feel-DRCS software using a digital X-ray detector is the digital X-ray image processing system designed to acquire images and process acquired images efficiently. The main features of this software are controlling and interfacing the detector, acquiring images after X-ray, storing acquired images, managing data, optimizing window level and width of acquired images, rotating images, zooming images, measuring images and so on. feel-DRCS is compatible with DICOM 3.0 standard. It can transfer images processed in PACS and print images with a film printer compatible with DICOM 3.0 by using DICOM and network systems. feel-DRCS is not approved for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data. feel-DRCS is not approved for the acquisition of mammographic image data.
Device Story
Software-based radiological image processing system (DROC); interfaces with digital X-ray detectors and generators. Inputs: patient study info (manual or DICOM worklist), raw image data from detector. Processing: raw image correction, noise reduction, enhancement, auto-windowing. Outputs: processed images displayed on monitor, transferred to PACS/DICOM printers. Used in clinical settings by radiologists/technicians. Enables image editing (crop, markers, window/level), measurement, and X-ray exposure parameter configuration (kVp, mA, sec) via generator interface. Benefits: efficient image acquisition, standardized DICOM workflow, improved image visualization for diagnosis.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Software-based image processing system; runs on Windows XP/7. Connectivity: DICOM 3.0, 10/100/1000 Ethernet. Database: MS-Access. Interfaces: Digital X-ray detectors and X-ray generators (protocol-dependent). Features: image enhancement, noise reduction, auto-windowing, measurement, annotation, DICOM storage/print/worklist. Standalone software application.
Indications for Use
Indicated for qualified medical personnel to acquire, process, store, and manage digital X-ray images. Not for mammographic image data. Users must be qualified to create and diagnose radiological image data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K110033
SEP 1 5 2011
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
Submitter Information: Imfou Co., Ltd #301, 3Fl, EnC Venture Dream Tower 5st, 197-13 Guro 3-dong, Guro-gu, Seoul 152-179, Korea.
Contact Person:
Ho Dong, Yang Onbix Corporation #821 Samil Plaza, 837-26 Yeuksam-dong Gangnam-gu, Seoul (135-768) Korea Tel: *82-2-566-3360 / Fax: *82-2-6280-3360 Email: onbix@naver.com
Date Summary Prepared: May 17, 2011
#### Device Name:
| Trade Name(s) : | feel-DRCS |
|-----------------------|----------------------------------------------|
| Classification Name : | Picture archiving and communications system. |
| Common Product Name : | Radiological Image Processing System. |
| Panel : | Radiology |
| Product Code: | LLZ |
Predicate Device Information:
K091364 / dicomPACS DX-R 1.6
### Device Description:
This medical imaging device software is composed of a combination of many functions. The functions include interfacing X-ray detector, integration X-ray generator, networking PACS, image enhancement processing and editing image/information. The name of this software is called DROC(Digital Radiography Operation Console)
The operational process of feel-DRCS is described below:
- 1 Input or select study information patient ID/Name, body part and projection, if you can use DICOM Modality Worklist Management SCP, you select in worklist.
- Setup exposure parameter of X-ray generator (2)
- After exposure, this software acquires raw images from X-ray Detector. (3)
- A After acquisition of raw image, this software corrects raw image.
203 (1)
{1}------------------------------------------------
- (5) After correction of raw image, this software processes with noise reduction, enhancement, auto windowing etc.
- (6) After processing of corrected image, this software displays an image on the monitor.
- (7) If this software is connected with PACS network, it transfers images to PACS Server or DICOM film printer.
## Intended Use:
feel-DRCS software using a digital X-ray detector is the digital X-ray image processing system designed to acquire images and process acquired images efficiently.
The main features of this software are controlling and interfacing the detector, acquiring images after X-ray, storing acquired images, managing data, optimizing window level and width of acquired images, rotating images, zooming images, measuring images and so on.
feel-DRCS is compatible with DICOM 3.0 standard. It can transfer images processed in PACS and print images with a film printer compatible with DICOM 3.0 by using DICOM and network systems.
feel-DRCS is not approved for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.
feel-DRCS is not approved for the acquisition of mammographic image data.
#### Comparison to Predicate Device(s):
This device is equivalent to the predicate devices in its intended use and technological characteristics, including:
- * indications for use
- * technological characteristics .
