitate the<br>introduction of guidewires, catheters and<br>other interventional medical devices into<br>the vascular system, and to minimize<br>blood loss associated with such<br>introduction | Introducer Sheath with Single Lumen<br>Dilator<br>Intended for use to facilitate the<br>introduction of catheters and other<br>medical devices into the vasculature<br>and to minimize blood loss associated<br>with such introduction<br><br>Introducer Sheath with Dual Lumen<br>Dilator<br>Intended for use in a two guidewire<br>procedure to facilitate the introduction<br>of catheters and other medical devices<br>into the vasculature and to minimize<br>blood loss associated with such<br>introduction |
# Table 7-6: Endologix AFX Introducer System Predicates Comparison Table
{5}------------------------------------------------
| Feature | Gore Introducer Sheath with<br>Silicone Pinch Valve<br>(K082356 and K032073) | Cook Medical Extra Large<br>Check-Flo Introducers<br>(K902469) | Endologix<br>Dual Lumen Catheter<br>(K991601) | Endologix IntuiTrak<br>Delivery System<br>- Introducer Sheath -<br>Single Lumen Dilator -<br>(P040002 / S021) | Endologix<br>AFX Introducer System<br>(K110090) |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Materials biocompatible for use in<br>the vasculature | Materials biocompatible for use<br>in the vasculature | Materials biocompatible for<br>use in the vasculature<br>including:<br>• Pebax (7233) components<br>• Polycarbonate luer<br>• PTFE Braided lumen | Materials biocompatible for use in the<br>vasculature including:<br>• Pebax for the sheath (72D) and<br>dilator (63D) components<br>• PTFE liner on the inner diameter of<br>the sheath<br>• Hydrophilic coating on the<br>Introducer sheath and dilator<br>• ABS handle<br>• Silicone gasket<br>• Butyl Rubber Trocar Valve<br>[K945719]<br>• DEHP free PVC Side port tubing<br>• Polycarbonate/silicone stopcock | Materials biocompatible for use in the<br>vasculature including:<br>• Pebax for the sheath (72D) and<br>dilator (63D) components<br>• PTFE liner on the inner diameter<br>of the sheath<br>• Hydrophilic coating on the<br>Introducer sheath and dilator<br>• ABS handle<br>• Silicone gasket<br>• Butyl Rubber Trocar Valve<br>[K945719]<br>• DEHP free PVC Side port tubing<br>• Polycarbonate and polyether<br>stopcock |
| Radiopaque<br>Marker | Radiopaque dilator for<br>fluoroscopic visibility | unknown | Gold | • 90% Platinum/10% Iridium marker<br>band embedded on the distal tip of<br>the Introducer Sheath<br>• Dilators are loaded with 30% BaSO4 | • 90% Platinum/10% Iridium marker<br>band embedded on the distal tip of<br>the Introducer Sheath<br>• Dilators are loaded with 20%<br>BaSO4 |
| Dimensions | Four configurations provided:<br>• 18 Fr - 22Fr<br>• 70 cm usable dilator length<br>• 30 cm sheath length<br>• Central dilator lumen<br>compatible with 0.035"<br>guidewire | Five configurations provided:<br>• 20 Fr - 24 Fr<br>• 56-81 cm usable dilator<br>length<br>• 25-65 cm outer sheath<br>length<br>• Central lumen compatible<br>with 0.035" guidewire | A single catheter<br>configuration provided:<br>9 Fr Catheter<br>• 90 cm usable length<br>• Central dilator lumen<br>compatible with 0.035"<br>guidewire | A single dilator configuration provided:<br>19 Fr Dilator<br>• 80.6 cm usable dilator length<br>• Central dilator lumen | A single dilatory configuration is<br>provided:<br>17 Fr Dilator<br>• 69 cm usable dilator length<br>• 3.1 cm tapered dilator tip<br>• 17 Fr dilator outer diameter<br>• 6.4 mm (19 Fr) outer sheath<br>diameter<br>• 45 cm sheath length<br>• 2 dilators included (single and<br>dual lumen)<br>• Central guidewire lumen<br>(dilators) |
| Feature | Gore Introducer Sheath with<br>Silicone Pinch Valve<br>(K082356 and K032073) | Cook Medical Extra Large<br>Check-Flo Introducers<br>(K902469) | Endologix<br>Dual Lumen Catheter<br>(K991601) | Endologix IntuiTrak<br>Delivery System<br>- Introducer Sheath -<br>Single Lumen Dilator -<br>(P040002 / S021) | Endologix<br>AFX Introducer System<br>(K110090) |
| Biocompatibility | Materials biocompatible for use in<br>the vasculature | Materials biocompatible for use<br>in the vasculature | Materials biocompatible for<br>use in the vasculature | Materials biocompatible for use in the<br>vasculature | Materials biocompatible for use in the<br>vasculature |
| Sterilization | Provided sterilized to a sterility<br>assurance level (SAL) of 10-6 by<br>ethylene oxide sterilization process | Provided sterilized to a sterility<br>assurance level (SAL) of 10-6 by<br>ethylene oxide sterilization<br>process | Provided sterilized to a<br>sterility assurance level (SAL)<br>of 10-6 by ethylene oxide<br>sterilization process | Provided sterilized to a sterility<br>assurance level (SAL) of 10-6 by<br>ethylene oxide sterilization process | Provided sterilized to a sterility<br>assurance level (SAL) of 10-6 by<br>ethylene oxide sterilization process |
| Introducer Sheath<br>Simulated Use<br>Testing | Compliant with ISO 10555-1 | Compliant with ISO 10555-1 | Compliant with ISO 10555-1 | All samples passed two 90° bends<br>without kinking All samples passed leak testing (up<br>to 8 psi) Crossing profile was measured and<br>was less than 6.47 mm | All samples passed two 90° bends<br>without kinking All samples passed leak testing<br>(up to 8 psi) Crossing profile was measured and<br>was less than 6.47 mm |
| Tensile Testing | Compliant with ISO 10555-1 | Compliant with ISO 10555-1 | Compliant with ISO 10555-1 | Compliant with ISO 10555-1 | Compliant with ISO 10555-1 |
| Shelf Life | Labeled with a 3-year shelf life. | Unknown | The Endologix Dual Lumen<br>Catheter is labeled with a<br>3-year shelf life. The device<br>functional, package<br>functional, and packaging<br>sterility has been qualified for<br>this shelf life. | The Endologix IntuiTrak Delivery<br>System is labeled with a 3-year shelf life.<br>The device functional, package<br>functional, and packaging sterility has<br>been qualified for this shelf life. | The Endologix AFX Introducer<br>System is labeled with a 3-year shelf<br>life. The device functional, package<br>functional, and packaging sterility has<br>been qualified for this shelf life. |
.
{6}------------------------------------------------
# Endologix, Inc.
K110090 510(k) Summary
May 2011
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 2 6 2011
Janet M. Fauls VP, Regulatory and Clinical Affairs Endologix, Inc. 11 Studebaker Irvine, CA 92618
Re: K110090
AFX Introducer System Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter. Introducer Regulatory Class: Class II Product Code: DYB Dated: May 23, 2011 Received: May 25, 2011
Dear Ms. Fauls:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Ms. Janet M. Fauls
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Bram Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{9}------------------------------------------------
## APPENDIX 2
### Indications for Use
510(k) Number: K110090
Device Name: Endologix AFX Introducer System
Indications for Use: The AFX Introducer System facilitates the introduction of guidewires, and other medical devices into the vasculature and minimizes blood loss associated with such introduction.
Prescription Use _______x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K110090
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.