- * performance properties
| Product name | Predicate device (K091364)<br>dicomPACS DX-R 1.6 | New device<br>feel-DRCS |
|--------------------------|---------------------------------------------------------------------------------|-------------------------------------------------------|
| Intended use | Acquiring, viewing, editing and storing radiographs and related patients images | same |
| Intended user | Radiologist | Radiologist |
| Operating system | Microsoft Windows XP/Vista/7 | Microsoft Windows XP/7 |
| Network | 10/100/1000 Ethernet | 10/100/1000 Ethernet |
| Monitor | 19 inch Monitor(1280 X 1024, etc) | Above 19inch Monitor(Above 1280 X<br>900) |
| User interaction / input | Mouse, Keyboard, Touch Monitor, | Mouse, Keyboard, Touch Monitor, |
| Multi-user | Available<br>But at a time, only one user can use<br>it. | Available<br>But at a time, only one user can use it. |
| Import / export images | Yes | Yes |
| Acquisition devices | Computed Radiography<br>Digital X-ray Detector | Computed Radiography<br>Digital X-ray Detector |
003 (2)
{2}------------------------------------------------
| Product name | Predicate device (K091364)<br>dicomPACS DX-R 1.6 | New device<br>feel-DRCS |
|----------------------------|--------------------------------------------------|------------------------------------|
| Imaging interfaces | Detector dependent | Detector dependent |
| Image organization | Yes | Yes |
| | Patient ID/Name/Study instance UID | Patient ID/Name/Study instance UID |
| Image search available | Yes | Yes |
| Image storage | Yes | Yes |
| Database storage | Yes | Yes |
| Database software | MS-SQL | MS-Access |
| Image viewing | Yes | Yes |
| Image measurement | Yes | Yes |
| Image annotation | Yes | Yes. |
| Image operations | Yes | Yes |
| security | Yes(Priority by User) | Yes(Priority by User) |
| DICOM 3.0 Compatibility | Yes | Yes |
| Generator Control | OK | OK |
| Generator Control protocol | Generator dependent | Generator dependent |
| Raw image data processing | OK | OK |
| Post image data processing | OK | OK |
| RIS Code manager | OK | OK |
# Summary of non-clinical testing:
This section is not applicable.
# Conclusion
-- |
: :
ン・・ : 一
Based on the information provided in this summary we conclude that this device is substantially equivalent to the predicate device K091364.
003 (3)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three wing-like shapes.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002
Imfou Co., Ltd. % Mr. Ho Dong Yang Cito Onbix Corporation #821 Smail Plaza, 837-26 Yeujsam-dong Gangnam-gu, Seoul (135-768) REPUBLIC OF KOREA
Re: KI10033
Trade/Device Name: Radiological Image Processing System/feel-DRCS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 8, 2011 Received: August 12, 2011
Dear Mr. Ho Dong Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
ff you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (01) 796 5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollit the (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
- 510(k) Number (if known): 110033
- Device Name: Radiological Image Processing System / feel-DRCS
- Indications for Use:
feel-DRCS software, used together with digital X-ray detector is the digital X-ray image processing system designed for acquiring images and processing acquired images efficiently. The main features of this software are controlling and interfacing the detector, acquiring images after X-ray, storing acquired images, managing data, optimizing window level and width of acquired images, rotating images, zooming images, measuring images and other features used for imaging processing.
feel-DRCS is compatible with DICOM 3.0 standard. It can transfer images processed in PACS and print images with a film printer compatible with DICOM 3.0 by using DICOM and network systems.
feel-DRCS is not approved for the acquisition of mammographic image data and is meant to be used by qualified medical personnel only. All users must be qualified to create and diagnose radiological image data.
Image /page/5/Figure/7 description: The image shows a diagram of a medical imaging system. The system includes a PACS server, a personal computer, a detector, and detector application software. The personal computer is running software called "feel-DRCS" which performs the following functions: Get DICOM worklist by network, Image acquisition, Image Processing, Image store, and DICOM Sending by network.
Following is the diagram of feel-DRCS software system.
The main functions of feel-DRCS as follows:
- A. Acquisition and storage of digital X-ray images from a digital X-ray Detector.
- B. Input Study information (patient information, exam information).
- C. Management of DataBase of stored (archived) images.
Page 1 of _2
{6}------------------------------------------------
- D. Image processing for enhancement of archived images.
- E. Review of stored images.
- F. Edit of image(Shutter or crop, insert Marker("L", "R"), Window width/level etc)
- G. DICOM Functions(DICOM Storage, DiCOM Worklist, DICOM Print, etc)
- H. In case of DR full system(X-ray machine and generator and Digital X-ray detector and etc) or need a interface with installed X-ray System in the feel-DRCS.
- (1) Ability to configure X-ray exposure condition (kVp, mA, Sec etc) for various body parts and positions.
- (2) Communication between Generator Console and feel-DRCS.
* This X-ray generator control function depends on the X-ray Generator company. Because feel-DRCS can only interface and control by the algorithm provided by the X-ray Company. And feel-DRCS can only select or change values of X-ray exposure parameters(kVp, mA, Second or kVp; mAs) according to defined value of each X-ray company. The feel-DRCS doesn't control exposure and electrical change and calibration X-ray. So, before exposure of Xray, radiological technician should check the X-ray exposure conditions in the console of X-ray Generator.
* If the X-ray generator doesn't allow interface with external software(ex. Feel-DRCS), the software cannot interface with X-ray Generator.
## Users
- A. For acquisition of Digital X-ray image by Digital X-ray Detector.
- B. For development or Production of Digital Radiography Retrofit system on the conventional X-ray system.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary S. Patel
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
| 510(k) Number | K110033 |
|---------------|---------|
|---------------|---------|
Page 2 of 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.